AzTrials
AzTrials

PregnancyNCT07223996

Use of In-Home Non-Stress Monitoring Device for Fetal Monitoring: A Nonsignificant Risk (NSR) Device Study

A study testing PregnaOne System in people with Pregnancy and Fetal Monitoring, run by Nestmedic Spolka Akcyjna at 3 sites in United States.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

This study evaluates the safety and performance of the PregnaOne System, a non-significant risk (NSR) investigational medical device for in-home fetal monitoring in pregnant women during the third trimester. Read more about this study

Who is it for?
Ages 22 years and older, women onlyHealthy volunteers can take part
Conditions the study is for
Pregnancy, Fetal Monitoring
What is being tested?
PregnaOne System
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
71 participants
Where
3 sitesUnited States
Study dates
30 Sep 2025to 30 Jan 2026
Phase
Interventional
Are there results?
Yes, posted in the registrySummary in the doctors' view

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 22 years and older
  • Women only
  • Healthy volunteers accepted
  • Adult, Older adult

Who may be able to join

  • Able to independently read and understand written and spoken English and willing to provide written informed consent and comply with all instructions required by the study protocol.
  • Female aged 22 years or older.
  • Estimated gestational age between 32 and 41 weeks.
  • Singleton gestation.
  • Pre-pregnancy or first prenatal visit BMI < 40 kg/m².
  • Able and willing to undergo fetal monitoring sessions in a simulated home environment at the study site.
  • Women who do not meet the gestational age criterion at the screening visit but will meet it by the study visit (within two weeks) may be enrolled.

Who cannot join

  • In active labor or in the delivery room during labor.
  • Requires hospitalization or life support.
  • Any contraindications for cardiotocography (CTG) monitoring.
  • Severe hypertension (≥160/110 mmHg measured twice, 15 minutes apart).
  • Known allergy to latex, ultrasound (US) gel, or ECG gel.
  • Skin conditions (e.g., edema, erythema, irritation, infection, lesions, or open wounds) at contact points for the Pregnabit Pro or comparator devices.
  • Implanted electronic devices (e.g., pacemaker, stimulator, defibrillator, pump).
  • Members of vulnerable populations other than pregnant women (e.g., minors, wards of the state, cognitively impaired individuals, prisoners, or institutionalized persons).

Trial locations and contacts

3 sites in 1 country.

United States3 sites
  • Christie Clinic, LLC

    Champaign, Illinois, 61820

  • The Iowa Clinic, P.C.

    West Des Moines, Iowa, 50266

  • Wilmington Health, PLLC

    Wilmington, North Carolina, 28401

Other trials in United States

About this study

This study evaluates the safety and performance of the PregnaOne System, a non-significant risk (NSR) investigational medical device for in-home fetal monitoring in pregnant women during the third trimester. The PregnaOne System consists of the Pregnabit Pro device and related software, designed to record fetal heart rate, maternal pulse, and uterine contractions in a home-like environment. The study compares the performance of the PregnaOne System with a similar FDA-cleared at-home monitoring system (the INVU system by Nuvo). Approximately 70 pregnant women will participate at three medical centers in the United States. Each participant will complete one training session and two fetal monitoring sessions during a single study visit.

Read the detailed description

This is a prospective, single-visit, non-randomized, comparative study designed to evaluate the PregnaOne System, an investigational Class II medical device intended for in-home fetal monitoring. The system includes the Pregnabit Pro device and associated software that collect and transmit data on fetal heart rate, fetal movement, maternal pulse rate, and uterine muscle activity for remote review by healthcare professionals.

The goal of this study is to assess the safety, usability, and signal performance of the PregnaOne System compared to the INVU system, an FDA-cleared home fetal monitoring device. The study will be conducted under a non-significant risk (NSR) designation.

Approximately 70 pregnant women in their third trimester will be enrolled at three U.S. medical centers. Each participant will attend one in-clinic visit simulating an at-home environment. The study visit includes:

Training session - participants are instructed on how to use the PregnaOne System.

First monitoring session - a 30-minute fetal monitoring period using both the PregnaOne System and the INVU system simultaneously.

Second monitoring session - a 30-minute self-conducted monitoring session using only the PregnaOne System without staff assistance.

During each session, participants remain in a seated or reclined position and record fetal movements. Study staff will be available to assist and ensure participant safety.

No known risks are associated with fetal heart rate monitoring, though mild discomfort from device straps may occur.

Participants receive compensation for their time and travel. Study data will support the submission of a 510(k) premarket notification to the U.S. Food and Drug Administration (FDA).

