Accuracy of Smartwatches in Measuring Oxygen Levels in Patients With Pulmonary Hypertension: A Pilot Study

An observational study of Pulmonary Hypertension, run by Golden Jubilee National Hospital at 1 site in United Kingdom.

Recruiting

Looking for participants now. Contact the study team or a site below to ask about taking part.

Sites and contacts
Official record
Search all trials

Trial summary

The goal of this study is to learn how accurate smartwatches are at measuring oxygen levels in patients with a condition called pulmonary hypertension. The main questions it aims to answer are: How accurate the smartwatches are at measuring oxygen levels when the patient is sitting at rest How accurate are… Read more about this study

Who is it for?
Ages 18 years and older
Conditions the study is for
Pulmonary Hypertension
How many people
20 participants
Where
1 siteUnited Kingdom
Study dates
17 Dec 2025to Dec 2026
Phase
Observational
Are there results?
Not yet: the study is under way

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 years and older
  • All sexes
  • No healthy volunteers
  • Adult, Older adult

Study population: Patients attending outpatient clinic at the Golden Jubilee National Hospital under the care of the Scottish Pulmonary Vascular Unit.

Who may be able to join

  • Aged 18-years old and above.
  • Diagnosed with pulmonary hypertension by right heart catheterisation showing a baseline mean pulmonary artery pressure > 20mmHg, PVR >2 Woods Units.
  • Able to perform a six-minute walk test (6MWT)
  • Able to give written informed consent.

Who cannot join

  • Peripheral arterial disease
  • Tattoos or skin markings which cover the dorsal wrist.
  • Resting SpO2 <88%
  • Severe concurrent medical condition that would prevent participation in study procedures or with life expectancy <3 months
  • Patients with bleeding disorders
  • Patients on anticoagulation
  • Except in circumstances where anticoagulation is given solely for historical perceived long term survival benefit for pulmonary arterial hypertension. This patient group could be given the option to come off their anticoagulation for 1 week prior to the study visit if they wished to take part in the study. It could then be restarted the day following the study visit providing there were no bleeding complications related to the arterial line.

Trial locations and contacts

1 site in 1 country. Contact the study team or a site near you to ask about taking part.

Martin Johnson, BA, MBChB(Hons), MD, FRCP contact

(0044) 0141 951 5497

martin.johnson@nhs.scot

Jamie Ingram, MBChB, BMSc, PGCert contact

(0044) 0141 951 5497

jamie.ingram2@nhs.scot

United Kingdom1 site, 1 recruiting
  • Golden Jubilee National Hospital

    Glasgow, G81 4DY

    Martin Johnson, BA, MBChB(Hons), MD, FRCP, (0044) 0141 951 5497, martin.johnson@nhs.scot

    Jamie Ingram, MBChB, BMSc, PGCert, (0044) 0141 951 5497, jamie.ingram2@nhs.scot

    Recruiting

Other trials in United Kingdom

About this study

Official title: Accuracy of Wrist-worn Pulse Oximeter SpO2 Measurements in Patients Diagnosed With Pulmonary Hypertension

The goal of this study is to learn how accurate smartwatches are at measuring oxygen levels in patients with a condition called pulmonary hypertension. The main questions it aims to answer are:

  • How accurate the smartwatches are at measuring oxygen levels when the patient is sitting at rest
  • How accurate are smartwatches at measuring oxygen levels after exercise
  • How accurate are smartwatches at measuring oxygen levels after breathing a lower level of oxygen through a mask.
Read more

Researchers will compare oxygen levels measured through a smartwatch with those checked through a finger probe oxygen monitor and also from a blood sample checked from the artery in the wrist to see if smartwatch oxygen measurements are similar.

Participants will:

-Attend the hospital once, just for a few hours to collect all the required data.

Read the detailed description

The goal of this pilot study is assess the accuracy of SpO2 measurements in smartwatches when compared to arterial oxygen levels and pulse oximeter readings at rest, following exercise and at simulated high altitude with the use of a hypoxic challenge test.

