Hysteroscopic SurgeryNCT04540952
Comparison of Fluid Capture Efficiency of the Total Capture Drape Versus a Standard Drape for Hysteroscopy
A study testing Total Capture Drape and Standard Hysteroscopy Drape in people with Hysteroscopic Surgery, run by Wake Forest University Health Sciences at 1 site in United States.
Trial summary
Safety during operative hysteroscopy requires a fluid management system to assist in gauging patient fluid absorption of media used for visualizing the uterine cavity. Serious patient complications, including hyponatremia (low serum sodium), heart failure, and pulmonary and cerebral edema, can result from over… Read more about this study
- Who is it for?
- Ages 18 years and older, women onlyHealthy volunteers can take part
- Conditions the study is for
- Hysteroscopic Surgery
- What is being tested?
- Total Capture Drape and Standard Hysteroscopy Drape
- Type of treatment
- Medical device
- Could I receive a placebo?
- No, everyone receives an active treatment
- How many people
- 68 participants
- Where
- 1 siteUnited States
- Study dates
- 9 Jan 2019to 10 Mar 2020
- Phase
- Interventional
- Are there results?
- Yes, posted in the registrySummary in the doctors' view
Key requirements
Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.
- Ages 18 years and older
- Women only
- Healthy volunteers accepted
- Adult, Older adult
Who may be able to join
- Women who are scheduled for a hysteroscopy that require a fluid management system at an Atrium Health operating room facility.
Who cannot join
- Women will be excluded if they are not acceptable candidates for a hysteroscopy procedure for any reason.
Trial locations and contacts
1 site in 1 country.
About this study
Official title: A Prospective, Randomized Trial to Compare the Fluid Capture Efficiency of the "Total Capture" Drape Versus a Standard Drape for Hysteroscopy: Innovation for Improved Patient Safety and Surgical Care.
Safety during operative hysteroscopy requires a fluid management system to assist in gauging patient fluid absorption of media used for visualizing the uterine cavity. Serious patient complications, including hyponatremia (low serum sodium), heart failure, and pulmonary and cerebral edema, can result from over absorption of this distending medium. Failure of surgical drapes to collect unabsorbed fluid causes inaccurate determination of the hysteroscopic fluid deficit (HD) thus preventing proper risk evaluation for patient fluid overload, could prompt premature procedure terminatio, and result in hazards for the OR team.
Specific Aims:
Read more
- To perform a comparative trial of the "Total Capture" hysteroscopy drape (TCD) versus the standard drape during patient surgeries to document improved, real-time determination of patient fluid absorption. Pilot testing of a prototype design of the "Total Capture drape" versus the conventional hysteroscopy drape indicated remarkable improvement in fluid capture and accurate fluid deficit determination in a plastic pelvic model experiment.
- To evaluate the clinical usefulness of the TCD compared to the Standard drape for hysteroscopy with the standardized metrics of: 1) the Technology Acceptance Model, and 2) The System Usability Scale. These metrics will allow us to quantitate clinical usefulness and usability of both the operating surgeons and operating room staff.
The investigators hypothesize the "Total Capture" Drape design provides an accurate recorded hysteroscopic fluid deficit in real-time so that patient safety and surgical parameters are optimized. The simple design of the TCD would add a very significant improvement for patient safety and care with minimal cost for the estimated 200,000 hysteroscopies performed per year in the U.S.
Read the detailed description
EXPERIMENTAL TRIAL WITH THE TCD VERSUS THE STANDARD HYSTEROSCOPY DRAPE IN A PELVIC MODEL SIMULATION
The investigators sought to document whether or not a prototype design of the "Total Capture" drape (TCD) would perform under replicated clinical conditions in a superior fashion compared to the currently marketed hysteroscopy drape. The experimental testing was performed at an operating room suite using the hysteroscopy fluid management system, conventional hysteroscopy drape, the Total Capture drape and the ZOE Gynecologic pelvic model. The pelvic model was configured so that all fluid introduced into the artificial vagina would not be absorbed by the model but would flow out of the artificial vagina.
A total of four trials per drape were conducted. One trial from each product had procedural difficulties making results invalid.
