Future Cardia™ Insertable Cardiac Monitor in Subjects With Paroxysmal Atrial Fibrillation (First-in-Human Study)

A study testing Future Cardia™ Insertable Cardiac Monitoring (ICM) System in people with Arrhythmias, Cardiac and Atrial Fibrillation, run by Future Cardia, Inc at 1 site in Croatia.

Active, not recruiting

Under way, not looking for participants. Everyone needed has enrolled. The study is still running.

Official record
Search all trials

Trial summary

This study is a first-in-human, prospective, multi-center, pre-market single-arm clinical trial to evaluate the Future Cardia™ ICM. Read more about this study

Who is it for?
Ages 18 to 75 years
What is being tested?
Future Cardia™ Insertable Cardiac Monitoring (ICM) System
Could I receive a placebo?
No, everyone receives an active treatment
How many people
30 participants
Where
1 siteCroatia
Study dates
1 Dec 2023to 1 Dec 2026
Phase
Interventional
Are there results?
Not yet: the study is under way

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 to 75 years
  • All sexes
  • No healthy volunteers
  • Adult, Older adult

Who may be able to join

  • Adult patients (≥18 years old and <75 years old)
  • Paroxysmal AF, defined as AF that terminates spontaneously or with intervention within 7 days of onset (ESC 2020 guidelines). In order to be included, patients must present one of the following conditions:
  • paroxysmal AF patients that are candidates for AF ablation;
  • patients hospitalized for symptomatic AF, who failed to convert in the emergency room (ER):
  • outpatients who require treatment with anti-arrhythmic drugs (AAD) (mostly propafenone or flecainide; more rarely sotalol or amiodaron are given for PAF).
  • Patient is willing and able to provide written informed consent.
  • Patient is willing and able to comply with the protocol, including follow-up visits and data transmissions.

Who cannot join

  • Patient who is scheduled to have MRI or is likely to require MR screening. The Future Cardia implant is not compatible with MRI.
  • Currently indicated for or implanted with a cardiovascular implantable electronic device (CIED) (e.g., ICM or ILR, pacemaker, ICD, CRT-D or CRT-P device) or hemodynamic monitoring system.
  • Compromised immune system or at high risk of developing an infection.
  • Active systemic infection or history of any infection within the last 30 days.
  • Subjects who are female must:
  • have a negative pregnancy test by β-hCG blood test.
Show all 16Show fewer
  • not breastfeeding
  • either be surgically sterile, postmenopausal (cessation of menses for at least 1 year), or if they have childbearing potential, agree to use a medically accepted, highly effective method of contraception during the entire duration of the study.
  • Subject is currently enrolled in another investigational study.
  • Any condition or abnormality (including clinical laboratory, physical examination, or vital sign abnormalities), current or past, that, in the opinion of the Investigator, would compromise the efficacy evaluation, safety of the patient, or would interfere with or complicate study procedures or assessments.
  • Any concomitant condition which, in the opinion of the investigator, would not allow safe participation in the study (e.g., drug addiction, alcohol abuse, emotional/psychological diagnosis).
  • Obesity, Class 2 (BMI ≥ 35-40) or Class 3 (BMI ≥ 40).
  • Subject is unwilling or unable to comply with the study procedures.
  • Subject is legally incapacitated and unable to provide written informed consent.
  • Patient with abnormal thoracic anatomy or scar tissue at the implant site that may adversely impact the insertion procedure.
  • For patients currently taking warfarin at the time of insertion, most recent INR value (within 7 days) is less than 3.5 for an acceptable risk of bleeding.

Trial locations and contacts

1 site in 1 country.

Croatia1 site
  • University Hospital of Split

    Split

Other trials in Croatia

About this study

Official title: Clinical Investigation of the Future Cardia™ Insertable Cardiac Monitor in Subjects With Paroxysmal Atrial Fibrillation Voice Of the Heart Trial (First-in-Human Study)

This study is a first-in-human, prospective, multi-center, pre-market single-arm clinical trial to evaluate the Future Cardia™ ICM.

Read the detailed description

The objective of this first-in-human, prospective, multi-center, pre-market single-arm clinical investigation is to evaluate the safety and performance of the Future Cardia™ ICM by assessing the insertion procedure, sensing and detection performance and data monitoring transmission success, and device- or procedure-related complication rates over a 6-month follow-up period.

