Posterior Capsule OpacificationCleaRingNCT06083025

Influence of CleaRing Open-capsule Device on Refractive Predictability in Cataract Patients (Part 2)

A study testing CleaRing in people with Posterior Capsule Opacification, run by Hanita Lenses at 1 site in Israel.

Unknown status

Status not confirmed. The sponsor has not verified this record for over two years, so it may be out of date.

Official record
Search all trials

Trial summary

The Hanita CleaRing device is an intraocular implant designed for inhibition of PCO leading to reduction in the Nd:YAG treatment rate. The device is implanted into the capsular bag after the extraction of the crystalline lens and prior to implantation of the IOL. Read more about this study

Who is it for?
Ages 50 to 80 years
Conditions the study is for
Posterior Capsule Opacification
What is being tested?
CleaRing
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
25 participants
Where
1 siteIsrael
Study dates
15 Apr 2023to 1 Apr 2025
Phase
Interventional
Are there results?
None posted in the registry

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 50 to 80 years
  • All sexes
  • No healthy volunteers
  • Adult, Older adult

Who may be able to join

  • Subject is between 50 and 80 years of age
  • Subject is designated for bilateral cataract surgery
  • Both eyes fulfil the following criteria:
  • Potentially able to achieve a good vision (by opinion of the investigator)
  • ACD is at least 2.5mm (from epithelium)
  • Corneal astigmatism is at most 2.0 dpt
Show all 10Show fewer
  • Axial length is between 22.0 and 26.0 mm
  • Able to obtain pupil dilation of at least 6.0 mm
  • Average keratometry values between 42.0 and 46.0 D
  • Subject understands the study requirements and the treatment procedures and provides written Informed Consent before any study-specific tests or procedures are performed

Who cannot join

  • Subject diagnosed with glaucoma of any type or having intraocular pressure above 24mmHg
  • Subject who has undergone previous intraocular surgery
  • Subject with corneal abnormality that would prevent stable and reliable refraction
  • Subject with weak or torn zonules
  • Subject with Pseudoexfoliation syndrome (PEX)
  • Subject with amblyopia
Show all 10Show fewer
  • Subject with retinal disease that in the opinion of the principal investigator would prevent stable and reliable refraction or might be worsened due to implantation of the device
  • Subject is diagnosed with active anterior segment intraocular inflammation
  • Subject is obligated to previous participation in another study with any investigational drug or device within the past 30 days
  • Subject is pregnant

Trial locations and contacts

1 site in 1 country.

Israel1 site, 1 recruiting
  • Wolfson Medical Center

    Holon, Tel Aviv

    Inbal Ratner, 035028702 ext. +972, inbalr@wmc.gov.il

    Guy Kleinmann, Prof.

    Recruiting

Other trials in Israel

About this study

The Hanita CleaRing device is an intraocular implant designed for inhibition of PCO leading to reduction in the Nd:YAG treatment rate. The device is implanted into the capsular bag after the extraction of the crystalline lens and prior to implantation of the IOL. The device is manufactured by Hanita Lenses from a hydrophilic acrylic co-polymer which is being used for the production of Intra-ocular lenses.

Read the detailed description

Cataract surgery has undergone significant improvement in terms of surgical technique, instrumentation and the quality of intraocular lenses (IOLs) over the past decades. Although the rate of posterior capsular opacification (PCO) has decreased, it is still the most common complication following uneventful cataract surgery, with the rate of laser capsulotomy ranging from 10% to 37%1-5 ND:YAG capsulotomy is a highly successful treatment, but it is not free of complications, such as increased intraocular pressure (IOP), pitting or dislocation of the IOL, cystoid macular edema, and retinal detachment. PCO also places an economic burden on the healthcare system. The importance of PCO prevention has increased in recent years, due to the expanding popularity of premium IOLs. Patients implanted with a premium IOL usually have high demands with regard to outcome, and PCO can also have an earlier effect on the performance of a premium IOL (i.e., multifocal IOLs).