Results of the Simulation Trial: In four trials of the total capture drape, there was no deficit observed. It was concluded that no major design iterations would be required to move the TCD prototype to the next phase of development
COMPARATIVE TRIAL FOR "PROOF OF CONCEPT" TESTING TO EVALUATE THE EFFICIENCY OF THE TOTAL CAPTURE DRAPE COMPARED TO THE CONVENTIONAL HYSTEROSCOPY DRAPE DURING HYSTEROSCOPIC SURGERY
All women undertaking hysteroscopy are required to have a sterile surgical drape with a fluid capture pouch system to collect the distention media not absorbed by the patient to calculate fluid deficits. Both the TCD and the standard hysteroscopy drape possess the characteristics described above, although design features of the standard drape are deficient. The TCD falls under Class II status, exempt from the 510(k) clearance process in consideration of FDA's approach to similar devices like the standard hysteroscopy drape (Regulation number 21 CRF 878.4370, product code KKX.
The hysteroscopy fluid deficit will be measured by the fluid management system during each hysteroscopy procedure. The uncollected fluid volume seen on the operating room floor and absorbed by the operation blankets and drapes will also be captured, as well as the estimated fluid loss from leaking hysteroscopy joints and gaskets not captured in the collection bag. Procedure time and alarm sounding for exceeding fluid deficit limit will also be recorded.
MEASURES OF DRAPE USABILITY, QUALITY, AND CONVENIENCE
Staff will be asked to complete the System Usability Scale (SUS) to evaluate their perceived clinical usefulness of the TCD and the standard drape. The SUS is a validated 10-item survey for assessing perceived usability that uses a 5-point Likert scale ranging from strongly disagree to strongly agree. To further subjectively and objectively assess the TCD, a human factors and design researcher will observe the surgeries. The human factors researcher will attend the first surgeries to ensure consistent rating with the design researcher who will subsequently observe the remaining surgeries until data saturation is reached. Staff will be asked to speak aloud using the system, which the researcher will record. Further, the researcher will complete an objective human factors assessment tool. These metrics will allow us to quantitate clinical usefulness and usability of both the operating surgeons and operating room staff. These results will not only provide this assessment of the TCD compared to the standard hysteroscopy drape (which is used across ORs within Atrium Health) but will also provide important information to improve the drape if deficiencies are identified.
DATA MANAGEMENT
Case report forms will be developed by the OBGYN clinical research nursing staff in collaboration with Dr. Marshburn and Dr. Zhao. A REDCap database will be created for secure electronic data capture. The data flow will consist of paper data collection for eligibility assessment, baseline data, randomization, operational procedure, fluid measures, surgery outcomes, adverse events, and protocol deviations. The data will be entered into the electronic database within 3 business days of the enrollment.
STATISTICAL METHODS AND SAMPLE SIZE JUSTIFICATION
The primary outcome for the study is the hysteroscopy fluid deficit (HFD) as determined by the fluid management system by subtracting the outflow volume (OV) from the inflow volume (IV) (HFD=IV-OV). Our null hypothesis is Ho: μStandard=μTCD where μ=mean fluid deficit. Our alternative hypothesis is Ho: μStandard>μTCD due to the underestimation of the outflow volume created by the loss of outflow fluid to floors and blankets. If it is assumed the true absorption rates by the patients are the same between the groups due to randomization, by estimating and testing the difference HFD between the two groups a more accurate measure of the uncollected fluid volume (UFV) will be estimated and tested as shown below:
μStandard - μTCD = (μStandard, true absorption + μStandard, UFV) - μTCD, true absorption
- (μ true absorption + μStandard, UFV) - μtrue absorption
- μStandard, UFV The primary analysis for the study will use a general linear model on the HFD with a fixed effect for study group (TCD versus Standard drape) controlling for surgeon and indication for hysteroscopy, the stratification factors for the randomization.
Secondary outcomes include estimated uncollected fluid volume from floor and blankets, occurrence of the fluid deficit alarm sounding, stopping of the operation due to concerns of overabsorption by the patient, and adverse events. Estimated uncollected fluid volume will be analyzed similar to the primary outcome. All other outcomes are categorical and will be compared using Chi-square analyses.