Being tested
Future Cardia™ Insertable Cardiac Monitoring (ICM) System
Keywords
Telemedicine, Cardiac Arrhythmia, ECG, PCG, Accelerometer, Cardiac monitor, Insertable cardiac monitor
Medical subject headings
Arrhythmias, Cardiac, Atrial Fibrillation, Bradycardia, Tachycardia, Ventricular, Heart Arrest

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Future Cardia™ Insertable Cardiac Monitoring (ICM) SystemDevice

    The Future Cardia™ ICM procedure will be performed as a stand-alone procedure for subcutaneous insertion.

Study groups in NCT06167434
Study groupRoleWhat they receive
Single ArmExperimentalInsertable Cardiac Monitor Implant.
Receives: Device: Future Cardia™ Insertable Cardiac Monitoring (ICM) System

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Characterization of Insertion Procedure - Implantation: Insertion Success

    Insertion Success. (Yes/No); Place Device into Pocket (Yes/No);

    Measured: up to 24 weeks

  2. Characterization of Insertion Procedure - Implantation: Incision Size

    Incision Size in millimeters (mm).

    Measured: up to 24 weeks

  3. Characterization of Insertion Procedure - Implantation: Device Orientation

    Device Orientation: Location between the suprasternal notch and the left nipple, (position A); or left parasternal region (position B) or left sub-mammary position can be used (position C).

    Measured: up to 24 weeks

  4. Characterization of Insertion Procedure - Implantation: Duration of Insertion Procedure.

    Duration of Insertion Procedure in minutes (min).

    Measured: up to 24 weeks

  5. Characterization of Explant Procedure - Device removal: Removal Success

    Removal Success (Yes/No)

    Measured: up to 24 weeks

  6. Characterization of Explant Procedure - Device removal: Incision Size

    Incision Size in millimeters (mm)

    Measured: up to 24 weeks

  7. Characterization of Explant Procedure - Device removal: Duration of explant procedure

    Duration of explant procedure in minutes (min).

    Measured: up to 24 weeks

  8. Device and Procedural Safety over a period of 180 days

    1. Number of Serious Adverse Events (SAE) related to the Device or insertion procedure.
    2. Number of Participants with Device-related or Procedure-related Serious Adverse Event that require additional invasive intervention (including need to for device removal, device replacement, surgical repositioning of the device, or another surgery related to the device or primary insertion procedure).

    Measured: up to 24 weeks

  9. Success of Wireless Transmissions over a period of 180 days

    Assess the percentage of successful wireless transmissions from the system throughout the study duration (transmission successful, the transmission failed or transmission re-try).

    Measured: up to 24 weeks

Secondary outcomes (4)
  1. Evaluation of the sensing quality of Future Cardia™ ICM, through assessment of the ECG signal by two independent electrophysiologists.

    ECG signal Assessment.

    Measured: up to 24 weeks

  2. Comparison of the Future Cardia™ ECG tracing and detection of arrhythmias with 7-day ECG recorder tracings or 24-hr Holter recording (optional, when available).

    Comparison of the Future Cardia™ ECG tracing and detection of arrhythmias with 7-day ECG recorder tracings or 24-hr Holter recording.

    Measured: up to 24 weeks

  3. Usability: survey of the implanting physicians regarding the ease of implantation and use of the system for monitoring during follow-up.

    Usability Survey

    Measured: up to 24 weeks

  4. Number of Serious Adverse Events (SAE) related to the Device or study procedure in patients who elect to continue monitoring using the ICM device after the 6-month study follow-up.

    Long-term safety in Post study follow-up (> 6 months up to 2 years).

    Measured: 2 Years

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted12 Dec 2023
  2. Study start1 Dec 2023
  3. Primary completion1 Jun 2026 (estimated)
  4. Study completion1 Dec 2026 (estimated)

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Diagnosis
Allocation
Not applicable
Intervention model
Single group
Masking
None (open label)
Enrolment
30 participants (estimated)
IDs, dates and publications
ClinicalTrials.gov ID
NCT06167434
Sponsor's protocol ID
FC-2022-01
Status last verified
Mar 2025
First submitted
21 Nov 2023
Last updated
8 Aug 2025
Results posted
No
FDA-regulated drug
No
FDA-regulated device
Yes
Individual data sharing
No

Questions about this trial

What is this clinical trial studying?

A study testing Future Cardia™ Insertable Cardiac Monitoring (ICM) System in people with Arrhythmias, Cardiac and Atrial Fibrillation, run by Future Cardia, Inc at 1 site in Croatia.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 7 inclusion and 16 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: under way, not looking for participants, as of the sponsor's last update on 8 Aug 2025. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in Croatia. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

This study is still under way. Results are not available in the source record yet.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Characterization of Insertion Procedure - Implantation: Insertion Success (time frame: up to 24 weeks). The study lists 9 primary outcomes in all. It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 8 Aug 2025.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in