HypoxiaNCT05532670
N600X Low Saturation Accuracy Validation
An observational study of Hypoxia, run by Medtronic - MITG at 1 site in United States.
Trial summary
To compare pulse oximetry saturation measurements, during normal to low saturation, to saturation measurements made by a multi-wavelength CO-oximeter, taken from arterial blood samples from healthy human subjects. Read more about this study
- Who is it for?
- Ages 18 to 50 yearsHealthy volunteers can take part
- Conditions the study is for
- Hypoxia
- Type of treatment
- Medical device
- How many people
- 15 participants
- Where
- 1 siteUnited States
- Study dates
- 29 Sep 2022to 1 Nov 2022
- Phase
- Observational
- Are there results?
- None posted in the registry
Key requirements
Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.
- Ages 18 to 50 years
- All sexes
- Healthy volunteers accepted
- Adult
Study population: Study Population The study population includes a 15 healthy non-smoking competent adults 18-50 years of age. With an aim of diversity across ethnicity, race and sex of varying skin tones.
Who may be able to join
- 1. Subject must have the ability to understand and provide written informed consent
- 2 Adult subjects 18 to 50 years of age.
- 3. Subject must be willing and able to comply with study procedures and duration.
- 4. Subject is a non-smoker or who has not smoked within 2 days prior to the study.
- 5. Any gender of any race.
- 6. Cleared same day Health Assessment form and health screening.
- 7. Cleared physical exam by a licensed physician, physician assistant, or advanced practice nurse (medical exam includes 12-lead ECG, medical history and blood test to show no sickle cell disease present)
- 8. Successful Perfusion Index Ulnar/Ulnar+Radial Ratio test showing adequate collateral blood flow.
Who cannot join
- Subject with sickle cell disease.
- Subject is considered as being morbidly obese (defined as BMI >39.5)
- Compromised circulation, injury, or physical malformation of fingers, hands, ears or forehead/skull or other sensor sites which would limit the ability to test sites needed for the study. (Note: Certain malformations may still allow subjects to participate if the condition is noted and would not affect the sites utilized.)
- Females of childbearing potential, who are pregnant, who are trying to get pregnant or who have a urine test positive for pregnancy on the day of the study
- Subjects with COHb levels >3% as assessed by CO-Oximetry during the procedure
- tHb < 10 g/dl as assessed by CO-Oximetry during the procedure
Show all 50Show fewer
- MetHb ≥ 2% as assessed by CO-Oximetry during the procedure
- Subjects with known respiratory conditions such as:
- uncontrolled / severe asthma,
- flu or influenza type infection
- pneumonia / bronchitis,
- shortness of breath / respiratory distress,
- unresolved respiratory or lung surgery,
- emphysema, COPD, lung disease
- Subjects with known heart or cardiovascular conditions such as:
- Hypertension: systolic >140mmHg, or Diastolic >90mmHg on 3 consecutive readings.
- have had cardiovascular surgery
- Chest pain (angina)
- heart rhythms other than a normal sinus rhythm or respiratory sinus arrhythmia
- previous heart attack
- blocked artery
- unexplained shortness of breath
- congestive heart failure (CHF)
- history of stroke
- transient ischemic attack
- carotid artery disease
- myocardial ischemia
- myocardial infarction
- cardiomyopathy
- Self-reported health conditions as identified in the Health Assessment Form
- diabetes,
- uncontrolled thyroid disease,
- kidney disease / chronic renal impairment,
- history of seizures (except childhood febrile seizures),
- epilepsy,
- history of unexplained syncope,
- recent history of frequent migraine headaches,
- recent symptomatic head injury, within the last 2 months
- Subjects with known clotting disorders
- history of bleeding disorders or personal history of prolonged bleeding from injury
- history of blood clots
- hemophilia
- current use of blood thinner: prescription or daily use of aspirin
- Subjects with Severe contact allergies to standard adhesives, latex or other materials found in pulse oximetry sensors, ECG electrodes, respiration monitor electrodes or other medical sensors
- Subjects with prior or known severe allergies to lidocaine (or similar pharmacological agents, e.g. Novocain) or heparin
- Arterial cannulation within the last 30 days prior to study date, (this may exclude only one radial artery site. Left or right)
- History of clinically significant complications from previous arterial cannulation.
