Diagnostic Accuracy of Carebot AI MMG in Mammography Screening: Multicenter MRMC Study

An observational study of Breast Neoplasms and Breast Cancer Detection, run by Carebot s.r.o. at 4 sites in 2 countries.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

This study evaluates the diagnostic performance of Carebot AI MMG, an artificial intelligence (AI)-enabled medical device for evaluating mammograms. The software analyzes standard full-field digital mammography (FFDM) images and classifies each examination as having no suspicious finding ("Low Risk"), a probably… Read more about this study

Who is it for?
Ages 18 years and older, women onlyHealthy volunteers can take part
What is being tested?
Carebot AI MMG software analysis
Type of treatment
Medical device
How many people
222 participants
Where
4 sitesin 2 countries
Study dates
1 Jan 2025to 3 Nov 2025
Phase
Observational
Are there results?
None posted in the registry

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 years and older
  • Women only
  • Healthy volunteers accepted
  • Adult, Older adult

Study population: Women undergoing routine screening mammography at four participating centers in Central Europe. The analytical dataset consists of a retrospectively assembled case-control subset of 222 anonymized 2D FFDM examinations (48 malignant and 174 non-malignant) selected from a source cohort of 4,729 screening studies.

Who may be able to join

  • Female sex
  • Age ≥ 18 years at the time of the screening mammogram
  • Screening full-field digital mammography (FFDM) examination with all four standard views (LCC, RCC, LMLO, RMLO) available
  • Sufficient image quality and complete DICOM metadata to allow retrospective analysis

Who cannot join

  • Male sex
  • Age < 18 years
  • Digital breast tomosynthesis (DBT/3D) examinations without a corresponding full 2D FFDM four-view series
  • Incomplete mammography series (missing one or more of LCC, RCC, LMLO, RMLO)
  • Corrupted or unreadable DICOM files
  • Missing or inconsistent key metadata (e.g., laterality, view, acquisition date)

Trial locations and contacts

4 sites in 2 countries.

4 sites in 2 countries

Slovakia3 sites
  • Dolnooravská nemocnica s poliklinikou MUDr. L. N. Jégého

    Dolný Kubín, 026 14

  • Nemocnica s poliklinikou Považská Bystrica

    Považská Bystrica, 017 01

  • Ľubovnianska nemocnica

    Stará Ľubovňa, 064 01

Other trials in Slovakia

Czechia1 site
  • Poliklinika MEDICON Budějovická

    Prague, 14000

Other trials in Czechia

About this study

Official title: Retrospective Multicenter Multi-Reader, Multi-Case Diagnostic Accuracy Study of Carebot AI MMG Compared With Radiologists on 2D Full-Field Digital Mammography in Breast Cancer Screening

This study evaluates the diagnostic performance of Carebot AI MMG, an artificial intelligence (AI)-enabled medical device for evaluating mammograms. The software analyzes standard full-field digital mammography (FFDM) images and classifies each examination as having no suspicious finding ("Low Risk"), a probably benign mass ("Medium Risk"), or a suspicious malignant mass ("High Risk").

The study is retrospective and observational. It uses anonymized mammography examinations from four screening centers, without any additional imaging or contact with patients. Three experienced breast radiologists independently read the same set of cases, and their assessments are used as the human benchmark. A histopathology-based reference standard, supplemented by radiologist consensus and follow-up information for negative cases, is used to determine whether cancer is present.

The main goal is to compare the AI system with human radiologists in terms of sensitivity and specificity for detecting breast cancer, and to assess whether the AI can achieve non-inferior performance at two predefined operating points: one favoring higher sensitivity and negative predictive value (rule-out) and one favoring higher specificity and positive predictive value (rule-in).

Read the detailed description

Design and setting This is a retrospective, multicenter, multi-reader, multi-case (MRMC) diagnostic accuracy study of Carebot AI MMG, conducted on anonymized 2D full-field digital mammography (FFDM) examinations acquired as part of routine breast cancer screening. Mammograms were collected from four screening centers over a defined time period. No additional imaging was performed for the purpose of this study, and no subjects were contacted.

Data source and population The source dataset consists of 4,729 screening mammography examinations from women aged 32 to 88 years (mean approximately 57 years). Only 2D FFDM studies with a complete set of standard projections (LCC, RCC, LMLO, RMLO) were included. Examinations with incomplete series, unreadable or corrupted DICOM files, or missing/inconsistent key metadata were excluded, as were tomosynthesis (DBT) studies, men, and women under 18 years of age. To ensure sufficient precision of performance estimates, the dataset was enriched with additional biopsy-proven cancers. The final analytical subset comprises 222 examinations, including 48 malignant and 174 non-malignant studies, with representation across three mammography devices (Hologic Selenia Dimensions, Hologic Lorad Selenia, Fujifilm FDR-3000AWS).

