Complex Abdominal Aortic AneurysmsfenCoNCT07529275
Clinical Performance and Safety Assessment of the Use of iCover Balloon Expandable Covered Stent (iVascular) Implanted as Bridging Stent in FEVAR (Fenestrated Endovascular Aortic Repair) for the Treatment of Complex Abdominal Aortic Aneurysms
A study testing iCover balloon-expandable covered peripheral stent system as bridging stent in people with Complex Abdominal Aortic Aneurysms, run by iVascular S.L.U. at 24 sites in 5 countries.
Trial summary
fenCo is a European multicentre, prospective study to evaluate the use of the iCover covered stents as bridging stents for reno-visceral target vessel during fEVAR for the treatment of complex abdominal aortic aneurysms. Read more about this study
- Who is it for?
- Ages 55 years and older
- Conditions the study is for
- Complex Abdominal Aortic Aneurysms
- What is being tested?
- iCover balloon-expandable covered peripheral stent system as bridging stent
- Type of treatment
- Medical device
- Could I receive a placebo?
- No, everyone receives an active treatment
- How many people
- 165 participants
- Where
- 24 sitesin 5 countries
- Study dates
- 16 Feb 2026to 30 Sep 2028
- Phase
- Interventional
- Are there results?
- Not yet: the study is under way
Key requirements
Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.
- Ages 55 years and older
- All sexes
- No healthy volunteers
- Adult, Older adult
Who may be able to join
- Patient suitable for treatment of a juxta, para-renal, suprarenal or thoracoabdominal aneurysm with a fenestrated endoprosthesis in accordance with current indications and guidelines for the management of abdominal aortic aneurysms (aneurysm > 5 centimetres for women and > 5.5 centimetres for men and/or growth > 5 mm in 6 months or 1 centimetre in 1 year).
- The fenestrated endograft must be designed in a way that positions the fenestrations directly in front of the orifice of the target vessel, ensuring that the distance between the fenestration and the orifice of the target vessel does not exceed 5 mm.
- Anatomy and patient strictly in accordance with the Instructions For Use (IFU) of the fenestrated endograft chosen to be implanted:
- Cook Medical: Zenith Fenestrated AAA Endovascular Graft
- Terumo Aortic: TREO and Anaconda Fenestrated Stent Grafts
- JOTEC (part of Artivion): E-xtra design engineering.
Show all 15Show fewer
- Additional iCover stent, if necessary, could be placed in the same target artery (2 iCovers stents max as bridging stent for each fenestration).
- Landing zone in the target vessel of at least 10 mm.
- Target arteries (renal arteries, superior mesenteric artery, celiac trunk) with a diameter between 5 and 10 mm.
- Angulation of the aorta at the level of the target vessels < 45 degrees.
- No early significant branching from the target vessel with a potential risk of coverage and subsequent significant risk of target organ infarctions.
- Age > 55 years.
- Patient having notified his consent to this study and willing to comply with specified follow-up evaluations at the specified times.
- Patient affiliated to or benefiting from a social security system.
- Patient with life expectancy > 12 months.
Who cannot join
- Patients contraindicated for anti-platelet therapy.
- Patients with uncontrolled haematological disorders or heparin-induced thrombocytopenia.
- Chronic or acute aortic dissection.
- Patients refusing treatment.
- Patients who are pregnant or wish to become pregnant.
- Patients scheduled for major or life-saving surgery within 30 days of the fenestrated stent procedure.
Show all 21Show fewer
- Patients considered hemodynamically unstable or requiring emergency treatment.
- Patients with severe arteriopathy leading to adverse outflow, which may impair bridging stent patency on the targeted artery.
- Thrombus in the aortic sealing zone and target arteries with thickness > to 3 mm.
- Stenosis (>50%) or occlusion of target arteries or distal disease resulting in poor iCover stent outflow.
- Patients allergic to stent materials (L605) and/or PTFE.
- Patients requiring a hybrid aortic technique with branches, semibranches or chimneys.
- Patient who has been implanted with any stent type or brand other than iCover as a bridging stent in any target artery.
- Patients with infectious/mycotic aneurysms.
- Angulation between renal artery and aortic wall < 30 degrees.
- Patients with complex iliac accesses out the IFUs of the fenestrated endograft chosen to be implanted (COOK Medical / Terumo Aortic/Jotec-Artivion) with the iCover Bridging stent.
- Patient is currently participating in another investigational drug or device study that has not completed the entire follow up period that may interfere with the results of this study.
