Blood PressurePPG ABPMNCT07240831

Accuracy of a Cuffless Photoplethysmography (PPG) Chest-Patch Monitor for 24-hour Ambulatory Blood Pressure Monitoring (ABPM)

A study testing Model BB-613WP in people with Blood Pressure and Heart Disease, run by Biobeat Technologies Ltd. at 11 sites in 3 countries.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

This study aims to validate the accuracy and reliability of blood pressure (BP) estimates obtained over 24 hours using a PPG-based chest-patch device compared to the gold standard ambulatory blood pressure monitoring (ABPM) method using an upper arm cuff-based oscillometric BP device, in both hypertensive and… Read more about this study

Who is it for?
Ages 18 years and older
Conditions the study is for
Blood Pressure, Heart Disease, Hypertension (HTN)
What is being tested?
Model BB-613WP
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
612 participants
Where
11 sitesin 3 countries
Study dates
15 Dec 2025to 2 Sep 2026
Phase
Interventional
Are there results?
None posted in the registry

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 years and older
  • All sexes
  • No healthy volunteers
  • Adult, Older adult

Who may be able to join

  • Females and Males subjects 18 years or older (from all skin tones/colors)
  • Scheduled for ABPM for hypertension screening, evaluation, or management
  • Subjects capable of using an application temporarily downloaded (for the duration of the study) on a personal smart phone device or on a device provided by the study sponsor
  • Subjects with Atrial fibrillation or Atrial flutter are eligible to participate if standard cuff-based ABPM monitor can capture blood pressure at initialization

Who cannot join

  • Indication for ABPM is orthostatic hypotension or Postural Orthostatic Tachycardiac Syndrome as indicated on the ABPM referral order
  • Any neurological, psychological, or other condition impairing the ability to provide informed consent or participate in the study
  • Inability to reliably access the Internet via a smart phone
  • Not able to perform standard cuff-based ABPM
  • Average absolute difference between the two arms is greater than 15 mmHg for Systolic Blood Pressure (SBP) or greater than 10 mmHg Diastolic Blood Pressure (DBP) during the initialization/calibration phase
  • Subjects with known allergy to adhesives or skin patches
Show all 9Show fewer
  • Subjects with solid dark tattoos on the area where the PPG-based device is attached
  • Subjects who are unwilling to have their chest hair shaved, if needed, to install the device (the device uses adhesives and needs to be in direct contact with the skin without hair)
  • Subjects who are unwilling to comply with the study instructions (avoid vigorous exercise and showering for 24 hours while the device is in place)

Trial locations and contacts

11 sites in 3 countries.

11 sites in 3 countries

United States6 sites
  • Cardiology, P.C.

    Birmingham, Alabama, 35211

  • Henry Ford Health System

    Detroit, Michigan, 48202-2689

  • Mayo Clinic

    Jacksonville, Florida, 32224

  • U Health-University of Miami Health System

    Miami, Florida, 33136

  • Mayo Clinic

    Rochester, Minnesota, 55905

  • Mayo Clinic

    Scottsdale, Arizona, 85259

Other trials in United States

Israel4 sites
  • HaEmek Hospital

    Afula, 1834111

  • The Holy Family Hospital Nazareth

    Nazareth, 1641100

  • Sheba Medical Center

    Ramat Gan

  • Maccabi Healthcare

    Tel Aviv

Other trials in Israel

Italy1 site
  • Instituto Auxologico Italiano

    Milan

Other trials in Italy

About this study

Official title: Accuracy of a Cuffless Photoplethysmography (PPG) Chest-Patch Monitor for 24-hour Ambulatory Blood Pressure Monitoring

This study aims to validate the accuracy and reliability of blood pressure (BP) estimates obtained over 24 hours using a PPG-based chest-patch device compared to the gold standard ambulatory blood pressure monitoring (ABPM) method using an upper arm cuff-based oscillometric BP device, in both hypertensive and normotensive individuals referred by their provider to undergo a 24-hours ABPM for clinical indication. The Awake/Asleep test, which is the primary test recommended for automated wearable cuffless BP devices that are cuff-calibrated (based on the 2023 European Society of Hypertension (ESH) recommendations for the validation of cuffless blood pressure measuring devices), will be conducted in this study. The secondary aim of the study is to assess the feasibility and convenience of the PPG-based device.

Read the detailed description

This study aims to evaluate the accuracy, validity, and efficacy of the Biobeat PPG-based cuffless chest-worn monitor for continuous and ambulatory blood pressure monitoring compared with a standard cuff-based ambulatory blood pressure monitor (ABPM). The study will follow the 2023 European Society of Hypertension (ESH) protocol for validating cuffless BP devices. The Biobeat device is a single-use, wireless, noninvasive patch applied to the chest that continuously measures blood pressure and other vital signs using photoplethysmography (PPG) technology.

