The accuracy and safety of the Wrist Ambulatory Blood Pressure Recorder (Model: LCA-B10) were assessed according to the requirements of EN ISO 81060-2:2018 + AMD1:2020. Read more about this study

Who is it for?
Ages 18 years and olderHealthy volunteers can take part
Conditions the study is for
Blood Pressure
What is being tested?
Wrist Ambulatory Blood Pressure Monitor
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
128 participants
Where
1 siteChina
Study dates
24 Apr 2024to 4 Jun 2024
Phase
Interventional
Are there results?
None posted in the registry

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 years and older
  • All sexes
  • Healthy volunteers accepted
  • Adult, Older adult

Who may be able to join

  • Subjects should voluntarily participate in the clinical trial and sign the informed consent form;
  • Subjects must be willing and able to follow the study procedures;
  • Subjects are over 18 years old whose wrist circumference should be within 13.0-21.0cm.

Who cannot join

  • Subjects with serious arrhythmias (atrial fibrillation, atrial flutter, ventricular flutter, ventricular fibrillation, ventricular tachycardia, borderline tachycardia, sinus arrest, pathological sinus syndrome, pacemaker rhythm) showed by electrocardiogram during screening period;
  • Subjects with sickle cell disease;
  • Subjects with prior mastectomy or arm lymph node dissection;
  • Subjects who are allergic to nylon, fluororubber or thermo-plastic polyurethane elastomer (TPU);
  • Subjects with wounds on the limb at the measurement site or inflammation of the wrist
  • Subjects who are undergoing intravenous fluid therapy, blood transfusion or other catheterization at the measurement site limb;
Show all 9Show fewer
  • Subjects who have eaten, consumed alcohol, smoked cigarettes, or exercised vigorously 30 minutes prior to the trial;
  • Subjects who have participated in other clinical trials within 30 days that may affect this trial;
  • Subjects who, in the opinion of the investigator, should not participate in this clinical trial.

Trial locations and contacts

1 site in 1 country.

China1 site
  • Bin Peng

    Chenzhou, Hunan, 423000

Other trials in China

About this study

Official title: Clinical Trial on the Effectiveness and Safety of Blood Pressure Function System of Wrist Ambulatory Blood Pressure Monitor

The accuracy and safety of the Wrist Ambulatory Blood Pressure Recorder (Model: LCA-B10) were assessed according to the requirements of EN ISO 81060-2:2018 + AMD1:2020.

Conditions
Blood Pressure
Being tested
Wrist Ambulatory Blood Pressure Monitor

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Wrist Ambulatory Blood Pressure MonitorDevice

    Comparing blood pressure measured by wrist blood pressure monitor with mercury sphygmomanometer .

Study groups in NCT06573216
Study groupRoleWhat they receive
Blood pressure measured by wrist ambulatory blood pressure monitor and mercury sphygmomanometerExperimentalNot described
Receives: Device: Wrist Ambulatory Blood Pressure Monitor

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Blood pressure mean deviation and standard deviation- evaluation criterion 1

    The mean deviation and standard deviation between the blood pressure values measured by the investigational device and those obtained using the auscultatory method with the reference device. This calculation is performed on valid data pairsof subjects (n≥255).

    Evaluation criterion 1: The blood pressure mean deviation should not exceed ±5mmHg, and the standard deviation should not exceed 8mmHg.

    Measured: 30 minutes

  2. Blood pressure mean deviation and standard deviation- evaluation criterion 2

    The mean deviation and standard deviation between the blood pressure values measured by the investigational device and those obtained using the auscultatory method with the reference device. This calculation is performed on the valid data group of all subjects (m≥85).

    Evaluation criterion 2: The blood pressure standard deviation within the valid data group (m ≥ 85) should not exceed the maximum allowable standard deviation corresponding to the mean deviation listed in Criterion 2 in section 5.2.4.1.2 b of ISO 81060-2:2018+AMD1:2020.

    Measured: 30 minutes

  3. The mean deviation and standard deviation of blood pressure measurements under ambulatory blood pressure monitoring

    The mean deviation and standard deviation of the blood pressure within the valid data group are calculated to evaluate whether the clinical accuracy of the investigational device meets the requirements outlined in ISO 81060-2:2018+AMD1:2020, specifically in section 5.2.4.1.2a).

    The blood pressure mean deviation should not exceed ±5mmHg, and the standard deviation should not exceed 8mmHg.

    Measured: 30 minutes

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted27 Aug 2024
  2. Study start24 Apr 2024
  3. Primary completion4 Jun 2024
  4. Study completion4 Jun 2024

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Diagnosis
Allocation
Not applicable
Intervention model
Single group
Masking
None (open label)
Enrolment
128 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT06573216
Sponsor's protocol ID
CTP-PPA0971CHN24011810332987
Status last verified
Aug 2024
First submitted
26 Aug 2024
Last updated
27 Aug 2024
Results posted
No
FDA-regulated drug
No
FDA-regulated device
No

Questions about this trial

What is this clinical trial studying?

A study testing Wrist Ambulatory Blood Pressure Monitor in people with Blood Pressure, run by Huawei Device Co., Ltd at 1 site in China.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 3 inclusion and 9 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 27 Aug 2024. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in China. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Blood pressure mean deviation and standard deviation- evaluation criterion 1 (time frame: 30 minutes). The study lists 3 primary outcomes in all. It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 27 Aug 2024.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in