Upper Limb Vascular ImagingNCT07916857

Comparison of VxWave and Standard Ultrasound Measurements of Blood Vessels in the Arm

A study testing VxWave Ultrasound Imaging System and Philips Lumify Ultrasound System With L12-4 Linear Probe in people with Upper Limb Vascular Imaging, run by Vexev Pty Ltd at 2 sites in Australia.

Not yet recruiting

Opening soon. The study is approved but enrolment has not started yet. Sites may already take enquiries.

Sites and contacts
Official record
Search all trials

Trial summary

This study compares blood-vessel measurements obtained with the VxWave Ultrasound Imaging System and a standard ultrasound system operated by a qualified sonographer. Adults aged 21 years or older undergo scanning of the same arm with both systems at one study visit. Read more about this study

Who is it for?
Ages 21 years and olderHealthy volunteers can take part
Conditions the study is for
Upper Limb Vascular Imaging
What is being tested?
VxWave Ultrasound Imaging System and Philips Lumify Ultrasound System With L12-4 Linear Probe
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
85 participants
Where
2 sitesAustralia
Study dates
15 Nov 2026to 20 Jan 2027
Phase
Interventional
Are there results?
Not yet: the study is under way

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 21 years and older
  • All sexes
  • Healthy volunteers accepted
  • Adult, Older adult

Who may be able to join

  • Age 21 years or older and able to provide written informed consent.
  • Complete upper limb(s).

Who cannot join

  • Unable to provide informed consent.
  • Unable or unwilling to place the limb onto the VxWave scanning bed.
  • Infection risk to site staff and/or other study participants.
  • Open or healing wound on the study limb.
  • Known abnormal upper-limb vascular anatomy that would make scanning at the protocol discrete locations infeasible.
  • Known allergy to parabens present in standard ultrasound gels.
Show all 11Show fewer
  • Any other reason at the discretion of the Principal Investigator.
  • Previous upper-limb vascular surgery.
  • Highly dependent on medical care.
  • Cognitive impairment or intellectual disability.
  • Vascular-anatomy and prior-surgery exclusions are assessed using self-reported history; review of medical records is not required for this assessment.

Trial locations and contacts

2 sites in 1 country. Contact the study team or a site near you to ask about taking part.

Michael Cartmill contact

+61 400 223 656

michael@vexev.com

Australia2 sites
  • Prince of Wales Private, A/Prof Shannon Thomas's Rooms (Recruitment and Enrollment Only)

    Randwick, New South Wales, 2031

    Michael Cartmill, +61 400 223 656, michael@vexev.com

  • Vexev Pty Ltd

    Surry Hills, New South Wales, 2010

    Michael Cartmill, +61 400 223 656, michael@vexev.com

Other trials in Australia

About this study

Official title: The VxWave Ultrasound Imaging System Upper Limb Clinical Performance Trial

This study compares blood-vessel measurements obtained with the VxWave Ultrasound Imaging System and a standard ultrasound system operated by a qualified sonographer. Adults aged 21 years or older undergo scanning of the same arm with both systems at one study visit. The researchers compare vessel diameter and blood-flow velocity at three specified locations in the upper arm and forearm. They also assess scan completion, scanning time, participant comfort, device deficiencies and adverse events. A subset undergoes a repeat standard ultrasound scan to assess the repeatability of the reference measurements.

Read the detailed description

This is an interventional, single-cohort, within-participant paired method-comparison study. Every participant undergoes VxWave scanning first, followed by Philips Lumify scanning of the same arm by a qualified vascular sonographer in a separate room. Three locations are examined: the brachial artery immediately proximal and immediately distal to the midpoint between the elbow crease and axillary fold, and the radial artery at the midpoint between the radial styloid process and elbow crease.

A qualified vascular sonographer who did not acquire the images performs all measurements offline using OsiriX MD for Lumify images and VxView for VxWave images. Random dataset identifiers conceal participant identity, pairing and repeat status. Measurements from the corresponding other-system dataset are unavailable during review. Imaging-system identity is not concealed by the use of separate software platforms.

