Brain DeathNCT07768215

Miniature In-line Respirator

A study testing Hope inVent in people with Brain Death and Respiration, Artificial, run by National Institutes of Health Clinical Center (CC) at 1 site in United States.

Not yet recruiting

Opening soon. The study is approved but enrolment has not started yet. Sites may already take enquiries.

Sites and contacts
Official record
Search all trials

Trial summary

This first-in-human pilot study will test an investigational miniature breathing device called the Hope inVent. The device is a small, single-use, in-line respirator that connects to a breathing tube and uses compressed oxygen to help move air in and out of the lungs. Read more about this study

Who is it for?
Ages 1 year to 100 years
Conditions the study is for
Brain Death, Respiration, Artificial
What is being tested?
Hope inVent
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
5 participants
Where
1 siteUnited States
Study dates
29 Oct 2026to 1 Dec 2028
Phase
Interventional
Are there results?
Not yet: the study is under way

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 1 year to 100 years
  • All sexes
  • No healthy volunteers
  • Child, Adult, Older adult

Who may be able to join

  • Diagnosis of brain death
  • Provision of signed and dated informed consent form for organ donation and research by donor family or surrogate decisionmaker.
  • Male or female, with no age restriction (see size exclusion).
  • Decedent transferred to the Gift of Life at Penn Medicine Donor Care Center (study site) for routine organ donor management and organ recovery procedures.

Who cannot join

  • An individual who meets any of the following criteria will be excluded from participation in this study:
  • Brain-dead research donor will be excluded if their baseline CO(2) or pulse oximetry is out of the accepted range for the study (defined as end tidal CO(2) between 35- and 45-mm Hg and the O(2) saturation measured by pulse oximetry > 94%) while on optimized ventilation.
  • Brain-dead research donors with severe respiratory disease. This may include:
  • Bullous lung disease - noted on pre-op imaging or by history.
  • h/o pneumothoraces
  • h/o severe asthma, COPD, or other significant lung disease
Show all 16Show fewer
  • Brain-dead research donors with morbid obesity (BMI >=40).
  • Brain-dead research donor body weight of < 25kg.
  • Potential Brain-dead lung transplant donors (as determined by Gift of Life Donor Program staff according to at the time of study site (Donor Care Center) admission)
  • Intraoperative arterial blood gas values outside the range of normal for pH, PaO2, PaCO2, HCO3, O2Sat as measured in pre-device ABG (while on optimized ventilation).
  • Acceptable range for Arterial Blood Gas (ABG values)
  • Parameter - pH / Range - 7.35-7.45
  • Parameter - PaO2 / Range - >=75 mm Hg
  • Parameter - PaCO2 / Range - 35-45 mm Hg
  • Parameter - HCO3 / Range - 22-26 mEq/L
  • Parameter - SpO2 / Range - 94-100

Trial locations and contacts

1 site in 1 country. Contact the study team or a site near you to ask about taking part.

Tracy S Williams, R.N. contact

(301) 448-5366

tracy.williams@nih.gov

Andrew J Mannes, M.D. contact

(301) 594-3427

amannes@cc.nih.gov

United States1 site
  • Hospital of the University of Pennsylvania

    Philadelphia, Pennsylvania, 19104

    Andrew Mannes, M.D., 301-594-3427, amannes@cc.nih.gov

Other trials in United States

About this study

Official title: First-In-Human Trial of Miniature In-Line, Pressure-Cycled, Emergency Respirator Device in Human Organ Donors

This first-in-human pilot study will test an investigational miniature breathing device called the Hope inVent. The device is a small, single-use, in-line respirator that connects to a breathing tube and uses compressed oxygen to help move air in and out of the lungs. It has no moving parts and is designed to provide short-term breathing support in situations where standard ventilators may not be available, such as emergencies, disasters, battlefield settings, transport, or ventilator shortages.

The study will enroll up to 5 brain-dead organ donors who are already receiving mechanical ventilation at the Gift of Life Donor Care Center at the Hospital of the University of Pennsylvania. Donors will only be included after authorization for organ donation and research has been obtained through standard procedures. Donors who may donate lungs or who have significant lung disease will not be included.

Read more

During the study, investigators will briefly replace the standard ventilator with the Hope inVent device for up to 15 minutes. The standard ventilator will remain immediately available, and the donor will be continuously monitored by the clinical and research teams. Investigators will measure whether the device can maintain acceptable breathing measures, including oxygen level and carbon dioxide level. They will also record breathing pressures and tidal volume, which is the amount of air delivered with each breath. If the device does not perform as expected or if any safety concern occurs, the donor will be returned immediately to the standard ventilator.

There is no direct benefit to the organ donor. Information from this study may help determine whether this type of simple, low-cost breathing device could be useful for future patients who need emergency or short-term ventilator support when standard ventilators are not available.

Read the detailed description

Study Description:

Proof of Concept pilot study to support use of a miniature in-line patient respirator for patient ventilation needs.

Objectives:

Primary Objective:

To assess the effectiveness of the miniature respirator to ventilate intubated braindead research donors while maintaining respiratory measures within acceptable range.

Secondary Objective

The miniature respirator performs as designed and tested in a lung simulator.

Endpoints:

Primary Endpoint:

A binary variable equal to 1 ("success") if both of the following criteria are met (during the placement of the test device):

  1. the respirator maintains the end tidal CO(2) between 35- and 45-mm Hg; and
  2. the O(2) saturation measured by pulse oximetry is > 94%.

