GlaucomaNCT07298356

Automated Applanation Tonometry - Updated

A study testing Standard GAT, Fixed force GAT and Upright applanating prototype in people with Glaucoma and Glaucoma, Suspect, run by Duke University at 1 site in United States.

Recruiting

Looking for participants now. Contact the study team or a site below to ask about taking part.

Sites and contacts
Official record
Search all trials

Trial summary

Goldmann Applanation Tonometry (GAT) is considered the clinical gold standard for eye pressure measurements and yet it is known to be a subjective measurement with limited repeatability and limited portability. Another clinical standard for checking eye pressure is known as the pneumotonometer. Read more about this study

Who is it for?
Ages 18 years and older
Conditions the study is for
Glaucoma, Glaucoma, Suspect
What is being tested?
Standard GAT, Fixed force GAT and Upright applanating prototype
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
200 participants
Where
1 siteUnited States
Study dates
2 Mar 2026to 29 Dec 2026
Phase
Interventional
Are there results?
Not yet: the study is under way

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 years and older
  • All sexes
  • No healthy volunteers
  • Adult, Older adult

Who may be able to join

  • Presenting for a routine eye exam
  • ≥ 18 years of age
  • Able and willing to give consent

Who cannot join

  • History of corneal scarring
  • Active infection of the eye
  • Ocular surface trauma or infection

Trial locations and contacts

1 site in 1 country. Contact the study team or a site near you to ask about taking part.

United States1 site, 1 recruiting
  • Duke Eye Center

    Durham, North Carolina, 27710

    Joanne Wen, MD, 919-684-2975, joanne.wen@duke.edu

    Joanne Wen, MD

    Recruiting

Other trials in United States

About this study

Goldmann Applanation Tonometry (GAT) is considered the clinical gold standard for eye pressure measurements and yet it is known to be a subjective measurement with limited repeatability and limited portability. Another clinical standard for checking eye pressure is known as the pneumotonometer. This method is more objective but not portable. The purpose of this study is to develop new methods of measuring eye pressure that are more objective, reproducible and portable. In this study, the investigators will be comparing the eye pressure measurements using 2 investigational methods to GAT and pneuumotonometer.

Read the detailed description

This is a prospective study. Patients presenting for their scheduled eye appointments will be recruited by verbal communication. Following informed consent, intraocular pressure (IOP) measurements will be made by the following methods:

Standard Goldmann Applanation Tonometry (GAT): this is the standard method for IOP measurement in clinical practice. The eye is given topical fluorescein/anesthetic, the GAT prism contacts the eye while the observer looks through the slit lamp machine ocular using blue light illumination to visual the applanation mires. The GAT dial is adjusted until mire alignment is achieved and the IOP measurement is read off the GAT dial.

Pneumotonometer: This is what is done in routine eye care. The participant will receive a topical anesthetic (numbing agent). A probe will touch the front of the eye while the machine measures the eye pressure.

Fixed force GAT: From the patient perspective, this method will feel identical to the standard GAT. The eye is given topical fluorescein/anesthetic. The GAT dial is set at 1.8 or 2.0, a complementary metal oxide semiconductor (C-MOS) camera is connected to one of the oculars of the slit lamp machine and under blue light illumination, the GAT prism contacts the eye while the CMOS camera makes a video of the mire appearance through the ocular. The diameters of the recorded mire images are measured and the IOP is calculated based on the mire diameter. A similar method using an iPod touch camera has been previously published by this PI.

Upright applanating prototype: With this prototype, an applanating prism (custom manufactured with medical grade acrylic in an ISO-13485 certified facility) is attached to a fixed-force spring that creates a force equivalent to 1.8 or 2.0 on the GAT dial. Blue LED lights on the prototype are used to create the blue illumination similar to the blue light used in clinical practice on the slit lamp or Perkins tonometer. A C-MOS camera is aligned with the GAT prism to image the applanation mires. The eye is given topical fluorescein/anesthetic. Like fixed-force GAT, the GAT prism contacts the eye while the CMOS camera makes a video of the mire appearance. The diameters of the recorded mire images are measured and the IOP is calculated based on the mire diameter.

The order of the above 4 measurements will be randomized.

Being tested
Standard GAT, Fixed force GAT, Upright applanating prototype, Pneumotonometer
Medical subject headings
Glaucoma, Ocular Hypertension

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Standard GATDevice

    Standard Goldmann Applanation Tonometry (GAT): this is the standard method for IOP measurement in clinical practice

  • Fixed force GATDevice

    An investigational device similar to standard GAT

  • Upright applanating prototypeDevice

    An investigational device, prism used in standard and fixed-force GAT is attached to a portable device

  • PneumotonometerDevice

    This is a standard for checking eye pressure. A topical anesthetic (numbing agent). A probe will touch the front of the eye while the machine measures the eye pressure

