GORE® TAG® Thoracic Branch Endoprosthesis Zone 0/1 Post-Approval Study
A study testing GORE® TAG® Thoracic Branch Endoprosthesis in people with Aortic Arch Dissection and Aortic Arch Aneurysm, run by W.L.Gore & Associates at 22 sites in United States.
Trial summary
A prospective multi-center post-market study collecting outcomes through at least 5-years and up to 10-years post procedure for subjects treated with the TBE Device in Zone 0/1 as part of routine clinical practice. Read more about this study
- Who is it for?
- Ages 18 years and older
- Conditions the study is for
- Aortic Arch Dissection, Aortic Arch Aneurysm
- What is being tested?
- GORE® TAG® Thoracic Branch Endoprosthesis
- Type of treatment
- Medical device
- Could I receive a placebo?
- No, everyone receives an active treatment
- How many people
- 125 participants
- Where
- 22 sitesUnited States
- Study dates
- 2 Mar 2026to Mar 2036
- Phase
- Interventional
- Are there results?
- Not yet: the study is under way
Key requirements
Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.
- Ages 18 years and older
- All sexes
- No healthy volunteers
- Adult, Older adult
Who may be able to join
- Patient or legally authorized representative (LAR) provides written authorization and/or consent per institution and geographical requirements.
- Patient has been or is intended to be treated with the TBE Device in Zone 0 or Zone 1.
- Patient is age ≥ 18 years at time of informed consent signature.
Who cannot join
- Patient who is, at the time of consent, unlikely to be available for defined follow-up visits.
- Patient with exclusion criteria required by local law.
- Patient is currently enrolled in or plans to enroll in any concurrent investigational drug and/or investigational device study within 12 months of study enrollment. Subjects cannot be enrolled in another Gore study.
Trial locations and contacts
22 sites in 1 country. Contact the study team or a site near you to ask about taking part.
22 sites in 1 country
United States22 sites, 22 recruiting
- Recruiting
University of Maryland School of Medicine
Baltimore, Maryland, 21201
Shahab Toursavadkohi, MD, 410-328-5840, stoursavadkohi@som.umaryland.edu
Shahab Toursavadkohi, MD
- Recruiting
Johns Hopkins University
Baltimore, Maryland, 21287
Mehrdad Ghoreishi, MD, 410-955-2800, mghorei1@jh.edu
Mehrdad Ghoreishi, MD
- Recruiting
University of North Carolina at Chapel Hill
Chapel Hill, North Carolina, 27599
Mark Farber, MD, 919-966-3391, mark_farber@med.unc.edu
Mark Faber, MD
- Recruiting
Duke University Medical Center
Durham, North Carolina, 27710
G. Chad Hughes IV, MD, 919-668-0903, gchad.hughes@duke.edu
G. Chad Hughes IV, MD
- Recruiting
University of Florida Health Heart & Vascular
Gainesville, Florida, 32608
Martin Back, MD, 352-273-5484, martin.back@surgery.ufl.edu
Martin Back, MD
- Recruiting
Hartford Hospital
Hartford, Connecticut, 06102
Akhilesh Jain, MD, 860-522-4158, akhilesh.jain@hhchealth.org
Akhilesh Jain, MD
- Recruiting
Baylor College of Medicine
Houston, Texas, 77030
Steven Maximus, MD, 832-355-9571, steven.maximus@bcm.edu
Steven Maximus, MD
- Recruiting
The Methodist Hospital-Houston
Houston, Texas, 77030
Marvin Atkins, MD, 713-441-5200, mdatkins@houstonmethodist.org
Marvin Atkins, MD
- Recruiting
Lakeland Regional Medical Center
Lakeland, Florida, 33805
Mohammed Hassan, MD, 863-413-5969, mohammed.hassan@mylrh.org
Mohammed Hassan, MD
- Recruiting
Keck Medical Center of USC
Los Angeles, California, 90033
Sukgu Han, MD, 323-442-5849, sukgu.han@med.usc.edu
Sukgu Han, MD
- Recruiting
University of Wisconsin - Madison
Madison, Wisconsin, 53792
Kyla Bennett, MD, 608-265-4420, bennettk@surgery.wisc.edu
