AmputationNCT04070105
No Power Bionic Lower Extremity Prosthesis
A study testing Compliant Adaptive Energy Storage and Return Foot and AllPro in people with Amputation and Amputation, Traumatic, run by University of Michigan at 1 site in United States.
Trial summary
Typically people need separate prosthetic feet for running and walking. To bridge the gap, this study will test the Compliant Adaptive Energy Storage and Return (CAESAR) foot. This foot can change from a walk mode to a run mode with the push of a button. Read more about this study
- Who is it for?
- Ages 18 years and older
- Conditions the study is for
- Amputation, Amputation, Traumatic, Amputation; Traumatic, Leg, Lower
- What is being tested?
- Compliant Adaptive Energy Storage and Return Foot and AllPro
- Type of treatment
- Medical device
- Could I receive a placebo?
- No, everyone receives an active treatment
- How many people
- 6 participants
- Where
- 1 siteUnited States
- Study dates
- 1 Sep 2020to 31 Dec 2022
- Phase
- Interventional
- Are there results?
- Yes, posted in the registrySummary in the doctors' view
Key requirements
Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.
- Ages 18 years and older
- All sexes
- No healthy volunteers
- Adult, Older adult
Who may be able to join
- At least 18 years old
- Transtibial amputation of one or both legs
- Classified as K3 or K4 on the Medicare Functional Classification Level
- Use of a well-fitting prosthetic socket and prosthetic foot for at least 6 months
- Able to walk and run without assistance
Who cannot join
- Significant injury to intact limb
- History of balance impairments or neurologic disorders
- Significant cardiovascular or respiratory disease
Trial locations and contacts
1 site in 1 country.
About this study
Typically people need separate prosthetic feet for running and walking. To bridge the gap, this study will test the Compliant Adaptive Energy Storage and Return (CAESAR) foot. This foot can change from a walk mode to a run mode with the push of a button. The investigators will test and improve this foot design mechanically, and then test this design on individuals with lower limb amputation in a lab setting. The goal of this project is to develop a passive prosthetic foot that can serve two purposes in someone's daily life: walking and running, to allow them to be more active.
Read the detailed description
Typically people need separate prosthetic feet for running and walking. To bridge the gap, this study will test the Compliant Adaptive Energy Storage and Return (CAESAR) foot. This foot can change from a walk mode to a run mode with the push of a button. The investigators will test and improve this foot design mechanically, and then test this design on individuals with lower limb amputation in a lab setting. The goal of this project is to develop a passive prosthetic foot that can serve two purposes in someone's daily life: walking and running, to allow them to be more active. Here we will test this device to determine how it affects the user's mechanics during running and walking. Participants will come to a research laboratory for three test sessions where they will run and walk in their current foot, a commercial high-activity foot, and the prototype CAESAR foot.
- Being tested
- Compliant Adaptive Energy Storage and Return Foot, AllPro
- Medical subject headings
- Amputation, Traumatic
What participants receive
What participants receive. In a randomised study each person is assigned to one group by chance.
Compliant Adaptive Energy Storage and Return FootDevice
Also called CAESAR Foot
The primary concept of the CAESAR foot is to switch from a walking foot to a running foot with the push of a button
AllProDevice
Participants will be fit with a high activity foot designed to be used for walking and running
| Study group | Role | What they receive |
|---|---|---|
| CAESAR Foot, Prescribed Feet, AllPro | Experimental | Participants will walk and run with different prostheses on different days in a laboratory. The order of testing is CAESAR Foot, Prescribed Feet, AllPro. Receives: Device: Compliant Adaptive Energy Storage and Return Foot; Device: AllPro |
| CAESAR Foot, AllPro, Prescribed Feet | Experimental | Participants will walk and run with different prostheses on different days in a laboratory. The order of testing is CAESAR Foot, AllPro, Prescribed Feet Receives: Device: Compliant Adaptive Energy Storage and Return Foot; Device: AllPro |
| Prescribed Feet, AllPro, CAESAR Foot | Experimental | Participants will walk and run with different prostheses on different days in a laboratory. The order of testing is Prescribed Feet, AllPro, CAESAR Foot Receives: Device: Compliant Adaptive Energy Storage and Return Foot; Device: AllPro |
| Prescribed Feet, CAESAR Foot, AllPro | Experimental | Participants will walk and run with different prostheses on different days in a laboratory. The order of testing is Prescribed Feet, CAESAR Foot, AllPro Receives: Device: Compliant Adaptive Energy Storage and Return Foot; Device: AllPro |
| AllPro, Prescribed Feet, CAESAR Foot | Experimental | Participants will walk and run with different prostheses on different days in a laboratory. The order of testing is AllPro, Prescribed Feet, CAESAR Foot Receives: Device: Compliant Adaptive Energy Storage and Return Foot; Device: AllPro |
| AllPro, CAESAR Foot, Prescribed Feet, | Experimental | Participants will walk and run with different prostheses on different days in a laboratory. The order of testing is AllPro, CAESAR Foot, Prescribed Feet Receives: Device: Compliant Adaptive Energy Storage and Return Foot; Device: AllPro |
What is being measured
The primary outcome is the main question the study is designed to answer.
