Oscillometric Blood Pressure Measurement: Evaluation of the Effect of Wearing a Sleeve and the Type of Measurement Algorithm

An observational study of Intensive Care Units, run by Nantes University Hospital at 1 site in France.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

For accurate measurements of arterial blood pressure (BP), international guidelines recommend placing the automated oscillometric cuff directly on the bare upper arm. However, for various reasons, cuffs are often applied over a layer of textile. Read more about this study

Who is it for?
Ages 18 to 99 years
Conditions the study is for
Intensive Care Units
What is being tested?
Patients of a surgical intensive care unit
How many people
75 participants
Where
1 siteFrance
Study dates
24 Jul 2025to 15 Apr 2026
Phase
Observational
Are there results?
None posted in the registry

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 to 99 years
  • All sexes
  • No healthy volunteers
  • Adult, Older adult

Study population: Patients of a surgical intensive care unit having an arterial catheter

Who may be able to join

  • - Patients having a catheter in the radial artery.
  • and with stable BP over a 5-min period (no change in invasive MBP >10% and no change in vasoactive drugs)

Who cannot join

  • - Cardiac arrhythmia.
  • Dysfunctional of the arterial catheter or waveform, suggesting signal over- or under-damping.
  • Arm circumference > 42 cm (measured at mid-arm).
  • Contraindication to BP measurement on the arm (fracture, wound, amputated limb, limb ischemia, infection, phlebitis, history of lymph node dissection, venous access on the limb preventing cuff inflation, other).
  • Patient's health status requiring urgent care incompatible with study protocol.
  • Asymmetry of MBP between the two arms (>5 mmHg) or inability to check it.
Show all 11Show fewer
  • Pregnant woman.
  • Age <18 years old.
  • Adult known to be under guardianship or legal supervision before inclusion.
  • Absence of social security coverage.
  • Non-French speaker patient (i.e. unable to receive study information)

Trial locations and contacts

1 site in 1 country.

France1 site
  • CHU de Nantes

    Nantes, 44093

Other trials in France

About this study

Official title: Measurements of Arterial Blood Pressure With an Automated Oscillometric Cuff: Evaluation of the Effect of Wearing a Sleeve and the Type of Measurement Algorithm. A Prospective Observational Single-center Study With an Invasive Reference

For accurate measurements of arterial blood pressure (BP), international guidelines recommend placing the automated oscillometric cuff directly on the bare upper arm. However, for various reasons, cuffs are often applied over a layer of textile. Whether this practice affects the reliability of the readings remains uncertain. Using a rigorous methodology, the CASBA study aims at addressing this issue.

Additionally, two different BP measurement algorithms are available in some oscillometric monitors. To our knowledge, no study has compared the performance of these two algorithms. This study will help determine which oscillometric algorithm should be given priority in intensive care facilities.

Read the detailed description

The automated oscillometric cuff is the predominant non-invasive method for measuring BP. For accurate measurements, international guidelines recommend placing the cuff directly on a bare upper arm. However, for reasons of convenience, hygiene, comfort, cultural or religious practices, or skin reactions, cuffs are often applied over a layer of textile. Some authors suggest this practice does not significantly affect BP readings, while other authors are more cautious or even advise against it. Prior studies are not devoid of limitations. Further research into sleeve effects on BP measurement accuracy is warranted. The present study project employs rigorous methodology and leverages invasive arterial catheters for reliable reference measurements. Additionally, the broad range of BP levels in surgical intensive care units ensures comprehensive data, including mean, systolic, and diastolic BP (MBP, SBP, and DBP).

Additionally, some oscillometric monitors incorporate two reference algorithms for BP measurement:

  • Auscultatory Reference Algorithm: Calibrates oscillometric measurements to align with manual cuff and stethoscope values. This is the default setting.
  • Invasive Reference Algorithm: Adjusts non-invasive BP values to approximate invasive measurements.

Health care providers are, in our experience, rarely aware of the existence of these two distinct algorithms. Although the reference auscultatory algorithm is generally recommended for routine non-invasive monitoring, such as at home or in a clinic, the invasive reference algorithm might be preferable in critical care settings. Importantly, in the absence of published studies on the subject, these considerations remain speculative.

However, it is important to determine which oscillometric algorithm should be prioritized in critical care settings, where invasive measurement serves as the reference. This will be a secondary objective of this study.

Being tested
Patients of a surgical intensive care unit
Keywords
Blood pressure determination (MeSH), Bare arm, Sleeve, Measurement algorithm, Catheterization, peripheral arterial (MeSH), Cuff, Oscillometry (MeSH)

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Patients of a surgical intensive care unitOther

    Patients of a surgical intensive care unit having an arterial catheter undergoing noninvasive measurements of BP in 3 situations (thick sleeve, thin sleeve, bare arm) and with 2 algorithms (for bare arm measurement only).