Kyla Bennett, MD
- Recruiting
Medical College of Wisconsin
Milwaukee, Wisconsin, 53226
Peter Rossi, MD, 414-955-1800, prossi@mcw.edu
Peter Rossi, MD
- Recruiting
Intermountain Medical Center
Murray, Utah, 84107
Jane He, MD, 801-507-4700, jane.he@imail2.org
Jane He, MD
- Recruiting
Stanford University Hospital and Clinics
Palo Alto, California, 94304-5407
A. Claire Watkins, MD, 650-723-5771, acwat@stanford.edu
A. Claire Watkins, MD
- Recruiting
Thomas Jefferson University
Philadelphia, Pennsylvania, 19107
Cecilia Chin, MD, 617-835-9038, cecillia.chin@jefferson.edu
Cecilia Chin, MD
- Recruiting
Allegheny General Hospital
Pittsburgh, Pennsylvania, 15212
Bart Chess, MD, 412-359-8820, bart.chess@ahn.org
Bart Chess, MD
- Recruiting
Baylor Scott & White Research Institute
Plano, Texas, 75093
Justin Schaffer, MD, 469-800-6200, justin.schaffer@bswhealth.org
Justin Schaffer, MD
- Recruiting
University of California Davis Health System
Sacramento, California, 95817
Andrew Barleben, MD, 916-734-6609, abarleben@health.ucdavis.edu
Andrew Barleben, MD
- Recruiting
HonorHealth
Scottsdale, Arizona, 85251
Hasan Aldailami, MD, 480-246-3000, haldailami@honorhealth.com
Hasan Aldailami, MD
- Recruiting
University of Washington Medical Center
Seattle, Washington, 98195
Matthew Sweet, MD, 206-598-8487, mpsweet@uw.edu
Matthew Sweet, MD
- Recruiting
University of South Florida
Tampa, Florida, 33606
Halim Yammine, MD, 813-821-8034, hyammine@usf.edu
Halim Yammine, MD
- Recruiting
MedStar Washington Hospital Center
Washington D.C., District of Columbia, 20010
William Yoon, MD, 202-877-5975, william.j.yoon@medstar.net
William Yoon, MD
About this study
A prospective multi-center post-market study collecting outcomes through at least 5-years and up to 10-years post procedure for subjects treated with the TBE Device in Zone 0/1 as part of routine clinical practice.
Read the detailed description
Minimum of 125 total subjects with a TBE Device implantation procedure initiated in up to 50 Sites in the United States. Subjects will have follow-up at 1 Month, 6 Month, 1 Year, and annually thereafter for a minimum of 5-years and up to a maximum of 10 years.
- Conditions
- Aortic Arch Dissection, Aortic Arch Aneurysm
- Being tested
- GORE® TAG® Thoracic Branch Endoprosthesis
- Keywords
- Zone 0, Zone 1
- Medical subject headings
- Dissection, Thoracic Aorta, Aneurysm, Aortic Arch
What participants receive
What participants receive. In a randomised study each person is assigned to one group by chance.
GORE® TAG® Thoracic Branch EndoprosthesisDevice
Lesions in aortic arch in Zone 0 or Zone 1
| Study group | Role | What they receive |
|---|---|---|
| GORE® TAG® Thoracic Branch Endoprosthesis (TBE Device) | Other | Not described Receives: Device: GORE® TAG® Thoracic Branch Endoprosthesis |
What is being measured
The primary outcome is the main question the study is designed to answer.
Number of subjects with index endovascular procedure initiated after (or following) the surgical revascularization procedure
Great vessel surgical revascularization procedure is completed, and the endovascular implantation procedure is initiated for the implantation of the TBE Device, without interim death or study discontinuation.
Measured: Index Endovascular Procedure Only
Device Technical Success (index procedure only)
- Successful access, delivery, and accurate deployment of the device to the intended location, and retrieval of the delivery system.
- Patency of the endovascular aortic and side branch components.
- The absence of any unanticipated additional corrective procedure related to the device, procedure, or withdrawal of the delivery system.