- Myocardial infarction or stroke within 3 months prior to the procedure.
- Patients with an unstable angina pectoris or heart insufficiency NYHA 3 or 4.
- Patients with ASA classification 5 or higher.
- Patients with physician modified endografts or in situ laser FEVAR.
Trial locations and contacts
24 sites in 5 countries. Contact the study team or a site near you to ask about taking part.
24 sites in 5 countries
France8 sites, 8 recruiting
- Recruiting
CHU BORDEAUX-Hopital Tripode
Bordeaux, 33076
Eric Prof. DUCASSE, 936724711 ext. +34, clinica@ivascular.global
- Not yet recruiting
CHU BREST- la Cavale Blanche
Brest, 29200
Bahaa Prof. NASR, 936724711 ext. +34, clinica@ivascular.global
- Not yet recruiting
CHU LILLE- Institut Cœur Poumon
Lille, 59000
Jonathan Prof. SOBOCINSKI, 936724711 ext. +34, clinica@ivascular.global
- Not yet recruiting
APHM- Hôpital De La Timone
Marseille, 13005
Virgile Dr. OMNES, 936724711 ext. +34, clinica@ivascular.global
- Not yet recruiting
Hôpital St Joseph
Marseille, 13008
Nicolas Dr. VALIERO, 936724711 ext. +34, clinica@ivascular.global
- Not yet recruiting
Nouvel Hôpital Privé les Franciscaines
Nîmes, 30000
Nicolas Dr. LOUIS, 936724711 ext. +34, clinica@ivascular.global
- Not yet recruiting
CHU TOULOUSE- Hôpital Rangueil
Toulouse, 31400
Xavier Prof. CHAUFOUR, 936724711 ext. +34, clinica@ivascular.global
- Not yet recruiting
Hôpital privé de Villeneuve d'Ascq
Villeneuve-d'Ascq, 59650
Martin Dr ROUER, 936724711 ext. +34, clinica@ivascular.global
Belgium7 sites, 7 recruiting
- Not yet recruiting
Imelda Ziekenhuis
Bonheiden, 2820
Wouter Dr. VAN DEN EYNDE, 936724711 ext. +34, clinica@ivascular.global
- Not yet recruiting
ZOL GENK
Genk, 3600
Wouter Dr. LANSINK, 936724711 ext. +34, clinica@ivascular.global
- Not yet recruiting
Maria Middelares GENT
Ghent, 9000
Willem Dr. WILLAERT, 936724711 ext. +34, clinica@ivascular.global
- Not yet recruiting
UZ GENT
Ghent, 9000
Natalie Dr. MOREELS, 936724711 ext. +34, clinica@ivascular.global
- Not yet recruiting
Grand hopital Saint Josep
Gilly, 6060
Philippe Dr. REMY, 936724711 ext. +34, clinica@ivascular.global
- Not yet recruiting
Jessa Hasselt
Hasselt, 3800
Frederic Dr. LIEBEER, 936724711 ext. +34, clinica@ivascular.global
- Not yet recruiting
Chu Liege
Liège, B-4000
Arnaud Dr KERZMANN, 936724711 ext. +34, clinica@ivascular.global
Germany6 sites, 6 recruiting
- Not yet recruiting
University Hospital Cologne
Cologne, 50937
Bernhard Prof. DORWEILER, 936724711 ext. +34, clinica@ivascular.global
- Not yet recruiting
Asklepios Klinik St. Georg
Hamburg, 20099
Thomas Prof. KOEPPEL, 936724711 ext. +34, clinica@ivascular.global
- Not yet recruiting
Universitätsklinik Hamburg Eppendorf
Hamburg, 20246
Tilo Prof. KOELBEL, 936724711 ext. +34, clinica@ivascular.global
- Not yet recruiting
Uni-Klinikum Leipzig
Leipzig, 04103
Andrej Prof. SCHMIDT, 936724711 ext. +34, clinica@ivascular.global
- Not yet recruiting
LMU München Campus Großhadern
München, 81377
Nikolaos Prof. TSILIMPARIS, 936724711 ext. +34, clinica@ivascular.global
- Not yet recruiting
TUM Munich
München, 81675
Daniela Dr BRANZAN, 936724711 ext. +34, clinica@ivascular.global
Netherlands2 sites, 2 recruiting
- Not yet recruiting
Amsterdam UMC
Amsterdam, 1105
Vincent Dr.JONGKIND, 936724711 ext. +34, clinica@ivascular.global
- Not yet recruiting
University Medical Center Groningen
Groningen, 9700
Jean-Paul Dr. DE VRIES, 936724711 ext. +34, clinica@ivascular.global
Spain1 site, 1 recruiting
- Not yet recruiting
Hospital Clinic - University of Barcelona
Barcelona, 08036
Viçenc Prof. RIAMBAU, 936724711 ext. +34, clinica@ivascular.global
About this study
fenCo is a European multicentre, prospective study to evaluate the use of the iCover covered stents as bridging stents for reno-visceral target vessel during fEVAR for the treatment of complex abdominal aortic aneurysms.
