Trial summary

This project addresses a key safety gap in High-Dose-Rate (HDR) brachytherapy, where very high, localized radiation doses mean that small geometric deviations in catheter placement or radioactive source positioning can cause clinically significant errors. Read more about this study

Who is it for?
Ages 18 years and older
Conditions the study is for
Prostate Cancer, Gynecologic Cancers
What is being tested?
Brachytherapy treatment monitoring using a plastic scintillation detector
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
25 participants
Where
No sites listed yet
Study dates
15 Jan 2027to 1 May 2028
Phase
Interventional

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 years and older
  • All sexes
  • No healthy volunteers
  • Adult, Older adult

Who may be able to join

  • Patients receiving prostate brachytherapy treatment, or female user receiving gynecological brachytherapy treatment for cervical cancer.

Who cannot join

  • Patients with a Clostridium difficile infection or any other high-risk positive infection.

Trial locations and contacts

No sites are listed yet. Contact the study team or a site near you to ask about taking part.

About this study

Official title: In Vivo Surveillance of Brachytherapy Using Plastic Scintillation Dosimetry: A Clinical Feasibility Study

This project addresses a key safety gap in High-Dose-Rate (HDR) brachytherapy, where very high, localized radiation doses mean that small geometric deviations in catheter placement or radioactive source positioning can cause clinically significant errors. Incident reports from the Radiation Oncology Incident Learning System (RO-ILS) of the American Society for Radiation Oncology (ASTRO) and guidelines from the American Association of Physicists in Medicine Task Group 59 (AAPM TG-59) identify these events-often occurring between imaging and treatment delivery-as among the most critical. Unlike external beam radiotherapy, brachytherapy lacks a prior phantom-based verification step, making in vivo dosimetry the only real-time method to confirm delivered dose and detect anomalies in dwell time or source position. International bodies such as the International Atomic Energy Agency (IAEA) recommend this approach for quality management and accident prevention.

Multipoint plastic scintillation detectors (mPSD) offer water-equivalent, real-time measurements and can enable three-dimensional (3D) radioactive source tracking, but their clinical use remains limited due to integration challenges and the absence of well-defined alert thresholds.

Read more

The project aims to implement a real-time in vivo monitoring solution using an mPSD at the Centre régional intégré de cancérologie (CRIC) of Lévis, Québec (QC) for HDR gynecological and prostate brachytherapy. Objectives include: (1) demonstrating clinical feasibility and integration within existing clinical workflows; (2) validating the accuracy and precision of dose, dwell-time, and source-position measurements under clinical conditions; and (3) developing practical alert thresholds to identify common deviations. The central hypothesis is that mPSD technology can provide reliable, reproducible real-time measurements without lengthening procedures or disrupting clinical workflow.

The methodology relies on adapting the HYPERSCINT® platform for HDR in vivo use. A custom mPSD, consisting of several scintillators along a single optical fiber, will be insertable into a 6-French (6F) catheter.

Phase 1 focuses on feasibility and integration, including HDR-specific calibration, definition of detector insertion scenarios, integration of acquisition hardware, and staff training.

Phase 2 validates performance in at least 20 patients. Planned treatment parameters and time-stamped detector signals acquired at 1-10 hertz (Hz) will be used to reconstruct delivered dose, dwell times, and source-to-detector distances. Measurement accuracy and reproducibility will be quantified using mixed-effects statistical models.

Phase 3 establishes preliminary alert thresholds based on deviations in dose, time, and position, along with visualization tools such as Bland-Altman plots, although these thresholds will not be applied clinically during the study.

Expected outcomes include demonstrated feasibility of in vivo mPSD monitoring, quantitative performance metrics with confidence intervals, identification of influential factors and operational recommendations, preliminary alert thresholds with reporting templates, and a roadmap for sustainable deployment in Lévis and other centers. Scientific outputs will also be produced for dissemination across the Santé Québec network.

Being tested
Brachytherapy treatment monitoring using a plastic scintillation detector
Medical subject headings
Prostatic Neoplasms

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Brachytherapy treatment monitoring using a plastic scintillation detectorDevice

    Typical brachytherapy treatments are conducted without dose or source position monitoring in real-time. This intervention will assess the feasibility of using a scintillation detector to that end.

Study groups in NCT07580651
Study groupRoleWhat they receive
In-vivo dosimetryExperimentalPatients in this treatment arm will receive the same treatment and care as patients outside of the protocol, but will have their treatment monitored by a plastic scintillation detector placed inside or at close proximity of the tumor.
Receives: Device: Brachytherapy treatment monitoring using a plastic scintillation detector

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Proportion of brachytherapy treatments with successful scintillation detector-based measurement

    Feasibility will be assessed by determining whether a scintillation detector successfully acquires a measurement during brachytherapy treatment. Radiation emitted from the brachytherapy source induces optical scintillation in the detector. The emitted light is transmitted via optical fiber and collected by a spectrometer located outside the treatment room. A successful measurement is defined as the acquisition of scintillation data that can be analyzed to infer clinically relevant parameters, including either the radiation dose rate as a function of time and/or the radioactive source position during treatment delivery. The outcome will be reported as the number and proportion of treatments with successful measurements.

