REmote CArdiac MOnitoring by the Corsano CardioWatch 287-2 Evaluation Study

An observational study of Atrial Fibrillation and Rhythm; Disorder, run by Corsano Health B.V. at 1 site in Netherlands.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

In this study, a wristband that monitors hearth rhythm and blood pressure is assessed in free living conditions. Two groups of patients are being recruited: patients receiving an EKG holter and patients receiving an automatic blood pressure monitor. Patients wear the wristband for 28 days. Read more about this study

Who is it for?
Ages 18 years and older
What is being tested?
Corsano CardioWatch 287-2
Type of treatment
Medical device
How many people
190 participants
Where
1 siteNetherlands
Study dates
15 Jun 2023to 28 Dec 2023
Phase
Observational
Are there results?
None posted in the registry

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 years and older
  • All sexes
  • No healthy volunteers
  • Adult, Older adult

Study population: One group (A) receiving an EKG monitoring holter for the duration of 24 hours as part of standard care for atrial fibrillation screening. Besides, one group (B) receiving an automatic blood pressure cuff for the duration of 24 hours as part of standard care for blood pressure evaluation.

Who may be able to join

  • ≥ 18 years old;
  • able to provide consent;
  • receiving EKG holter or automatic blood pressure cuff for home monitoring per doctor prescription

Who cannot join

  • Unable to wear the Corsano CardioWatch 287 due to reasons such as allergic reactions, wounds, amputations etc.;
  • Unable to receive blood pressure measurements per cuff due to lymphedema, amputation, dialysis shunt, wounds, etc.;
  • Pregnant women;
  • Breastfeading women;
  • Upper arm circumference not within the cuff range (22-42 cm)
  • Unable or not willing to sign informed consent;
  • Significant mental or cognitive impairment.

Trial locations and contacts

1 site in 1 country.

Netherlands1 site
  • Reinier de Graaf Gasthuis

    Delft, South Holland, 2625 AD

Other trials in Netherlands

About this study

Official title: Remote Cardiac Monitoring by the Corsano CardioWatch 287-2 Evaluation Study

In this study, a wristband that monitors hearth rhythm and blood pressure is assessed in free living conditions.

Two groups of patients are being recruited: patients receiving an EKG holter and patients receiving an automatic blood pressure monitor. Patients wear the wristband for 28 days. During this time, the number of atrial fibrillation episodes detected by the wristband is being counted, which is to be compared with the number of episodes detected by the holter. Also, blood pressure is measured continuously, which is to be compared with daily blood pressure measurements using a conventional blood pressure cuff.

Read the detailed description

Background of the study:

Wearables have the potential to monitor patients remotely. The Corsano CardioWatch 287-2 is such a medical device that can monitor atrial fibrillation and long-term blood pressure. The device has been validated using clinical trials in hospitals, but validation in the intended remote setting is lacking.

Objective of the study:

Primary objective: To compare the number of episodes of atrial fibrillation detected by the Corsano CardioWatch 287-2 during 28 days of use with the number of episodes of atrial fibrillation detected by conventional Holter monitoring during 48 hours of use (standard care).

Secondary objective: To assess the difference in blood pressure measurements obtained by the Corsano CardioWatch 287-2 and the conventional cuff blood pressure monitor over a period of 28 days; to assess the usability of the Corsano CardioWatch 287-2 from a patient perspective.

Study design:

Group A:

Conventional 24 hour EKG holter monitoring will be compared to 28 day continuous monitoring by the Corsano CardioWatch 287-2 to assess whether the percentage of patients in whom atrial fibrillation is detected will increase. Detection of atrial fibrillation with the Corsano CardioWatch 287-2 is done by optical photoplethysmography (PPG), after which atrial fibrillation is confirmed with a 30-second ECG measurement.

Additionally, conventional cuff blood pressure will be compared to blood pressure measured by the Corsano CardioWatch 287-2 to validate free-living blood pressure monitoring. Patients will measure their blood pressure daily with a cuff blood pressure measurement device. The measurements at day 7, 14, 21 will be used to reinforce the Corsano blood pressure algorithm. The remaining measurements will be used to compare the blood pressure with the values measured by the Corsano CardioWatch 287-2 to assess accuracy.

Group B:

Conventional automatic cuff blood pressure measurement will be compared with the Corsano CardioWatch 287-2 for 24-48 hours. The automatic cuff blood pressure monitor will take a blood pressure reading every 30 minutes. Each reading will be compared with the Corsano CardioWatch 287-2 reading.

Study population:

Patients receiving a 24 hour EKG holter or an ambulant blood pressure device after visiting the cardiologist's outpatient clinic.

Primary study parameters/outcome of the study:

Absolute percentage increase of patients in whom at least one event of atrial fibrillation is detected by the Corsano CardioWatch 287-2 during 28 days of monitoring, compared to a conventional EKG holter during 24-48 hours of monitoring.

