Atrioventricular BlockNCT00187278

Biopace Study: Biventricular Pacing for Atrioventricular Block to Prevent Cardiac Desynchronization

A study testing Biventricular Pacing and RV Pacing in people with Atrioventricular Block and Ventricular Dysfunction, run by Abbott Medical Devices at 1 site in Germany.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record

Search all trials

Trial summary

The primary purpose of the study is to evaluate if patients with a standard indication for permanent ventricular pacing, left ventricular ejection fraction without limit, or any QRS duration will profit from the prevention of ventricular desynchronisation. Read more about this study

Who is it for?
Ages 18 years and older
What is being tested?
Biventricular Pacing and RV Pacing
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
1,833 participants
Where
1 siteGermany
Study dates
May 2003to Oct 2014
Phase
Interventional

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 years and older
  • All sexes
  • No healthy volunteers
  • Adult, Older adult

Who may be able to join

  • Presence of an indication for ventricular pacing according to the actual guidelines for the implantation of cardiac pacemakers and a need for frequent (or even permanent) ventricular pacing for:
  • Permanent 3rd degree atrioventricular (AV)-block or
  • Intermittent 3rd degree AV-block in combination with 1st degree AV-block with a pQ-interval ≥ 220 ms or
  • 2nd degree AV-block type Mobitz II in combination with 1st degree AV-block with a pQ-interval ≥ 220 ms or
  • 2nd degree AV-block type Mobitz I (if indicated) in combination with 1st degree AV-block with a pQ-interval ≥ 220 ms or
  • 1st degree AV-block with a pQ-interval ≥ 220 ms and indication for ventricular pacing (includes indication for ventricular pacing based on long HV interval measured during invasive electrophysiological testing) or
Show all 15Show fewer
  • Sick-sinus-syndrome with symptomatic sinus bradycardia or sinus arrest as primary indication for device implantation in combination with long 1st degree AV-block with a pQ-interval ≥ 220 ms or
  • Chronic (permanent) atrial fibrillation (flutter or tachycardia) with a spontaneous heart (ventricular) rate at rest ≤ 60/min or
  • Chronic (permanent) atrial fibrillation (flutter or tachycardia) with a spontaneous heart (ventricular) rate at rest ≤ 75/min, if initiation or increase of pharmacological treatment with a relevant heart rate lowering effect (negative chronotropic effect) is planned for the time after pacemaker implantation (i.e. ß-blockers for heart failure and rate control)
  • Patients scheduled for AV node ablation
  • Any QRS duration and morphology
  • Left ventricular ejection fraction (LVEF) without limit as measured by echocardiography (in at least one plane, either 4- or 2-chamber or apical long axis view)
  • Signed written informed consent of the patient or a first-degree relative for study participation after informing the patient/relative about the risks and the aim of the study
  • Willingness and ability to comply with the prescribed follow-up tests and schedule of evaluations.
  • Absence of an implanted ventricular pacing device (patients with atrial pacemakers and new need for ventricular pacing may be included)

Who cannot join

  • Implanted Cardioverter Defibrillator or consideration for implantation of an ICD due to arrhythmia indication. However, ICD implant for primary prevention of sudden cardiac death in patients with LVEF ≤ 35 % (in accordance with the actual guidelines for the implantation of arrhythmia devices[LVEF < 30%] and in accordance with the results of the SCD-Heft study [LVEF < 35%) will be allowed.
  • Implanted ventricular pacing device
  • Status 1 for cardiac transplantation and likelihood to receive transplantation within 2 years (these patients would not be expected to fulfill the follow-up requirements as outlined in this protocol)
  • Evidence of acute left ventricular dysfunction and high probability for its reversibility (e.g. acute myocarditis, tachycardiomyopathy)
  • Implanted prosthetic tricuspid valve
  • Severe musculoskeletal disorder(s)
Show all 11Show fewer
  • Age below 18 years
  • Current or planned pregnancy in the next 6 months
  • Current or recent (within the past 30 days) participation in any other clinical investigation
  • Life expectancy of less than 6 months
  • Patient's inability to independently comprehend and complete the Quality of Life (QoL) questionnaire

Trial locations and contacts

1 site in 1 country.

Germany1 site
  • Klinikum der Philipps-Universität Marburg

    Marburg, 35033

Other trials in Germany

About this study

Official title: Biventricular Pacing for Atrioventricular Block in Left Ventricular Dysfunction to Prevent Cardiac Desynchronization

The primary purpose of the study is to evaluate if patients with a standard indication for permanent ventricular pacing, left ventricular ejection fraction without limit, or any QRS duration will profit from the prevention of ventricular desynchronisation.

Read the detailed description

The study will be performed as a controlled, single-blind, international, multicenter, prospective, randomized, parallel group design.

In order to pragmatically examine the effectiveness of biventricular pacing in patients with an indication for ventricular pacing, the study group with biventricular pacing is compared to a control group with standard pacemakers which only allow univentricular (right ventricular) stimulation, as it has been the standard outside of clinical studies until so far.