Peak Propulsive Ground Reaction Force on the Prosthetic Side During Walking
Participants will walk over sensors that measure the force under their foot. Peak ground reaction force in the anterior direction is measured in Newtons and then divided by the individual's body weight (BW). The resultant force is measured in Body-weights.
Measured: All conditions were complete during 3 lab visits, approximately four hours long
Participants Perceived Comfort When Using Different Prosthetic Feet During Walking
Participants were asked their perceived comfort on a visual analog scale from 0 (most uncomfortable) to 100 (most comfortable)
Measured: All conditions were complete during 3 lab visits, approximately four hours long
Peak Propulsive Ground Reaction Force on the Prosthetic Side During Running
Participants will run on a treadmill with force sensors below the treadmill belt. Data will be measured in Newtons and then normalized by dividing by body weight.
Measured: All conditions were complete during 3 lab visits, approximately four hours long
Participants Perceived Comfort When Using Different Prosthetic Feet During Running
Participants were asked their perceived comfort on a visual analog scale from 0 (most uncomfortable) to 100 (most comfortable)
Measured: All conditions were complete during 3 lab visits, approximately four hours long
Secondary outcomes (1)
Net V02 (Metabolic Cost) During Running
Participants will wear a mask that measures their rate of oxygen consumption while they run for 5 minutes at a comfortable pace
Measured: All conditions were complete during 3 lab visits, approximately four hours long
Timeline
From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.
- First posted28 Aug 2019
- Study start1 Sep 2020
- Primary completion31 Dec 2022
- Study completion31 Dec 2022
Resources
- Official record on ClinicalTrials.govThe registry entry this page is built from, with its full change history.
- Posted resultsParticipant flow, outcome measures and adverse events reported by the sponsor.
- Other trials for AmputationStudies of the same condition, open and under way.
- More trials run by University of MichiganThe sponsor's other registered studies.
- What the trial phases meanHow a treatment moves from first tests in people to approval.
More details
For doctors and researchers: how the study is designed, who runs it, and its registry IDs.
Study design
- Study type
- Interventional
- Phase
- Phase N/A
- Primary purpose
- Basic science
- Allocation
- Randomised
- Intervention model
- Crossover
- Model description
- Participants will walk with their standard prosthesis, a commercial high activity foot, and then the new design of CAESAR foot
- Masking
- None (open label)
- Enrolment
- 6 participants (actual)
Sponsor and investigators
- Lead sponsor
- University of Michigan
- Sponsor type
- University, hospital or other
- Responsible party
- Principal investigator, Deanna H Gates, University of Michigan
IDs, dates and publications
- ClinicalTrials.gov ID
- NCT04070105
- Sponsor's protocol ID
- HUM00150407
- Status last verified
- Feb 2024
- First submitted
- 20 Aug 2019
- Last updated
- 27 Mar 2024
- Results posted
- Yes, on ClinicalTrials.gov
- Data monitoring committee
- No
- FDA-regulated drug
- No
- FDA-regulated device
- Yes
Study at a glance
From the registry record, last updated by the sponsor on 27 Mar 2024. Recruitment status is as the sponsor reported it: confirm with the study team.
- Official title
- No Power Bionic Lower Extremity Prosthesis
- ClinicalTrials.gov ID
- NCT04070105
- Acronym / sponsor ID
- HUM00150407
- Study type
- Interventional
- Recruitment status
- Completed
- Enrolment
- 6 participants (actual)
- Lead sponsor
- University of MichiganUniversity, hospital or other
- Study start
- 1 Sep 2020 (actual)
- Primary completion
- 31 Dec 2022 (actual)
- Study completion
- 31 Dec 2022 (actual)
- First posted
- 28 Aug 2019 (actual)
- Results
- Posted 27 Mar 2024
Clinical summary
The research question as the sponsor registered it. This describes what the study plans to test, not what it found.