Measured: Index Endovascular Procedure Only
Lesion-related mortality
Death meeting at least one of the following characteristics:
- Death during the hospitalization or within 30 days following the surgical revascularization procedure.
- Death during the hospitalization or within 30 days following the index endovascular procedure or following conversion to open repair.
- Death during the hospitalization or within 30 days following a complication from a secondary procedure associated with the index lesion or study device.
- Any death where the treated disease / index lesion or study device caused or significantly contributed to the death, including lesion-rupture, aortic-related complications, disease progression involving the index lesion.
Measured: Within 30 days of surgical revascularization or within 30 days of index endovascular procedure
Disabling Stroke
Stroke is the acute onset of symptoms consistent with focal or multifocal central nervous system (CNS) injury caused by vascular blockage resulting in ischemia or vascular rupture resulting in hemorrhage, that:
- Is identified within 30-days following the surgical revascularization procedure or within 30-days following the index endovascular procedure -AND-
- Persists for ≥ 24 hours or until death, or has symptoms lasting < 24 hours, with pathology or neuroimaging confirmation of CNS infarction -AND-
- Has an mRS ≥ 2 with an increase from baseline of at least one grade due to neurological deficits at no more than 120 days post index endovascular procedure.
Measured: Within 30 days of surgical revascularization or within 30 days of index endovascular procedure
Permanent Paraplegia (within 30-days)
Paraplegia secondary to Spinal Cord Ischemia identified within 30 days following the index endovascular procedure combined with Spinal Cord Ischemia scale grade = 3 at the one-month follow-up visit. Paraplegia will be assessed using the SVS spinal cord ischemia grading system
Measured: Within 30 days of index endovascular procedure
Permanent Paraparesis (within 30-days)
Paraparesis secondary to Spinal Cord Ischemia identified within 30 days following the index endovascular procedure, combined with Spinal Cord Ischemia scale grade = 2 at the one-month follow-up visit. Paraparesis will be assessed using the SVS spinal cord ischemia grading system.
Measured: Within 30 days of index endovascular procedure
New onset renal failure (within 30-days) requiring on-going dialysis
New onset of sustained renal failure identified within 30 days following the index endovascular procedure, combined with requiring dialysis at the 1-Month follow-up visit.
Measured: Within 30 days of index endovascular procedure
Lesion rupture (treated area)
Rupture in the treated segment of the vessel (e.g., aorta or branch) verified with direct observation or CT / CTA scan.
Measured: Minimum of 60 months and up to a maximum 120 months
Loss of aortic or branch patency
No flow or contrast detected through the implanted aortic component or endovascular branch component confirmed with imaging and/or direct observation.
Measured: Minimum of 60 months and up to a maximum of 120 months
Endoleaks
Perfusion of a treated lesion via:
- Type I endoleak is defined as a sealing failure at one of the attachment zones of the graft to the vessel wall, whereby arterial flow perfuses the treated lesion via a perigraft channel:
- Type Ia: Proximal aortic attachment zone
- Type Ib: Distal aortic attachment zone
- Type Ic: Distal attachment zone for branch component.
- Type II: Retrograde flow from native aortic branch arteries.
- Type III endoleak: Component disconnection or fabric disruption resulting in arterial flow into the perigraft space:
- Type IIIa: Attachment of aortic components (aortic-aortic)
- Type IIIb: Fabric tear or disruption
- Type IIIc: Attachment of aortic component-side-branch or side-branch-side-branch.
- Type IV: Late endoleak due to flow through porous fabric.
- Type V/Endotension: Aneurysm sac enlargement > 5 mm with no imaging evidence of an endoleak.
- Indeterminate: Endoleak is identified, but source cannot be ascertained.
Measured: Minimum of 60 months and up to a maximum of 120 months
Reintervention
Additional surgical or interventional procedure related to the treated disease / index lesion, the study device, or to the treatment / procedure.
This may include surgical or interventional treatment for endoleaks, disease progression, procedures on the branch to improve or restore patency, spinal drains for Spinal Cord Injury (SCI) management, or conversion to open surgery.
Events will not include access site(s) complications, interventions to address issues related to non-treated area of the index lesion, such as bare stent implantation to address bowel ischemia associated with aortic dissection.