    Measured: From the patient anesthesia at the beginning of the brachytherapy procedure to the end of brachytherapy radiation delivery on the same day.

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted12 May 2026
  2. Study start15 Jan 2027 (estimated)
  3. Primary completion31 Dec 2027 (estimated)
  4. Study completion1 May 2028 (estimated)

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Other
Allocation
Not applicable
Intervention model
Single group
Masking
None (open label)
Enrolment
25 participants (estimated)
IDs, dates and publications
ClinicalTrials.gov ID
NCT07580651
Sponsor's protocol ID
2027-1305
Other IDs
2025
Acronym
(CurieInVivo)
Status last verified
Oct 2026
First submitted
31 Mar 2026
Last updated
21 Oct 2026
Results posted
No
Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
No
Individual data sharing
No: Sharing of individual participant data, under Quebec Law, is heavily regulated, especially for data shared outside of the Province. For regulatory simplicity, the investigators would prefer not sharing the individual participant data.

Publications

  • Fonseca GP, Johansen JG, Smith RL, Beaulieu L, Beddar S, Kertzscher G, Verhaegen F, Tanderup K. In vivo dosimetry in brachytherapy: Requirements and future directions for research, development, and clinical practice. Phys Imaging Radiat Oncol. 2020 Sep 28;16:1-11. doi: 10.1016/j.phro.2020.09.002. eCollection 2020 Oct.

    PubMed 33458336
  • Linares Rosales HM, Duguay-Drouin P, Archambault L, Beddar S, Beaulieu L. Optimization of a multipoint plastic scintillator dosimeter for high dose rate brachytherapy. Med Phys. 2019 May;46(5):2412-2421. doi: 10.1002/mp.13498. Epub 2019 Apr 2.

    PubMed 30891803
  • Linares Rosales HM, Johansen JG, Kertzscher G, Tanderup K, Beaulieu L, Beddar S. 3D source tracking and error detection in HDR using two independent scintillator dosimetry systems. Med Phys. 2021 May;48(5):2095-2107. doi: 10.1002/mp.14607. Epub 2021 Mar 28.

    PubMed 33222208
  • Therriault-Proulx F, Beddar S, Beaulieu L. On the use of a single-fiber multipoint plastic scintillation detector for 192Ir high-dose-rate brachytherapy. Med Phys. 2013 Jun;40(6):062101. doi: 10.1118/1.4803510.

    PubMed 23718599
  • Kubo HD, Glasgow GP, Pethel TD, Thomadsen BR, Williamson JF. High dose-rate brachytherapy treatment delivery: report of the AAPM Radiation Therapy Committee Task Group No. 59. Med Phys. 1998 Apr;25(4):375-403. doi: 10.1118/1.598232.

    PubMed 9571605
  • Hoopes DJ, Dicker AP, Eads NL, Ezzell GA, Fraass BA, Kwiatkowski TM, Lash K, Patton GA, Piotrowski T, Tomlinson C, Ford EC. RO-ILS: Radiation Oncology Incident Learning System: A report from the first year of experience. Pract Radiat Oncol. 2015 Sep-Oct;5(5):312-318. doi: 10.1016/j.prro.2015.06.009. Epub 2015 Jun 25.

    PubMed 26362705
  • Therriault-Proulx F, Archambault L, Beaulieu L, Beddar S. Development of a novel multi-point plastic scintillation detector with a single optical transmission line for radiation dose measurement. Phys Med Biol. 2012 Nov 7;57(21):7147-59. doi: 10.1088/0031-9155/57/21/7147. Epub 2012 Oct 12.

    PubMed 23060069
  • Beaulieu L, Beddar S. Review of plastic and liquid scintillation dosimetry for photon, electron, and proton therapy. Phys Med Biol. 2016 Oct 21;61(20):R305-R343. doi: 10.1088/0031-9155/61/20/R305. Epub 2016 Oct 3.

    PubMed 27694714
  • Yamada Y, Rogers L, Demanes DJ, Morton G, Prestidge BR, Pouliot J, Cohen GN, Zaider M, Ghilezan M, Hsu IC; American Brachytherapy Society. American Brachytherapy Society consensus guidelines for high-dose-rate prostate brachytherapy. Brachytherapy. 2012 Jan-Feb;11(1):20-32. doi: 10.1016/j.brachy.2011.09.008.

    PubMed 22265435

Free account

Keep searching with a free account

You’ve used your 3 free searches for this session. Log in or create a free AzTrials account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in