Study groups in NCT07298356
Study groupRoleWhat they receive
Standard GATActive comparatorThis is the gold-standard method for IOP measurement in clinical practice
Receives: Device: Standard GAT
Fixed-force GATExperimentalFrom the patient perspective, this method will feel identical to the standard GAT. The eye is given topical fluorescein/anesthetic. The GAT dial is set at 1.8 or 2.0, a C-MOS camera is connected to one of the oculars of the slit lamp machine and under blue light illumination, the GAT prism contacts the eye while the CMOS camera makes a video of the mire appearance through the ocular. The diameters of the recorded mire images are measured and the IOP is calculated based on the mire diameter
Receives: Device: Fixed force GAT
Upright applanating prototypeExperimentalWith this prototype, an applanating prism (custom manufactured with medical grade acrylic in an ISO-13485 certified facility) is attached to a fixed-force spring that creates a force equivalent to 1.8 or 2.0 on the GAT dial. Blue LED lights on the prototype are used to create the blue illumination similar to the blue light used in clinical practice on the slit lamp or Perkins tonometer. A C-MOS camera is aligned with the GAT prism to image the applanation mires. The eye is given topical fluorescein/anesthetic. Like fixed-force GAT, the GAT prism contacts the eye while the CMOS camera makes a video of the mire appearance. The diameters of the recorded mire images are measured and the IOP is calculated based on the mire diameter
Receives: Device: Upright applanating prototype
PneumotonometerActive comparatorThis is another standard method for IOP measurement in clinical practice
Receives: Device: Pneumotonometer

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Intraocular pressure (IOP) as measured by standard GAT

    this is the standard method for IOP measurement in clinical practice. The eye is given topical fluorescein/anesthetic, the GAT prism contacts the eye while the observer looks through the slit lamp machine ocular using blue light illumination to visual the applanation mires. The GAT dial is adjusted until mire alignment is achieved and the IOP measurement is read off the GAT dial.

    Measured: Baseline

  2. Intraocular pressure (IOP) as measured by fixed force GAT

    From the patient perspective, this method will feel identical to the standard GAT. The eye is given topical fluorescein/anesthetic. The GAT dial is set at 1.8 or 2.0, a C-MOS camera is connected to one of the oculars of the slit lamp machine and under blue light illumination, the GAT prism contacts the eye while the CMOS camera makes a video of the mire appearance through the ocular.

    Measured: Baseline

  3. Intraocular pressure as measured by upright applanating

    With this prototype, an applanating prism (custom manufactured with medical grade acrylic in an ISO-13485 certified facility) is attached to a fixed-force spring that creates a force equivalent to 1.8 or 2.0 on the GAT dial. Blue LED lights on the prototype are used to create the blue illumination similar to the blue light used in clinical practice on the slit lamp or Perkins tonometer. A C-MOS camera is aligned with the GAT prism to image the applanation mires. The eye is given topical fluorescein/anesthetic. Like fixed-force GAT, the GAT prism contacts the eye while the CMOS camera makes a video of the mire appearance. The diameters of the recorded mire images are measured and the IOP is calculated based on the mire diameter.

    Measured: Baseline

  4. Intraocular pressure as measured by supine Pneumotonometer

    Measured: Baseline

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted23 Dec 2025
  2. Study start2 Mar 2026
  3. Primary completion29 Dec 2026 (estimated)
  4. Study completion29 Dec 2026 (estimated)

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Diagnosis
Allocation
Randomised
Intervention model
Parallel groups
Model description
In this 2 investigational methods (Fixed-force GAT and handheld tonometer) will be compared to the 2 standard methods (Standard GAT and pneumotonometer).
Masking
None (open label)
Enrolment
200 participants (estimated)
IDs, dates and publications
ClinicalTrials.gov ID
NCT07298356
Sponsor's protocol ID
Pro00106111_1
Status last verified
Dec 2025
First submitted
19 Dec 2025
Last updated
6 Mar 2026
Results posted
No
Data monitoring committee
Yes
FDA-regulated drug
No
FDA-regulated device
Yes
Individual data sharing
No: No Currently, there is no plan to share data with other researchers.

Questions about this trial

What is this clinical trial studying?

A study testing Standard GAT, Fixed force GAT and Upright applanating prototype in people with Glaucoma and Glaucoma, Suspect, run by Duke University at 1 site in United States.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 3 inclusion and 3 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: looking for participants now, as of the sponsor's last update on 6 Mar 2026. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in United States. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

This study is still under way. Results are not available in the source record yet.

How can a doctor refer a patient or contact the research team?

Contact Sarah Jones (919 681 6584, sarah.jones1@duke.edu), or a site near the patient. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Intraocular pressure (IOP) as measured by standard GAT (time frame: Baseline). The study lists 4 primary outcomes in all. It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 6 Mar 2026.

Free account

Keep searching with a free account

You’ve used your 3 free searches for this session. Log in or create a free AzTrials account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in
Automated Applanation Tonometry - Updated | AzTrials