Read the detailed description
In terms of complex endovascular repair, fenestrated stent grafts should be considered the preferred treatment option when feasible. The main advantage of Fenestrated Endovascular Aneurysm Repair (fEVAR) lies in the avoidance of aortic cross clamping and subsequent lower risk of renal dysfunction, less surgical trauma and faster recovery, which may be advantageous for patients at high risk for open surgery. fenCo study aims to demonstrate the safety and efficacy of iCover when used as bridging stent in fEVAR procedures for the treatment of complex abdominal aortic aneurysms.
- Conditions
- Complex Abdominal Aortic Aneurysms
- Being tested
- iCover balloon-expandable covered peripheral stent system as bridging stent
- Keywords
- FEVAR, bridging stent, complex abdominal aortic aneurysms
What participants receive
What participants receive. In a randomised study each person is assigned to one group by chance.
iCover balloon-expandable covered peripheral stent system as bridging stentDevice
This device will be used as bridging stent during FEVAR (fenestrated endovascular aortic repair) procedures.
| Study group | Role | What they receive |
|---|---|---|
| iCover balloon-expandable covered peripheral stent system | Experimental | iCover balloon-expandable covered peripheral stent system Receives: Device: iCover balloon-expandable covered peripheral stent system as bridging stent |
What is being measured
The primary outcome is the main question the study is designed to answer.
Bridging stent patency at 12-month visit
Defined as rate of stents with an absence of binary restenosis (≥ 50% stenosis) based on CT Angio at 12 months visit (+/-2 months).
Measured: 12 months visit
Absence of target vessel instability between implantation and 12-month visit.
Defined as rate of patients with an absence of target vessel instability described as a composite endpoint of any branch-related complication leading to: aneurysm rupture/ death/ occlusion/ component separation/ or reintervention to maintain branch patency or to treat a branch-related component separation or endoleak between implantation and 12 months visit (+/-2 months).
Measured: starting at implant day through the study completion, with an average of 12Months up to 14Months per window.
Secondary outcomes (15)
Technical success: Successful introduction and deployment of the iCover stent implanted as bridging stent in FEVAR procedure.
Defined as rate of stents with successful introduction and deployment of the iCover (iVascular) implanted as bridging stent in FEVAR.
Measured: During the index procedure (from stent introduction to final deployment, up to procedure completion)
Bridging stent patency at the first postoperative imaging, at 6 and at 12 months visits.
Defined as rate of stents without binary restenosis > 50% or occlusion on duplex scan and/or angioCT scan examination, or angiography at discharge at 6 and 12 months visits.
Measured: at discharge visit (up to 3 months, average of 7days) and at 6 and 12 months visits ( can be up to 8 and 14months from implant day)
Absence of endoleaks linked to iCover covered stent defects at the first postoperative imaging, at 6 and 12 months visits.
Defined as rate of stents without endoleaks linked to iCover covered stent defects (Malposition, rupture or disconnection) on Duplex scan, angio CTscan or angiography at discharge visit, at 6 months visits and at 12 months visit.
Measured: discharge visit , 6 and 12 months visits.
Absence of re-intervention due to a technical defect in iCover covered stents at discharge, at 6 months visit, and at 12 months visit.
Defined as rate of stents without re-intervention due to a technical defect in iCover covered stents post-procedure between surgery and post operative to discharge period, between surgery and 6 months visit, between surgery and 12 months visit.
Measured: between surgery and post operative to discharge period, between surgery and 6 months visit, between surgery and 12 months visit.
Absence of type I or III endoleaks post-procedure at the first postoperative imaging, at 6 and 12 months visits.
Defined as rate of stents without type I or III endoleaks post-procedure on duplex scan, angio CTscan or angiography at discharge period, 6 months visit and at 12 months visit.