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. To assess whether the non-invasive MBP measurement error is acceptable in either situation: bare arm, thick and thin sleeve

    The measurement error will be considered acceptable if the current international standard, the AAMI/ESH/ISO standard, is met: mean error (bias) ≤ 5.0 mmHg and its standard deviation ≤ 8.0 mmHg.

    The reference BP will be measured using an arterial catheter, taking advantage of the fact that many critically ill patients already have one

    Measured: Day 1

Secondary outcomes (13)
  1. To assess whether the non-invasive SBP and DBP measurement error is acceptable in either situation: bare arm, thick and thin sleeve

    The measurement error will be considered acceptable if the current international standard, the AAMI/ESH/ISO standard, is met: mean error (bias) ≤ 5.0 mmHg and its standard deviation ≤ 8.0 mmHg.

    The reference BP will be measured using an arterial catheter, taking advantage of the fact that many critically ill patients already have one.

    Measured: Day 1

  2. To perform statistical comparisons of measurement errors based on sleeve presence, sleeve thickness, and reference algorithm.

    A p-value < 0.05 will be used as the threshold for statistical significance in comparisons.

    Measured: Day 1

  3. To assess the risk associated with measurement errors for each condition.

    Via a dedicated error grid, the risk associated with measurement errors was classified into one of the 5 risk levels ranging from "no risk" to "dangerous risk". Distribution across the different risk levels was compared by Fisher exact tests.

    Measured: Day 1

  4. To compare the ability of non-invasive measurements to detect hypotension (MBP or SBP <65 mmHg or <90 mmHg, respectively) and hypertension (MBP or SBP >90 mmHg or >140 mmHg, respectively).

    The ability of averaged noninvasive measurements to detect hypotension (invasive MBP <65 mmHg, SBP <90 mmHg), hypertension (invasive MBP>100 mmHg, SBP >140 mmHg), and a significant therapy-induced change in invasive MBP (>10%) was determined through area under the receiver operating characteristic curve (AUCROC) analysis.

    Measured: Day 1

  5. To evaluate variability in cuff measurements within a given condition.

    Measurement variability will be assessed using the coefficient of variation. The clinical performance of an average of multiple measurements will also be compared to that of a single measurement.

    Measured: Day 1

  6. Analysis of subgroup differences based on BP levels (above vs. below median values).

    The primary evaluation criteria will be applied to subgroup analyses for each BP component.

    Measured: Day 1

  7. Analysis of subgroup differences based on measurement error levels (above vs. below median values)

    The primary evaluation criteria will be applied to subgroup analyses for each BP component.

    Measured: Day 1

  8. Analysis of subgroup differences based on BMI (above vs. below median and above vs. below 30 kg/m²)

    The primary evaluation criteria will be applied to subgroup analyses for different BMI ranges.

    Measured: Day 1

  9. Analysis of subgroup differences based on arm circumference (above vs. below median)

    The primary evaluation criteria will be applied to subgroup analyses for varying arm circumferences.

    Measured: Day 1

  10. Analysis of subgroup differences based on age (above vs. below 60 years)

    The primary evaluation criteria will be applied to subgroup analyses for different classes of ages

    Measured: Day 1

  11. Analysis of subgroup differences based on age (above vs. below 60 years)

    The primary evaluation criteria will be applied to subgroup analyses for different classes of ages.

    Measured: Day 1

  12. Analysis of subgroup differences based on presence of acute circulatory failure or not

    The primary evaluation criteria will be applied to subgroup analyses according to acute circulatory failure or not.

    Measured: Day 1

  13. Analysis of subgroup differences based on ventilation mode (mechanical vs. non-mechanical)

    The primary evaluation criteria will be applied to subgroup analyses according to the ventilation mode

    Measured: Day 1

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted31 May 2025
  2. Study start24 Jul 2025
  3. Primary completion15 Apr 2026
  4. Study completion15 Apr 2026

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Observational
Observational model
Case-control
Time perspective
Prospective
Enrolment
75 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT06999863
Sponsor's protocol ID
RC25_0025
Acronym
CASBA
Status last verified
Apr 2026
First submitted
29 Apr 2025
Last updated
6 May 2026
Results posted
No
Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
No
Individual data sharing
No

Questions about this trial

What is this clinical trial studying?

An observational study of Intensive Care Units, run by Nantes University Hospital at 1 site in France.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 2 inclusion and 11 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 6 May 2026. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in France. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

To assess whether the non-invasive MBP measurement error is acceptable in either situation: bare arm, thick and thin sleeve (time frame: Day 1). It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 6 May 2026.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in