Measured: Minimum of 60 months and up to a maximum of 120 months
Secondary outcomes (17)
Access-related complications (index endovascular procedure only)
Complications associated with the access sites used during treatment and any subsequent unplanned access-related reinterventions. This may include pseudoaneurysm, thrombosis, or complications associated with percutaneous closure devices.
Measured: Index endovascular procedure only
Life threatening bleed (index endovascular procedure only)
An event associated with the index endovascular procedure leading to any of the following events as prior to discharge:
- Overt bleeding requiring a transfusion of ≥5 units of whole blood/red blood cells within 48 hours of index endovascular procedure.
- Cardiac tamponade
- Overt bleeding requiring reoperation, surgical exploration, or reintervention for the purpose of controlling bleeding
- Bleeding requiring intravenous vasoactive drugs
- Fatal bleeding is bleeding that directly causes death with no other explainable cause
Measured: Index endovascular procedure only
Stroke
Stroke is the acute onset of symptoms consistent with focal or multifocal central nervous system (CNS) injury caused by vascular blockage resulting in ischemia or vascular rupture resulting in hemorrhage, that:
- Persists for ≥ 24 hours or until death -OR-
- Symptoms lasting < 24 hours, with pathology or neuroimaging confirmation of CNS infarction.
Disabling stroke will be determined based on mRS ≥ 2 with an increase from baseline of at least one grade due to neurological deficits at no more than 120 days following the suspected stroke event.
Measured: Minimum of 60 months and up to a maximum of 120 months
Transient Ischemic Attack (TIA)
Transient focal neurological signs or symptoms (lasting < 24 h) presumed to be due to focal brain ischemia but without evidence of acute infarction by neuroimaging or pathology (or in the absence of imaging).
Measured: Minimum of 60 months and up to a maximum of 120 months
Paraplegia (within 30-days)
Paraplegia secondary to Spinal Cord Ischemia identified within 30 days following the index endovascular procedure. Paraplegia will be assessed using the SVS spinal cord ischemia grading system.
Measured: Within 30 days of index endovascular procedure
Paraparesis (within 30-days)
Paraparesis secondary to Spinal Cord Ischemia identified within 30 days following the index endovascular procedure. Paraparesis will be assessed using the SVS spinal cord ischemia grading system.
Measured: Within 30 days of index endovascular procedure
Distal device-related thromboembolic adverse event requiring intervention or surgery (through 1-Month)
Device-related ischemic events in tissues distal to the device implantation site resulting in bypass, open surgical or transcatheter intervention, limb amputation, or leading to death.
Measured: Through 1 Month of index endovascular procedure
Myocardial infarction (within 30-days)
Elevated CK-MB (≥ 10x ULN) measured within 48 hours following the surgical revascularization or index endovascular procedure or need for urgent or emergent PCI/CABG within 30-days following the surgical revascularization or index endovascular procedure as confirmed at the 1-Month follow-up visit.
Measured: Within 30 days of surgical revascularization procedure or within 30 days of index endovascular procedure
Laryngeal or Phrenic Nerve injury (within 30-days)
Irreversible laryngeal nerve injury or phrenic nerve injury requiring unplanned surgery or intervention within 30-days following the surgical revascularization or index endovascular procedure as confirmed at the 1-Month follow-up visit.
Measured: Within 30 days of surgical revascularization or within 30 days of index endovascular procedure
Fistula formation
Fistula detected between the aorta and any adjacent structure associated with the area treated with the endovascular stent graft system verified with direct observation or CT scan.
Measured: Minimum of 60 months and up to a maximum of 120 months
Upper extremity ischemia
Clinically reported, device-related ischemia (Serious Adverse Event) not evident at the time of the index procedure with CTA-verified branch vessel compromise (i.e., malperfusion).
Measured: Minimum of 60 months and up to a maximum of 120 months
Lesion enlargement (treated area)
An increase in maximum vessel (e.g., aorta or treated branch) diameter of > 5 mm in the treated region as compared to baseline using orthogonal (i.e., perpendicular to the centerline) measurements on CT / CTA scans
Measured: Minimum of 60 months and up to maximum of 120 months
Device migration
Longitudinal movement of all or part of the device for a distance ≥ 10 mm, as confirmed by CTA scan, relative to anatomical landmarks and device positioning at the first post-operative CTA scan.
