GlaucomaNCT03912584
Study Comparing the P200TE and the P200TxE in Glaucoma Patients
A study testing P200TE and P200TxE in people with Glaucoma, run by Optos, PLC at 1 site in United States.
Trial summary
The study evaluates the image quality between two OCT devices. Read more about this study
- Who is it for?
- Ages 22 years and older
- Conditions the study is for
- Glaucoma
- What is being tested?
- P200TE and P200TxE
- Type of treatment
- Medical device
- Could I receive a placebo?
- No, everyone receives an active treatment
- How many people
- 10 participants
- Where
- 1 siteUnited States
- Study dates
- 29 Apr 2019to 30 Apr 2019
- Phase
- Interventional
- Are there results?
- Yes, posted in the registrySummary in the doctors' view
Key requirements
Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.
- Ages 22 years and older
- All sexes
- No healthy volunteers
- Adult, Older adult
Who may be able to join
- Male or female subjects 22 years of age or older who have full legal capacity to volunteer on the date the informed consent is signed;
- Subjects who can follow the instructions by the clinical staff at the clinical site,
- Subjects who agree to participate in the study;
- Subjects who have been diagnosed with glaucoma in the study eye as confirmed by the investigator;
Who cannot join
- Subjects unable to tolerate ophthalmic imaging;
- Subjects with ocular media not sufficiently clear to obtain acceptable OCT images;
- Subjects with any clinically significant ocular pathology except glaucoma, as determined by self-report and/or investigator assessment on the day of the study visit;
- Subjects who are not reliably tested with at least one Humphrey Field Analyzer (HFA) visual field (24-2 or 30-2, white on white) measured on the day of the study visit or within the previous year from the study visit, defined as fixation losses > 20% or false positives > 33%, or false negatives > 33%;
- Subjects with history of dementia or multiple sclerosis
Trial locations and contacts
1 site in 1 country.
United States1 site
Medical Center Ophthalmology Associates
San Antonio, Texas, 78240
About this study
Official title: Qualitative OCT Image Grading Study Comparing the P200TE and the P200TxE in Glaucoma Patients
The study evaluates the image quality between two OCT devices.
Read the detailed description
The objective of this study is to collect OCT scans to compare the qualitative B scan images produced by the P200TxE to the P200TE device in glaucoma subjects.
- Conditions
- Glaucoma
- Being tested
- P200TE, P200TxE
- Medical subject headings
- Glaucoma
What participants receive
What participants receive. In a randomised study each person is assigned to one group by chance.
P200TEDevice
The P200TE Ophthalmoscope provides widefield Scanning Laser Ophthalmoscope (SLO) fundus imaging and Optical Coherence Tomography (OCT) imaging in one device.
P200TxEDevice
The P200TxE Ophthalmoscope provides widefield Scanning Laser Ophthalmoscope (SLO) fundus imaging and Optical Coherence Tomography (OCT) imaging in one device.
| Study group | Role | What they receive |
|---|---|---|
| Optical Coherence Tomographer | Other | Not described Receives: Device: P200TE; Device: P200TxE |
What is being measured
The primary outcome is the main question the study is designed to answer.
Qualitative B-Scan Assessment
Scores will be based on a 4 point grading scale (0 to 3), with 0 corresponding to failure, 1 is poor, 2 is fair, and 3 is good.
The four point scale is defined as follows:
0 indicates the scan is a failure, and contains no clinically useful information. No relevant structures are visible in the scan.
- indicates the condition of the structure or feature being assessed is poor, and it contains very limited clinical utility.
- indicates the condition of the structure or feature being assessed is fair or average. It contains some useful clinical information and therefore provides adequate clinical utility.
- indicates the condition of the structure or feature being assessed is good. It contains useful clinical information. The clinical utility is high or good.
Measured: 1 day (start: April 29, 2019 - completed: April 30, 2019)
Secondary outcomes (1)
Number of Participants With Adverse Events
Analysis of safety will be assessed through adverse events reported by the site and documented on each subject's case report form. No AEs were recorded during this study.
