Pediatric Sleep Disordered BreathingNCT07851402

Pediatric/Adolescent WatchPAT 400 Reliability Pilot Study

A study testing WatchPAT 400 in people with Pediatric Sleep Disordered Breathing and Obstructive Sleep Apnea, run by Milton S. Hershey Medical Center at 1 site in United States.

Not yet recruiting

Opening soon. The study is approved but enrolment has not started yet. Sites may already take enquiries.

Sites and contacts
Official record
Search all trials

Trial summary

The goal of this clinical trial is to learn whether a newer sleep-testing device gives results that closely match the standard overnight sleep study, in children and adolescents ages 6-18 who are being evaluated for possible sleep apnea. Read more about this study

Who is it for?
Ages 6 to 18 years
What is being tested?
WatchPAT 400
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
35 participants
Where
1 siteUnited States
Study dates
1 Dec 2026to 1 Dec 2028
Phase
Interventional
Are there results?
Not yet: the study is under way

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 6 to 18 years
  • All sexes
  • No healthy volunteers
  • Child, Adult

Who may be able to join

  • Age 6-18 years at time of enrollment
  • Referred to the sleep clinic for Sleep disordered breathing symptoms.
  • Scheduling for a technician-attended in-laboratory polysomnogram (PSG)
  • Obstructive Sleep Apnea-18 (OSA-18) questionnaire score ≥60
  • No neurodevelopmentally disability
  • Parent/legal guardian able and willing to provide written informed consent (participants who are 18 years old provide their own consent); participant able to provide age-appropriate assent where applicable
  • Finger, hand, and chest anatomy able to accommodate the WatchPAT 400 finger probe, wrist unit, and chest sensor patch

Who cannot join

  • Known peripheral vascular disease, Raynaud phenomenon, or other condition affecting digital arterial pulsation
  • Current use of alpha-adrenergic blocking agents or other medications known to significantly affect peripheral vasomotor tone
  • Permanent cardiac pacemaker, atrial fibrillation, or other clinically significant arrhythmia at the time of study
  • Known neurodevelopmental disability (Trisomy 21, autism spectrum disorder, or similar), as above
  • Prior amputation, digital deformity, or injury to the fingers precluding probe placement
  • Inability or unwillingness to tolerate simultaneous placement of both the PSG and WatchPAT sensors
  • Known latex or adhesive allergy relevant to device sensor materials
  • Pregnancy, for adolescent participants where applicable

Trial locations and contacts

1 site in 1 country. Contact the study team or a site near you to ask about taking part.

United States1 site
  • Penn State Hershey Medical Center

    Hummelstown, Pennsylvania, 17036

Other trials in United States

About this study

Official title: Diagnostic Accuracy and Inter-Rater Reliability of the WatchPAT® 400 Home Sleep Apnea Testing Device Compared to Attended In-Laboratory Polysomnography in Children and Adolescents Aged 6-18 Years

The goal of this clinical trial is to learn whether a newer sleep-testing device gives results that closely match the standard overnight sleep study, in children and adolescents ages 6-18 who are being evaluated for possible sleep apnea. The main questions it aims to answer are:

Does the new device (WatchPAT 400) give breathing and oxygen level results that agree with the standard overnight sleep study (polysomnography)? How reliably can two doctors, each looking at only one test's results, arrive at similar conclusions about a child's sleep apnea?

Read more

Participants will:

Have their regular, already-scheduled overnight sleep study performed exactly as usual Wear the WatchPAT 400 device (a small sensor on one finger, a wrist unit, and one adhesive patch on the chest) at the same time, during that same night Have both sets of results reviewed separately by two different study doctors

Researchers will compare the results from the WatchPAT 400 device to the results from the standard sleep study to see how closely they agree, and how well the new device performs at detecting sleep apnea in children and teenagers.

Read the detailed description

This is a prospective, single-center, blinded, within-subject diagnostic-accuracy pilot study. Each participant undergoes simultaneous overnight recording with both attended, technician-monitored polysomnography (PSG) - the clinically indicated standard-of-care test they are already scheduled for - and the WatchPAT 400 device, on the same single night. This within-subject design was chosen specifically to eliminate night-to-night physiological variability as a confounder in the agreement analysis, a limitation common to studies that compare sequential rather than simultaneous recordings.

Obstructive sleep apnea (OSA) affects an estimated 1-6% of children and is associated with meaningful health consequences if left undiagnosed and untreated, including behavioral and cognitive effects (attention, learning, hyperactivity), cardiovascular strain, and impaired growth. Because current diagnostic capacity is constrained by the resource intensity of PSG, many children with symptoms suggestive of OSA experience substantial delays before a formal diagnostic sleep study can be obtained, during which these consequences may accrue. A validated, less burdensome diagnostic alternative could meaningfully shorten that diagnostic gap.