Investigational device and comparator The investigational device is Carebot AI MMG (software version 2.9, deep-learning models v2.3), a stand-alone AI system that analyzes 2D FFDM exams and outputs a case-level classification into three categories, together with an internal risk score. Two predefined operating points are evaluated: a high-sensitivity (HSe) threshold, where both benign and malignant masses are treated as "positive" (rule-out setting), and a high-specificity (HSp) threshold, where only malignant masses are counted as "positive" (rule-in setting).

As a human comparator, three experienced radiologists (RAD 1-3) independently read the same anonymized studies using a dedicated DICOM viewer integrated with a labeling application. Radiologists were blinded to AI outputs, clinical information, and outcomes and recorded a case-level classification into the same three categories (Negative/Benign/Malignant). For primary analyses, their binary decisions are derived using the same HSe and HSp rules. In addition, a "random reader" benchmark is constructed for balanced accuracy by repeatedly sampling one radiologist's decision per case in a bootstrap framework (20,000 iterations).

Reference standard The reference standard is established at the study level. A case is labeled "Malignant" if there is histopathological confirmation of breast cancer from biopsy performed in temporal association with the index mammogram. A case is labeled "Non-malignant" if there is consensus between two local radiologists that the finding is negative or stably benign, typically corroborated by at least 2 years of imaging follow-up. Tumor staging (e.g., TNM) is not used in the present analysis. All 48 malignant cases from the participating centers are included in the analytical subset; no cancer-positive examinations were excluded.

Objectives and endpoints The primary objectives are: (1) to demonstrate that the balanced accuracy (BA) of Carebot AI MMG is at least 0.80 in both HSe and HSp operating points; (2) to demonstrate non-inferiority of the AI's balanced accuracy compared with the MRMC "random reader" benchmark with a non-inferiority margin of 0.05; and (3) to demonstrate non-inferiority of sensitivity (Se) of AI versus each of the three radiologists in both HSe and HSp, with a non-inferiority margin of 0.07. Secondary objectives are to describe specificity (Sp), positive predictive value (PPV), negative predictive value (NPV), and to characterize patterns of false-negative and false-positive decisions and their potential implications for clinical risk management.

Statistical analysis Diagnostic performance metrics (Se, Sp, PPV, NPV, BA) are calculated at the case level for Carebot AI MMG and each radiologist in both HSe and HSp. Wilson 95% confidence intervals are used for proportions. Paired McNemar tests are used to compare AI and individual readers in terms of Se and Sp. For balanced accuracy, a MRMC bootstrap procedure with 20,000 iterations is used to construct the distribution of the random reader and to estimate the probability that BA_AI is greater than or equal to BA_random minus the pre-specified margin. Non-inferiority in sensitivity is assessed using a Nam-Blackwelder-type framework on discordant pairs. False-negative and false-positive cases are reviewed qualitatively with emphasis on lesion conspicuity, breast density, and typical error patterns (e.g., dense parenchyma, lesions near the pectoral muscle, benign vascular structures, asymmetries).

Risk, ethics, and data protection The study is non-interventional and entirely retrospective. All mammography examinations were acquired as part of routine care before the study, and all DICOM data were irreversibly anonymized at the site level in compliance with GDPR and applicable national law before transfer to the sponsor. No additional radiation exposure or patient contact occurs, and no adverse events are expected. Given this design, the study does not meet the MDR definition of a clinical investigation under Article 62 and is not subject to prior notification under Article 74(1); individual informed consent is not required. Results are intended to support the clinical evaluation of Carebot AI MMG as a decision-support tool in organized mammography screening.

Being tested
Carebot AI MMG software analysis
Keywords
Breast cancer screening, Mammography, Full-field digital mammography, Artificial intelligence, Deep learning, Diagnostic accuracy, Carebot AI MMG
Medical subject headings
Breast Neoplasms

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Carebot AI MMG software analysisDevice

    Retrospective stand-alone AI analysis of anonymized 2D full-field digital mammography (FFDM) examinations. The AI system (Carebot AI MMG, version 2.9) processes existing images and outputs case-level risk classifications; no additional imaging, randomization, or changes to patient management occur as part of this study.