The investigators will sample 10 procedures(or until data saturation) from each arm (TCD vs standard) to conduct qualitative assessments on Usability using the Human Factors Assessment Tool /Acceptance for the for the surgeons and operating team members.For the SUS, the design strategist will collect data on attending physicians only on their first and fifth procedures (the study team may change it to 1st and 3rd if deemed appropriate). The investigators will estimate means or medians with standard deviations (or interquartile ranges) for the items and total scores for the measuresSUS. Both within (1st vs. 5th) and between arms (TCD vs. standard) comparison on total scores of SUS will be conducted. The investigators will compare the two groups using linear mixed models or generalized linear mixed models (depending on the distribution) with a random effect for the procedure to control for correlation of the measures among team members participating in the same procedure.
There is no prior data on HFD in patients to estimate the variability need for the sample size estimation. There is preliminary data from the simulated experiment showing a 100% recovering of loss deficit by the TCD. If it is conservatively assumed the loss deficit will decrease from 40% to 20% of the inflow volume, this would result in a clinically meaningful improvement with an approximate effect size of 0.92 based on the variability of loss deficit in the simulation. Given this, effect size (relative to standard deviation) is used as the primary justification for the number of patients needed for the study. With a sample size of n=34 per arm, there would be 81% power to detect an effect size of 0.7 (0.7*standard deviation of the HFD) with a two sided, two sample t-test (α=0.05). To account for potential missing data of 10%, it is proposed randomizing 76 women (n=38 per arm). With n=34 women in each group, there would also be 80% power to detect an absolute 33% reduction in the proportion of procedures with the alarm sounding due to fluid deficit threshold being reached assuming the proportion in the standard drape arm is 50%. This power was calculated using a Fisher's Exact test for testing proportions between two independent samples. The same power would apply for detecting differences in the proportion of procedures terminated.
DATA SAFETY AND MONITORING
The PI and the study team will meet biweekly to review the study procedures, enrollment (screening and randomization), implementation, protocol deviations, data collection, and adverse events. The biostatistical team will generate monthly reports to monitor screening, enrollment, completeness of data for intervention implementation and outcomes, adverse events, and protocol deviations.
- Conditions
- Hysteroscopic Surgery
- Being tested
- Total Capture Drape, Standard Hysteroscopy Drape
- Keywords
- hysteroscopy, fluid management, safety, surgical complications, surgical drape
What participants receive
What participants receive. In a randomised study each person is assigned to one group by chance.
Total Capture DrapeDevice
Surgical drape created by Principal Investigator to adequately collect fluid during hysteroscopy procedure
Standard Hysteroscopy DrapeDevice
Standard surgical drape used to adequately collect fluid during hysteroscopy procedure
| Study group | Role | What they receive |
|---|---|---|
| Total Capture Drape | Experimental | Surgical drape created by Principal Investigator to adequately collect fluid during hysteroscopy procedure Receives: Device: Total Capture Drape |
| Control | Active comparator | Standard surgical drape used to adequately collect fluid during hysteroscopy procedure Receives: Device: Standard Hysteroscopy Drape |
What is being measured
The primary outcome is the main question the study is designed to answer.
Hysteroscopic Fluid Deficit (HFD) Measured by the Fluid Management System by Subtracting the Outflow Volume (OV) Returned to the Fluid Management System From the Inflow Volume (IV).
Surgery duration, and average of 30 minutes was used to measure hysteroscopic fluid deficit measured by the fulid management system by subtracting the outflow volume returned to the fluid management system from the inflow volume.
Measured: Surgery Duration
Secondary outcomes (2)
Number of Participants With Uncollected Distention Fluid Volume Found on Operating Room Floor and Bedding
Uncollected fluid on operating room floor was documented as present or absent.
Measured: Surgery Duration
Number of Participants With Hysteroscopy Fluid Management Alarm Activation
Observed if fluid management alarm indicated unsafe level of patient fluid absorption
Measured: 1 day (day of procedure)
Timeline
From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.
- First posted7 Sep 2020
- Study start9 Jan 2019
- Primary completion10 Mar 2020
- Study completion10 Mar 2020
Why it stopped: COVID-19
Resources
- Official record on ClinicalTrials.govThe registry entry this page is built from, with its full change history.
