Aneurysm of Aortic ArchNCT02264977
Early Feasibility of the Branched TAG® Device in the Treatment of Aortic Arch Aneurysms
A study testing GORE® TAG® Thoracic Branch Endoprosthesis in people with Aneurysm of Aortic Arch, run by W.L.Gore & Associates at 6 sites in United States.
Trial summary
The purpose of this study is to assess the early feasibility of the use of the GORE® TAG® Thoracic Branch Endoprosthesis to treat aneurysms involving the aortic arch Read more about this study
- Who is it for?
- Ages 18 years and older
- Conditions the study is for
- Aneurysm of Aortic Arch
- What is being tested?
- GORE® TAG® Thoracic Branch Endoprosthesis
- Type of treatment
- Medical device
- Could I receive a placebo?
- No, everyone receives an active treatment
- How many people
- 9 participants
- Where
- 6 sitesUnited States
- Study dates
- Apr 2015to Mar 2022
- Phase
- Interventional
Key requirements
Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.
- Ages 18 years and older
- All sexes
- No healthy volunteers
- Adult, Older adult
Who may be able to join
- Presence of aortic aneurysm involving the aortic arch deemed to warrant surgical repair which requires proximal graft placement in Zone 0 or Zone 1:
- Fusiform (≥ 55 mm), or
- Fusiform (>2 times native aortic diameter), or
- Saccular (no diameter criteria)
- Subject is considered a high risk candidate for conventional open surgical repair at the discretion of the Investigator
- Age ≥18 years at time of informed consent signature
Show all 22Show fewer
- Subject is capable of complying with protocol requirements, including follow-up
- Informed Consent Form (ICF) is signed by Subject or legal representative
- Must have appropriate proximal aortic landing zone, defined as:
- Must require placement of the proximal extent of the Aortic Component in Zone 0 or Zone 1 for exclusion of the lesion
- Acceptable proximal landing zone outer curvature length for the required device
- Landing zone inner diameters between 16-48 mm in Zone 1 Subjects and 24-48mm in Zone 0 Subjects
- Landing zone, which must include either the brachiocephalic or left common carotid native ostium, cannot be aneurysmal, heavily calcified, or heavily thrombosed
- Must have appropriate distal aortic landing zone, defined as:
- Outer curvature length must be ≥2cm proximal to the celiac artery
- Aortic inner diameters between 16-48 mm (diameter must be between 16-42mm if using distal TAG® Device extension)
- Landing zone cannot be aneurysmal, heavily calcified, or heavily thrombosed
- Landing zone in native aorta or previously implanted GORE® TAG® Device
- Must have appropriate target branch vessel landing zone, defined as:
- Length of ≥3 cm proximal to first major branch vessel if using Aortic Component with 8 mm portal diameter, or length of ≥2.5 cm proximal to first major branch vessel if using Aortic Component with 12 mm portal diameter (required for Zone 0 Subjects)
- Target branch vessel inner diameters of 6-15 mm if using Aortic Component with 8mm portal diameter, or inner diameters of 11-18 mm if using Aortic Component with 12mm portal diameter (required for Zone 0 Subjects)
- Target branch vessel landing zone must be in native aorta that cannot be aneurysmal, heavily calcified, or heavily thrombosed
Who cannot join
- Concomitant aneurysm/disease of the ascending aorta, or abdominal aorta requiring repair
- Previous endovascular repair of the ascending aorta
- Previous endovascular repair of the DTA with a non-Gore device
- Surgery within 30 days of treatment
- Infected aorta
- Dissection of the aorta
Show all 24Show fewer
- Intramural hematoma of the aortic arch or DTA without aneurysm
- Life expectancy <2 years
- Myocardial infarction or stroke within 6 weeks prior to treatment
- Patient has a systemic infection and may be at increased risk of endovascular graft infection
- Pregnant female at time of informed consent signature
- Degenerative connective tissue disease, e.g. Marfan's or Ehler-Danlos Syndrome
- Participation in another drug or medical device study within one year of study enrollment
- Known history of drug abuse within one year of treatment
- Significant thrombus or atheroma in the aortic arch
- Tortuous or stenotic iliac and/or femoral arteries preventing introducer sheath insertion and the inability to use a conduit for vascular access
- Planned coverage of celiac artery
- Patient has known sensitivities or allergies to the device materials
- Patient has known hypersensitivity or contraindication to anticoagulants or contrast media, which is not amenable to pre-treatment
- Previous instance of Heparin Induced Thrombocytopenia type 2 (HIT-2) or known hypersensitivity to heparin
- Diameter taper outside of the device sizing range between proximal and distal landing zones of aorta and the inability to use additional devices of different diameters to compensate for the taper
- Mycotic aneurysm
- Persistent refractory shock (systolic blood pressure <90 mm Hg)
- Patient has body habitus or other medical condition which prevents adequate visualization of the aorta
Trial locations and contacts
6 sites in 1 country.