Measured: Minimum of 60 months and up to a maximum of 120 months
New dissection (with / without treatment)
A new arterial tear that was not present at the time of initial presentation caused either by a stent graft, natural disease progression or iatrogenic injury during the endovascular procedure. This can include the propagation, or extension, of a previous dissection.
Measured: Minimum of 60 months and up to maximum of 120 months
Device integrity events (e.g., fracture, kinking, compression)
Defined as any of the following:
- Wire fractures
- Stent kinking: Narrowing of the stent graft associated with demonstrable angulation in any of the stent components, with demonstrable flow
- Disruption / tears in the graft component of the stent graft
- Stent compression or invagination: Transient or permanent stent graft collapse following complete device deployment, resulting in an overall reduction in the vessel luminal diameter
Measured: Minimum of 60 months and up to maximum of 120 months
False lumen status (dissection pathology)
Status of the false lumen both within and outside of the segment of the aorta initially treated with an endovascular stent graft:
- Patent: Flow present throughout the entire aortic false lumen (absence of thrombus) on arterial-phase or delayed contrast-enhanced imaging.
- Partial thrombosis: Thrombus or clot within the aortic false lumen but with a residual patent flow channel on arterial-phase or delayed contrast-enhanced imaging.
- Complete thrombosis: Complete thrombosis of the aortic false lumen on arterial- and delayed-phase imaging.
Measured: Minimum of 60 months and up to maximum of 120 months
False lumen perfusion (dissection pathology)
Flow into the false lumen via:
- Type IA entry flow: Flow between the proximal endograft and aortic wall allowing systemic pressure antegrade flow into the primary entry tear and proximal false lumen.
- Type IB entry flow: Distal entry tear adjacent to endograft due to septal fenestration or a new intimal tear at the distal aspect of the stent graft (dSINE) allowing systemic pressure direct flow into the false lumen.
- Type II entry flow: Retrograde entry flow through arch vessel branches (innominate, carotid, subclavian) or thoracic bronchial and intercostal arteries into the false lumen.
- Type R entry flow: Antegrade entry flow from the true lumen into the false lumen through distal branch fenestrations (uncovered intercostal arteries, visceral or renal arteries, lumbar arteries, iliac branches) or septal fenestrations (excluding SINE).
Measured: Minimum of 60 months and up to maximum of 120 months
Timeline
From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.
- First posted10 Sep 2025
- Study start2 Mar 2026
- Primary completionJul 2028 (estimated)
- Study completionMar 2036 (estimated)
Resources
- Official record on ClinicalTrials.govThe registry entry this page is built from, with its full change history.
- Other trials for Aortic Arch DissectionStudies of the same condition, open and under way.
- More trials run by W.L.Gore & AssociatesThe sponsor's other registered studies.
- What the trial phases meanHow a treatment moves from first tests in people to approval.
More details
For doctors and researchers: how the study is designed, who runs it, and its registry IDs.
Study design
- Study type
- Interventional
- Phase
- Phase N/A
- Primary purpose
- Treatment
- Allocation
- Not applicable
- Intervention model
- Single group
- Model description
- GORE® TAG® Thoracic Branch Endoprosthesis (TBE)
- Masking
- None (open label)
- Enrolment
- 125 participants (estimated)
Sponsor and investigators
- Lead sponsor
- W.L.Gore & Associates
- Sponsor type
- Industry (pharma, biotech, devices)
- Responsible party
- Sponsor
IDs, dates and publications
- ClinicalTrials.gov ID
- NCT07166133
- Sponsor's protocol ID
- TBE 25-03
- Status last verified
- Oct 2026
- First submitted
- 2 Sep 2025
- Last updated
- 23 Oct 2026
- Results posted
- No
- Data monitoring committee
- No
- FDA-regulated drug
- No
- FDA-regulated device
- Yes
- Individual data sharing
- Yes: Study Protocol
Study at a glance
From the registry record, last updated by the sponsor on 23 Oct 2026. Recruitment status is as the sponsor reported it: confirm with the study team.