Typically people need separate prosthetic feet for running and walking. To bridge the gap, this study will test the Compliant Adaptive Energy Storage and Return (CAESAR) foot. This foot can change from a walk mode to a run mode with the push of a button. The investigators will test and improve this foot design mechanically, and then test this design on individuals with lower limb amputation in a lab setting. The goal of this project is to develop a passive prosthetic foot that can serve two purposes in someone's daily life: walking and running, to allow them to be more active.
Detailed description
Typically people need separate prosthetic feet for running and walking. To bridge the gap, this study will test the Compliant Adaptive Energy Storage and Return (CAESAR) foot. This foot can change from a walk mode to a run mode with the push of a button. The investigators will test and improve this foot design mechanically, and then test this design on individuals with lower limb amputation in a lab setting. The goal of this project is to develop a passive prosthetic foot that can serve two purposes in someone's daily life: walking and running, to allow them to be more active. Here we will test this device to determine how it affects the user's mechanics during running and walking. Participants will come to a research laboratory for three test sessions where they will run and walk in their current foot, a commercial high-activity foot, and the prototype CAESAR foot.
Study design
- Primary purpose
- Basic science
- Allocation
- Randomised
- Intervention model
- Crossover
- Masking
- None (open label)
- Number of arms
- 6
- Model description
- Participants will walk with their standard prosthesis, a commercial high activity foot, and then the new design of CAESAR foot
| Arm | Type | Receives |
|---|---|---|
| CAESAR Foot, Prescribed Feet, AllPro | Experimental | Device: Compliant Adaptive Energy Storage and Return Foot, Device: AllProParticipants will walk and run with different prostheses on different days in a laboratory. The order of testing is CAESAR Foot, Prescribed Feet, AllPro. |
| CAESAR Foot, AllPro, Prescribed Feet | Experimental | Device: Compliant Adaptive Energy Storage and Return Foot, Device: AllProParticipants will walk and run with different prostheses on different days in a laboratory. The order of testing is CAESAR Foot, AllPro, Prescribed Feet |
| Prescribed Feet, AllPro, CAESAR Foot | Experimental | Device: Compliant Adaptive Energy Storage and Return Foot, Device: AllProParticipants will walk and run with different prostheses on different days in a laboratory. The order of testing is Prescribed Feet, AllPro, CAESAR Foot |
| Prescribed Feet, CAESAR Foot, AllPro | Experimental | Device: Compliant Adaptive Energy Storage and Return Foot, Device: AllProParticipants will walk and run with different prostheses on different days in a laboratory. The order of testing is Prescribed Feet, CAESAR Foot, AllPro |
| AllPro, Prescribed Feet, CAESAR Foot | Experimental | Device: Compliant Adaptive Energy Storage and Return Foot, Device: AllProParticipants will walk and run with different prostheses on different days in a laboratory. The order of testing is AllPro, Prescribed Feet, CAESAR Foot |
| AllPro, CAESAR Foot, Prescribed Feet, | Experimental | Device: Compliant Adaptive Energy Storage and Return Foot, Device: AllProParticipants will walk and run with different prostheses on different days in a laboratory. The order of testing is AllPro, CAESAR Foot, Prescribed Feet |
Who may qualify
- Ages
- 18 years and older
- Sex
- All
- Healthy volunteers
- Not accepted
- Age groups
- Adult, Older adult
Key inclusion criteria (5 listed)
- At least 18 years old
- Transtibial amputation of one or both legs
- Classified as K3 or K4 on the Medicare Functional Classification Level
- Use of a well-fitting prosthetic socket and prosthetic foot for at least 6 months
- Able to walk and run without assistance
Key exclusion criteria (3 listed)
- Significant injury to intact limb
- History of balance impairments or neurologic disorders
- Significant cardiovascular or respiratory disease
Eligibility is decided by the study team after screening; this overview does not establish that a patient qualifies. Full inclusion and exclusion criteria.