- Posted resultsParticipant flow, outcome measures and adverse events reported by the sponsor.
- Other trials for Hysteroscopic SurgeryStudies of the same condition, open and under way.
- More trials run by Wake Forest University Health SciencesThe sponsor's other registered studies.
- What the trial phases meanHow a treatment moves from first tests in people to approval.
More details
For doctors and researchers: how the study is designed, who runs it, and its registry IDs.
Study design
- Study type
- Interventional
- Phase
- Phase N/A
- Primary purpose
- Other
- Allocation
- Randomised
- Intervention model
- Parallel groups
- Model description
- Prospective Randomized Early Clinical Trial
- Masking
- Quadruple blind
- Who is masked
- Participant, Care provider, Investigator, Outcomes assessor
- Enrolment
- 68 participants (actual)
Sponsor and investigators
- Lead sponsor
- Wake Forest University Health Sciences
- Sponsor type
- University, hospital or other
- Collaborators
- North Carolina Biotechnology Center
- Responsible party
- Sponsor
- Investigators
- Paul B. Marshburn, MD, Principal investigator, Wake Forest University Health Sciences
IDs, dates and publications
- ClinicalTrials.gov ID
- NCT04540952
- Sponsor's protocol ID
- 04-16-07B
- Status last verified
- Nov 2021
- First submitted
- 29 Jun 2020
- Last updated
- 21 Apr 2022
- Results posted
- Yes, on ClinicalTrials.gov
- Data monitoring committee
- No
- FDA-regulated drug
- No
- FDA-regulated device
- Yes
- Individual data sharing
- No: Patients personal health information are confidential. All patient data are de-identified for analysis and reporting
Study at a glance
From the registry record, last updated by the sponsor on 21 Apr 2022. Recruitment status is as the sponsor reported it: confirm with the study team.
- Official title
- A Prospective, Randomized Trial to Compare the Fluid Capture Efficiency of the "Total Capture" Drape Versus a Standard Drape for Hysteroscopy: Innovation for Improved Patient Safety and Surgical Care.
- ClinicalTrials.gov ID
- NCT04540952
- Acronym / sponsor ID
- 04-16-07B
- Conditions
- Hysteroscopic Surgery
- Keywords
- hysteroscopy, fluid management, safety, surgical complications, surgical drape
- Study type
- Interventional
- Recruitment status
- TerminatedReason stopped: COVID-19
- Enrolment
- 68 participants (actual)
- Lead sponsor
- Wake Forest University Health SciencesUniversity, hospital or other
- Collaborators
- North Carolina Biotechnology Center
- Study start
- 9 Jan 2019 (actual)
- Primary completion
- 10 Mar 2020 (actual)
- Study completion
- 10 Mar 2020 (actual)
- First posted
- 7 Sep 2020 (actual)
- Results
- Posted 30 Nov 2021
Clinical summary
The research question as the sponsor registered it. This describes what the study plans to test, not what it found.
Safety during operative hysteroscopy requires a fluid management system to assist in gauging patient fluid absorption of media used for visualizing the uterine cavity. Serious patient complications, including hyponatremia (low serum sodium), heart failure, and pulmonary and cerebral edema, can result from over absorption of this distending medium. Failure of surgical drapes to collect unabsorbed fluid causes inaccurate determination of the hysteroscopic fluid deficit (HD) thus preventing proper risk evaluation for patient fluid overload, could prompt premature procedure terminatio, and result in hazards for the OR team.
Specific Aims:
- To perform a comparative trial of the "Total Capture" hysteroscopy drape (TCD) versus the standard drape during patient surgeries to document improved, real-time determination of patient fluid absorption. Pilot testing of a prototype design of the "Total Capture drape" versus the conventional hysteroscopy drape indicated remarkable improvement in fluid capture and accurate fluid deficit determination in a plastic pelvic model experiment.
- To evaluate the clinical usefulness of the TCD compared to the Standard drape for hysteroscopy with the standardized metrics of: 1) the Technology Acceptance Model, and 2) The System Usability Scale. These metrics will allow us to quantitate clinical usefulness and usability of both the operating surgeons and operating room staff.
