No Specific Medical Conditions or Disease StatesNCT07223931

Validating Dozee for Contactless Monitoring of Heart, Breathing, and Movemen

An observational study of No Specific Medical Conditions or Disease States and No Specific Conditions, run by Turtle Shell Technologies Pvt. Ltd. at 1 site in United States.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

The Dozee VS system (K221555) is a contactless device that tracks heart rate and breathing rate using a sensor sheet placed under the mattress. It captures body vibrations and converts them into vital signs, which can be monitored remotely in real time. Read more about this study

Who is it for?
Ages 18 to 100 yearsHealthy volunteers can take part
What is being tested?
Device: Telemetry monitoring
Type of treatment
Medical device
How many people
91 participants
Where
1 siteUnited States
Study dates
16 Sep 2025to 18 Dec 2025
Phase
Observational
Are there results?
None posted in the registry

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 to 100 years
  • All sexes
  • Healthy volunteers accepted
  • Adult, Older adult

Study population: The study will be conducted on atleast 60 subjects in the age group above 18 years regardless of gender. The patients/volunteers be informed about the study and a written informed consent form will be obtained from them. Dozee will be placed under the mattress of the patient's bed and the data will be collected through the device whenever a patient is on the bed.

Who may be able to join

  • Subjects will be included if they meet the following inclusion criteria:
  • Adults Male or female. (Above 18 years of age)
  • Provide written informed consent.
  • Weight between 40kg/88.18lbs and 120kg/264.55lbs
  • Agree to not eat during the testing period
  • Able to move and exit the bed independently and safely.

Who cannot join

  • Subjects will be excluded if they have any of the following inclusion criteria:
  • Are connected to a device that may interfere with the device monitoring in this study.
  • Are receiving any bedside care which may be incompatible with the study procedures.
  • A likely need to receive or undergo a procedure during the testing period.
  • Cannot accept a nasal cannula, or have an ECG leads placed on the chest.
  • Have a significant medical condition in the judgement of the investigator, which may compromise the study testing procedures.

Trial locations and contacts

1 site in 1 country.

United States1 site
  • ERA Health Research

    Midland, Texas, 79707

Other trials in United States

About this study

Official title: Clinical Validation of the Dozee System for Contactless Monitoring of Heart Rate, Respiration Rate, Motion, and Bed-Exits

The Dozee VS system (K221555) is a contactless device that tracks heart rate and breathing rate using a sensor sheet placed under the mattress. It captures body vibrations and converts them into vital signs, which can be monitored remotely in real time. The system also supports add-ons like a blood pressure cuff, pulse oximeter, and thermometer, offering a complete monitoring solution. This study aims to validate the updated AI-powered heart rate and improved breathing rate algorithms across different sensor sheet types to confirm clinical accuracy

Read the detailed description

The Dozee VS system (K221555), developed by Turtle Shell Technologies, is a non-invasive, contactless device that continuously monitors heart rate (HR) and breathing rate (RR). A thin sensor sheet placed under the mattress detects tiny body vibrations from the heart and lungs, which are processed by smart algorithms and displayed on a secure online dashboard for real-time monitoring.

Unlike traditional methods such as ECGs, nasal tubes, or thermistors-which are invasive, require skilled placement, and are uncomfortable for long-term use-Dozee is simple, comfortable, and doesn't rely on patient compliance like wearables do. The system can also connect to add-on devices, including a blood pressure cuff, pulse oximeter, and thermometer, giving a full picture of a patient's vital signs and helping spot early signs of deterioration.

Hardware upgrades now include dual-sensor sheets that improve accuracy across different mattresses and patient types. The new HR algorithm uses AI/ML for better accuracy, while the improved RR algorithm performs more reliably even with movement or weak signals.

Study Goal: This study aims to validate the updated algorithms (Gen 2) for HR and RR across the three sensor sheet types (DS, S1, and S1-I) to ensure they meet clinical accuracy standards and to benchmark against the previous versions of the algorithms

Being tested
Device: Telemetry monitoring

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Device: Telemetry monitoringDevice

    All participants will be on standard telemetry monitoring for Heart Rate, Respiration Rate and movements using the gold standard (Electrocardiography and Capnography respectively).

