Clinical Relevance of the HD sEMG Signals in the Post-stroke Rehabilitation

A study testing Acquisition of HD sEMG signals using the WPM-SEMG device in people with Stroke, run by Alain Kaelin at 1 site in Switzerland.

Unknown status

Status not confirmed. The sponsor has not verified this record for over two years, so it may be out of date.

Official record
Search all trials

Trial summary

The goal of this clinical trial is to learn about the WPM-SEMG-V2 device (Wireless Portable and Multi-Channel, WPM) designated to acquire high-density (HD) surface electromyography (sEMG) signals in healthy volunteers and post-stroke patients. Read more about this study

Who is it for?
Ages 18 to 99 yearsHealthy volunteers can take part
Conditions the study is for
Stroke
What is being tested?
Acquisition of HD sEMG signals using the WPM-SEMG device
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
40 participants
Where
1 siteSwitzerland
Study dates
13 Mar 2023to Apr 2026
Phase
Interventional
Are there results?
None posted in the registry

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 to 99 years
  • All sexes
  • Healthy volunteers accepted
  • Adult, Older adult

Who may be able to join

  • Healthy volunteers
  • Written Informed Consent
  • Female or male
  • 18 years or older
  • Ability to understand the investigation
  • Willingness to complete all the investigation assessments
Show all 21Show fewer
  • Ability to perform the procedures of the investigation
  • Ability to perform maximal voluntary muscle extension of the target muscle and limb
  • Additional inclusion criteria only for the recreationally active or athletic, healthy volunteers :
  • Subjects who participated in at least 150 min of moderate activity per week over the last six months.
  • between 18 and 20 years (preferably)
  • Post-stroke patients
  • Written informed consent
  • Female or male
  • 18 years or older
  • Have a first-time stroke
  • Ability to perform maximal voluntary muscle extension of the target muscle and limb
  • Preserved cognitive capacity to perform the task
  • Ability to understand the investigation
  • Willingness to complete all the investigation assessments
  • Ability to perform the procedures of the investigation

Who cannot join

  • Healthy volunteers
  • Any significant acute disease state
  • Skin disorders/allergies at the site of contact with the investigational device
  • History of skin disease
  • Known hypersensitivity or allergy to component of the investigational product (e.g. metal alloys) or other materials used (e.g., abrasive gel, foam with polystyrene markers)
  • Chronic use of medications or treatment
Show all 15Show fewer
  • Post-stroke patients
  • History of spinal cord injury or traumatic brain damage
  • Serious medical illness that precludes performing the task
  • Severe locomotion disorder due to other causes
  • Severe neurological disease other than stroke
  • Any significant acute disease state
  • Skin disorders/allergies at the site of contact with the investigational device
  • History of skin disease
  • Known hypersensitivity or allergy to component of the investigational product (e.g. metal alloys) or other materials used (e.g., abrasive gel, foam with polystyrene markers)

Trial locations and contacts

1 site in 1 country.

Alain Kaelin, Prof. contact

+41 91 811 62 57

alain.kaelin@eoc.ch

Switzerland1 site, 1 recruiting
  • Istituto di Neuroscienze Cliniche della Svizzera Italiana

    Lugano

    Alain Kaelin, Prof., +41 091 811 62 57, alain.kaelin@eoc.ch

    Recruiting

Other trials in Switzerland

About this study

Official title: Assistance and Evaluation of the Post-stroke Rehabilitation Process Via the Analysis of High-Density Surface EMG Signals Acquired With the WPM-SEMG-V2 Device: A Single Centre, Prospective, Feasibility Investigation in Healthy Volunteers and Post-stroke Patients

The goal of this clinical trial is to learn about the WPM-SEMG-V2 device (Wireless Portable and Multi-Channel, WPM) designated to acquire high-density (HD) surface electromyography (sEMG) signals in healthy volunteers and post-stroke patients. The main objectives are: 1) to technically and clinically validate the WPM-SEMG-V2 device in healthy volunteers; 2) to demonstrate the clinical relevance of the HD sEMG signals to support the process of rehabilitation of post-stroke patients. Participants will be asked to perform simple exercises using the muscles of the hand or the leg in static or dynamic contractions while wearing the WPM-SEMG-V2 device.

Read the detailed description

The participant will be instructed how to execute static or the dynamic exercises while wearing the WPM-SEMG-V2 device. The WPM-SEMG-V2 device will be initially tested in a small number of healthy volunteers to evaluate the technical and clinical performance of the device in different settings (i.e., static and dynamic muscle contractions). Preliminary information gathered in the first part of the investigation may eventually guide further device modifications. The second part will capture preliminary clinical performance of the device in post-stroke patients. The main aim is to study the behaviour of the neuromuscular system in voluntary muscle activations and demonstrate the clinical relevance of the HD sEMG signals to support the process of rehabilitation of post-stroke subjects.

Conditions
Stroke
Being tested
Acquisition of HD sEMG signals using the WPM-SEMG device
Medical subject headings
Stroke

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Acquisition of HD sEMG signals using the WPM-SEMG deviceDevice

    The WPM-SEMG-V2 device will be placed over the muscle of the participant to be investigated. A reference electrode over a bone close to the muscle of investigation will be applied. A sensor combing an accelerometer, gyroscope and magnetometer integrated in the WPM-SEMG-V2 device will be also used for the acquisition of body movement measures. An additional component for trapezoidal force measurement will be used during the acquisition in static contractions to measure the force. The participant will be instructed by the investigator how to execute the static or the dynamic exercises. Visual feedback will help the subject to track the trajectory.