The investigators hypothesize the "Total Capture" Drape design provides an accurate recorded hysteroscopic fluid deficit in real-time so that patient safety and surgical parameters are optimized. The simple design of the TCD would add a very significant improvement for patient safety and care with minimal cost for the estimated 200,000 hysteroscopies performed per year in the U.S.
Detailed description
EXPERIMENTAL TRIAL WITH THE TCD VERSUS THE STANDARD HYSTEROSCOPY DRAPE IN A PELVIC MODEL SIMULATION
The investigators sought to document whether or not a prototype design of the "Total Capture" drape (TCD) would perform under replicated clinical conditions in a superior fashion compared to the currently marketed hysteroscopy drape. The experimental testing was performed at an operating room suite using the hysteroscopy fluid management system, conventional hysteroscopy drape, the Total Capture drape and the ZOE Gynecologic pelvic model. The pelvic model was configured so that all fluid introduced into the artificial vagina would not be absorbed by the model but would flow out of the artificial vagina.
A total of four trials per drape were conducted. One trial from each product had procedural difficulties making results invalid.
Results of the Simulation Trial: In four trials of the total capture drape, there was no deficit observed. It was concluded that no major design iterations would be required to move the TCD prototype to the next phase of development
COMPARATIVE TRIAL FOR "PROOF OF CONCEPT" TESTING TO EVALUATE THE EFFICIENCY OF THE TOTAL CAPTURE DRAPE COMPARED TO THE CONVENTIONAL HYSTEROSCOPY DRAPE DURING HYSTEROSCOPIC SURGERY
All women undertaking hysteroscopy are required to have a sterile surgical drape with a fluid capture pouch system to collect the distention media not absorbed by the patient to calculate fluid deficits. Both the TCD and the standard hysteroscopy drape possess the characteristics described above, although design features of the standard drape are deficient. The TCD falls under Class II status, exempt from the 510(k) clearance process in consideration of FDA's approach to similar devices like the standard hysteroscopy drape (Regulation number 21 CRF 878.4370, product code KKX.
The hysteroscopy fluid deficit will be measured by the fluid management system during each hysteroscopy procedure. The uncollected fluid volume seen on the operating room floor and absorbed by the operation blankets and drapes will also be captured, as well as the estimated fluid loss from leaking hysteroscopy joints and gaskets not captured in the collection bag. Procedure time and alarm sounding for exceeding fluid deficit limit will also be recorded.
MEASURES OF DRAPE USABILITY, QUALITY, AND CONVENIENCE
Staff will be asked to complete the System Usability Scale (SUS) to evaluate their perceived clinical usefulness of the TCD and the standard drape. The SUS is a validated 10-item survey for assessing perceived usability that uses a 5-point Likert scale ranging from strongly disagree to strongly agree. To further subjectively and objectively assess the TCD, a human factors and design researcher will observe the surgeries. The human factors researcher will attend the first surgeries to ensure consistent rating with the design researcher who will subsequently observe the remaining surgeries until data saturation is reached. Staff will be asked to speak aloud using the system, which the researcher will record. Further, the researcher will complete an objective human factors assessment tool. These metrics will allow us to quantitate clinical usefulness and usability of both the operating surgeons and operating room staff. These results will not only provide this assessment of the TCD compared to the standard hysteroscopy drape (which is used across ORs within Atrium Health) but will also provide important information to improve the drape if deficiencies are identified.
DATA MANAGEMENT
Case report forms will be developed by the OBGYN clinical research nursing staff in collaboration with Dr. Marshburn and Dr. Zhao. A REDCap database will be created for secure electronic data capture. The data flow will consist of paper data collection for eligibility assessment, baseline data, randomization, operational procedure, fluid measures, surgery outcomes, adverse events, and protocol deviations. The data will be entered into the electronic database within 3 business days of the enrollment.
