Cardiac PacingABPNCT07173777

Cardiac Anodal Biphasic Pacing

A study testing Cardiac Waveform Generator in people with Cardiac Pacing, run by Boston Medical Center at 1 site in United States.

Not yet recruiting

Opening soon. The study is approved but enrolment has not started yet. Sites may already take enquiries.

Sites and contacts
Official record
Search all trials

Trial summary

The goal of this study is to test a new pacing method called anodal biphasic pacing (ABP) to determine if this pacing works as well-or better-than current pacing methods. This new method may improve how the heart works and reduce some of the problems caused by regular pacing. Read more about this study

Who is it for?
Ages 18 years and older
Conditions the study is for
Cardiac Pacing
What is being tested?
Cardiac Waveform Generator
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
108 participants
Where
1 siteUnited States
Study dates
Dec 2026to Dec 2027
Phase
Interventional
Are there results?
Not yet: the study is under way

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 years and older
  • All sexes
  • No healthy volunteers
  • Adult, Older adult

Who may be able to join

  • Cohort A
  • Planned interventional cardiac procedure
  • Cohort B
  • Planned generator exchange of dual chamber cardiac implantable electronic device (CIED)
  • Functioning atrial lead
  • Cohort C
  • Planned de novo implant or generator exchange of CIED with cardiac resynchronization therapy
  • Functioning atrial lead

Who cannot join

  • Permanent atrial fibrillation
  • Third degree AV block without stable escape rhythm
  • Ischemic heart disease or coronary disease > 40%
  • Unable to receive heparin
  • Are not fluent in English
  • Unable to read in English
  • Not able to provide informed consent
  • Women who are pregnant

Trial locations and contacts

1 site in 1 country. Contact the study team or a site near you to ask about taking part.

Robert Helm, MD contact

617-638-8985

robert.helm@bmc.org

Denise Fine, BS contact

617-638-8716

denise.fine@bmc.org

United States1 site
  • Boston Medical Center

    Boston, Massachusetts, 02118

    Robert Helm, MD, robert.helm@bmc.org

    Denise Fine, BS, denise.fine@bmc.org

Other trials in United States

About this study

The goal of this study is to test a new pacing method called anodal biphasic pacing (ABP) to determine if this pacing works as well-or better-than current pacing methods. This new method may improve how the heart works and reduce some of the problems caused by regular pacing.

Current implantable pacemakers use a monophasic cathodal waveform to stimulate the heart. Monophasic cathodal pacing (MCP) waveforms slow conduction, impair contractility, cause inflammation, increase risk of atrial fibrillation, heart failure, and mortality. Anodal biphasic pacing (ABP) is an alternative waveform that can stimulate the heart. ABP preconditions the heart and then initiates cardiac contraction. ABP may address the limitations of MCP.

Read the detailed description

This study is a single-center, prospective, investigator-initiated, non-randomized, study that will investigate ABP in patients with structurally normal hearts and those with non-ischemic cardiomyopathy who are undergoing interventional cardiac procedure, generator exchange of dual chamber Cardiac implantable electronic device (CIED), or de novo implant or generator exchange of CIED with cardiac resynchronization therapy.

Eligible participants, without heart disease and those with nonischemic cardiomyopathy, undergoing CIED implant or generator exchange or interventional cardiac procedure at Boston Medical Center will be screened and prospectively enrolled. Participants will be stratified by left ventricular ejection fraction (EF): those with severely reduced EF (≤35%), mid-range EF (> 35%-49%) and normal EF (EF≥ 50%).

Primary efficacy objectives:

  1. To identify which patient populations have the greatest and most consistent hemodynamic benefit from ABP as compared with cathodal pacing.
  2. To confirm that ABP does not negatively impact non-responders.
  3. To define the ideal anodal biphasic waveform characteristics (amplitude and pulse width) that maximizes the positive effect in subjects who demonstrate hemodynamic improvement ABP
  4. To assess differences in capture thresholds between anodal biphasic and cathodal pacing.

Secondary safety objectives:

  1. To assess and characterize any ventricular arrhythmias associated with anodal biphasic in comparison to cathodal pacing.
  2. To assess device safety
  3. To assess procedure safety
Conditions
Cardiac Pacing
Being tested
Cardiac Waveform Generator
Keywords
Anodal biphasic pacing (ABP), Monophasic cathodal pacing (MCP), Ejection fraction, Cardiac implantable electronic device, Cardiac contractility modulation, Implantable cardiac defibrillator

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Cardiac Waveform GeneratorDevice

    A pacing device that allows for programmable pulse waveforms to generate a predefined set of anodal biphasic waveforms. It possesses a battery-powered floating point gate array (FPGA) using software that allows flexibility in waveform configuration.