- A radial artery with ten or more arterial cannulations right or left, excludes that site.
- Unwillingness or inability to remove colored nail polish or colored artificial nails other than clear from test digits.
- Other known health condition, should be considered upon disclosure in Health Assessment form
Trial locations and contacts
1 site in 1 country.
About this study
Official title: SpO2 Accuracy Validation of N600x Sensors Via Reference CO-Oximetry in Healthy, Well-Perfused Subjects
To compare pulse oximetry saturation measurements, during normal to low saturation, to saturation measurements made by a multi-wavelength CO-oximeter, taken from arterial blood samples from healthy human subjects.
Read the detailed description
The study utilizes a single-site, physiology laboratory (Clinimark) to conduct this Post Market, non-randomized, prospective studies. The goal is to enroll a minimum of fifteen subjects. Subject pool goal of diversity across ethnicity, race and sex of varying skin tones to meet the study design requirements.
The general purpose of Invasive Controlled Desaturation Studies is to support the SpO2 accuracy of the post market pulse oximetry sensors and paired monitor in comparison to reference-standard measurements of blood SaO2 by a CO-oximeter during low saturation conditions. This is achieved through paired observations of SpO2 and SaO2 values over the specified SpO2 accuracy range of 60% to 100% SaO2 of the pulse oximeter on a group of healthy adult volunteers. The fraction of inspired oxygen (FiO2) delivered to test subjects is varied to achieve a series of targeted steady-state saturation periods. Arterial blood samples are periodically taken from an indwelling arterial catheter for use in the comparison. Pulse rate accuracy will be evaluated during the same data collection period as SpO2. Pulse rate will be compared to reference ECG heart rate.
- Conditions
- Hypoxia
- Being tested
- Non-Invasive Controlled Hypoxia
- Keywords
- Oxygen Saturation, Invasive Controlled Desaturation, Pulse Oximetry, Desaturation of Blood
- Medical subject headings
- Hypoxia
What participants receive
What participants receive. In a randomised study each person is assigned to one group by chance.
Non-Invasive Controlled HypoxiaDevice
Also called Desaturation, Blood Sampling
Following arterial cannulation, hypoxia is induced by reducing the percent of inspired oxygen the subject breathes while simultaneous measurements are recorded from the test pulse oximeters and arterial blood samples are collected at targeted levels of saturation.
What is being measured
The primary outcome is the main question the study is designed to answer.
Comparison of pulse rate (BPM) by a pulse oximeter monitor and sensor, compared to ECG pulse rate measurement
To validate the proposed claims for pulse rate (BPM) accuracy in diverse subject population during Low Saturation over a specified saturation ranges.
Measured: October 2022
Comparison of SPO2 saturation levels by a pulse oximeter to saturation measurements made by a multi-wavelength CO-oximeter.
To validate the proposed claims for SPO2 saturation accuracy in diverse subject population during Low Saturation over a specified saturation ranges.
Measured: October 2022
Timeline
From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.
- First posted8 Sep 2022
- Study start29 Sep 2022
- Primary completion7 Oct 2022
- Study completion1 Nov 2022
Resources
- Official record on ClinicalTrials.govThe registry entry this page is built from, with its full change history.
- Other trials for HypoxiaStudies of the same condition, open and under way.
- What the trial phases meanHow a treatment moves from first tests in people to approval.
More details
For doctors and researchers: how the study is designed, who runs it, and its registry IDs.