Patient recruitment:

Patients will be recruited prospectively from the Scottish Pulmonary Vascular Unit at the Golden Jubilee National Hospital, Glasgow, Scotland.

Patients attending outpatient clinic will be screened for eligibility. Additionally, a study poster will be displayed in the clinic waiting area. If eligible, patients will be asked if they would like to learn about the study. After discussing the study overview in person, the patient will be offered the patient information sheet to take away with them and advised that if they are interested in taking part or have any further questions about the study, they can either reply using the opt in form provided or by phoning the study team. Those patients who opt in to the trial will be contacted to provide them with a date to attend for a study visit.

On the day of the study visit, prior to commencing any study activities, participants will see a different doctor from the research team who will clarify their full understanding of the study and will give them the opportunity to complete the written study consent form.

Procedure:

During a study visit, participants will have an arterial line inserted into their radial artery by an experienced intensive care unit doctor to allow easy access to arterial blood sampling. Each paired sample will include: smartwatch oxygen saturations, arterial oxygen saturations, finger probe pulse oximeter measurements taken at the same time. Paired samples will be taken at baseline (while the patient is in a seated, rested state), post-exercise in the recovery period after a six-minute walk test and in simulated high altitude during a hypoxic challenge.

The six minute walk test will comprise shuttles of 25 metres continuously for 6 minutes. 10 paired samples will be taken during the recovery period after a six-minute walk test. The hypoxic challenge will be undertaken by asking the patient to breathe through a 40% Venturi mask driven by 100% nitrogen at a flow rate of 10L/min which will deliver an inhaled oxygen level of 15%. If SpO2 falls below 85% for over 30 seconds the test will be stopped and if required, inhaled oxygen will be delivered to correct hypoxia. The participant will breathe the air/nitrogen mixture for 20 minutes unless the procedure is aborted due to hypoxia prior to that. 10 paired samples will be taken during the hypoxic challenge test. Following completion the arterial line will be removed and pressure will be applied until any bleeding stops. Participants will be monitored for 30 minutes prior to them leaving.

Data collection/management:

Each study participant will be assigned a study number which will be used on all study documentation to avoid using patient identifiable information. All data will be stored on an NHS password encrypted computer. Data will be collected on pseudonymised data capture sheets with a study ID number and all research measurements will be transferred to an Excel spreadsheet for analysis.

Sample size:

The study is a cross sectional study of validity requiring one study visit per participant. Regarding sample size, this is a pilot study to gain initial estimates of the mean bias and limits of agreement of smartwatch measurements compared to the gold standard arterial oxygen levels. The planned sample size of 20 is not based on a sample size calculation, and is considered sufficient to obtain initial estimates, and to allow planning of larger studies if required. The investigators will aim to recruit 20 patients and will recruit up to a maximum of 25 in total if required in the case of incomplete data sets.

Statistical analysis:

The bias of smartwatches and standard pulse oximeter oxygen saturation measurements relative to the gold standard, arterial blood gas, will be compared using Bland-Altman analysis.

Plan for missing data:

In the event that an arterial sample cannot be analysed, the smartwatch and fingerprobe pulse oximeter samples will still be noted and compared. The investigators will aim to obtain 10 complete paired samples per patient after exercise and also 10 full sets during the hypoxic challenge. The investigators will aim to recruit 20 patients but will recruit up to 25 patients as required to achieve complete data sets. In the event that a smartwatch reading is not obtained during a paired sample, this will be documented as 'no measurement obtained'. This will still represent a complete paired sample as inability to obtain a measurement is useful to note as an outcome. In the event that the fingerprobe pulse oximeter is unable to obtain a reading the data will be documented as 'no measurement obtained' but will not require any increased requirement for recruitment as long as smartwatch and arterial gas measurements have been acquired as a paired sample.

