Blood Pressure Treatment by the Corsano CardioWatch 287-2 Evaluation Study

A study testing Corsano CardioWatch 287-2 non-invasive blood pressure (NIBP) monitoring in people with High Blood Pressure, run by Corsano Health B.V. at 1 site in Netherlands.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

Rationale: Wearables have the potential to monitor patients remotely. The Corsano CardioWatch 287-2 is such a medical device that can monitor long-term blood pressure. The device has been validated using clinical trials in hospitals, but evaluation in the intended remote setting during treatment is lacking. Read more about this study

Who is it for?
Ages 18 to 80 years
Conditions the study is for
High Blood Pressure
What is being tested?
Corsano CardioWatch 287-2 non-invasive blood pressure (NIBP) monitoring
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
80 participants
Where
1 siteNetherlands
Study dates
2 Jul 2024to 1 Jun 2026
Phase
Interventional
Are there results?
None posted in the registry

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 to 80 years
  • All sexes
  • No healthy volunteers
  • Adult, Older adult

Who may be able to join

  • aged between 18 and 80 years old;
  • able to provide consent;
  • untreated or treated for hypertension, with uncontrolled BP and medical indication for antihypertensive drug initiation or uptitration or change in antihypertensive drug per doctor prescription.

Who cannot join

  • Unable to wear the Corsano CardioWatch 287 due to reasons such as allergic reactions, wounds, amputations etc.;
  • Unable to receive blood pressure measurements per cuff due to lymphedema, amputation, dialysis shunt, wounds, etc.;
  • Pregnant women;
  • Breastfeeding women;
  • Upper arm circumference not within the cuff range (22-42 cm)
  • Unable or not willing to sign for informed consent;
  • Significant mental or cognitive impairment.

Trial locations and contacts

1 site in 1 country.

Netherlands1 site
  • Reinier de Graaf Gasthuis

    Delft, South Holland, 2625 AD

Other trials in Netherlands

About this study

Rationale: Wearables have the potential to monitor patients remotely. The Corsano CardioWatch 287-2 is such a medical device that can monitor long-term blood pressure. The device has been validated using clinical trials in hospitals, but evaluation in the intended remote setting during treatment is lacking.

Primary objective: To assess the ability to track blood pressure decrease measured by the Corsano CardioWatch 287-2 after a period of 28 days of antihypertensive drug treatment initiation, uptitration or change in antihypertensive drugs.

Read more

Study population: A group, untreated or treated for hypertension, with uncontrolled BP and medical indication for antihypertensive drug treatment initiation, uptitration or change in antihypertensive drugs.

Study design: Blood pressure will be measured by the investigative device and a reference device at pre-treatment and after 28 days. The investigative device is the Corsano CardioWatch 287-2, which measures blood pressure through optical photoplethysmography (PPG). The reference method involves automatic blood pressure cuff measurements.

Main study parameters/endpoints: Absolute systolic and diastolic blood pressure decrease after change of treatment measured by the Corsano Cardiowatch 287-2 after 28 days, compared to automatic blood pressure cuff measurements.

Read the detailed description

Background and Introduction:

Blood pressure (BP) is a cardiovascular risk factor that affects millions of people worldwide. BP is a major determinant of stroke, heart failure, coronary artery disease, and chronic kidney disease [1]. Hypertension may be asymptomatic or episodic. Accurate monitoring of BP is required to tailor medication to the patient's specific needs to maintain blood pressure control [2]. Therefore, accurate and continuous monitoring of BP is crucial for optimal diagnosis, treatment, and prevention of cardiovascular events.

However, conventional methods for measuring BP have several limitations. Most BP monitors use an occluding cuff that inflates periodically to obtain intermittent readings. Cuff measurements may not reflect the true BP variations that occur throughout the day and night due to various factors such as physical activity, stress level or sleep quality [3].