Measured: discharge visit, 6 and 12 months visits.
30-day post-procedure mortality.
Defined as rate of post-procedure mortality (all causes) within 30-day post procedure.
Measured: within 30 days post procedure
In-hospital post-procedure mortality (all causes).
Defined as rate of In-hospital post-procedure mortality (all causes).
Measured: between procedure and 12 months visit
Absence of aortic endograft migration at 6 months visit and at 12 months visit.
Defined as rate of endografts without aortic migration > 10 mm at 6 Months visit and 12 months visit.
Measured: at 6 and 12 months visits
No increase in maximum aneurysmal sac diameter measured by CT angiography at 6 and at 12 months visits.
Defined as rate of patients without increase of > 5 mm in maximum aneurysmal sac diameter compared with post-procedure angio CT scan measured by angio CT scan at 6 and 12 months visits.
Measured: at 6 and 12 months visits
Absence of post-procedural aortic surgical conversion at discharge, at 6 and 12 months visits.
Defined as absence of post-procedural aortic surgical conversion between the implantation and the follow-up visits
Measured: between implant and 6 and 12 months visits
Absence of aneurysm-related mortality post-procedure at discharge, at 6 months and 12 months visits.
Defined as absence of aneurysm-related mortality post-procedure between the implantation and the follow-up visits.
Measured: between implant and 6 and 12 months visits
Absence of aneurysm related secondary endovascular procedures at discharge, at 6 and 12 months visits.
Defined as absence of aneurysm related secondary endovascular procedures between the implantation and the follow-up visits.
Measured: between implant 6 and 12 months visits.
Absence of major adverse events (MAEs) at discharge, at 6 and 12 months visits.
Defined as aortic rupture, aneurysm-related death, in-hospital death, decline in renal function (by at least 20%), lower extremity ischemia, mesenteric ischemia, or unplanned reintervention between the implantation and the follow-up visits.
Measured: between implant and 6, 12 months vists.
Absence of target vessel instability at 12 months visit.
Absence of target vessel instability at 12 months visit.
Measured: at 12 months visit
Target vessel patency at at 6 and 12 months visits
Defined as proportion of target vessel patency at 6 and 12 months.
Measured: 6 and 12 months visits
Timeline
From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.
- First posted14 Apr 2026
- Study start16 Feb 2026
- Primary completion30 Jun 2028 (estimated)
- Study completion30 Sep 2028 (estimated)
Resources
- Official record on ClinicalTrials.govThe registry entry this page is built from, with its full change history.
- Other trials for Complex Abdominal Aortic AneurysmsStudies of the same condition, open and under way.
- What the trial phases meanHow a treatment moves from first tests in people to approval.
More details
For doctors and researchers: how the study is designed, who runs it, and its registry IDs.
Study design
- Study type
- Interventional
- Phase
- Phase N/A
- Primary purpose
- Treatment
- Allocation
- Not applicable
- Intervention model
- Single group
- Masking
- None (open label)
- Enrolment
- 165 participants (estimated)
Sponsor and investigators
- Lead sponsor
- iVascular S.L.U.
- Sponsor type
- Industry (pharma, biotech, devices)
- Responsible party
- Sponsor
IDs, dates and publications
- ClinicalTrials.gov ID
- NCT07529275
- Sponsor's protocol ID
- FenCo-01
- Acronym
- fenCo
- Status last verified
- Apr 2026
- First submitted
- 26 Feb 2026
- Last updated
- 14 Apr 2026
- Results posted
- No
- Data monitoring committee
- Yes
- FDA-regulated drug
- No
- FDA-regulated device
- No
Study at a glance
From the registry record, last updated by the sponsor on 14 Apr 2026. Recruitment status is as the sponsor reported it: confirm with the study team.
- Official title
- Clinical Performance and Safety Assessment of the Use of iCover Balloon Expandable Covered Stent (iVascular) Implanted as Bridging Stent in FEVAR (Fenestrated Endovascular Aortic Repair) for the Treatment of Complex Abdominal Aortic Aneurysms
- ClinicalTrials.gov ID
- NCT07529275
- Acronym / sponsor ID
- fenCo / FenCo-01
- Conditions
- Complex Abdominal Aortic Aneurysms
- Keywords
- FEVAR, bridging stent, complex abdominal aortic aneurysms
- Study type
- Interventional
- Recruitment status
- Recruiting
- Enrolment
- 165 participants (estimated)
- Lead sponsor
- iVascular S.L.U.Industry (pharma, biotech, devices)
- Study start
- 16 Feb 2026 (actual)
- Primary completion
- 30 Jun 2028 (estimated)
- Study completion
- 30 Sep 2028 (estimated)
- First posted
- 14 Apr 2026 (actual)
- Results
- Not posted in the source record
Clinical summary
The research question as the sponsor registered it. This describes what the study plans to test, not what it found.
fenCo is a European multicentre, prospective study to evaluate the use of the iCover covered stents as bridging stents for reno-visceral target vessel during fEVAR for the treatment of complex abdominal aortic aneurysms.