Study design
- Study type
- Observational
- Observational model
- Other
- Time perspective
- Prospective
- Enrolment
- 15 participants (actual)
Sponsor and investigators
- Lead sponsor
- Medtronic - MITG
- Sponsor type
- Industry (pharma, biotech, devices)
- Responsible party
- Sponsor
- Investigators
- Arthur Cabrera, MD, Principal investigator, Clinimark, LLC
IDs, dates and publications
- ClinicalTrials.gov ID
- NCT05532670
- Sponsor's protocol ID
- MDT22013LSOXMN
- Status last verified
- Nov 2022
- First submitted
- 31 Aug 2022
- Last updated
- 23 Nov 2022
- Results posted
- No
- Data monitoring committee
- No
- FDA-regulated drug
- No
- FDA-regulated device
- Yes
- Individual data sharing
- No: All participant data will be analyzed and outlined in final study report. Individual participant data will only be made available to the participant and or participant primary care physician as applicable.
Study at a glance
From the registry record, last updated by the sponsor on 23 Nov 2022. Recruitment status is as the sponsor reported it: confirm with the study team.
- Official title
- SpO2 Accuracy Validation of N600x Sensors Via Reference CO-Oximetry in Healthy, Well-Perfused Subjects
- ClinicalTrials.gov ID
- NCT05532670
- Acronym / sponsor ID
- MDT22013LSOXMN
- Conditions
- Hypoxia
- Keywords
- Oxygen Saturation, Invasive Controlled Desaturation, Pulse Oximetry, Desaturation of Blood
- Study type
- Observational
- Recruitment status
- Completed
- Enrolment
- 15 participants (actual)
- Lead sponsor
- Medtronic - MITGIndustry (pharma, biotech, devices)
- Study start
- 29 Sep 2022 (actual)
- Primary completion
- 7 Oct 2022 (actual)
- Study completion
- 1 Nov 2022 (actual)
- First posted
- 8 Sep 2022 (actual)
- Results
- Not posted in the source record
Clinical summary
The research question as the sponsor registered it. This describes what the study plans to test, not what it found.
To compare pulse oximetry saturation measurements, during normal to low saturation, to saturation measurements made by a multi-wavelength CO-oximeter, taken from arterial blood samples from healthy human subjects.
Detailed description
The study utilizes a single-site, physiology laboratory (Clinimark) to conduct this Post Market, non-randomized, prospective studies. The goal is to enroll a minimum of fifteen subjects. Subject pool goal of diversity across ethnicity, race and sex of varying skin tones to meet the study design requirements.
The general purpose of Invasive Controlled Desaturation Studies is to support the SpO2 accuracy of the post market pulse oximetry sensors and paired monitor in comparison to reference-standard measurements of blood SaO2 by a CO-oximeter during low saturation conditions. This is achieved through paired observations of SpO2 and SaO2 values over the specified SpO2 accuracy range of 60% to 100% SaO2 of the pulse oximeter on a group of healthy adult volunteers. The fraction of inspired oxygen (FiO2) delivered to test subjects is varied to achieve a series of targeted steady-state saturation periods. Arterial blood samples are periodically taken from an indwelling arterial catheter for use in the comparison. Pulse rate accuracy will be evaluated during the same data collection period as SpO2. Pulse rate will be compared to reference ECG heart rate.
Study design
- Observational model
- Other
- Time perspective
- Prospective
Who may qualify
- Ages
- 18 to 50 years
- Sex
- All
- Healthy volunteers
- Accepted
- Age groups
- Adult
- Study population
- Study Population The study population includes a 15 healthy non-smoking competent adults 18-50 years of age. With an aim of diversity across ethnicity, race and sex of varying skin tones.
Key inclusion criteria (8 listed)
- 1. Subject must have the ability to understand and provide written informed consent
- 2 Adult subjects 18 to 50 years of age.
- 3. Subject must be willing and able to comply with study procedures and duration.
- 4. Subject is a non-smoker or who has not smoked within 2 days prior to the study.
- 5. Any gender of any race.
Key exclusion criteria (50 listed)
- Subject with sickle cell disease.
- Subject is considered as being morbidly obese (defined as BMI >39.5)
- Compromised circulation, injury, or physical malformation of fingers, hands, ears or forehead/skull or other sensor sites which would limit the ability to test sites needed for the study. (Note: Certain malformations may still allow subjects to participate if the condition is noted and would not affect the sites utilized.)