Wearable devices are emerging as a promising alternative to conventional methods for measuring BP [4, 5, 6]. They can measure BP continuously and non-invasively by using novel sensors and algorithms that do not require a cuff or electrodes. Wearable devices can also enable remote monitoring and data transmission to healthcare providers or researchers. However, wearable devices need to be evaluated against standard methods in different clinical settings and patient populations.

The Corsano CardioWatch 287-1 conforms with the European Medical Device Regulation (CE-MDR) and is a clinically validated vital signs monitoring bracelet. It is able to continuously measure pulse rate [6], inter-beat intervals [6], breathing rate [7], sleep and activity. The Corsano CardioWatch 287-2, an iteration on the 287-1, adds electrocardiogram (ECG), oxygen saturation (SpO2), galvanic skin response (GSR), core body temperature and non-invasive blood pressure (NIBP).

The Corsano CardioWatch Corsano 287-2 has been submitted to laboratory tests against internationally recognized standards for the monitoring of pulse rate, respiratory rate, SpO2, GSR, temperature and ECG. Additionally, the MULTI-VITAL study (NL80236.000.22) has clinically validated Corsano 287-2 in a controlled intramural healthcare setting. The Remote Cardiac Monitoring (RECAMO) study is underway to assess the free-living remote care setting long-term monitoring validity using a static test. Clinical evaluation in a non-static remote care setting is still lacking, however.

Antihypertensive drugs reduce BP through blood volume reduction and vasodilatation which are very different physiological mechanisms. This consequently changes the PPG waveform in distinct ways which constitutes a necessarily challenging test for cuffless devices. The accuracy of the Corsano CardioWatch 287-2 during a change of antihypertensive treatment has not yet been evaluated. However it is an important target for future use of the CardioWatch 287-2. Additionally, it is a recommended research in the European Society of Hypertension (ESH) recommendations for the validation of cuffless blood pressure measuring devices [8].

The current trial aims to clinically assess the Corsano CardioWatch 287-2 on blood pressure monitoring during antihypertensive treatment in a remote care setting.

Objectives:

The main objective is to assess the Corsano CardioWatch 287-2's ability to track blood pressure decrease over a period of 28 days after antihypertensive drug treatment initiation, uptitration or change in antihypertensive drug.

The assessment is done by comparing the blood pressure measurements of the Corsano CardioWatch 287-2 with those of an automatic cuff.

Study Design:

Patients who are untreated or treated for hypertension, with uncontrolled BP and medical indication for antihypertensive drug treatment initiation, uptitration, or change in antihypertensive drugs will be approached for participation in the trial. The ability of the CardioWatch 287-2 to track decreases in blood pressure will be assessed. Blood pressure measured by the CardioWatch 287-2 at the beginning and end of the study period will be compared with the blood pressure measured by a reference device, which is an automatic blood pressure cuff. Blood pressure will be measured continuously by the Corsano CardioWatch 287-2 during 28 days. These measurements are only meant for study purposes and not part of standard care.

Blood pressure measurements with the Corsano cuff (an automatic oscillometric blood pressure monitoring device) will be used for initialization of the Corsano CardioWatch 287-2 on day 0. On day 1, the antihypertensive drug treatment initiation, uptitration, or change in antihypertensive drug will be started. The automatic blood pressure cuff measurements at day 0 and 28 will be used to assess the CardioWatch's accuracy by comparing it to the closest available blood pressure measurement by the CardioWatch.

The mean difference in blood pressure decrease between the Corsano CardioWatch 287-2 and reference automatic blood pressure cuff will be analyzed.

Study population:

The study will involve a patient group visiting the outpatient cardiology clinic at the Reinier de Graaf Gasthuis. This group has an uncontrolled BP and therefore medical indication for antihypertensive drug treatment initiation, uptitration, or change in antihypertensive drugs.