Measured: 1 day
Timeline
From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.
- First posted11 Apr 2019
- Study start29 Apr 2019
- Primary completion30 Apr 2019
- Study completion30 Apr 2019
Resources
- Official record on ClinicalTrials.govThe registry entry this page is built from, with its full change history.
- Posted resultsParticipant flow, outcome measures and adverse events reported by the sponsor.
- Other trials for GlaucomaStudies of the same condition, open and under way.
- What the trial phases meanHow a treatment moves from first tests in people to approval.
More details
For doctors and researchers: how the study is designed, who runs it, and its registry IDs.
Study design
- Study type
- Interventional
- Phase
- Phase N/A
- Primary purpose
- Diagnosis
- Allocation
- Not applicable
- Intervention model
- Single group
- Masking
- None (open label)
- Enrolment
- 10 participants (actual)
Sponsor and investigators
- Lead sponsor
- Optos, PLC
- Sponsor type
- Industry (pharma, biotech, devices)
- Responsible party
- Sponsor
IDs, dates and publications
- ClinicalTrials.gov ID
- NCT03912584
- Sponsor's protocol ID
- OPT1038
- Status last verified
- Sep 2022
- First submitted
- 9 Apr 2019
- Last updated
- 23 Sep 2022
- Results posted
- Yes, on ClinicalTrials.gov
- FDA-regulated drug
- No
- FDA-regulated device
- Yes
Study at a glance
From the registry record, last updated by the sponsor on 23 Sep 2022. Recruitment status is as the sponsor reported it: confirm with the study team.
- Official title
- Qualitative OCT Image Grading Study Comparing the P200TE and the P200TxE in Glaucoma Patients
- ClinicalTrials.gov ID
- NCT03912584
- Acronym / sponsor ID
- OPT1038
- Conditions
- Glaucoma
- Study type
- Interventional
- Recruitment status
- Completed
- Enrolment
- 10 participants (actual)
- Lead sponsor
- Optos, PLCIndustry (pharma, biotech, devices)
- Study start
- 29 Apr 2019 (actual)
- Primary completion
- 30 Apr 2019 (actual)
- Study completion
- 30 Apr 2019 (actual)
- First posted
- 11 Apr 2019 (actual)
- Results
- Posted 23 Sep 2022
Clinical summary
The research question as the sponsor registered it. This describes what the study plans to test, not what it found.
The study evaluates the image quality between two OCT devices.
Detailed description
The objective of this study is to collect OCT scans to compare the qualitative B scan images produced by the P200TxE to the P200TE device in glaucoma subjects.
Study design
- Primary purpose
- Diagnosis
- Allocation
- Not applicable
- Intervention model
- Single group
- Masking
- None (open label)
- Number of arms
- 1
| Arm | Type | Receives |
|---|---|---|
| Optical Coherence Tomographer | Other | Device: P200TE, Device: P200TxE |
Who may qualify
- Ages
- 22 years and older
- Sex
- All
- Healthy volunteers
- Not accepted
- Age groups
- Adult, Older adult
Key inclusion criteria (4 listed)
- Male or female subjects 22 years of age or older who have full legal capacity to volunteer on the date the informed consent is signed;
- Subjects who can follow the instructions by the clinical staff at the clinical site,
- Subjects who agree to participate in the study;
- Subjects who have been diagnosed with glaucoma in the study eye as confirmed by the investigator;
Key exclusion criteria (5 listed)
- Subjects unable to tolerate ophthalmic imaging;
- Subjects with ocular media not sufficiently clear to obtain acceptable OCT images;
- Subjects with any clinically significant ocular pathology except glaucoma, as determined by self-report and/or investigator assessment on the day of the study visit;
- Subjects who are not reliably tested with at least one Humphrey Field Analyzer (HFA) visual field (24-2 or 30-2, white on white) measured on the day of the study visit or within the previous year from the study visit, defined as fixation losses > 20% or false positives > 33%, or false negatives > 33%;
- Subjects with history of dementia or multiple sclerosis
Eligibility is decided by the study team after screening; this overview does not establish that a patient qualifies. Full inclusion and exclusion criteria.