STATISTICAL METHODS AND SAMPLE SIZE JUSTIFICATION
The primary outcome for the study is the hysteroscopy fluid deficit (HFD) as determined by the fluid management system by subtracting the outflow volume (OV) from the inflow volume (IV) (HFD=IV-OV). Our null hypothesis is Ho: μStandard=μTCD where μ=mean fluid deficit. Our alternative hypothesis is Ho: μStandard>μTCD due to the underestimation of the outflow volume created by the loss of outflow fluid to floors and blankets. If it is assumed the true absorption rates by the patients are the same between the groups due to randomization, by estimating and testing the difference HFD between the two groups a more accurate measure of the uncollected fluid volume (UFV) will be estimated and tested as shown below:
μStandard - μTCD = (μStandard, true absorption + μStandard, UFV) - μTCD, true absorption
- (μ true absorption + μStandard, UFV) - μtrue absorption
- μStandard, UFV The primary analysis for the study will use a general linear model on the HFD with a fixed effect for study group (TCD versus Standard drape) controlling for surgeon and indication for hysteroscopy, the stratification factors for the randomization.
Secondary outcomes include estimated uncollected fluid volume from floor and blankets, occurrence of the fluid deficit alarm sounding, stopping of the operation due to concerns of overabsorption by the patient, and adverse events. Estimated uncollected fluid volume will be analyzed similar to the primary outcome. All other outcomes are categorical and will be compared using Chi-square analyses.
The investigators will sample 10 procedures(or until data saturation) from each arm (TCD vs standard) to conduct qualitative assessments on Usability using the Human Factors Assessment Tool /Acceptance for the for the surgeons and operating team members.For the SUS, the design strategist will collect data on attending physicians only on their first and fifth procedures (the study team may change it to 1st and 3rd if deemed appropriate). The investigators will estimate means or medians with standard deviations (or interquartile ranges) for the items and total scores for the measuresSUS. Both within (1st vs. 5th) and between arms (TCD vs. standard) comparison on total scores of SUS will be conducted. The investigators will compare the two groups using linear mixed models or generalized linear mixed models (depending on the distribution) with a random effect for the procedure to control for correlation of the measures among team members participating in the same procedure.
There is no prior data on HFD in patients to estimate the variability need for the sample size estimation. There is preliminary data from the simulated experiment showing a 100% recovering of loss deficit by the TCD. If it is conservatively assumed the loss deficit will decrease from 40% to 20% of the inflow volume, this would result in a clinically meaningful improvement with an approximate effect size of 0.92 based on the variability of loss deficit in the simulation. Given this, effect size (relative to standard deviation) is used as the primary justification for the number of patients needed for the study. With a sample size of n=34 per arm, there would be 81% power to detect an effect size of 0.7 (0.7*standard deviation of the HFD) with a two sided, two sample t-test (α=0.05). To account for potential missing data of 10%, it is proposed randomizing 76 women (n=38 per arm). With n=34 women in each group, there would also be 80% power to detect an absolute 33% reduction in the proportion of procedures with the alarm sounding due to fluid deficit threshold being reached assuming the proportion in the standard drape arm is 50%. This power was calculated using a Fisher's Exact test for testing proportions between two independent samples. The same power would apply for detecting differences in the proportion of procedures terminated.
DATA SAFETY AND MONITORING
The PI and the study team will meet biweekly to review the study procedures, enrollment (screening and randomization), implementation, protocol deviations, data collection, and adverse events. The biostatistical team will generate monthly reports to monitor screening, enrollment, completeness of data for intervention implementation and outcomes, adverse events, and protocol deviations.
Study design
- Primary purpose
- Other
- Allocation
- Randomised
- Intervention model
- Parallel groups
- Masking
- Quadruple blind (participant, care provider, investigator, outcomes assessor)
- Number of arms
- 2
- Model description
- Prospective Randomized Early Clinical Trial
| Arm | Type | Receives |
|---|---|---|
| Total Capture Drape | Experimental | Device: Total Capture DrapeSurgical drape created by Principal Investigator to adequately collect fluid during hysteroscopy procedure |
| Control | Active comparator | Device: Standard Hysteroscopy DrapeStandard surgical drape used to adequately collect fluid during hysteroscopy procedure |
Who may qualify
- Ages
- 18 years and older
- Sex
- Female only
- Healthy volunteers
- Accepted
- Age groups
- Adult, Older adult
Key inclusion criteria (1 listed)
- Women who are scheduled for a hysteroscopy that require a fluid management system at an Atrium Health operating room facility.