Sample size calculation:

The participants will be included with the aim to assess the long-term variation in blood pressure during antihypertensive treatment. Sample size calculation for the assessment of blood pressure monitoring is based on Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO) consensus that at least 35 patients are required for the validation of a 'moderate accuracy' device and recommends the following criteria to be met [8]:

  • Males ≥ 30 % and females ≥ 30 %
  • Subjects with an added diuretic drug ≥ 35%
  • Subjects with an added long-acting dihydropyridine calcium channel blocker ≥ 35%
  • ≥ 60% of the subjects need a reference systolic BP decrease ≥ 10 mmHg
  • ≥ 60% of the subjects need a reference diastolic BP decrease ≥ 5 mmHg
  • ≥ 30% of the subjects need a reference systolic BP decrease < 10 mmHg
  • ≥ 30% of the subjects need a reference diastolic BP decrease < 5 mmHg Based on prior studies conducted by Corsano, e.g. RECAMO Study, the investigators experienced difficulty in obtaining a diverse study group in terms of inclusion requirements. To meet these criteria the investigators will include at least 80 patients.
Being tested
Corsano CardioWatch 287-2 non-invasive blood pressure (NIBP) monitoring
Keywords
Remote monitoring, Photoplethysmography, Blood pressure, Blood pressure lowering medication, Wearable technology, Evaluation
Medical subject headings
Hypertension

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Corsano CardioWatch 287-2 non-invasive blood pressure (NIBP) monitoringDevice

    Patients receive Corsano CardioWatch 287-2 for continous NIBP monitoring and perform several spot-checks using an automated BP cuff as a reference during the start or uptitration of BP lowering medication

Study groups in NCT06508619
Study groupRoleWhat they receive
Patients receiving blood pressure lowering medication (uptitration) as standard of careExperimentalPatients receiving blood pressure lowering medication (uptitration) as standard of care
Receives: Device: Corsano CardioWatch 287-2 non-invasive blood pressure (NIBP) monitoring

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Baseline average BP for both CardioWatch 287-2 and reference

    Baseline average blood pressure per patient and its standard deviation measured by automatic blood pressure cuff reference device and CardioWatch 287-2, separately.

    Measured: Measured in each patient on 1st study day

  2. Post-treatment average BP for both CardioWatch 287-2 and reference

    Post-treatment average blood pressure and its standard deviation measured by automatic blood pressure cuff reference device and CardioWatch 287-2, separately.

    Measured: Measured in each patient on 28th/last study day

  3. Average BP change comparing CardioWatch 287-2 and reference

    Absolute systolic and diastolic BP decrease and its standard deviation after change of treatment measured by the Corsano CardioWatch 287-2 after 28 days of monitoring, compared to automatic blood pressure cuff measurements.

    Measured: Measured in each patient on 28th/last study day

Secondary outcomes (2)
  1. Bland-Altman for CardioWatch 287-2 vs. reference BP

    Mean bias and limits of agreement between blood pressure measurements measured by the Corsano CardioWatch 287-2 and blood pressure measured by an automatic blood pressure cuff using a Bland Altman analysis.

    Measured: Measured in each patient on 28th/last study day

  2. Pearson correlation coefficient between CardioWatch 287-2 vs. reference BP

    Pearson correlation coefficient between blood pressure decrease measured by the Corsano CardioWatch 287-2 and blood pressure decrease measured by an automatic blood pressure cuff.

    Measured: Measured in each patient on 28th/last study day

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted18 Jul 2024
  2. Study start2 Jul 2024
  3. Primary completion1 Jun 2026
  4. Study completion1 Jun 2026

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Supportive care
Allocation
Not applicable
Intervention model
Single group
Masking
None (open label)
Enrolment
80 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT06508619
Sponsor's protocol ID
BPTreat
Acronym
BPTreat
Status last verified
Jun 2026
First submitted
7 Jun 2024
Last updated
11 Jun 2026
Results posted
No
Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
No
Individual data sharing
No

Publications

  • Michaud GF, Stevenson WG. Atrial Fibrillation. N Engl J Med. 2021 Jan 28;384(4):353-361. doi: 10.1056/NEJMcp2023658. No abstract available.