Otherwise, the primary endpoint is equal to 0 ("failure").

Secondary Endpoints:

Binary variables equal to 1 ("success") if tidal volume and airway pressure settings (PEEP and maximum peak airway pressure) are within 30% of predicted values. Otherwise, the secondary endpoints are equal to 0 ("failure").

Being tested
Hope inVent
Keywords
Hope inVent Respirator, Miniature In-line Respirator, Pressure-Cycled Ventilation
Medical subject headings
Brain Death, Respiratory Aspiration

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Hope inVentDevice

    The Hope inVent is an investigational, single-use, miniature in-line respirator designed to provide short-term ventilatory support for intubated individuals. The device is a 3D-printed, pressure-cycled, gas-powered respirator with no moving parts. It connects to the endotracheal tube breathing circuit and uses compressed oxygen as its power source. In this study, only the Green Hope inVent model will be used. The device cycles between fixed pressure settings: peak inspiratory pressure of 18 cm H2O and positive end-expiratory pressure of 6 cm H2O. Enrolled brain-dead research donors will be placed on the device once for up to 15 minutes, with continuous respiratory and physiologic monitoring. The donor will then be returned to the standard mechanical ventilator.

Study groups in NCT07768215
Study groupRoleWhat they receive
Hope inVent Miniature In-line RespiratorExperimentalAll enrolled brain-dead organ donors who meet eligibility criteria will receive the investigational Green Hope inVent miniature in-line, pressure-cycled respirator for up to 15 minutes. Respiratory and physiologic measures will be collected before device placement while the donor is on the standard mechanical ventilator and during use of the Hope inVent device. Each donor serves as their own control; there is no separate comparator arm. The standard ventilator will remain available, and donors will be returned to standard ventilation after testing or earlier if safety or performance criteria require.
Receives: Device: Hope inVent

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Proportion of donors with successful ventilation and oxygenation during Hope inVent testing.

    Success is defined as maintaining both end-tidal carbon dioxide between 35 and 45 mm Hg and oxygen saturation by pulse oximetry greater than 94% during placement of the Hope inVent device. If either criterion is not met, or if device use is stopped early because of inadequate ventilation or oxygenation, the outcome is classified as failure.

    Measured: During device use, from connection of the Hope inVent device to removal, up to 15 minutes.

Secondary outcomes (3)
  1. Proportion of donors with tidal volume within 30% of predicted value

    Tidal volume will be measured using an in-line respiratory monitor while the donor is ventilated with the Hope inVent device. Each donor will be classified as a success if tidal volume is within 30% of the protocol-specified predicted value; otherwise, the donor will be classified as a failure.

    Measured: During device use, from connection of the Hope inVent device to removal, up to 15 minutes.

  2. Proportion of donors with positive end-expiratory pressure within 30% of predicted value

    Positive end-expiratory pressure will be measured using an in-line respiratory monitor during Hope inVent ventilation. Each donor will be classified as a success if positive end-expiratory pressure is within 30% of the protocol-specified predicted value; otherwise, the donor will be classified as a failure.

    Measured: During device use, from connection of the Hope inVent device to removal, up to 15 minutes

  3. Proportion of donors with maximum peak airway pressure within 30% of predicted value

    Maximum peak airway pressure will be measured using an in-line respiratory monitor during Hope inVent ventilation. Each donor will be classified as a success if maximum peak airway pressure is within 30% of the protocol-specified predicted value; otherwise, the donor will be classified as a failure.

    Measured: During device use, from connection of the Hope inVent device to removal, up to 15 minutes

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted17 Aug 2026
  2. Study start29 Oct 2026 (estimated)
  3. Primary completion1 Dec 2028 (estimated)
  4. Study completion1 Dec 2028 (estimated)

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Device feasibility
Allocation
Not applicable
Intervention model
Single group
Masking
None (open label)
Enrolment
5 participants (estimated)
IDs, dates and publications
ClinicalTrials.gov ID
NCT07768215
Sponsor's protocol ID
10002654
Other IDs
002654-CC
Status last verified
12 Aug 2026
First submitted
14 Aug 2026
Last updated
26 Oct 2026
Results posted
No
FDA-regulated drug
No
FDA-regulated device
No
Individual data sharing
Yes: De-identified or anonymized individual donor-level data may be shared with collaborating investigators or institutions at NIH, the University of Pennsylvania, or outside institutions, and/or submitted to NIH-designated databases, consistent with donor authorization, applicable IRB approvals, NIH policies, and federal regulations. Shared data will not include direct identifiers. If coded data are shared, the code key will not be provided to collaborators. No biospecimens will be stored or shared for future use.

Questions about this trial

What is this clinical trial studying?

A study testing Hope inVent in people with Brain Death and Respiration, Artificial, run by National Institutes of Health Clinical Center (CC) at 1 site in United States.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 4 inclusion and 16 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: opening soon, as of the sponsor's last update on 26 Oct 2026. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in United States. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

This study is still under way. Results are not available in the source record yet.

How can a doctor refer a patient or contact the research team?

Contact Tracy S Williams, R.N. ((301) 448-5366, tracy.williams@nih.gov), or a site near the patient. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Proportion of donors with successful ventilation and oxygenation during Hope inVent testing. (time frame: During device use, from connection of the Hope inVent device to removal, up to 15 minutes.). It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 26 Oct 2026.

Free account

Keep searching with a free account

You’ve used your 3 free searches for this session. Log in or create a free AzTrials account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in