- Official title
- GORE® TAG® Thoracic Branch Endoprosthesis Zone 0/1 Post-Approval Study
- ClinicalTrials.gov ID
- NCT07166133
- Acronym / sponsor ID
- TBE 25-03
- Conditions
- Aortic Arch Dissection, Aortic Arch Aneurysm
- Keywords
- Zone 0, Zone 1
- Study type
- Interventional
- Recruitment status
- Recruiting
- Enrolment
- 125 participants (estimated)
- Lead sponsor
- W.L.Gore & AssociatesIndustry (pharma, biotech, devices)
- Study start
- 2 Mar 2026 (actual)
- Primary completion
- Jul 2028 (estimated)
- Study completion
- Mar 2036 (estimated)
- First posted
- 10 Sep 2025 (actual)
- Results
- Not posted in the source record
Clinical summary
The research question as the sponsor registered it. This describes what the study plans to test, not what it found.
A prospective multi-center post-market study collecting outcomes through at least 5-years and up to 10-years post procedure for subjects treated with the TBE Device in Zone 0/1 as part of routine clinical practice.
Detailed description
Minimum of 125 total subjects with a TBE Device implantation procedure initiated in up to 50 Sites in the United States. Subjects will have follow-up at 1 Month, 6 Month, 1 Year, and annually thereafter for a minimum of 5-years and up to a maximum of 10 years.
Study design
- Primary purpose
- Treatment
- Allocation
- Not applicable
- Intervention model
- Single group
- Masking
- None (open label)
- Number of arms
- 1
- Model description
- GORE® TAG® Thoracic Branch Endoprosthesis (TBE)
| Arm | Type | Receives |
|---|---|---|
| GORE® TAG® Thoracic Branch Endoprosthesis (TBE Device) | Other | Device: GORE® TAG® Thoracic Branch Endoprosthesis |
Who may qualify
- Ages
- 18 years and older
- Sex
- All
- Healthy volunteers
- Not accepted
- Age groups
- Adult, Older adult
Key inclusion criteria (3 listed)
- Patient or legally authorized representative (LAR) provides written authorization and/or consent per institution and geographical requirements.
- Patient has been or is intended to be treated with the TBE Device in Zone 0 or Zone 1.
- Patient is age ≥ 18 years at time of informed consent signature.
Key exclusion criteria (3 listed)
- Patient who is, at the time of consent, unlikely to be available for defined follow-up visits.
- Patient with exclusion criteria required by local law.
- Patient is currently enrolled in or plans to enroll in any concurrent investigational drug and/or investigational device study within 12 months of study enrollment. Subjects cannot be enrolled in another Gore study.
Eligibility is decided by the study team after screening; this overview does not establish that a patient qualifies. Full inclusion and exclusion criteria.
Interventions
| Intervention | Type | Arms |
|---|---|---|
| GORE® TAG® Thoracic Branch EndoprosthesisLesions in aortic arch in Zone 0 or Zone 1 | Device | GORE® TAG® Thoracic Branch Endoprosthesis (TBE Device) |
Primary outcome
Number of subjects with index endovascular procedure initiated after (or following) the surgical revascularization procedure
Time frame: Index Endovascular Procedure Only
Great vessel surgical revascularization procedure is completed, and the endovascular implantation procedure is initiated for the implantation of the TBE Device, without interim death or study discontinuation.
Device Technical Success (index procedure only)
Time frame: Index Endovascular Procedure Only
* Successful access, delivery, and accurate deployment of the device to the intended location, and retrieval of the delivery system. * Patency of the endovascular aortic and side branch components. * The absence of any unanticipated additional corrective procedure related to the device, procedure, or withdrawal of the delivery system.
