Intraocular PressureNCT07152808
Clinical Study for TONOREF III Tested to ANSI Z80.10-2014
A study testing NIDEK TONOREF III and Haag-Streit Goldmann Manual Tonometer in people with Intraocular Pressure, run by Nidek Co. LTD..
Trial summary
The primary objective of this clinical study was to show that tonometry values for TONOREF III, as the representative device of the NIDEK tonometer series, were comparable to the predicate device. The secondary objective was to demonstrate that the test device was as safe as the predicate device. Read more about this study
- Who is it for?
- Ages 22 years and olderHealthy volunteers can take part
- Conditions the study is for
- Intraocular Pressure
- What is being tested?
- NIDEK TONOREF III and Haag-Streit Goldmann Manual Tonometer
- Type of treatment
- Medical device
- Could I receive a placebo?
- No, everyone receives an active treatment
- How many people
- 52 participants
- Where
- No sites listed yet
- Study dates
- 2 Sep 2022to 19 Jun 2023
- Phase
- Interventional
- Are there results?
- Yes, posted in the registrySummary in the doctors' view
Key requirements
Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.
- Ages 22 years and older
- All sexes
- Healthy volunteers accepted
- Adult, Older adult
Who may be able to join
- Each subject had to have:
- been at least 22 years of age of either sex and any race or ethnicity;
- been willing and able to provide written informed consent prior to any study procedures being performed;
- been willing and able to follow all instructions and attend all study visits;
- corneal astigmatism of ≤3D and corneal thickness of 500-600 μm in at least one eye.
Who cannot join
- Each subject had to not to :
- have only one functional eye;
- have poor or eccentric fixation in the study eye;
- have corneal scarring or have had corneal surgery, including corneal laser surgery in the study eye
- have microphthalmos in the study eye;
- have buphthalmos in the study eye;
Show all 15Show fewer
- be a contact lens wearer in the study eye;
- have dry eyes;
- be a lid squeezer - blepharospasm;
- have nystagmus in the study eye;
- have keratoconus in the study eye;
- have any other corneal or conjunctival pathology or infection in the study eye;
- use medication that affect IOP;
- current participation in other clinical trials;
- have a condition or a situation, which in the Investigator's opinion, may put the subject at increased risk, confound study data, or interfere significantly with the subject's study participation.
About this study
Official title: Clinical Pilot Study to Provide Research Data for TONOREF III Tested to ANSI Z80.10-2014
The primary objective of this clinical study was to show that tonometry values for TONOREF III, as the representative device of the NIDEK tonometer series, were comparable to the predicate device.
The secondary objective was to demonstrate that the test device was as safe as the predicate device.
- Conditions
- Intraocular Pressure
- Being tested
- NIDEK TONOREF III, Haag-Streit Goldmann Manual Tonometer
What participants receive
What participants receive. In a randomised study each person is assigned to one group by chance.
NIDEK TONOREF IIIDevice
The auto ref/kerato/tono/pachymeter TONOREF III is a medical device which measures objective refractive errors, corneal curvature radius, intraocular pressure and corneal thickness of the patient's eye.
Haag-Streit Goldmann Manual TonometerDevice
Haag-Streit Goldmann Manual Tonometer measures intraocular pressure to aid in the screening and diagnosis of glaucoma
| Study group | Role | What they receive |
|---|---|---|
| TONOREF III and Goldmann Tonometer Measurements | Experimental | Subjects will undergo IOP measurements using both the investigational device (TONOREF III) and the predicate device (Goldmann Applanation Tonometer). Receives: Device: NIDEK TONOREF III; Device: Haag-Streit Goldmann Manual Tonometer |
What is being measured
The primary outcome is the main question the study is designed to answer.
Agreement of Intraocular Pressure
Agreement of the measured IOP between the test device and the predicate device (GAT) for the tonometer function of the TONOREF III. (Agreement means the similarity of results when measurements from the test device and the predicate device are compared against each other.)
Measured: 1-2 study visits per subject over the course of 8 days
Secondary outcomes (1)
The Numbers of Adverse Events
To demonstrate that the test device is as safe as GAT
Measured: 1-2 study visits per subject over the course of 8 days
Timeline
From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.
- First posted3 Sep 2025
- Study start2 Sep 2022
- Primary completion19 Jun 2023
- Study completion19 Jun 2023
Resources
- Official record on ClinicalTrials.govThe registry entry this page is built from, with its full change history.
- Posted resultsParticipant flow, outcome measures and adverse events reported by the sponsor.
- Other trials for Intraocular PressureStudies of the same condition, open and under way.
- What the trial phases meanHow a treatment moves from first tests in people to approval.
More details
For doctors and researchers: how the study is designed, who runs it, and its registry IDs.
Study design
- Study type
- Interventional
- Primary purpose
- Diagnosis
- Allocation
- Not applicable
- Intervention model
- Single group
- Masking
- None (open label)
- Enrolment
- 52 participants (actual)
Sponsor and investigators
- Lead sponsor
- Nidek Co. LTD.
- Sponsor type
- Industry (pharma, biotech, devices)
- Responsible party
- Sponsor
IDs, dates and publications
- ClinicalTrials.gov ID
- NCT07152808
- Sponsor's protocol ID
- NIDEK-TONOREF-PILOT-0002
- Status last verified
- Apr 2026
- First submitted
- 26 Aug 2025
- Last updated
- 24 Apr 2026
- Results posted
- Yes, on ClinicalTrials.gov
- Data monitoring committee
- No
Study at a glance
From the registry record, last updated by the sponsor on 24 Apr 2026. Recruitment status is as the sponsor reported it: confirm with the study team.