The first 20 consecutive participants after Participant ID 030 who complete comparator acquisition undergo one repeat comparator scan after probe and gel removal and at least two minutes of rest. The first comparator acquisition remains the reference for primary agreement analysis; repeat acquisition is used only for reference-pathway repeatability.

Enrollment stops upon reaching either 50 participants with successful unilateral scans in the per-protocol set or 85 participants in the full analysis set. The first unilateral visit supplies the primary participant record. Optional contralateral-arm visits provide exploratory observations. No routine follow-up visit is scheduled; any adverse event is followed until resolution or 60 days, whichever occurs first.

Participants may provide consent and enroll at Prince of Wales Private or remotely. All ultrasound scanning and study imaging data acquisition are performed at Vexev Pty Ltd, Suite 501, 78 Kippax Street, Surry Hills NSW 2010. The Prince of Wales Private location is used for recruitment, consent and enrollment only; no scanning or imaging data acquisition takes place there.

Being tested
VxWave Ultrasound Imaging System, Philips Lumify Ultrasound System With L12-4 Linear Probe

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • VxWave Ultrasound Imaging SystemDevice

    Robotically controlled acquisition of duplex B-mode and Doppler ultrasound data of the upper-limb vasculature, excluding the hand. The participant rests the limb on the scan bed while the probe moves beneath it. Images are measured offline using VxView. VxWave scanning occurs before the comparator scan.

  • Philips Lumify Ultrasound System With L12-4 Linear ProbeDevice

    Qualified vascular-sonographer acquisition of ultrasound data from the same upper-limb vascular locations after VxWave scanning. Images are measured offline using OsiriX MD. The specified repeatability subset undergoes a separate repeat comparator acquisition after probe lift-off and at least two minutes of rest.

Study groups in NCT07916857
Study groupRoleWhat they receive
VxWave and Comparator UltrasoundExperimentalAll participants undergo VxWave scanning followed by Philips Lumify scanning of the same arm during one study visit. The first 20 consecutive participants after Participant ID 030 who complete comparator acquisition also undergo a repeat Lumify scan after probe lift-off and at least two minutes of rest.
Receives: Device: VxWave Ultrasound Imaging System; Device: Philips Lumify Ultrasound System With L12-4 Linear Probe

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Inter-System Agreement in Arterial Diameter

    Mean bias and 95% limits of agreement for VxWave minus comparator arterial diameter in millimetres at three pre-specified brachial/radial artery locations. Each location uses the mean of three diameter measurements. Multilevel Bland-Altman analysis accounts for measurements within participants, with 95% confidence intervals for bias and limits of agreement. The +/-0.5 mm margin is a descriptive benchmark and does not define a binary study-success criterion. Only diagnostic-quality measurements are included in agreement analysis.

    Measured: During the first study visit (Day 1; approximately 2 hours, or 2 hours 15 minutes with repeat comparator scanning).

  2. Inter-System Agreement in Peak Systolic Velocity

    Mean percentage bias and 95% limits of agreement in symmetric relative percentage differences between VxWave and comparator peak systolic velocity at the three pre-specified brachial/radial artery locations. Mean of three representative peak systolic velocity measurements per location. Multilevel Bland-Altman analysis accounts for measurements within participants, with 95% confidence intervals for bias and limits of agreement. Absolute-scale agreement in cm/s is a sensitivity analysis. No PSV acceptance criterion is specified. Only diagnostic-quality measurements are included.

    Measured: During the first study visit (Day 1; approximately 2 hours, or 2 hours 15 minutes with repeat comparator scanning).

Secondary outcomes (5)
  1. Successful VxWave Scan Completion

    Percentage of participants for whom VxWave completes the peripheral vascular imaging examination, with scan data judged by the reporting sonographer to be of diagnostic quality for measurement and able to confirm measurements. Reasons for failure include inability to perform both B-mode and Doppler measurements at two of the three discrete locations, or failure to complete scan data acquisition within a reasonable time (less than 55 minutes).