Lesion-related mortality
Time frame: Within 30 days of surgical revascularization or within 30 days of index endovascular procedure
Death meeting at least one of the following characteristics: * Death during the hospitalization or within 30 days following the surgical revascularization procedure. * Death during the hospitalization or within 30 days following the index endovascular procedure or following conversion to open repair. * Death during the hospitalization or within 30 days following a complication from a secondary procedure associated with the index lesion or study device. * Any death where the treated disease / index lesion or study device caused or significantly contributed to the death, including lesion-rupture, aortic-related complications, disease progression involving the index lesion.
Disabling Stroke
Time frame: Within 30 days of surgical revascularization or within 30 days of index endovascular procedure
Stroke is the acute onset of symptoms consistent with focal or multifocal central nervous system (CNS) injury caused by vascular blockage resulting in ischemia or vascular rupture resulting in hemorrhage, that: * Is identified within 30-days following the surgical revascularization procedure or within 30-days following the index endovascular procedure -AND- * Persists for ≥ 24 hours or until death, or has symptoms lasting < 24 hours, with pathology or neuroimaging confirmation of CNS infarction -AND- * Has an mRS ≥ 2 with an increase from baseline of at least one grade due to neurological deficits at no more than 120 days post index endovascular procedure.
Permanent Paraplegia (within 30-days)
Time frame: Within 30 days of index endovascular procedure
Paraplegia secondary to Spinal Cord Ischemia identified within 30 days following the index endovascular procedure combined with Spinal Cord Ischemia scale grade = 3 at the one-month follow-up visit. Paraplegia will be assessed using the SVS spinal cord ischemia grading system
Permanent Paraparesis (within 30-days)
Time frame: Within 30 days of index endovascular procedure
Paraparesis secondary to Spinal Cord Ischemia identified within 30 days following the index endovascular procedure, combined with Spinal Cord Ischemia scale grade = 2 at the one-month follow-up visit. Paraparesis will be assessed using the SVS spinal cord ischemia grading system.
New onset renal failure (within 30-days) requiring on-going dialysis
Time frame: Within 30 days of index endovascular procedure
New onset of sustained renal failure identified within 30 days following the index endovascular procedure, combined with requiring dialysis at the 1-Month follow-up visit.
Lesion rupture (treated area)
Time frame: Minimum of 60 months and up to a maximum 120 months
Rupture in the treated segment of the vessel (e.g., aorta or branch) verified with direct observation or CT / CTA scan.
Loss of aortic or branch patency
Time frame: Minimum of 60 months and up to a maximum of 120 months
No flow or contrast detected through the implanted aortic component or endovascular branch component confirmed with imaging and/or direct observation.
Endoleaks
Time frame: Minimum of 60 months and up to a maximum of 120 months
Perfusion of a treated lesion via: * Type I endoleak is defined as a sealing failure at one of the attachment zones of the graft to the vessel wall, whereby arterial flow perfuses the treated lesion via a perigraft channel: * Type Ia: Proximal aortic attachment zone * Type Ib: Distal aortic attachment zone * Type Ic: Distal attachment zone for branch component. * Type II: Retrograde flow from native aortic branch arteries. * Type III endoleak: Component disconnection or fabric disruption resulting in arterial flow into the perigraft space: * Type IIIa: Attachment of aortic components (aortic-aortic) * Type IIIb: Fabric tear or disruption * Type IIIc: Attachment of aortic component-side-branch or side-branch-side-branch. * Type IV: Late endoleak due to flow through porous fabric. * Type V/Endotension: Aneurysm sac enlargement > 5 mm with no imaging evidence of an endoleak. * Indeterminate: Endoleak is identified, but source cannot be ascertained.
Reintervention
Time frame: Minimum of 60 months and up to a maximum of 120 months
Additional surgical or interventional procedure related to the treated disease / index lesion, the study device, or to the treatment / procedure. This may include surgical or interventional treatment for endoleaks, disease progression, procedures on the branch to improve or restore patency, spinal drains for Spinal Cord Injury (SCI) management, or conversion to open surgery. Events will not include access site(s) complications, interventions to address issues related to non-treated area of the index lesion, such as bare stent implantation to address bowel ischemia associated with aortic dissection.