Study groups in NCT07223931
Study groupRoleWhat they receive
Single Arm Study (Cohort)Group/Cohort Description: All participants will be on standard telemetry monitoring for Heart Rate and Respiration Rate using the gold standard (Electrocardiography and Capnography respectively). In addition, a contactless monitoring devices will be placed on the patient bed to measure data. This device is Dozee VS (Investigational Device). Data from all devices will be measured simultaneously.
Receives: Device: Device: Telemetry monitoring

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Mean Absolute Error (MAE) of Dozee's Heart Rate a from the Gold Standard ECG

    Heart Rate collected at 1 Min intervals from gold standard and Dozee. Mean Absolute Error (MAE) of the Dozee Heart Rate will be calculated.

    Measured: 6 months

  2. Accuracy (%) of Dozee in identfying subject movement

    Time stamps of subject movement during the study will be recorded manullay. Time stamps of Dozee's movement will be recorded. Accuracy of the Dozee device in detecting movement compared to manual observation will be computed in terms of percentage.

    Measured: 6 months

  3. Accuracy(%) of of Dozee's bed-exits

    Time stamps of bed-exits will be recorded manually during the study will be recorded. Time stamps of Dozee's bed-exit will be computed. Accuracy of the Dozee device in detecting bed-exit compared to manual observation as gold standard will be computed in terms of percentage.

    Measured: 6 months

  4. Mean Absolute Error (MAE) of Dozee's Respiratory Rate from the Gold Standard Capnography

    Respiratory Rate collected at 1 Min intervals from gold standard and Dozee. Mean Absolute Error (MAE) of the Dozee Respiratory Rate will be calculated.

    Measured: 6 Months

Secondary outcomes (5)
  1. Range of values (minimum, maximum, and spread) of Heart Rate in beats per min measured by the gold standard and the Dozee

    The outcome will be assessed by comparing the range of values (minimum, maximum, and spread) of Heart Rate ( beats per minute) measured by the gold standard and the Dozee over the study period.

    Measured: 6 months

  2. Range of values (minimum, maximum, and spread) of Respiratory Rate in breaths per min measured by the gold standard and the Dozee

    The outcome will be assessed by comparing the range of values (minimum, maximum, and spread) of Respiratory Rate ( breaths per minute) measured by the gold standard and the Dozee over the study period.

    Measured: 6 Months

  3. Bland- Altman Level of Agreement between Dozee Heart Rate and Respiratory Rate with the Gold Standard

    Level of Agreement between Dozee's HR and RR with reference device computed by Bland Altman.

    Measured: 6 months

  4. Detection rate (%) of the Dozee's Heart Rate and Respiratory Rate

    The time-stamps of Dozee's Heart Rate and Respiratory Rate and the gold standard Heart Rate and Respiratory Rate will be recorded. DR is the proportion of gold standard points for which the DS successfully provides a corresponding measured value, relative to the total number of gold standard points.

    Measured: 6 months

  5. Level of Agreement between Dozee Heart Rate and Respiratory Rate with the Gold Standard

    Level of Agreement between Dozee Heart Rate and Respiratory Rate with the Gold Standard by using Deming Regression with Pearson Co-efficient

    Measured: 6 months

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted3 Nov 2025
  2. Study start16 Sep 2025
  3. Primary completion25 Nov 2025
  4. Study completion18 Dec 2025

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Observational
Observational model
Cohort
Time perspective
Prospective
Enrolment
91 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT07223931
Sponsor's protocol ID
DZ_VS_VAL
Status last verified
Jul 2026
First submitted
23 Sep 2025
Last updated
21 Sep 2026
Results posted
No
Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
Yes
Individual data sharing
No: Individual Participant Data (IPD) are protected under HIPPA Act of 1996.

Questions about this trial

What is this clinical trial studying?

An observational study of No Specific Medical Conditions or Disease States and No Specific Conditions, run by Turtle Shell Technologies Pvt. Ltd. at 1 site in United States.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 6 inclusion and 6 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 21 Sep 2026. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in United States. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Mean Absolute Error (MAE) of Dozee's Heart Rate a from the Gold Standard ECG (time frame: 6 months). The study lists 4 primary outcomes in all. It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 21 Sep 2026.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in