Secundary study parameters/outcome of the study (if applicable):

  1. Mean blood pressure difference and its SD between blood pressure measured by the Corsano CardioWatch 287-2 and blood pressure measured by a conventional oscillometric blood pressure cuff.
  2. Usability of the Corsano CardioWatch 287-2 in a remote care setting as determined by a questionnaire.

Nature and extent of the burden and risks associated with participation, benefit and group relatedness:

At the cardiologist's office, Group A and B patients receive 3 blood pressure measurements. These are performed by a nurse. Group A will take daily blood pressure readings of themselves at home. If atrial fibrillation is detected by the wristband, Group A patients will be asked to take an ECG. Finally, Group A patients will be asked once to complete a questionnaire. Group B only has to wear the wristband for 24 to 48 hours as an additional burden in addition to the aforementioned initial measurements.

Being tested
Corsano CardioWatch 287-2
Keywords
wristband, wearable, blood pressure, ekg, atrial fibrillation, hypertension, comparison
Medical subject headings
Atrial Fibrillation, Arrhythmias, Cardiac, Hypertension, Hypotension

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Corsano CardioWatch 287-2Device

    Patients receive either an EKG holter or an automatic blood pressure cuff, as prescribed by their cardiologist. The readings from these devices (atrial fibrillation detection and blood pressure) will be compared to the readings from the CardioWatch 287-2.

Study groups in NCT05899959
Study groupRoleWhat they receive
Patients receiving an EKG holterpatients receiving an EKG holter at the outpatient clinic for 24-48 hours of remote monitoring
Receives: Device: Corsano CardioWatch 287-2
Patients receiving an automatic blood pressure cuffpatients receiving an automatic blood pressure cuff at the outpatient clinic for 24 hours of remote monitoring
Receives: Device: Corsano CardioWatch 287-2

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Atrial fibrillation detection rate

    Absolute percentage increase of patients in whom at least one event of atrial fibrillation is detected by the Corsano CardioWatch 287-2 during 28 days of monitoring, compared to a conventional EKG holter during 24-48 hours of monitoring.

    Measured: 28 days

  2. Blood pressure accuracy

    Mean systolic and diastolic blood pressure difference and its SD between blood pressure measured by the Corsano CardioWatch 287-2 and blood pressure measured by a conventional oscillometric blood pressure cuff.

    Measured: 28 days (EKG holter group) or 1-2 days (Automatic blood pressure group)

Secondary outcomes (2)
  1. Blood pressure accuracy

    Bias and limits of agreement between systolic and diastolic blood pressure measured by the Corsano CardioWatch 287-2 and blood pressure measured by a conventional oscillometric blood pressure cuff.

    Measured: 28 days (EKG holter group) or 1-2 days (Automatic blood pressure group)

  2. Usability questionnaire

    Usability of the Corsano CardioWatch 287-2 in a remote care setting as assessed by a self-developed questionnaire

    Measured: 28 days (EKG holter group) or 1-2 days (Automatic blood pressure group)

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted12 Jun 2023
  2. Study start15 Jun 2023
  3. Primary completion28 Dec 2023
  4. Study completion28 Dec 2023

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Observational
Observational model
Cohort
Time perspective
Prospective
Enrolment
190 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT05899959
Sponsor's protocol ID
RECAMO
Acronym
RECAMO
Status last verified
Aug 2024
First submitted
2 Jun 2023
Last updated
26 Aug 2024
Results posted
No
Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
No
Individual data sharing
No

Publications

  • van Vliet M, Aalberts JJJ, Hamelinck C, Hauer AD, Hoftijzer D, Monnink SHJ, Schipper JC, Constandse JC, Peters NS, Lip GYH, Steinhubl SR, Ronner E. Assessment of photoplethysmography-based blood pressure determinations during long-term and short-term remote cardiac monitoring: the RECAMO study. Eur Heart J Digit Health. 2025 Mar 27;6(4):763-771. doi: 10.1093/ehjdh/ztaf027. eCollection 2025 Jul.

    PubMed 40703134
  • van Vliet M, Aalberts JJJ, Hamelinck C, Hauer AD, Hoftijzer D, Monnink SHJ, Schipper JC, Constandse JC, Peters NS, Lip GYH, Steinhubl SR, Ronner E. Ambulatory atrial fibrillation detection and quantification by wristworn AI device compared to standard holter monitoring. NPJ Digit Med. 2025 Mar 25;8(1):177. doi: 10.1038/s41746-025-01555-9.

    PubMed 40133622

Questions about this trial

What is this clinical trial studying?

An observational study of Atrial Fibrillation and Rhythm; Disorder, run by Corsano Health B.V. at 1 site in Netherlands.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 3 inclusion and 7 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 26 Aug 2024. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in Netherlands. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Atrial fibrillation detection rate (time frame: 28 days). The study lists 2 primary outcomes in all. It is what the study was designed to measure, not a finding.

Where can I read the original study information?

On the ClinicalTrials.gov record for NCT05899959, which also lists 2 linked publications.

Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 26 Aug 2024.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in