17 secondary outcomes
Access-related complications (index endovascular procedure only)
Time frame: Index endovascular procedure only
Life threatening bleed (index endovascular procedure only)
Time frame: Index endovascular procedure only
Stroke
Time frame: Minimum of 60 months and up to a maximum of 120 months
Transient Ischemic Attack (TIA)
Time frame: Minimum of 60 months and up to a maximum of 120 months
Paraplegia (within 30-days)
Time frame: Within 30 days of index endovascular procedure
Paraparesis (within 30-days)
Time frame: Within 30 days of index endovascular procedure
Distal device-related thromboembolic adverse event requiring intervention or surgery (through 1-Month)
Time frame: Through 1 Month of index endovascular procedure
Myocardial infarction (within 30-days)
Time frame: Within 30 days of surgical revascularization procedure or within 30 days of index endovascular procedure
Laryngeal or Phrenic Nerve injury (within 30-days)
Time frame: Within 30 days of surgical revascularization or within 30 days of index endovascular procedure
Fistula formation
Time frame: Minimum of 60 months and up to a maximum of 120 months
Upper extremity ischemia
Time frame: Minimum of 60 months and up to a maximum of 120 months
Lesion enlargement (treated area)
Time frame: Minimum of 60 months and up to maximum of 120 months
Device migration
Time frame: Minimum of 60 months and up to a maximum of 120 months
New dissection (with / without treatment)
Time frame: Minimum of 60 months and up to maximum of 120 months
Device integrity events (e.g., fracture, kinking, compression)
Time frame: Minimum of 60 months and up to maximum of 120 months
False lumen status (dissection pathology)
Time frame: Minimum of 60 months and up to maximum of 120 months
False lumen perfusion (dissection pathology)
Time frame: Minimum of 60 months and up to maximum of 120 months
Results and safety
This study is still under way. Results are not available in the source record yet.
Research links
- Official record on ClinicalTrials.govLast updated by the sponsor 23 Oct 2026.
- Individual participant dataSharing planned: Yes. Study Protocol
Related trials for Aortic Arch Dissection
Enrolling by invitation
Endovascular Treatment of Thoracic Aortic Disease
Thoracic Aortic Aneurysms, Dissecting, Aneurysm
Not yet recruiting
A Novel Wire-Reinforced Fenestration of a Hybrid Frozen Elephant Trunk Device for Supra-Aortic Branch Vessel Stenting During Aortic Arch Replacement
Aortic Aneurysm, Aortic Dissection
Recruiting
Clinical Study to Evaluate the Safety and Effectiveness of Arcevo LSA
Aortic Arch Aneurysm, Aortic Arch Dissection
Recruiting
MBA Early Feasibility Study
Thoracic Aortic Aneurysm, Penetrating Aortic Ulcers (PAUs)
Recruiting
Hemiarch vs Extended Arch in Type 1 Aortic Dissection
Aortic Dissection
Referral and contact
AzTrials does not enrol participants or take referrals. Contact the study team or a site listed below; they decide who can take part.
22 sites in 1 country: see every site and its status.
Questions about this trial
What is this clinical trial studying?
A study testing GORE® TAG® Thoracic Branch Endoprosthesis in people with Aortic Arch Dissection and Aortic Arch Aneurysm, run by W.L.Gore & Associates at 22 sites in United States.
The full registered description is in the About this study section and the official record.
Who may be eligible to take part?
The sponsor lists 3 inclusion and 3 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.
Is this clinical trial currently recruiting?
The record says: looking for participants now, as of the sponsor's last update on 23 Oct 2026. Status can change between updates, so confirm with the study contact before making plans.
Where is the study taking place?
The record lists 22 sites in United States. Whether a site is taking part right now should be confirmed with that site. See every site.
What are the possible benefits and risks?
They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.
Are results available for this study?
This study is still under way. Results are not available in the source record yet.
How can a doctor refer a patient or contact the research team?
Contact DeAnne Hart (800-437-8181, tbe25-03@wlgore.com), or a site near the patient. AzTrials does not enrol participants or take referrals.
What does the primary outcome measure?
Number of subjects with index endovascular procedure initiated after (or following) the surgical revascularization procedure (time frame: Index Endovascular Procedure Only). The study lists 11 primary outcomes in all. It is what the study was designed to measure, not a finding.