- Females of childbearing potential, who are pregnant, who are trying to get pregnant or who have a urine test positive for pregnancy on the day of the study
- Subjects with COHb levels >3% as assessed by CO-Oximetry during the procedure
Eligibility is decided by the study team after screening; this overview does not establish that a patient qualifies. Full inclusion and exclusion criteria.
Interventions
| Intervention | Type | Arms |
|---|---|---|
| Non-Invasive Controlled HypoxiaAlso: Desaturation, Blood SamplingFollowing arterial cannulation, hypoxia is induced by reducing the percent of inspired oxygen the subject breathes while simultaneous measurements are recorded from the test pulse oximeters and arterial blood samples are collected at targeted levels of saturation. | Device | – |
Primary outcome
Comparison of pulse rate (BPM) by a pulse oximeter monitor and sensor, compared to ECG pulse rate measurement
Time frame: October 2022
To validate the proposed claims for pulse rate (BPM) accuracy in diverse subject population during Low Saturation over a specified saturation ranges.
Comparison of SPO2 saturation levels by a pulse oximeter to saturation measurements made by a multi-wavelength CO-oximeter.
Time frame: October 2022
To validate the proposed claims for SPO2 saturation accuracy in diverse subject population during Low Saturation over a specified saturation ranges.
Results and safety
The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.
Research links
- Official record on ClinicalTrials.govLast updated by the sponsor 23 Nov 2022.
- Individual participant dataSharing planned: No. All participant data will be analyzed and outlined in final study report. Individual participant data will only be made available to the participant and or participant primary care physician as applicable.
Related trials for Hypoxia
Recruiting
Influence of Intermittent Hypoxia on Loop Gain in Healthy Subjects
Intermittent Hypoxia, Healthy Volunteers
Active, not recruitingPhase 1/2
A Study of Belzutifan (MK-6482) in Combination With Palbociclib Versus Belzutifan Monotherapy in Participants With Advanced Renal Cell Carcinoma (MK-6482-024/LITESPARK-024)
Renal Cell Carcinoma
Not yet recruiting
Levetiracetam as an Adjuvant Therapy to Hypothermia in Neonatal Asphyxia
Hypoxic Ischemic Encephalopathy of Newborn, Neuro-protection
RecruitingPhase 1/2
Substudy 03C: A Study of Combination Therapies in Participants With Renal Cell Carcinoma With Recurrent Disease During or After Anti-PD-(L)1 Therapy (MK-3475-03C/KEYMAKER-U03)
Renal Cell Carcinoma
Recruiting
Amnioinfusion for Chorioamnionitis: Targeting Neonatal Brain Injury Biomarkers
Chorioamnionitis Affecting Fetus or Newborn, Amnioinfusion
Referral and contact
AzTrials does not enrol participants or take referrals. Contact the study team or a site listed below; they decide who can take part. This study is not recruiting at the moment.
Arthur Cabrera, MD principal investigator
Clinimark, LLC
1 site in 1 country: see every site and its status.
Questions about this trial
What is this clinical trial studying?
An observational study of Hypoxia, run by Medtronic - MITG at 1 site in United States.
The full registered description is in the About this study section and the official record.
Who may be eligible to take part?
The sponsor lists 8 inclusion and 50 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.
Is this clinical trial currently recruiting?
The record says: finished, as of the sponsor's last update on 23 Nov 2022. Status can change between updates, so confirm with the study contact before making plans.
Where is the study taking place?
The record lists 1 site in United States. Whether a site is taking part right now should be confirmed with that site. See every site.
What are the possible benefits and risks?
They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.
Are results available for this study?
The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.
How can a doctor refer a patient or contact the research team?
The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.
What does the primary outcome measure?
Comparison of pulse rate (BPM) by a pulse oximeter monitor and sensor, compared to ECG pulse rate measurement (time frame: October 2022). The study lists 2 primary outcomes in all. It is what the study was designed to measure, not a finding.
Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?
No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.
How often is this information updated?
AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 23 Nov 2022.