OSA in children is conventionally diagnosed using attended, in-laboratory PSG, which requires a multi-sensor montage including EEG, EOG, chin/leg EMG, nasal airflow sensors, thoracoabdominal respiratory belts, and pulse oximetry. While PSG remains the diagnostic reference standard, it is resource-intensive, requires specialized pediatric sleep laboratory capacity that is not universally available, and can be poorly tolerated by some children due to the number and placement of sensors. Peripheral arterial tonometry (PAT)-based home sleep apnea testing has emerged as a substantially less burdensome alternative, using far fewer sensors, but has historically lacked adequate validation in pediatric populations. This study investigates whether the newest-generation PAT device can close that evidence gap.

The WatchPAT 400 derives its outputs from three primary physiological signals: a peripheral arterial tonometry (PAT) signal measured at the finger, which reflects sympathetic nervous system activity and vasomotor tone; pulse oximetry, which reflects peripheral oxygen saturation; and actigraphy, which distinguishes sleep from wake states and detects body position. An additional chest-worn sensor measures respiratory effort, snoring intensity, and body position. The device's proprietary algorithm combines these signals to automatically generate a PAT-derived apnea-hypopnea index (pAHI), respiratory disturbance index (pRDI), estimated sleep staging, oxygen desaturation metrics, hypoxic burden (a measure of the cumulative depth and duration of oxygen desaturation over the recording, distinct from simple event counts), and, in this newest device generation, a central apnea-hypopnea index (pAHIc) and automated flagging of candidate cardiac arrhythmia/premature-beat events. This algorithmic processing is a fixed, FDA-cleared feature of the commercial device and is not modified or developed by the study investigators; its output is independently reviewed by a blinded study physician rather than accepted as a final result.

The WatchPAT 400 builds on an earlier-generation device platform (WatchPAT 300, FDA 510(k) K180775) that has an existing, though limited, body of pediatric and adolescent validation literature reporting reasonable correlation with PSG-derived respiratory indices in patients generally 12 years of age and older. The WatchPAT 400 itself, however, was only FDA-cleared in September 2025 (510(k) K250460), and its newly added features - the central apnea index and hypoxic burden reporting in particular - have essentially no published pediatric validation data at the time this study was designed. This pilot is intended to generate some of the first such data.

Outcome interpretation in this study is blinded and bidirectional: one sleep physician scores the PSG without access to the WatchPAT recording or its automated report, while a second sleep physician independently scores the WatchPAT recording without access to the PSG results. This separation is intended to prevent either physician's interpretation from being influenced by the comparator, which is a common source of bias in prior device-validation literature.

Because most children referred for snoring do not, in fact, have clinically significant OSA, enrollment is enriched using a validated parent-reported symptom questionnaire (OSA-18) score threshold indicating at least moderate probability of pediatric OSA. This threshold was deliberately set to preserve a clinically meaningful spread of underlying OSA severity (rather than restricting to only the highest-probability referrals), so that the resulting cohort is expected to include participants with no OSA, mild OSA, moderate OSA, and severe OSA on PSG. This approach is intended to reduce the spectrum-bias risk common to diagnostic-accuracy pilot studies that enroll only clearly symptomatic, high-probability cases.

The study's statistical approach reflects its explicit framing as a Phase 1 pilot: the primary analysis (intraclass correlation coefficient with 95% confidence interval, and Bland-Altman limits of agreement between PSG-derived AHI and WatchPAT-derived pAHI) is powered to generate a preliminary but not definitive estimate of agreement, along with the population's actual OSA prevalence and device technical-failure rate. These pilot estimates are intended to inform the sample size of a subsequent, adequately powered Phase 2 confirmatory study (anticipated enrollment of approximately 100-150 participants), rather than to establish definitive diagnostic-accuracy claims on their own.

Secondary analyses include sensitivity, specificity, positive predictive value, and negative predictive value at standard AHI severity thresholds; Cohen's weighted kappa for inter-rater agreement on OSA severity category between the two blinded physicians; and an exploratory analysis of the device's central apnea index, which is analyzed within FDA labeling only for the subset of participants aged 17 and older, and as an explicitly investigational, off-label endpoint for younger participants.

Because the WatchPAT 400 automatically flags candidate cardiac arrhythmia/premature-beat events - a feature not present in the prior device generation, and one the manufacturer explicitly states is not intended as a cardiac diagnostic tool - any such flag generated during a study recording is reviewed by the principal investigator, who determines whether referral for further cardiac evaluation is clinically warranted, independent of and in addition to the participant's usual clinical care. The principal investigator also reviews accumulating safety and feasibility data at defined enrollment intervals throughout the study and has established criteria for suspending enrollment if safety or technical-performance concerns arise.

