Agreement and Precision Study of UNITY® DX and a Comparator Biometer

An observational study of Normal Eyes and Abnormal Eyes, run by Cylite Pty Ltd at 1 site in United States.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

The purpose of this study is to validate the clinical measurement performance of an investigational biometer. Subjects will undergo multiple measurements of the eye, and the agreement, repeatability, and reproducibility of the biometric measurements will be evaluated. Read more about this study

Who is it for?
Ages 22 years and olderHealthy volunteers can take part
Conditions the study is for
Normal Eyes, Abnormal Eyes
What is being tested?
Investigational Biometer and IOLMaster 700 Biometer
Type of treatment
Medical device
How many people
160 participants
Where
1 siteUnited States
Study dates
18 Jan 2025to 30 Mar 2025
Phase
Observational
Are there results?
None posted in the registry

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 22 years and older
  • All sexes
  • Healthy volunteers accepted
  • Adult, Older adult

Study population: Cohorts will be selected from approximately one investigative site located in the United States.

Who may be able to join

  • Able to understand and sign an Institution Review Board approved Informed Consent form.
  • Willing and able to attend all scheduled study visits as required by the protocol.
  • Identified by the investigator as eligible for at least one of the 4 eye groups identified in the protocol.

Who cannot join

  • Unable to fixate, for example, due to nystagmus or other eye movement abnormality.
  • Any ocular disease and/or condition that, in the investigator's clinical judgment, may put the subject at significant risk, may compromise study results, or may interfere significantly with the subject's participation in the study.
  • Active ocular infection or inflammation.
  • Rigid or contact lens wear during the previous two weeks prior to Screening.

Trial locations and contacts

1 site in 1 country.

United States1 site
  • NorthEast Eye Research Associates LLC

    Woburn, Massachusetts, 01801

Other trials in United States

About this study

The purpose of this study is to validate the clinical measurement performance of an investigational biometer. Subjects will undergo multiple measurements of the eye, and the agreement, repeatability, and reproducibility of the biometric measurements will be evaluated.

Read the detailed description

This is intended to be a 1-visit study approximately 3 hours in length with planned duration of exposure of up to 20 minutes per device (investigational biometer and IOLMaster 700). Four cohorts will be enrolled. One eye per subject will be assessed.

Being tested
Investigational Biometer, IOLMaster 700 Biometer
Keywords
Biometric measurements

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Investigational BiometerDevice

    Investigational non-contact, non-invasive ophthalmic imaging and analysis device used for visualization of ocular structures and measurement of anterior segment and biometric parameters

  • IOLMaster 700 BiometerDevice

    Commercially available non-contact instrument used to capture optical biometry measurements

Study groups in NCT06826599
Study groupRoleWhat they receive
Group 1 - Normal EyesThe eye will be imaged up to seven times with the biometer to achieve three acceptable scans
Receives: Device: Investigational Biometer; Device: IOLMaster 700 Biometer
Group 2 - Eyes with cataractsThe eye will be imaged up to seven times with the biometer to achieve three acceptable scans
Receives: Device: Investigational Biometer; Device: IOLMaster 700 Biometer
Group 3 - Eyes with significant refractive errorsThe eye will be imaged up to seven times with the biometer to achieve three acceptable scans
Receives: Device: Investigational Biometer; Device: IOLMaster 700 Biometer
Group 4 - Eyes with abnormal corneal topographyThe eye will be imaged up to seven times with the biometer to achieve three acceptable scans
Receives: Device: Investigational Biometer; Device: IOLMaster 700 Biometer

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Central corneal thickness (CCT)

    Central corneal thickness will be measured in microns. This outcome measure will be analyzed using Bland-Altman Limits of Agreement, repeatability, reproducibility, and percent coefficient of variation (repeatability and reproducibility) methods.

    Measured: Day 1

  2. Lens thickness (LT)

    Lens thickness will be measured in microns and analyzed using Bland-Altman Limits of Agreement, repeatability, reproducibility, and percent coefficient of variation (repeatability and reproducibility) methods.

    Measured: Day 1

  3. Anterior chamber depth (regression) (ACD-R)

    Anterior chamber depth (the physical length from the anterior cornea to the anterior lens) will be measured in microns and analyzed using Bland-Altman Limits of Agreement, repeatability, reproducibility, and percent coefficient of variation (repeatability and reproducibility) methods.

    Measured: Day 1

  4. Axial length (regression) (AL-R)

    Axial length (the distance between the back and front of the eye) will be measured in millimeters and analyzed using Bland-Altman Limits of Agreement, repeatability, reproducibility, and percent coefficient of variation (repeatability and reproducibility) methods.

    Measured: Day 1

Secondary outcomes (2)
  1. Anterior Chamber Depth Sum of Segments (ACD-SoS)

    ACD-SoS is calculated by calculating the individual corneal thickness and aqueous depth segments, applying the known refractive indices for each segment, and summing the resultant measurement of each segment. ACD-SoS will be measured in microns and analyzed using Bland-Altman Limits of Agreement, repeatability, and reproducibility methods.

    Measured: Day 1

  2. Axial Length Sum of Segments (AL-SoS)

    AL-SoS is calculated by calculating the individual corneal thickness, aqueous depth, lens thickness and vitreous length segments, applying the known refractive indices for each segment, and summing the resultant measurement of each segment. AL-SoS will be measured in millimeters and analyzed using using Bland-Altman Limits of Agreement, repeatability, and reproducibility methods.

    Measured: Day 1

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted14 Feb 2025
  2. Study start18 Jan 2025
  3. Primary completion30 Mar 2025
  4. Study completion30 Mar 2025

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Observational
Observational model
Cohort
Time perspective
Prospective
Enrolment
160 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT06826599
Sponsor's protocol ID
CYC-CL-SOP-004
Status last verified
Aug 2025
First submitted
4 Feb 2025
Last updated
5 Aug 2025
Results posted
No
Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
Yes

Questions about this trial

What is this clinical trial studying?

An observational study of Normal Eyes and Abnormal Eyes, run by Cylite Pty Ltd at 1 site in United States.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 3 inclusion and 4 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 5 Aug 2025. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in United States. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Central corneal thickness (CCT) (time frame: Day 1). The study lists 4 primary outcomes in all. It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 5 Aug 2025.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in