Study groups in NCT05849896
Study groupRoleWhat they receive
Static contractions in healthy volunteersExperimentalAcquisition of HD SEMG signals of an upper limb muscle and a lower limb muscle while the participant executes simple static exercises.
Receives: Device: Acquisition of HD sEMG signals using the WPM-SEMG device
Dynamic contractions in healthy volunteersExperimentalAcquisition of HD SEMG signals of a leg muscle while the participant executes simple dynamic exercises.This arm will recruited recreationally active or athletic volunteers, preferably between 18 and 20 years
Receives: Device: Acquisition of HD sEMG signals using the WPM-SEMG device
Static contractions in post-stroke patientsExperimentalAcquisition of HD SEMG signals of a lower limb muscle while the participant executes simple static exercises.
Receives: Device: Acquisition of HD sEMG signals using the WPM-SEMG device

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Technical performance of the device (0 to 2)

    Assessed during signal acquisitions in static and/or dynamic contractions in healthy volunteers by checking the following parameters:

    1. Interaction (0 or 1): when the WPM-SEMG-V2 is connected to the component for trapezoidal force measurement, both the sEMG signals and force data are recorded by the device on the device's SD card.
    2. Availability (0 or 1): all data recordings are stored separately in the SD card.
    3. Integrity of the final data (0 or 1): the recorded signals have the same duration of the recordings after post-processing of the data timestamp.

    The technical performance of the device is defined as Interaction*(availability+integrity) in static contractions.

    The technical performance of the device is the sum of the scores for availability and integrity in dynamic contractions.

    Measured: 1 day

  2. Quality of the WPM-SEMG-V2 device in the acquisition of HD sEMG signals (0 to 2)

    Assessed during signal acquisitions in static and/or dynamic contractions, both in healthy volunteers and post-stroke patients by measuring the following parameters:

    1. Signal amplitude (0 or 1): the peak-to-peak sEMG signal amplitude is within the anatomical range of few-tens of µV to 2 mV.
    2. Signal spectrum (0 or 1): the sEMG signal spectrum is within the anatomical range (sEMG power is mostly accounted for by the frequency components < 350 Hz).

    Quality of the WPM-SEMG-V2 device in the acquisition of HD sEMG signals is the sum of the scores given to signal amplitude and signal spectrum.

    Measured: 30 days

  3. Correlation between the properties of populations of Motor Unit (MU) with the type of the stroke.

    Assessed during signal acquisitions in static contractions in post-stroke patients.

    The following properties will be extracted from the MUs populations: discharge rate (pulses/seconds), recruitment threshold (percentage of the maximum voluntary force contraction), motor unit conduction velocity (meter/seconds), Muscle Fiber Conduction Velocity (meter/seconds. Type of stroke: acute or subacute.

    Measured: 30 days

  4. Correlation between the properties of populations of Motor Unit (MU) with the severity of the stroke.

    Assessed during signal acquisitions in static contractions in post-stroke patients.

    The following properties will be extracted from the MUs populations: discharge rate (pulses/seconds), recruitment threshold (percentage of the maximum voluntary force contraction), motor unit conduction velocity (meter/seconds), Muscle Fiber Conduction Velocity (meter/seconds. The severity of the stroke will be measured by the National Institutes of Health Stroke Scale (score 0-42).

    Measured: 30 days

  5. Correlation between the properties of populations of Motor Unit (MU) with the location of the stroke.

    Assessed during signal acquisitions in static contractions in post-stroke patients.

    The following properties will be extracted from the MUs populations: discharge rate (pulses/seconds), recruitment threshold (percentage of the maximum voluntary force contraction), motor unit conduction velocity (meter/seconds), Muscle Fiber Conduction Velocity (meter/seconds. The location of the stroke is obtained from MRI imaging (left hemispheric, right hemisferic, cortical, subcortical, brain stem).

    Measured: 30 days

  6. Percentage of tracked MU

    Defined as the number of tracked MU / number of detected MUs in the baseline session, among the baseline, medium term, and long term sessions. Assessed during signal acquisitions in static contractions, both in healthy volunteers and post-stroke patients

    Measured: 30 days

Secondary outcomes (1)
  1. Incidence of serious and non-serious adverse event, adverse device effect, device deficiency and malfunction

    As assessed by reported adverse event.

    Measured: 30 days

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted9 May 2023
  2. Study start13 Mar 2023
  3. Primary completionFeb 2026 (estimated)
  4. Study completionApr 2026 (estimated)

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Other
Allocation
Non-randomised
Intervention model
Sequential
Masking
None (open label)
Enrolment
40 participants (estimated)
IDs, dates and publications
ClinicalTrials.gov ID
NCT05849896
Sponsor's protocol ID
SUPSI-SEMG-002
Status last verified
Sep 2024
First submitted
29 Mar 2023
Last updated
5 Sep 2024
Results posted
No
FDA-regulated drug
No
FDA-regulated device
No
Individual data sharing
Undecided

Questions about this trial

What is this clinical trial studying?

A study testing Acquisition of HD sEMG signals using the WPM-SEMG device in people with Stroke, run by Alain Kaelin at 1 site in Switzerland.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 21 inclusion and 15 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: status not confirmed, as of the sponsor's last update on 5 Sep 2024. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in Switzerland. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

Results are not available in the source record.

How can a doctor refer a patient or contact the research team?

Contact Alain Kaelin, Prof. (+41 91 811 62 57, alain.kaelin@eoc.ch), or a site near the patient. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Technical performance of the device (0 to 2) (time frame: 1 day). The study lists 6 primary outcomes in all. It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 5 Sep 2024.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in