Interventions
| Intervention | Type | Arms |
|---|---|---|
| Compliant Adaptive Energy Storage and Return FootAlso: CAESAR FootThe primary concept of the CAESAR foot is to switch from a walking foot to a running foot with the push of a button | Device | AllPro, CAESAR Foot, Prescribed Feet,, AllPro, Prescribed Feet, CAESAR Foot, CAESAR Foot, AllPro, Prescribed Feet, CAESAR Foot, Prescribed Feet, AllPro, Prescribed Feet, AllPro, CAESAR Foot, Prescribed Feet, CAESAR Foot, AllPro |
| AllProParticipants will be fit with a high activity foot designed to be used for walking and running | Device | AllPro, CAESAR Foot, Prescribed Feet,, AllPro, Prescribed Feet, CAESAR Foot, CAESAR Foot, AllPro, Prescribed Feet, CAESAR Foot, Prescribed Feet, AllPro, Prescribed Feet, AllPro, CAESAR Foot, Prescribed Feet, CAESAR Foot, AllPro |
Primary outcome
Peak Propulsive Ground Reaction Force on the Prosthetic Side During Walking
Time frame: All conditions were complete during 3 lab visits, approximately four hours long
Participants will walk over sensors that measure the force under their foot. Peak ground reaction force in the anterior direction is measured in Newtons and then divided by the individual's body weight (BW). The resultant force is measured in Body-weights.
Participants Perceived Comfort When Using Different Prosthetic Feet During Walking
Time frame: All conditions were complete during 3 lab visits, approximately four hours long
Participants were asked their perceived comfort on a visual analog scale from 0 (most uncomfortable) to 100 (most comfortable)
Peak Propulsive Ground Reaction Force on the Prosthetic Side During Running
Time frame: All conditions were complete during 3 lab visits, approximately four hours long
Participants will run on a treadmill with force sensors below the treadmill belt. Data will be measured in Newtons and then normalized by dividing by body weight.
Participants Perceived Comfort When Using Different Prosthetic Feet During Running
Time frame: All conditions were complete during 3 lab visits, approximately four hours long
Participants were asked their perceived comfort on a visual analog scale from 0 (most uncomfortable) to 100 (most comfortable)
1 secondary outcome
Net V02 (Metabolic Cost) During Running
Time frame: All conditions were complete during 3 lab visits, approximately four hours long
Results and safety
Loading the posted results from ClinicalTrials.gov…
Research links
- Official record on ClinicalTrials.govLast updated by the sponsor 27 Mar 2024.
- Posted resultsParticipant flow, baseline, all outcome measures and adverse events.
Related trials for Amputation
Recruiting
Bio Leg: Advancing Mobility (BAM)
Amputation of Lower Limb Above Knee, Amputee
Recruiting
Prosthetic Socket Fit Testing System
Amputation
Recruiting
Impact of Footwear on Gait in Unilateral Transtibial Amputees
Transtibial Amputation, Lower Limb Amputation
Recruiting
TENS for Restoring Somatotopic Sensory Feedback in Lower Limb Amputees and Improving Phantom Limb Perception
Lower Limb Amputation Above Knee, Lower Limb Amputation Below Knee
Recruiting
Clinical Feasibility of the IMES Transradial Prosthesis
Amputation
Referral and contact
AzTrials does not enrol participants or take referrals. Contact the study team or a site listed below; they decide who can take part. This study is not recruiting at the moment.
The record lists no contact. Use the official record to reach the sponsor.
1 site in 1 country: see every site and its status.
Questions about this trial
What is this clinical trial studying?
A study testing Compliant Adaptive Energy Storage and Return Foot and AllPro in people with Amputation and Amputation, Traumatic, run by University of Michigan at 1 site in United States.
The full registered description is in the About this study section and the official record.
Who may be eligible to take part?
The sponsor lists 5 inclusion and 3 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.
Is this clinical trial currently recruiting?
The record says: finished, as of the sponsor's last update on 27 Mar 2024. Status can change between updates, so confirm with the study contact before making plans.
Where is the study taking place?
The record lists 1 site in United States. Whether a site is taking part right now should be confirmed with that site. See every site.
What are the possible benefits and risks?
They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.
Are results available for this study?
Yes. Results are posted in the registry (first posted 27 Mar 2024). A summary is in the For doctors view, and the full results are on ClinicalTrials.gov.
How can a doctor refer a patient or contact the research team?
The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.
What does the primary outcome measure?
Peak Propulsive Ground Reaction Force on the Prosthetic Side During Walking (time frame: All conditions were complete during 3 lab visits, approximately four hours long). The study lists 4 primary outcomes in all. It is what the study was designed to measure, not a finding.