The trial is a prospective, multicenter, single-arm validation study in adults undergoing 24-hour ABPM for any clinical indication. Participants will simultaneously wear both the Biobeat device and a standard ABPM for 24 hours. The Biobeat device will be calibrated with three reference cuff measurements at the start of monitoring. Data from both devices will be compared across awake, asleep, and total 24-hour periods to determine agreement in systolic and diastolic blood pressure.

The primary objective is to demonstrate equivalence between the Biobeat PPG-based system and standard ABPM, with accuracy defined as a mean difference ≤5 mmHg and standard deviation ≤8 mmHg. Secondary endpoints include assessment of device performance across different blood pressure ranges and demographic subgroups, as well as participant comfort and usability. The study represents minimal risk to participants; the only potential adverse effect is mild skin irritation from the adhesive patch.

Successful validation will confirm that the Biobeat cuffless PPG-based device provides accurate, reliable 24-hour blood pressure monitoring comparable to existing ABPM systems, supporting its use as a comfortable, patient-friendly alternative for both clinical and ambulatory settings.

Being tested
Model BB-613WP
Keywords
BLOOD PRESSURE, BLOOD PRESSURE MONITORING, Hypertension (HTN), CUFFLESS PHOTOPLETHYSMOGRAPHY, PPG, CUFFLESS, vital signs, ABPM, 24 hour ABPM, Ambulatory Blood Pressure Monitoring, Ambulatory Blood Pressure, BP
Medical subject headings
Heart Diseases, Hypertension

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Model BB-613WPDevice

    Also called Biobeat Chest-Patch, PPg-based monitor

    During the study period, participants will undergo standard ABPM monitoring per clinical indication and wear a PPG-based monitor

Study groups in NCT07240831
Study groupRoleWhat they receive
Device: Non-invasive monitoringExperimentalThe study is designed as a prospective, multicenter, single arm, validation study that follows the 2023 European Society of Hypertension (ESH) recommendations for the validation of cuffless blood pressure measuring devices.27 There is no primary hypothesis. The study will validate 24-hours blood pressure monitoring using the investigational FDA-cleared device by comparing the level of agreement with parallel measurements obtained by the gold-standard upper arm. The Awake/Asleep test is the first and primary test recommended for this type of device to assess validity.
Receives: Device: Model BB-613WP

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Primary Endpoint

    The accuracy and reliability of the PPG-based chest-patch device will be evaluated by comparing its blood pressure measurements with those obtained through the gold-standard upper arm cuff oscillometric ABPM method. Both systolic and diastolic blood pressure values will be assessed.

    Measured: 24 hours monitoring

Secondary outcomes (2)
  1. Secondary Endpoint

    Comparison of the compliance and completeness of the 2 measurement methods. The investigators will report the percentage of participants with at least 30% missed data (defining non-compliance or incomplete study).

    Measured: 24 hours

  2. Secondary Endpoint

    Convenience of the device (participants' questionnaire) will be assessed.

    Participants will rate the experience on a scale of 1 to 5, with 5 indicating a higher satisfaction score for the following questions:

    1. Comfort level during placement of the monitor
    2. Comfort level during monitoring
    3. Interference with sleep
    4. Interference with daytime activity (i.e., dressing, ambulating, eating)
    5. General satisfaction with the monitoring Additional questions-
    6. Which diagnostic option would you recommend to a friend / family member? Options- Biobeat, cuff, either is fine. Answering "Biobeat" will be considered a positive response.
    7. In the future, if you undergo monitoring with Biobeat- how would you prefer to receive the monitor? Options- Mailed directly home, applied at the clinic. Either response is acceptable

    Measured: 24 hours

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted21 Nov 2025
  2. Study start15 Dec 2025
  3. Primary completion2 Sep 2026
  4. Study completion2 Sep 2026

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Diagnosis
Allocation
Not applicable
Intervention model
Single group
Masking
None (open label)
Enrolment
612 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT07240831
Sponsor's protocol ID
Biobeat_CLP_001
Acronym
PPG ABPM
Status last verified
Sep 2026
First submitted
24 Sep 2025
Last updated
16 Sep 2026
Results posted
No
Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
Yes
Individual data sharing
No: This is proprietary company information.

Questions about this trial

What is this clinical trial studying?

A study testing Model BB-613WP in people with Blood Pressure and Heart Disease, run by Biobeat Technologies Ltd. at 11 sites in 3 countries.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 4 inclusion and 9 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 16 Sep 2026. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 11 sites in United States, Israel, Italy. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Primary Endpoint (time frame: 24 hours monitoring). It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 16 Sep 2026.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in