Key exclusion criteria (1 listed)
- Women will be excluded if they are not acceptable candidates for a hysteroscopy procedure for any reason.
Eligibility is decided by the study team after screening; this overview does not establish that a patient qualifies. Full inclusion and exclusion criteria.
Interventions
| Intervention | Type | Arms |
|---|---|---|
| Total Capture DrapeSurgical drape created by Principal Investigator to adequately collect fluid during hysteroscopy procedure | Device | Total Capture Drape |
| Standard Hysteroscopy DrapeStandard surgical drape used to adequately collect fluid during hysteroscopy procedure | Device | Control |
Primary outcome
Hysteroscopic Fluid Deficit (HFD) Measured by the Fluid Management System by Subtracting the Outflow Volume (OV) Returned to the Fluid Management System From the Inflow Volume (IV).
Time frame: Surgery Duration
Surgery duration, and average of 30 minutes was used to measure hysteroscopic fluid deficit measured by the fulid management system by subtracting the outflow volume returned to the fluid management system from the inflow volume.
2 secondary outcomes
Number of Participants With Uncollected Distention Fluid Volume Found on Operating Room Floor and Bedding
Time frame: Surgery Duration
Number of Participants With Hysteroscopy Fluid Management Alarm Activation
Time frame: 1 day (day of procedure)
Results and safety
Loading the posted results from ClinicalTrials.gov…
Research links
- Official record on ClinicalTrials.govLast updated by the sponsor 21 Apr 2022.
- Posted resultsParticipant flow, baseline, all outcome measures and adverse events.
- Individual participant dataSharing planned: No. Patients personal health information are confidential. All patient data are de-identified for analysis and reporting
Related trials for Hysteroscopic Surgery
Not yet recruitingObservational study
Predictors of Successful Office Operative Hysteroscopy Using a Mini-Resectoscope
Retained Products of Conception, Hysteroscopy
Not yet recruiting
Yiwu Cold Knife Technique Versus Electrocision for Multiple Endometrial Polyps
Yiwu Cold Knife Technique for Multiple Endometrial Polyps
Not yet recruitingPhase 4
Remimazolam-Flumazenil Versus Sevoflurane for Hysteroscopic Day Surgery
Hysteroscopy, Ambulatory Surgery
Not yet recruitingObservational study
"Evaluating The Applicability of LASMAR Scoring in Predicting Surgical Outcomes Of Hysteroscopic Myomectomy"
LAZMAR Scoring , Hysteroscopic Myomectomy
Enrolling by invitation
Supraglottic Jet Oxygenation Ventilation During Hysteroscopic Surgery
Hysteroscopic Surgery
Referral and contact
AzTrials does not enrol participants or take referrals. Contact the study team or a site listed below; they decide who can take part. This study is not recruiting at the moment.
Paul B. Marshburn, MD principal investigator
Wake Forest University Health Sciences
1 site in 1 country: see every site and its status.
Questions about this trial
What is this clinical trial studying?
A study testing Total Capture Drape and Standard Hysteroscopy Drape in people with Hysteroscopic Surgery, run by Wake Forest University Health Sciences at 1 site in United States.
The full registered description is in the About this study section and the official record.
Who may be eligible to take part?
The sponsor lists 1 inclusion and 1 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.
Is this clinical trial currently recruiting?
The record says: stopped early, as of the sponsor's last update on 21 Apr 2022. Status can change between updates, so confirm with the study contact before making plans.
Where is the study taking place?
The record lists 1 site in United States. Whether a site is taking part right now should be confirmed with that site. See every site.
What are the possible benefits and risks?
They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.
Are results available for this study?
Yes. Results are posted in the registry (first posted 30 Nov 2021). A summary is in the For doctors view, and the full results are on ClinicalTrials.gov.
How can a doctor refer a patient or contact the research team?
The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.
What does the primary outcome measure?
Hysteroscopic Fluid Deficit (HFD) Measured by the Fluid Management System by Subtracting the Outflow Volume (OV) Returned to the Fluid Management System From the Inflow Volume (IV). (time frame: Surgery Duration). It is what the study was designed to measure, not a finding.
Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?
No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.
How often is this information updated?
AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 21 Apr 2022.