- Official title
- Clinical Pilot Study to Provide Research Data for TONOREF III Tested to ANSI Z80.10-2014
- ClinicalTrials.gov ID
- NCT07152808
- Acronym / sponsor ID
- NIDEK-TONOREF-PILOT-0002
- Conditions
- Intraocular Pressure
- Study type
- Interventional
- Recruitment status
- Completed
- Enrolment
- 52 participants (actual)
- Lead sponsor
- Nidek Co. LTD.Industry (pharma, biotech, devices)
- Study start
- 2 Sep 2022 (actual)
- Primary completion
- 19 Jun 2023 (actual)
- Study completion
- 19 Jun 2023 (actual)
- First posted
- 3 Sep 2025 (actual)
- Results
- Posted 4 Feb 2026
Clinical summary
The research question as the sponsor registered it. This describes what the study plans to test, not what it found.
The primary objective of this clinical study was to show that tonometry values for TONOREF III, as the representative device of the NIDEK tonometer series, were comparable to the predicate device.
The secondary objective was to demonstrate that the test device was as safe as the predicate device.
Study design
- Primary purpose
- Diagnosis
- Allocation
- Not applicable
- Intervention model
- Single group
- Masking
- None (open label)
- Number of arms
- 1
| Arm | Type | Receives |
|---|---|---|
| TONOREF III and Goldmann Tonometer Measurements | Experimental | Device: NIDEK TONOREF III, Device: Haag-Streit Goldmann Manual TonometerSubjects will undergo IOP measurements using both the investigational device (TONOREF III) and the predicate device (Goldmann Applanation Tonometer). |
Who may qualify
- Ages
- 22 years and older
- Sex
- All
- Healthy volunteers
- Accepted
- Age groups
- Adult, Older adult
Key inclusion criteria (5 listed)
- been at least 22 years of age of either sex and any race or ethnicity;
- been willing and able to provide written informed consent prior to any study procedures being performed;
- been willing and able to follow all instructions and attend all study visits;
- corneal astigmatism of ≤3D and corneal thickness of 500-600 μm in at least one eye.
Key exclusion criteria (15 listed)
- have only one functional eye;
- have poor or eccentric fixation in the study eye;
- have corneal scarring or have had corneal surgery, including corneal laser surgery in the study eye
- have microphthalmos in the study eye;
- have buphthalmos in the study eye;
Eligibility is decided by the study team after screening; this overview does not establish that a patient qualifies. Full inclusion and exclusion criteria.
Interventions
| Intervention | Type | Arms |
|---|---|---|
| NIDEK TONOREF IIIThe auto ref/kerato/tono/pachymeter TONOREF III is a medical device which measures objective refractive errors, corneal curvature radius, intraocular pressure and corneal thickness of the patient's eye. | Device | TONOREF III and Goldmann Tonometer Measurements |
| Haag-Streit Goldmann Manual TonometerHaag-Streit Goldmann Manual Tonometer measures intraocular pressure to aid in the screening and diagnosis of glaucoma | Device | TONOREF III and Goldmann Tonometer Measurements |
Primary outcome
Agreement of Intraocular Pressure
Time frame: 1-2 study visits per subject over the course of 8 days
Agreement of the measured IOP between the test device and the predicate device (GAT) for the tonometer function of the TONOREF III. (Agreement means the similarity of results when measurements from the test device and the predicate device are compared against each other.)
1 secondary outcome
The Numbers of Adverse Events
Time frame: 1-2 study visits per subject over the course of 8 days
Results and safety
Loading the posted results from ClinicalTrials.gov…
Research links
- Official record on ClinicalTrials.govLast updated by the sponsor 24 Apr 2026.
- Posted resultsParticipant flow, baseline, all outcome measures and adverse events.
Related trials for Intraocular Pressure
Active, not recruiting
Effect of IOP Lowering on Progressive HM
High Myopia, Intraocular Pressure
Not yet recruiting
Glaucoma Drop Aids Part 2
Glaucoma, Intraocular Pressure
Active, not recruiting
Clinical Investigation of Eye C Better's Photobionic Intraocular Pressure Biomonitor Tonometer
Elevated Intraocular Pressure (IOP)
Recruiting
Impact of Home Intraocular Pressure Telemonitoring on Intraocular Pressure Control and Glaucoma Progression
Glaucoma, Intraocular Pressure
Not yet recruitingPhase 1
Eplerenone and IOP in Healthy Adults
Intraocular Pressure (IOP), Healthy Adult
Referral and contact
AzTrials does not enrol participants or take referrals. Contact the study team or a site listed below; they decide who can take part. This study is not recruiting at the moment.
The record lists no contact. Use the official record to reach the sponsor.
Questions about this trial
What is this clinical trial studying?
A study testing NIDEK TONOREF III and Haag-Streit Goldmann Manual Tonometer in people with Intraocular Pressure, run by Nidek Co. LTD..
The full registered description is in the About this study section and the official record.
Who may be eligible to take part?
The sponsor lists 5 inclusion and 15 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.
Is this clinical trial currently recruiting?
The record says: finished, as of the sponsor's last update on 24 Apr 2026. Status can change between updates, so confirm with the study contact before making plans.
What are the possible benefits and risks?
They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.
Are results available for this study?
Yes. Results are posted in the registry (first posted 4 Feb 2026). A summary is in the For doctors view, and the full results are on ClinicalTrials.gov.
How can a doctor refer a patient or contact the research team?
The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.
What does the primary outcome measure?
Agreement of Intraocular Pressure (time frame: 1-2 study visits per subject over the course of 8 days). It is what the study was designed to measure, not a finding.
Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?
No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.
How often is this information updated?
AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 24 Apr 2026; AzTrials last refreshed its copy on 7 Nov 2026.