    PubMed 33503344
  • Sheppard JP, Lown M, Burt J, Ford GA, Hobbs FDR, Little P, Mant J, Payne RA, McManus RJ; OPTiMISE Investigators. Blood Pressure Changes Following Antihypertensive Medication Reduction, by Drug Class and Dose Chosen for Withdrawal: Exploratory Analysis of Data From the OPTiMISE Trial. Front Pharmacol. 2021 Apr 20;12:619088. doi: 10.3389/fphar.2021.619088. eCollection 2021.

    PubMed 33959004
  • Meidert AS, Saugel B. Techniques for Non-Invasive Monitoring of Arterial Blood Pressure. Front Med (Lausanne). 2018 Jan 8;4:231. doi: 10.3389/fmed.2017.00231. eCollection 2017.

    PubMed 29359130
  • Sessler DI, Saugel B. Beyond 'failure to rescue': the time has come for continuous ward monitoring. Br J Anaesth. 2019 Mar;122(3):304-306. doi: 10.1016/j.bja.2018.12.003. Epub 2019 Jan 3. No abstract available.

    PubMed 30770047
  • McGillion MH, Duceppe E, Allan K, Marcucci M, Yang S, Johnson AP, Ross-Howe S, Peter E, Scott T, Ouellette C, Henry S, Le Manach Y, Pare G, Downey B, Carroll SL, Mills J, Turner A, Clyne W, Dvirnik N, Mierdel S, Poole L, Nelson M, Harvey V, Good A, Pettit S, Sanchez K, Harsha P, Mohajer D, Ponnambalam S, Bhavnani S, Lamy A, Whitlock R, Devereaux PJ; PROTECT Network Investigators. Postoperative Remote Automated Monitoring: Need for and State of the Science. Can J Cardiol. 2018 Jul;34(7):850-862. doi: 10.1016/j.cjca.2018.04.021. Epub 2018 Apr 25.

    PubMed 29960614
  • Blok S, Piek MA, Tulevski II, Somsen GA, Winter MM. The accuracy of heartbeat detection using photoplethysmography technology in cardiac patients. J Electrocardiol. 2021 Jul-Aug;67:148-157. doi: 10.1016/j.jelectrocard.2021.06.009. Epub 2021 Jul 2.

    PubMed 34256184
  • Gehring JM, Saeijs-van Niel LC, Ten Bosch-Paniagua LP, Frank MH. Continuous respiration rate monitoring using photoplethysmography technology in patients with Obstructive Sleep Apnea. Under peer review

  • Stergiou GS, Avolio AP, Palatini P, Kyriakoulis KG, Schutte AE, Mieke S, Kollias A, Parati G, Asmar R, Pantazis N, Stamoulopoulos A, Asayama K, Castiglioni P, De La Sierra A, Hahn JO, Kario K, McManus RJ, Myers M, Ohkubo T, Shroff SG, Tan I, Wang J, Zhang Y, Kreutz R, O'Brien E, Mukkamala R. European Society of Hypertension recommendations for the validation of cuffless blood pressure measuring devices: European Society of Hypertension Working Group on Blood Pressure Monitoring and Cardiovascular Variability. J Hypertens. 2023 Dec 1;41(12):2074-2087. doi: 10.1097/HJH.0000000000003483. Epub 2023 Jun 22.

    PubMed 37303198

Questions about this trial

What is this clinical trial studying?

A study testing Corsano CardioWatch 287-2 non-invasive blood pressure (NIBP) monitoring in people with High Blood Pressure, run by Corsano Health B.V. at 1 site in Netherlands.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 3 inclusion and 7 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 11 Jun 2026. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in Netherlands. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Baseline average BP for both CardioWatch 287-2 and reference (time frame: Measured in each patient on 1st study day). The study lists 3 primary outcomes in all. It is what the study was designed to measure, not a finding.

Where can I read the original study information?

On the ClinicalTrials.gov record for NCT06508619, which also lists 8 linked publications.

Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 11 Jun 2026; AzTrials last refreshed its copy on 8 Nov 2026.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in