Keywords
smartwatch, pulmonary hypertension, oxygen saturations, SpO2
Medical subject headings
Hypertension, Pulmonary

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

Study groups in NCT07311135
Study groupRoleWhat they receive
Adult patients with pulmonary hypertensionPre-capillary pulmonary hypertension, able to complete a six-minute walk test

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Agreement of oxygen saturations (SpO2) measured by the wrist-worn pulse oximeter when compared with arterial blood gas oxygen saturations (SaO2) at rest

    SpO2 (measured by the wrist-worn pulse oximeter (smartwatch) and standard fingerprobe pulse oximeter) and SaO2 (measured by the arterial blood gas analysis) measured as a percentage.

    Measured: 1 study visit, anticipated to be around 3 hours.

Secondary outcomes (2)
  1. 1. Agreement of oxygen saturations (SpO2) measured by the wrist-worn pulse oximeter when compared with arterial blood gas oxygen saturations (SaO2) after exercise

    SpO2 (measured by the wrist-worn pulse oximeter (smartwatch) and standard fingerprobe pulse oximeter) and SaO2 (measured by arterial blood gas analysis) measured as a percentage after exercise

    Measured: 1 study visit, estimated 3 hours duration

  2. 1. Agreement of oxygen saturations (SpO2) measured by the wrist-worn pulse oximeter when compared with arterial blood gas oxygen saturations (SaO2) after a hypoxic challenge

    SpO2 (measured by the wrist-worn pulse oximeter (smartwatch) and standard fingerprobe pulse oximeter) and SaO2 (measured by arterial blood gas analysis) measured as a percentage, after a hypoxic challenge test.

    Measured: 1 study visit, estimated 3 hours duration

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted30 Dec 2025
  2. Study start17 Dec 2025
  3. Primary completionDec 2026 (estimated)
  4. Study completionDec 2026 (estimated)

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Observational
Observational model
Other
Time perspective
Cross-sectional
Enrolment
20 participants (estimated)
IDs, dates and publications
ClinicalTrials.gov ID
NCT07311135
Sponsor's protocol ID
25/CARD/03
Status last verified
Feb 2026
First submitted
16 Dec 2025
Last updated
17 Feb 2026
Results posted
No
Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
No
Individual data sharing
No

Publications

  • Spaccarotella C, Polimeni A, Mancuso C, Pelaia G, Esposito G, Indolfi C. Assessment of Non-Invasive Measurements of Oxygen Saturation and Heart Rate with an Apple Smartwatch: Comparison with a Standard Pulse Oximeter. J Clin Med. 2022 Mar 8;11(6):1467. doi: 10.3390/jcm11061467.

    PubMed 35329793
  • Pipek LZ, Nascimento RFV, Acencio MMP, Teixeira LR. Comparison of SpO2 and heart rate values on Apple Watch and conventional commercial oximeters devices in patients with lung disease. Sci Rep. 2021 Sep 23;11(1):18901. doi: 10.1038/s41598-021-98453-3.

    PubMed 34556765
  • Arslan B, Sener K, Guven R, Kapci M, Korkut S, Sutasir MN, Tekindal MA. Accuracy of the Apple Watch in measuring oxygen saturation: comparison with pulse oximetry and ABG. Ir J Med Sci. 2024 Feb;193(1):477-483. doi: 10.1007/s11845-023-03456-w. Epub 2023 Jul 13.

    PubMed 37440093

Questions about this trial

What is this clinical trial studying?

An observational study of Pulmonary Hypertension, run by Golden Jubilee National Hospital at 1 site in United Kingdom.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 4 inclusion and 7 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: looking for participants now, as of the sponsor's last update on 17 Feb 2026. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in United Kingdom. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

This study is still under way. Results are not available in the source record yet.

How can a doctor refer a patient or contact the research team?

Contact Martin Johnson, BA, MBChB(Hons), MD, FRCP ((0044) 0141 951 5497, martin.johnson@nhs.scot), or a site near the patient. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Agreement of oxygen saturations (SpO2) measured by the wrist-worn pulse oximeter when compared with arterial blood gas oxygen saturations (SaO2) at rest (time frame: 1 study visit, anticipated to be around 3 hours.). It is what the study was designed to measure, not a finding.

Where can I read the original study information?

On the ClinicalTrials.gov record for NCT07311135, which also lists 3 linked publications.

Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 17 Feb 2026.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in