Detailed description
In terms of complex endovascular repair, fenestrated stent grafts should be considered the preferred treatment option when feasible. The main advantage of Fenestrated Endovascular Aneurysm Repair (fEVAR) lies in the avoidance of aortic cross clamping and subsequent lower risk of renal dysfunction, less surgical trauma and faster recovery, which may be advantageous for patients at high risk for open surgery. fenCo study aims to demonstrate the safety and efficacy of iCover when used as bridging stent in fEVAR procedures for the treatment of complex abdominal aortic aneurysms.
Study design
- Primary purpose
- Treatment
- Allocation
- Not applicable
- Intervention model
- Single group
- Masking
- None (open label)
- Number of arms
- 1
| Arm | Type | Receives |
|---|---|---|
| iCover balloon-expandable covered peripheral stent system | Experimental | Device: iCover balloon-expandable covered peripheral stent system as bridging stentiCover balloon-expandable covered peripheral stent system |
Who may qualify
- Ages
- 55 years and older
- Sex
- All
- Healthy volunteers
- Not accepted
- Age groups
- Adult, Older adult
Key inclusion criteria (15 listed)
- Patient suitable for treatment of a juxta, para-renal, suprarenal or thoracoabdominal aneurysm with a fenestrated endoprosthesis in accordance with current indications and guidelines for the management of abdominal aortic aneurysms (aneurysm > 5 centimetres for women and > 5.5 centimetres for men and/or growth > 5 mm in 6 months or 1 centimetre in 1 year).
- The fenestrated endograft must be designed in a way that positions the fenestrations directly in front of the orifice of the target vessel, ensuring that the distance between the fenestration and the orifice of the target vessel does not exceed 5 mm.
- Anatomy and patient strictly in accordance with the Instructions For Use (IFU) of the fenestrated endograft chosen to be implanted:
- Additional iCover stent, if necessary, could be placed in the same target artery (2 iCovers stents max as bridging stent for each fenestration).
- Landing zone in the target vessel of at least 10 mm.
Key exclusion criteria (21 listed)
- Patients contraindicated for anti-platelet therapy.
- Patients with uncontrolled haematological disorders or heparin-induced thrombocytopenia.
- Chronic or acute aortic dissection.
- Patients refusing treatment.
- Patients who are pregnant or wish to become pregnant.
Eligibility is decided by the study team after screening; this overview does not establish that a patient qualifies. Full inclusion and exclusion criteria.
Interventions
| Intervention | Type | Arms |
|---|---|---|
| iCover balloon-expandable covered peripheral stent system as bridging stentThis device will be used as bridging stent during FEVAR (fenestrated endovascular aortic repair) procedures. | Device | iCover balloon-expandable covered peripheral stent system |
Primary outcome
Bridging stent patency at 12-month visit
Time frame: 12 months visit
Defined as rate of stents with an absence of binary restenosis (≥ 50% stenosis) based on CT Angio at 12 months visit (+/-2 months).
Absence of target vessel instability between implantation and 12-month visit.
Time frame: starting at implant day through the study completion, with an average of 12Months up to 14Months per window.
Defined as rate of patients with an absence of target vessel instability described as a composite endpoint of any branch-related complication leading to: aneurysm rupture/ death/ occlusion/ component separation/ or reintervention to maintain branch patency or to treat a branch-related component separation or endoleak between implantation and 12 months visit (+/-2 months).
15 secondary outcomes
Technical success: Successful introduction and deployment of the iCover stent implanted as bridging stent in FEVAR procedure.
Time frame: During the index procedure (from stent introduction to final deployment, up to procedure completion)
Bridging stent patency at the first postoperative imaging, at 6 and at 12 months visits.