Interventions
| Intervention | Type | Arms |
|---|---|---|
| P200TEThe P200TE Ophthalmoscope provides widefield Scanning Laser Ophthalmoscope (SLO) fundus imaging and Optical Coherence Tomography (OCT) imaging in one device. | Device | Optical Coherence Tomographer |
| P200TxEThe P200TxE Ophthalmoscope provides widefield Scanning Laser Ophthalmoscope (SLO) fundus imaging and Optical Coherence Tomography (OCT) imaging in one device. | Device | Optical Coherence Tomographer |
Primary outcome
Qualitative B-Scan Assessment
Time frame: 1 day (start: April 29, 2019 - completed: April 30, 2019)
Scores will be based on a 4 point grading scale (0 to 3), with 0 corresponding to failure, 1 is poor, 2 is fair, and 3 is good. The four point scale is defined as follows: 0 indicates the scan is a failure, and contains no clinically useful information. No relevant structures are visible in the scan. 1. indicates the condition of the structure or feature being assessed is poor, and it contains very limited clinical utility. 2. indicates the condition of the structure or feature being assessed is fair or average. It contains some useful clinical information and therefore provides adequate clinical utility. 3. indicates the condition of the structure or feature being assessed is good. It contains useful clinical information. The clinical utility is high or good.
1 secondary outcome
Number of Participants With Adverse Events
Time frame: 1 day
Results and safety
Loading the posted results from ClinicalTrials.gov…
Research links
- Official record on ClinicalTrials.govLast updated by the sponsor 23 Sep 2022.
- Posted resultsParticipant flow, baseline, all outcome measures and adverse events.
Related trials for Glaucoma
Recruiting
Implementing AI-based Glaucoma Screening Within Federally Qualified Health Centers
Glaucoma, Ocular Hypertension
Not yet recruiting
AI-Assisted vs Conventional Teleglaucoma Referral
Glaucoma
Recruiting
Brain Stimulation Effects on Orientation and Mobility Skills in Adults With Vision Impairment
Retinitis Pigmentosa (RP), Rod Cone Dystrophy
Recruiting
Automated Applanation Tonometry - Updated
Glaucoma, Glaucoma, Suspect
RecruitingPhase 2
A Phase 2a Study Comparing the Efficacy, Safety and Tolerability of Pregabalin Eye Drops Comparing With Timolol Eye Drops in Adults With Glaucoma or High Eye Pressure.
Primary Open - Angle Glaucoma (POAG), Ocular Hypertension (OHT)
Referral and contact
AzTrials does not enrol participants or take referrals. Contact the study team or a site listed below; they decide who can take part. This study is not recruiting at the moment.
The record lists no contact. Use the official record to reach the sponsor.
1 site in 1 country: see every site and its status.
Questions about this trial
What is this clinical trial studying?
A study testing P200TE and P200TxE in people with Glaucoma, run by Optos, PLC at 1 site in United States.
The full registered description is in the About this study section and the official record.
Who may be eligible to take part?
The sponsor lists 4 inclusion and 5 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.
Is this clinical trial currently recruiting?
The record says: finished, as of the sponsor's last update on 23 Sep 2022. Status can change between updates, so confirm with the study contact before making plans.
Where is the study taking place?
The record lists 1 site in United States. Whether a site is taking part right now should be confirmed with that site. See every site.
What are the possible benefits and risks?
They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.
Are results available for this study?
Yes. Results are posted in the registry (first posted 23 Sep 2022). A summary is in the For doctors view, and the full results are on ClinicalTrials.gov.
How can a doctor refer a patient or contact the research team?
The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.
What does the primary outcome measure?
Qualitative B-Scan Assessment (time frame: 1 day (start: April 29, 2019 - completed: April 30, 2019)). It is what the study was designed to measure, not a finding.