Being tested
Biventricular Pacing, RV Pacing
Keywords
Standard pacing indication
Medical subject headings
Atrioventricular Block, Ventricular Dysfunction

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Biventricular PacingDevice

    Biventricular Pacemaker implant

  • RV PacingDevice

    Standard Pacemaker implant

Study groups in NCT00187278
Study groupRoleWhat they receive
RV PacingActive comparatorStandard Pacemaker implant
Receives: Device: RV Pacing
Biventricular PacingExperimentalBiventricular Pacemaker implant
Receives: Device: Biventricular Pacing

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Total Mortality

    Number of deaths observed

    Measured: Study duration (5.7 years mean follow-up)

  2. Death or Heart Failure Hospitalization

    Number of subjects who experienced death or hospitalization where heart failure is a primary cause of hospitalization

    Measured: Study duration (5.7 years mean follow-up)

Secondary outcomes (15)
  1. Death Due to Cardiovascular Causes

    Number of subjects who died due to cardiovascular causes, as classified by the independent Event Adjudication Committee (IEAC)

    Measured: Study duration (5.7 years mean follow-up)

  2. Functional Capacity as Measured by the Distance Covered in the 6-minute Walk Test

    Measured: 12 months post-implant

  3. Health Related Quality of Life Measured by the Minnesota Living With Heart Failure Questionnaire©

    Minnesota Living with Heart Failure (MLWHF) questionnaire score at 12 months

    MLWHF score: the total scale is from 0 to 105, with higher scores indicating worse health status.

    Measured: 12 months post-implant

  4. Incidence of Hospitalizations for Deterioration of Heart Failure

    Measured: Study duration (5.7 years mean follow-up)

  5. Incidence of Hospitalizations for Cardiovascular Events

    Measured: Study duration (5.7 years mean follow-up)

  6. Incidence of Hospitalizations for Any Reason

    Measured: Study duration (5.7 years mean follow-up)

  7. Cardiac Structure and Function

    Measured: 12 & 24 months

  8. Successful Implantation of the Left Ventricular Lead

    Measured: Implantation

  9. Chronic Atrial Fibrillation (Defined as Presence of Atrial Fibrillation in Two Subsequent ECG´s/Visits)

    Chronic atrial fibrillation is defined as presence of atrial fibrillation in two subsequent ECG´s/visits.

    Measured: Study duration (5.7 years mean follow-up)

  10. Functional Capacity as Measured by the Distance Covered in the 6-minute Walk Test

    Measured: 24 months post-implant

  11. Health Related Quality of Life Measured by the Minnesota Living With Heart Failure Questionnaire©

    Minnesota Living with Heart Failure (MLWHF) questionnaire score at 24 months.

    MLWHF score: the total scale is from 0 to 105, with higher scores indicating worse health status.

    Measured: 24 months post-implant

  12. Duration of Hospitalizations for Deterioration of Heart Failure

    Measured: Study Duration (5.7 years mean follow-up)

  13. Duration of Hospitalizations for Cardiovascular Events

    Measured: Study Duration (5.7 years mean follow-up)

  14. Duration of Hospitalizations for Any Reason

    Measured: Study duration (5.7 years mean follow-up)

  15. Adverse Events Related to Left Ventricular Lead

    Only serious adverse events related to the left ventricular lead are included.

    Measured: Study Duration (5.7 years mean follow-up)

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted16 Sep 2005 (estimated)
  2. Study startMay 2003
  3. Primary completionMay 2014
  4. Study completionOct 2014

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Treatment
Allocation
Randomised
Intervention model
Parallel groups
Masking
Single blind
Who is masked
Participant
Enrolment
1,833 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT00187278
Sponsor's protocol ID
CR03006HF
Status last verified
Jan 2021
First submitted
13 Sep 2005
Last updated
25 Jan 2021
Data monitoring committee
Yes
FDA-regulated drug
No
FDA-regulated device
No

Publications

  • Funck RC, Blanc JJ, Mueller HH, Schade-Brittinger C, Bailleul C, Maisch B; BioPace Study Group. Biventricular stimulation to prevent cardiac desynchronization: rationale, design, and endpoints of the 'Biventricular Pacing for Atrioventricular Block to Prevent Cardiac Desynchronization (BioPace)' study. Europace. 2006 Aug;8(8):629-35. doi: 10.1093/europace/eul075.

    PubMed 16864616
  • Funck RC, Mueller HH, Lunati M, Piorkowski C, De Roy L, Paul V, Wittenberg M, Wuensch D, Blanc JJ; BioPace study group. Characteristics of a large sample of candidates for permanent ventricular pacing included in the Biventricular Pacing for Atrio-ventricular Block to Prevent Cardiac Desynchronization Study (BioPace). Europace. 2014 Mar;16(3):354-62. doi: 10.1093/europace/eut343. Epub 2013 Nov 7.

    PubMed 24200715
  • Funck RC, Muller HH, Lunati M, De Roy L, Klein N, Meisel E, Milasinovic G, Carlson MD, Wittenberg M, Hindricks G, Blanc JJ. Biventricular vs. right ventricular pacing devices in patients anticipated to require frequent ventricular pacing (BioPace). Europace. 2025 Mar 5;27(3):euaf029. doi: 10.1093/europace/euaf029.

    PubMed 40105785