Study groups in NCT07173777
Study groupRoleWhat they receive
Cohort AExperimentalCardiac patients who are undergoing interventional cardiac procedure including electrophysiology (EP) with planned retrograde left ventricular access or diagnostic coronary catheterization.
Receives: Device: Cardiac Waveform Generator
Cohort BExperimentalPatients who have pacing indication and are undergoing routine generator exchange of dual chamber cardiac implantable electronic device (CIED).
Receives: Device: Cardiac Waveform Generator
Cohort CExperimentalPatients who are undergoing new implant or generator exchange of CIED with cardiac resynchronization therapy.
Receives: Device: Cardiac Waveform Generator

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Clinically significant maximum rate of pressure change maximum rate of pressure change within the left ventricle during its contraction phase- dP/dtmax.

    Response will be expressed as a percent change in these measures with anodal biphasic pacing (ABP) as compared with cathodal pacing. A clinically significant hemodynamic response to pacing will be defined as a >10% increase in dP/dtmax.

    Measured: about 30 minutes

  2. Clinically significant stroke work

    Response will be expressed as a percent change in these measures with anodal biphasic pacing (ABP) as compared with cathodal pacing. A clinically significant hemodynamic response to pacing will be defined as a >10% increase in stroke work.

    Measured: about 30 minutes

  3. Clinically significant left ventricular end-diastolic pressure (LVEDP)

    Response will be expressed as a percent change in these measures with anodal biphasic pacing (ABP) as compared with cathodal pacing. A clinically significant hemodynamic response to pacing will be defined as a >10% increase in LVEDP.

    Measured: about 30 minutes

  4. Clinically significant diastolic relaxation (tau)

    Response will be expressed as a percent change in these measures with anodal biphasic pacing (ABP) as compared with cathodal pacing. A clinically significant hemodynamic response to pacing will be defined as a >10% increase in tau.

    Measured: about 30 minutes

  5. Clinically significant volume measurements

    Response will be expressed as a percent change in these measures with anodal biphasic pacing (ABP) as compared with cathodal pacing. A clinically significant hemodynamic response to pacing will be defined as a >10% increase in volume measurements.

    Measured: about 30 minutes

  6. Capture threshold

    This outcome will be measured with decremental pacing threshold testing where pacing output (voltage or pulse width) is decremented until there is a loss of ventricular capture. The minimum output prior to loss of capture is defined as the capture threshold.

    Measured: about 30 minutes

Secondary outcomes (3)
  1. Waveform safety concerns

    Defined as the development of any of the following with ABP: ventricular tachycardia > 3 beats, premature ventricular contractions at frequency greater than baseline, ventricular couplet, significant drop (>5%) in invasive hemodynamic measures or blood pressure; or any cardioversion for atrial or ventricular arrhythmia.

    Measured: about 30 minutes

  2. Device safety issues

    Defined as the occurrence of device related adverse event including: device malfunction, failure to output programmed pulse waveform, or failure to output set voltage

    Measured: about 30 minutes

  3. Procedural safety issues

    Defined as the occurrence of procedure related adverse event including: vascular complication, cardiac complication including cardiac perforation, valvular injury, atrioventricular (AV) or bundle branch block, or thromboembolism.

    Measured: about 30 minutes

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted15 Sep 2025
  2. Study startDec 2026 (estimated)
  3. Primary completionDec 2027 (estimated)
  4. Study completionDec 2027 (estimated)

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Diagnosis
Allocation
Non-randomised
Intervention model
Parallel groups
Masking
None (open label)
Enrolment
108 participants (estimated)
IDs, dates and publications
ClinicalTrials.gov ID
NCT07173777
Sponsor's protocol ID
H-46078
Acronym
ABP
Status last verified
Sep 2026
First submitted
8 Sep 2025
Last updated
29 Sep 2026
Results posted
No
Data monitoring committee
Yes
FDA-regulated drug
No
FDA-regulated device
Yes
Individual data sharing
No

Questions about this trial

What is this clinical trial studying?

A study testing Cardiac Waveform Generator in people with Cardiac Pacing, run by Boston Medical Center at 1 site in United States.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 8 inclusion and 8 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: opening soon, as of the sponsor's last update on 29 Sep 2026. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in United States. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

This study is still under way. Results are not available in the source record yet.

How can a doctor refer a patient or contact the research team?

Contact Robert Helm, MD (617-638-8985, robert.helm@bmc.org), or a site near the patient. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Clinically significant maximum rate of pressure change maximum rate of pressure change within the left ventricle during its contraction phase- dP/dtmax. (time frame: about 30 minutes). The study lists 6 primary outcomes in all. It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 29 Sep 2026.

Free account

Keep searching with a free account

You’ve used your 3 free searches for this session. Log in or create a free AzTrials account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in