Being tested
PregnaOne System
Keywords
fetal heart rate monitoring, non-stress test, prenatal care, telemedicine, maternal health, home monitoring, PregnaOne System

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • PregnaOne SystemDevice

    Also called Pregnabit Pro, PregnaOne Platform

    The PregnaOne System is a non-significant risk (NSR) investigational medical device designed for in-home fetal monitoring. It includes the Pregnabit Pro device and the cloud-based PregnaOne Platform software. The system records fetal heart rate, maternal pulse, and uterine activity, transmitting data to a remote monitoring center for clinical evaluation.

    During the study, participants will perform two monitoring sessions:

    1. Simultaneous use of the PregnaOne and INVU systems.
    2. Independent monitoring using only the PregnaOne System. The study evaluates the usability, safety, and performance of the PregnaOne System in comparison with a reference device.
Study groups in NCT07223996
Study groupRoleWhat they receive
PregnaOne SystemExperimentalParticipants will use the PregnaOne System for fetal monitoring during a single study visit. Each participant will complete two monitoring sessions: 1. Simultaneous use of the PregnaOne System and the INVU system. 2. Self-monitoring using only the PregnaOne System. The study aims to evaluate the usability, safety, and performance of the PregnaOne System for non-stress fetal monitoring in a home-like setting.
Receives: Device: PregnaOne System

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Agreement Between PregnaOne System and Comparator Device (INVU System) for Fetal Heart Rate (fHR) Measurements

    Agreement between fetal heart rate (fHR) measurements obtained with the PregnaOne System and the comparator device was assessed using Bland-Altman analysis. The reported values are the mean difference (PregnaOne minus INVU) and the lower and upper 95% limits of agreement, each presented with its corresponding 95% confidence interval.

    Measured: At single study visit (Day 1)

  2. Positive Percent Agreement Between PregnaOne System and Comparator Device (INVU System) for Uterine Activity Events

    Positive percent agreement (PPA) for uterine activity events was calculated as the proportion of uterine activity events identified by the comparator device (INVU) that were also identified by the PregnaOne System, using the prespecified locked event-detection and matching algorithm. A two-sided bias-corrected and accelerated (BCa) 95% confidence interval was calculated.

    Measured: At single study visit (Day 1)

  3. Percent of Participants Who Successfully Complete a Self-Monitoring Session

    To assess the ability of participants to successfully complete a self-monitoring session in a simulated home environment. Successful self-monitoring was evaluated using two components: successful device placement as determined by an HCP and successful acquisition of a quality recording of fHR, uterine muscle activity, and mPR as determined by an HCP. Results for both components are reported separately.

    Measured: At single study visit (Day 1)

Secondary outcomes (3)
  1. Percent of Participants Who Obtain a Quality Maternal Pulse Rate (mPR) Recording

    Percentage of participants whose PregnaOne recording session contained maternal pulse rate (mPR) values in at least 90% of recordings during simultaneous monitoring. This binary criterion was used to define a quality mPR recording for the final analysis.

    Measured: At single study visit (Day 1)

  2. Percent of Participants Who Successfully Place the Device

    To determine the percentage of participants who successfully place the PregnaOne device as determined by a healthcare professional (HCP) during self-monitoring.

    Measured: At single study visit (Day 1)

  3. Percent of Participants Who Obtain a Quality Recording of fHR, Uterine Muscle Activity, and mPR During Self-Monitoring

    To determine the percentage of participants who obtain a quality recording of fetal heart rate (fHR), uterine muscle activity, and maternal pulse rate (mPR) during self-monitoring, as assessed by a healthcare professional (HCP).

    Measured: At single study visit (Day 1)

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted3 Nov 2025
  2. Study start30 Sep 2025
  3. Primary completion30 Jan 2026
  4. Study completion30 Jan 2026

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Other
Allocation
Not applicable
Intervention model
Single group
Model description
Participants will complete one study visit including two monitoring sessions: (1) simultaneous use of the PregnaOne System and INVU system, and (2) self-monitoring using only the PregnaOne System. No randomization or parallel assignment is applied.
Masking
None (open label)
Enrolment
71 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT07223996
Sponsor's protocol ID
004
Status last verified
Oct 2026
First submitted
24 Oct 2025
Last updated
22 Oct 2026
Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
Yes
Individual data sharing
No

Questions about this trial

What is this clinical trial studying?

A study testing PregnaOne System in people with Pregnancy and Fetal Monitoring, run by Nestmedic Spolka Akcyjna at 3 sites in United States.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 7 inclusion and 8 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 22 Oct 2026. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 3 sites in United States. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

Yes. Results are posted in the registry (first posted 22 Oct 2026). A summary is in the For doctors view, and the full results are on ClinicalTrials.gov.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Agreement Between PregnaOne System and Comparator Device (INVU System) for Fetal Heart Rate (fHR) Measurements (time frame: At single study visit (Day 1)). The study lists 3 primary outcomes in all. It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 22 Oct 2026.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in