Previous attempts to prevent PCO have included investigations of various IOL materials and designs, surgical techniques, and pharmacological materials. Unexpectedly lower PCO rates were recently noticed in eyes that were implanted with a special type of IOL, e.g., the Synchrony IOL (AMO, Santa Ana, CA, USA)8 and the FluidVision IOL (PowerVision, Belmont, CA, USA). It was hypothesized that the capsular bag stayed open due to the special design of these IOLs, and that this should reduce the PCO rate. Based upon this concept, a special open capsule ring was designed for intracapsular implantation in an attempt to maintain the capsular bag open and to reduce PCO rate. Safety and efficacy of the device was demonstrated through several animal studies. The tested devices were effective in reducing posterior capsule opacification and Soemmering's ring formation by 69% to 77% compared to control group. The Hanita CleaRing device is designed to be implanted in the capsular bag together with a standard posterior chamber IOL in a minimally invasive cataract surgery. The device is intended to reduce posterior capsular opacification leading to reduced Nd:YAG treatment rate. The raw material of the device is a hydrophilic co-polymer, considered to be biocompatible and practically inert to the eye tissue. The objective of this study is to evaluate the safety and the efficacy of CleaRing device in cataract patients.

Being tested
CleaRing
Keywords
Posterior capsule opacification, Cataract, Ring
Medical subject headings
Cataract

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • CleaRingDevice

    The Hanita CleaRing device is an intraocular implant designed for inhibition of PCO leading to reduction in the Nd:YAG treatment rate. The device is implanted into the capsular bag after the extraction of the crystalline lens and prior to implantation of the IOL. The device is manufactured by Hanita Lenses from a hydrophilic acrylic co-polymer which is being used for the production of Intra-ocular lenses.

Study groups in NCT06083025
Study groupRoleWhat they receive
CleaRing device for cataract patientsExperimentalAll subjects who meet eligibility requirements will be asked to participate. Subjects will be considered enrolled into the study after: 1. A signed Informed Consent has been obtained. 2. The subject has met all of the inclusion and none of the exclusion criteria. 3. The ability of the subject to safely undergo cataract surgery under local anesthesia was confirmed by the investigator, according to standard procedure of the medical center.
Receives: Device: CleaRing

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. The first primary objective of this study is to evaluate the influence of Clearing open-capsule device on refractive predictability in cataract patients.

    Refractive will be evaluated by Automated refraction

    Measured: 1, 3 and 5 years

  2. The second primary objective of this study is to evaluate the influence of Clearing open-capsule device on refractive predictability in cataract patients.

    Refractive will be evaluated by subjective refraction

    Measured: 1, 3 and 5 years

  3. The third primary objective of this study is to evaluate the influence of Clearing open-capsule device on refractive predictability in cataract patients.

    Refractive will be evaluated by visual acuity measurement.

    Measured: 1, 3 and 5 years

Secondary outcomes (1)
  1. Secondary objective of this study is to evaluate the influence of the device on PCO rate

    PCO will be evaluated by using Biomicroscopy (Slit Lamp)

    Measured: 1, 3 and 5 years

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted13 Oct 2023
  2. Study start15 Apr 2023
  3. Primary completion1 Mar 2025 (estimated)
  4. Study completion1 Apr 2025 (estimated)

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Treatment
Allocation
Not applicable
Intervention model
Single group
Model description
This study is a prospective, single group, single center, open label study. Each study subject will be implanted bilaterally with IOL of surgeon's choice (same IOL in both eyes) and CleaRing device in one eye.
Masking
None (open label)
Enrolment
25 participants (estimated)
IDs, dates and publications
ClinicalTrials.gov ID
NCT06083025
Sponsor's protocol ID
CR-0719 Part 2
Acronym
CleaRing
Status last verified
Oct 2023
First submitted
5 Oct 2023
Last updated
13 Oct 2023
Results posted
No
Data monitoring committee
Yes
FDA-regulated drug
No
FDA-regulated device
No
Individual data sharing
No

Questions about this trial

What is this clinical trial studying?

A study testing CleaRing in people with Posterior Capsule Opacification, run by Hanita Lenses at 1 site in Israel.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 10 inclusion and 10 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: status not confirmed, as of the sponsor's last update on 13 Oct 2023. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in Israel. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

Results are not available in the source record.

How can a doctor refer a patient or contact the research team?

Contact Inbal Ratner (035028702 ext. +972, inbalr@wmc.gov.il), or a site near the patient. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

The first primary objective of this study is to evaluate the influence of Clearing open-capsule device on refractive predictability in cataract patients. (time frame: 1, 3 and 5 years). The study lists 3 primary outcomes in all. It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 13 Oct 2023.

Free account

Keep searching with a free account

You’ve used your 3 free searches for this session. Log in or create a free AzTrials account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in