Evaluate the Performance of the Dexcom G7 Continuous Glucose Monitoring (CGM) System in Critically Ill Patients Undergoing Major Abdominal Surgery and Solid Organ Transplantation Which Require Blood Glucose Monitoring for Intensive Insulin Therapy.

A study testing Continuous glucose monitor in people with CGM and Intensive Care (ICU), run by Institute for Clinical and Experimental Medicine at 1 site in Czechia.

Recruiting

Looking for participants now. Contact the study team or a site below to ask about taking part.

Sites and contacts
Official record
Search all trials

Trial summary

Main objective: The goal is to to determine the accuracy of Dexcom G7 Continuous glucose monitor system in critically ill patients, undergoing major abdominal surgery and solid organ transplantation, in the intensive care unit. Read more about this study

Who is it for?
Ages 18 years and older
What is being tested?
Continuous glucose monitor
Could I receive a placebo?
Yes, some participants may receive a placebo
How many people
62 participants
Where
1 siteCzechia
Study dates
14 Jan 2026to Apr 2027
Phase
Interventional
Are there results?
Not yet: the study is under way

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 years and older
  • All sexes
  • No healthy volunteers
  • Adult, Older adult

Who may be able to join

  • Age 18 years or over
  • Intensive care unit admission
  • Requiring blood glucose monitoring for insulin therapy

Who cannot join

  • High-dose ascorbic acid or acetaminophen > 4g/day
  • Physical or mental health condition precluding continuous wear of the continuous glucose monitor, as judged by the physician

Trial locations and contacts

1 site in 1 country. Contact the study team or a site near you to ask about taking part.

Marek Protus, MD contact

+420602278062

marek.protus@ikem.cz

Alexander Macek, MD contact

+421910631876

alexander.macek@ikem.cz

Czechia1 site, 1 recruiting
  • Institute of Clinical and Experimental Medice

    Prague

    Marek Protus, MD, +420602278062, marek.protus@ikem.cz

    Alexander Macek, MD, +421910631876, alexander.macek@ikem.cz

    Recruiting

Other trials in Czechia

About this study

Official title: Determining the Reliability and Accuracy of the Dexcom G7 Continuous Glucose Monitoring System in the ICU

Main objective:

The goal is to to determine the accuracy of Dexcom G7 Continuous glucose monitor system in critically ill patients, undergoing major abdominal surgery and solid organ transplantation, in the intensive care unit.

Read more

On admission to the intensive care unit or at the end of surgery, two Dexcom G7 sensors are placed in the infraclavicular position.

Sensor number 1 will be blinded and will only be active when paired with the receiver, without the ability to monitor glucose levels.

Sensor number 2 is used for continuous glucose monitoring and can be used to adjust the insulin dose.

Patients will wear the continuous glucose monitoring sensor for a maximum of 10 days, including the time spent in the intensive care unit and the immediate care unit. The study will not interfere with any usual care clinical workflows.

Calibration is only performed on an unblinded sensor. The reference method for calibration is the glucose from the blood glucose meter.

Calibrate on the first day of monitoring: 4 times a day after approximately 6 hours, or if the blood glucose reading from the blood glucose meter is not within the estimated error range of the continuous glucose monitor (difference greater than 1,5mmol/L).

Calibration on days 2 and 3 of monitoring: once in the morning, or when the meter's glucose reading is not within the estimated error range of the continuous glucose monitoring (difference greater than 1,5mmol/L) On subsequent days when the blood glucose meter reading is not within the estimated error range of continuous glucose monitoring (difference greater than 1,5mmol/L).

Hypothesis:

  • Ho: No difference in mean absolute relative difference (MARD) between calibrated and non-calibrated sensors.
  • Ha: Calibrated sensors have lower mean absolute relative difference (δ = 4%, SD = 5%).

Further objectives:

To evaluate further aspects regarding the Dexcom G7 continuous glucose monitor system's performance in the intensive care unit (ICU), particularly in terms of sensitivity and specificity to detect hypo- and hyperglycaemic events, data availability, device deficiencies and safety.

To identify predictive factors of the accuracy of the Dexcom G7 continuous glucose monitor system in critically ill patients in the ICU.

Read the detailed description

Design:

  • Each intensive care unit patient wears two sensors (calibrated + non-calibrated).
  • 4 glucose measurements/day × 6 days = 24 measurements/subject.
  • Sequential Organ Failure Assessment (SOFA) score recorded daily.
Being tested
Continuous glucose monitor
Medical subject headings
Diabetes Mellitus

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Continuous glucose monitorDevice

    Continuous measurement of glucose from subcutaneous tissue

Study groups in NCT07611721
Study groupRoleWhat they receive
Open labeled CGMActive comparatorAdditional calibrations performed on CGM device during monitoring period
Receives: Device: Continuous glucose monitor
Control labeled CGMSham comparatorNo additional calibrations performed on CGM device during monitoring period
Receives: Device: Continuous glucose monitor

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Calculate overall mean absolute relative difference between arms

    Analyse the difference in mean absolute relative difference (MARD) of Dexcom G7 continuous glucose monitor between calibrated and uncalibrated sensor, in patients undergoing major abdominal surgery and solid organ transplantation.

    Measured: From enrolment to discharge from ICU care or up-to day 10

Secondary outcomes (1)
  1. Accuracy metrics

    To evaluate further aspects regarding the Dexcom G7 continuous glucose monitor system's performance in the intensive care unit (ICU), particularly in terms of sensitivity and specificity to detect hypo- and hyperglycaemic events, data availability, device deficiencies and safety.

    To identify predictive factors of the accuracy of the Dexcom G7 continuous glucose monitor system in critically ill patients in the ICU.

    Measured: From enrolment to discharge from ICU care or up-to day 10

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted28 May 2026
  2. Study start14 Jan 2026
  3. Primary completionSep 2026 (estimated)
  4. Study completionApr 2027 (estimated)

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Device feasibility
Allocation
Non-randomised
Intervention model
Single group
Masking
Triple blind
Who is masked
Participant, Care provider, Investigator
Enrolment
62 participants (estimated)
IDs, dates and publications
ClinicalTrials.gov ID
NCT07611721
Sponsor's protocol ID
G-25-71
Other IDs
ID Project No. LX22NPO5104
Acronym
DREAM-ICU
Status last verified
Apr 2026
First submitted
14 May 2026
Last updated
28 May 2026
Results posted
No
Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
No

Questions about this trial

What is this clinical trial studying?

A study testing Continuous glucose monitor in people with CGM and Intensive Care (ICU), run by Institute for Clinical and Experimental Medicine at 1 site in Czechia.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 3 inclusion and 2 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: looking for participants now, as of the sponsor's last update on 28 May 2026. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in Czechia. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. Some participants in this study may receive a placebo. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

This study is still under way. Results are not available in the source record yet.

How can a doctor refer a patient or contact the research team?

Contact Marek Protus, MD (+420602278062, marek.protus@ikem.cz), or a site near the patient. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Calculate overall mean absolute relative difference between arms (time frame: From enrolment to discharge from ICU care or up-to day 10). It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 28 May 2026.

Free account

Keep searching with a free account

You’ve used your 3 free searches for this session. Log in or create a free AzTrials account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in