Time frame: at discharge visit (up to 3 months, average of 7days) and at 6 and 12 months visits ( can be up to 8 and 14months from implant day)
Absence of endoleaks linked to iCover covered stent defects at the first postoperative imaging, at 6 and 12 months visits.
Time frame: discharge visit , 6 and 12 months visits.
Absence of re-intervention due to a technical defect in iCover covered stents at discharge, at 6 months visit, and at 12 months visit.
Time frame: between surgery and post operative to discharge period, between surgery and 6 months visit, between surgery and 12 months visit.
Absence of type I or III endoleaks post-procedure at the first postoperative imaging, at 6 and 12 months visits.
Time frame: discharge visit, 6 and 12 months visits.
30-day post-procedure mortality.
Time frame: within 30 days post procedure
In-hospital post-procedure mortality (all causes).
Time frame: between procedure and 12 months visit
Absence of aortic endograft migration at 6 months visit and at 12 months visit.
Time frame: at 6 and 12 months visits
No increase in maximum aneurysmal sac diameter measured by CT angiography at 6 and at 12 months visits.
Time frame: at 6 and 12 months visits
Absence of post-procedural aortic surgical conversion at discharge, at 6 and 12 months visits.
Time frame: between implant and 6 and 12 months visits
Absence of aneurysm-related mortality post-procedure at discharge, at 6 months and 12 months visits.
Time frame: between implant and 6 and 12 months visits
Absence of aneurysm related secondary endovascular procedures at discharge, at 6 and 12 months visits.
Time frame: between implant 6 and 12 months visits.
Absence of major adverse events (MAEs) at discharge, at 6 and 12 months visits.
Time frame: between implant and 6, 12 months vists.
Absence of target vessel instability at 12 months visit.
Time frame: at 12 months visit
Target vessel patency at at 6 and 12 months visits
Time frame: 6 and 12 months visits
Results and safety
This study is still under way. Results are not available in the source record yet.
Research links
- Official record on ClinicalTrials.govLast updated by the sponsor 14 Apr 2026.
Related trials for Complex Abdominal Aortic Aneurysms
Not yet recruitingObservational study
Real-World Data Collection of GORE Devices Used In the Endovascular Treatment of Aortic Aneurysms Which Incorporate Visceral Arteries.
Complex Aortic Aneurysms, Thoracoabdominal Aneurysm
Recruiting
Clinical Trial of BAO-G Technique in Complex Aortic Aneurysm Endovascular Repair (BAO-G-02)
Thoracoabdominal Aortic Aneurysm, Endovascular Repair
RecruitingObservational study
Patient-Reported Quality of Life After Open Surgery for Complex Abdominal Aortic Aneurysms: A Prospective Study
Complex Abdominal Aortic Aneurysms
Recruiting
CAAAs, TAAAs, Aortic Arch Aneurysms or Dissections With Fenestrated/Branched Stent Graft
Complex Abdominal Aortic Aneurysm, Thoracoabdominal Aortic Aneurysm
RecruitingObservational study
European Aortic Data Collection Project
Complex Abdominal Aortic Aneurysm, Thoracoabdominal Aortic Aneurysm
Referral and contact
AzTrials does not enrol participants or take referrals. Contact the study team or a site listed below; they decide who can take part.
24 sites in 5 countries: see every site and its status.
Questions about this trial
What is this clinical trial studying?
A study testing iCover balloon-expandable covered peripheral stent system as bridging stent in people with Complex Abdominal Aortic Aneurysms, run by iVascular S.L.U. at 24 sites in 5 countries.
The full registered description is in the About this study section and the official record.
Who may be eligible to take part?
The sponsor lists 15 inclusion and 21 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.
Is this clinical trial currently recruiting?
The record says: looking for participants now, as of the sponsor's last update on 14 Apr 2026. Status can change between updates, so confirm with the study contact before making plans.
Where is the study taking place?
The record lists 24 sites in France, Belgium, Germany, Netherlands, Spain. Whether a site is taking part right now should be confirmed with that site. See every site.
What are the possible benefits and risks?
They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.
Are results available for this study?
This study is still under way. Results are not available in the source record yet.
How can a doctor refer a patient or contact the research team?
Each site lists its own contact: choose the nearest one under Trial locations. AzTrials does not enrol participants or take referrals.
What does the primary outcome measure?
Bridging stent patency at 12-month visit (time frame: 12 months visit). The study lists 2 primary outcomes in all. It is what the study was designed to measure, not a finding.
Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?
No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.
How often is this information updated?
AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 14 Apr 2026.