Study groups in NCT07298252
Study groupRoleWhat they receive
Malignant casesWomen with biopsy-proven breast cancer included in the analytical subset (n = 48). Each case corresponds to a screening full-field digital mammography (FFDM) examination with all four standard views (LCC, RCC, LMLO, RMLO), retrospectively identified from participating screening centers.
Receives: Device: Carebot AI MMG software analysis
Non-malignant casesWomen without histopathological evidence of breast cancer, classified as negative or stably benign by two independent local radiologists with at least 2 years of imaging follow-up (n = 174). Each case corresponds to a screening FFDM examination with all four standard views (LCC, RCC, LMLO, RMLO), retrospectively selected from the same screening population.
Receives: Device: Carebot AI MMG software analysis

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Balanced accuracy (BA) of Carebot AI MMG for detecting malignant versus non-malignant examinations

    Balanced accuracy (BA) is defined as the average of sensitivity and specificity for classifying each mammography examination as malignant or non-malignant. BA will be calculated at two pre-specified operating points of the AI system: a high-sensitivity (HSe) setting and a high-specificity (HSp) setting. Performance will be estimated with 95% confidence intervals and compared to a multi-reader benchmark constructed from three experienced radiologists and a bootstrap-based "random reader" reference.

    Measured: Baseline (index mammography examination; examinations acquired between 01-01-2025 and 14-11-2025; retrospective assessment

  2. Sensitivity (Se) of Carebot AI MMG versus histopathology-based reference standard

    Sensitivity is defined as the proportion of malignant examinations correctly classified as positive by the AI system. Sensitivity will be calculated at both the HSe and HSp operating points and compared pairwise with the sensitivity of each of the three radiologists using the same case-level ground truth.

    Measured: Baseline (index mammography examination; examinations acquired between 01-01-2025 and 14-11-2025; retrospective assessment

Secondary outcomes (3)
  1. Specificity (Sp) of Carebot AI MMG versus histopathology-based reference standard

    Specificity is defined as the proportion of non-malignant examinations correctly classified as negative by the AI system. Specificity will be calculated at both the HSe and HSp operating points and compared pairwise with the specificity of each of the three radiologists.

    Measured: Baseline (index mammography examination; examinations acquired between 01-01-2025 and 14-11-2025; retrospective assessment

  2. Positive predictive value (PPV) of Carebot AI MMG

    Positive predictive value is defined as the proportion of AI-positive examinations that are truly malignant according to the reference standard. PPV will be reported for both HSe and HSp operating points and compared descriptively with PPV values for each radiologist.

    Measured: Baseline (index mammography examination; examinations acquired between 01-01-2025 and 14-11-2025; retrospective assessment

  3. Negative predictive value (NPV) of Carebot AI MMG

    Negative predictive value is defined as the proportion of AI-negative examinations that are truly non-malignant according to the reference standard. NPV will be reported for both HSe and HSp operating points and compared descriptively with NPV values for each radiologist.

    Measured: Baseline (index mammography examination; examinations acquired between 01-01-2025 and 14-11-2025; retrospective assessment

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted23 Dec 2025
  2. Study start1 Jan 2025
  3. Primary completion3 Nov 2025
  4. Study completion3 Nov 2025

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Observational
Observational model
Case-control
Time perspective
Retrospective
Enrolment
222 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT07298252
Sponsor's protocol ID
CB-MMG-02-MC
Other IDs
CB-MMG-02-MC
Acronym
CARE-MMG-MRMC
Status last verified
Jan 2026
First submitted
14 Nov 2025
Last updated
14 Jan 2026
Results posted
No
Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
No
Individual data sharing
No: This is a retrospective, multicenter diagnostic accuracy study using anonymized full-field digital mammography (FFDM) images and associated metadata obtained under local data use agreements. Individual-level imaging data are not planned to be shared outside the participating institutions due to contractual, privacy, and regulatory constraints. Aggregate, de-identified summary results (including performance metrics and key subgroup analyses) may be shared in publications and upon reasonable request, but no IPD repository is planned.

Questions about this trial

What is this clinical trial studying?

An observational study of Breast Neoplasms and Breast Cancer Detection, run by Carebot s.r.o. at 4 sites in 2 countries.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 4 inclusion and 6 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 14 Jan 2026. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 4 sites in Slovakia, Czechia. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Balanced accuracy (BA) of Carebot AI MMG for detecting malignant versus non-malignant examinations (time frame: Baseline (index mammography examination; examinations acquired between 01-01-2025 and 14-11-2025; retrospective assessment). The study lists 2 primary outcomes in all. It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 14 Jan 2026.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in