    Measured: During the VxWave scan at the first study visit (Day 1).

  2. Time Required to Complete VxWave Scanning

    Elapsed time in minutes from initiation of the first acquisition run on the VxWave console to scan status complete on the final acquisition run, including time between runs. A scan may contain more than one acquisition run, including restarts after participant repositioning.

    Measured: During the VxWave scan at the first study visit (Day 1).

  3. Participant Comfort and Satisfaction With VxWave Scanning

    Four questionnaire items assess overall comfort, satisfaction with scanning time, satisfaction with audible noise and overall satisfaction. Each item has five ordered response categories: Excellent, Very good, Good, Fair and Poor, with Excellent representing the most favorable response. Ordinal responses will be summarised using median/interquartile range or proportions. No composite score is specified in the protocol or supplied assessment form.

    Measured: Immediately after the VxWave scan during the first study visit (Day 1).

  4. Repeatability of Comparator Arterial Diameter Measurements

    For the first 20 consecutive participants after Participant ID 030 who complete comparator acquisition, assess mean difference (bias), 95% limits of agreement and repeatability coefficient between first and repeat comparator arterial diameter measurements in millimetres at each pre-specified location. Multilevel Bland-Altman analysis accounts for measurements within participants, with 95% confidence intervals for bias and limits of agreement. The first comparator acquisition remains the reference for the primary endpoint; repeat acquisition is used for this secondary endpoint only.

    Measured: During the first study visit (Day 1); repeat comparator acquisition follows probe lift-off and at least two minutes of rest.

  5. Repeatability of Comparator Peak Systolic Velocity Measurements

    For the same 20-participant repeatability cohort, assess mean difference (bias), 95% limits of agreement and repeatability coefficient between first and repeat comparator peak systolic velocity measurements, expressed as symmetric relative percentage differences, at each pre-specified location. Multilevel Bland-Altman analysis accounts for measurements within participants, with 95% confidence intervals. This descriptive endpoint characterises reference-pathway repeatability and does not define an acceptance test.

    Measured: During the first study visit (Day 1); repeat comparator acquisition follows probe lift-off and at least two minutes of rest.

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted4 Nov 2026
  2. Study start15 Nov 2026 (estimated)
  3. Primary completion15 Jan 2027 (estimated)
  4. Study completion20 Jan 2027 (estimated)

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Other
Allocation
Not applicable
Intervention model
Single group
Model description
Single-cohort, within-participant paired method-comparison study. Every participant undergoes VxWave scanning followed by Philips Lumify scanning of the same arm during the same visit. The repeatability subset undergoes an additional comparator acquisition. There is no allocation between intervention arms.
Masking
None (open label)
Enrolment
85 participants (estimated)
IDs, dates and publications
ClinicalTrials.gov ID
NCT07916857
Sponsor's protocol ID
CIP-2
Status last verified
Oct 2026
First submitted
29 Oct 2026
Last updated
4 Nov 2026
Results posted
No
Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
No
Individual data sharing
Yes: Individual participant data will be shared on request.

Questions about this trial

What is this clinical trial studying?

A study testing VxWave Ultrasound Imaging System and Philips Lumify Ultrasound System With L12-4 Linear Probe in people with Upper Limb Vascular Imaging, run by Vexev Pty Ltd at 2 sites in Australia.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 2 inclusion and 11 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: opening soon, as of the sponsor's last update on 4 Nov 2026. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 2 sites in Australia. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

This study is still under way. Results are not available in the source record yet.

How can a doctor refer a patient or contact the research team?

Contact Michael Cartmill (+61 400 223 656, michael@vexev.com), or a site near the patient. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Inter-System Agreement in Arterial Diameter (time frame: During the first study visit (Day 1; approximately 2 hours, or 2 hours 15 minutes with repeat comparator scanning).). The study lists 2 primary outcomes in all. It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 4 Nov 2026.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in