This is an investigator-initiated study, conducted without external industry sponsorship. Because WatchPAT 400 is used outside its FDA-cleared age indication for part of this study population, and its central apnea parameter is used outside labeling for participants younger than 17, this study is being conducted under a Non-Significant Risk (NSR) device determination per 21 CFR 812.2(b), with device results used for research comparison purposes only and not to guide any participant's clinical care. If this pilot demonstrates acceptable reliability, findings will directly inform the design of a larger confirmatory study and the longer-term goal of expanding validated, well-tolerated home sleep testing options for children, including populations who currently have particular difficulty completing conventional PSG.

Being tested
WatchPAT 400
Keywords
Peripheral arterial tonometry, Home sleep apnea testing, Pediatric sleep study, Apnea-hypopnea index, Central sleep apnea. Inter-rater reliability
Medical subject headings
Sleep Apnea, Obstructive

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • WatchPAT 400Device

    A peripheral arterial tonometry (PAT)-based home sleep apnea testing device (Itamar Medical/ZOLL Itamar; FDA 510(k) K250460) consisting of a finger probe, wrist-worn recording unit, and adhesive chest sensor patch. The device is worn overnight, simultaneously with standard polysomnography, to generate an apnea-hypopnea index, respiratory disturbance index, sleep-stage estimates, and, in this device generation, a central apnea index and hypoxic burden measure, for comparison against polysomnography results.

Study groups in NCT07851402
Study groupRoleWhat they receive
WatchPAT 400 + Standard PSG (Simultaneous)ExperimentalAll enrolled participants receive this single study arm. During their already-scheduled, clinically indicated overnight polysomnogram (PSG), participants simultaneously wear the WatchPAT 400 device (finger probe, wrist unit, and adhesive chest sensor patch) for the same overnight recording. The PSG is scored by one study physician, blinded to the WatchPAT results, for standard clinical care; the WatchPAT recording is independently scored by a second study physician, blinded to the PSG results, for research comparison purposes only. No randomization, dose, or treatment assignment is involved.
Receives: Device: WatchPAT 400

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Agreement Between WatchPAT 400-Derived Apnea-Hypopnea Index (pAHI) and Polysomnography-Derived Apnea-Hypopnea Index (AHI)

    Agreement between the WatchPAT 400 pAHI and the polysomnography (PSG)-derived AHI, both obtained from the same simultaneous overnight recording, assessed using the intraclass correlation coefficient (two-way mixed effects, absolute agreement) with 95% confidence interval, and Bland-Altman limits of agreement (mean bias and 95% limits of agreement, in events/hour).

    Measured: Single overnight recording (one night)

Secondary outcomes (1)
  1. Sensitivity and Specificity of WatchPAT 400 for Detection of Obstructive Sleep Apne

    Sensitivity, specificity, positive predictive value, and negative predictive value of the WatchPAT 400 pAHI for detecting obstructive sleep apnea and classifying its severity (none, mild, moderate, severe), using polysomnography-derived AHI as the reference standard, calculated at standard clinical thresholds (AHI ≥1, ≥5, and ≥10 events/hour).

    Measured: Single overnight recording (one night)

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted1 Oct 2026
  2. Study start1 Dec 2026 (estimated)
  3. Primary completion30 Nov 2027 (estimated)
  4. Study completion1 Dec 2028 (estimated)

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Diagnosis
Allocation
Not applicable
Intervention model
Single group
Model description
Each participant undergoes simultaneous overnight recording with both the standard-of-care polysomnogram (PSG) and the investigational WatchPAT 400 device, worn together during the same single night. There is no randomization and no separate study arms; every participant receives both measurements concurrently, allowing each participant to serve as their own comparison and eliminating night-to-night variability as a confound in the agreement analysis.
Masking
None (open label)
Enrolment
35 participants (estimated)
IDs, dates and publications
ClinicalTrials.gov ID
NCT07851402
Sponsor's protocol ID
29948
Status last verified
Sep 2026
First submitted
25 Sep 2026
Last updated
5 Oct 2026
Results posted
No
Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
Yes
Individual data sharing
Yes: De-identified individual participant data underlying the primary and secondary outcome measures reported in the resulting publication (including AHI, pAHI, RDI, pAHIc, OSA severity category, and relevant demographic/anthropometric variables) may be shared with qualified researchers upon request.

Questions about this trial

What is this clinical trial studying?

A study testing WatchPAT 400 in people with Pediatric Sleep Disordered Breathing and Obstructive Sleep Apnea, run by Milton S. Hershey Medical Center at 1 site in United States.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 7 inclusion and 8 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: opening soon, as of the sponsor's last update on 5 Oct 2026. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in United States. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

This study is still under way. Results are not available in the source record yet.

How can a doctor refer a patient or contact the research team?

Contact Ahmet Tanritanir, MD (+17175318977, atanritanir@pennstatehealth.psu.edu), or a site near the patient. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Agreement Between WatchPAT 400-Derived Apnea-Hypopnea Index (pAHI) and Polysomnography-Derived Apnea-Hypopnea Index (AHI) (time frame: Single overnight recording (one night)). It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 5 Oct 2026.