6 sites in 1 country
United States6 sites
University of Michigan
Ann Arbor, Michigan, 48109
The Hitchcock Foundation (Dartmouth Hitchcock Medical Center)
Lebanon, New Hampshire, 03756
Hospital at University of Pennsylvania
Philadelphia, Pennsylvania, 19104
Univerisity of Pittsburgh Medical Center
Pittsburgh, Pennsylvania, 15232
Mayo Clinic Rochester
Rochester, Minnesota, 55905
Leland Stanford Junior University
Stanford, California, 94305-5407
About this study
Official title: Evaluation of the GORE® TAG® Thoracic Branch Endoprosthesis (Branched TAG® Device) in the Treatment of Aortic Arch Aneurysms
The purpose of this study is to assess the early feasibility of the use of the GORE® TAG® Thoracic Branch Endoprosthesis to treat aneurysms involving the aortic arch
Read the detailed description
The GORE® TAG® Thoracic Endoprosthesis (TAG® Device) received premarket approval (PMA) for use in endovascular aneurysm repair of the descending thoracic aorta (DTA) on 23-Mar-2005 under P040043, and design changes that resulted in the conformable GORE® TAG® Thoracic Endoprosthesis (CTAG) received premarket approval for the treatment of aneurysms of the DTA on 23-Aug-2011 (P040043/S039). The TAG® Device and CTAG are intended to exclude an aneurysm from the blood circulation in patients diagnosed with DTA aneurysms. However, endovascular treatment options for patients with aortic arch aneurysms are limited as current stent graft technology would require coverage of aortic arch vessels. This prompted the creation of the GORE® TAG® Thoracic Branch Endoprosthesis (previously known as Branched TAG® Device), which was being evaluated under IDE G130120 for Zone 2 aneurysms when this clinical investigation was initiated to evaluate the GORE® TAG® Thoracic Branch Endoprosthesis for the treatment of Zone 0 and Zone 1 aneurysms of the aortic arch.
- Conditions
- Aneurysm of Aortic Arch
- Being tested
- GORE® TAG® Thoracic Branch Endoprosthesis
- Medical subject headings
- Aneurysm, Aortic Arch
What participants receive
What participants receive. In a randomised study each person is assigned to one group by chance.
GORE® TAG® Thoracic Branch EndoprosthesisDevice
| Study group | Role | What they receive |
|---|---|---|
| Branched TAG® Device | Experimental | Treatment with the GORE® TAG® Thoracic Branch Endoprosthesis Receives: Device: GORE® TAG® Thoracic Branch Endoprosthesis |
What is being measured
The primary outcome is the main question the study is designed to answer.
Number of Participants With Successful Study Device Access
Access to the aneurysm and target landing zone location is obtained via conventional vascular access and endovascular techniques.
Measured: During treatment procedure (day 0)
Number of Participants With Successful Study Device Deployment
Absence of deployment failure will be considered a successful deployment. Deployment failure will be considered the failure of any Branched TAG® Device component (Aortic Component, Aortic Extender, or SB Component) to be released from the delivery catheter resulting in a serious adverse event (SAE) due to mechanical failure or use error.
Measured: During treatment procedure (day 0)
Number of Participants With Primary Procedural Side Branch Patency
The presence of forward flow through the implanted Side Branch Component into the target branch vessel.
Measured: At conclusion of the treatment procedure (day 0)
Secondary outcomes (2)
Number of Participants With 1 Month Side Branch Primary Patency Assessed by an Independent Core Lab
Measured: 1 Month
Number of Participants Without 1 Month Device Related Endoleaks Assessed by an Independent Core Lab
Device-related endoleaks are defined as the presence of contrast within the aneurysm sac originating from the junction between any Branched TAG® Device component and the landing zone (Type IA or IB) or the junction between the Aortic Component and either the Side Branch Component or the Aortic Extender (Type III).
Measured: 1 Month
Timeline
From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.
- First posted15 Oct 2014 (estimated)
- Study startApr 2015
- Primary completionJul 2016
- Study completionMar 2022
Resources
- Official record on ClinicalTrials.govThe registry entry this page is built from, with its full change history.
- Posted resultsParticipant flow, outcome measures and adverse events reported by the sponsor.
- Other trials for Aneurysm of Aortic ArchStudies of the same condition, open and under way.
- More trials run by W.L.Gore & AssociatesThe sponsor's other registered studies.
- What the trial phases meanHow a treatment moves from first tests in people to approval.
More details
For doctors and researchers: how the study is designed, who runs it, and its registry IDs.
Study design
- Study type
- Interventional
- Phase
- Phase N/A
- Primary purpose
- Device feasibility
- Allocation
- Not applicable
- Intervention model
- Single group
- Masking
- None (open label)
- Enrolment
- 9 participants (actual)
Sponsor and investigators
- Lead sponsor
- W.L.Gore & Associates
- Sponsor type
- Industry (pharma, biotech, devices)
- Responsible party
- Sponsor
- Investigators
- Michael D Dake, MD, Principal investigator, Stanford University
IDs, dates and publications
- ClinicalTrials.gov ID
- NCT02264977
- Sponsor's protocol ID
- SSB 11-03
- Status last verified
- Aug 2022
- First submitted
- 8 Oct 2014
- Last updated
- 14 Sep 2022
- Results posted
- Yes, on ClinicalTrials.gov
- Data monitoring committee
- Yes