Hypothermia; AnesthesiaNCT05516771
Evaluation of a Licensed Double-sensor-heat-flux (DHF) Non-invasive Core Temperature Sensor in Small Children and Toddlers Undergoing Surgery
An observational study of Hypothermia; Anesthesia, run by Oliver Kimberger at 1 site in Austria.
Trial summary
Title: Evaluation of a licensed double-sensor-heat-flux (DHF) non-invasive core temperature sensor in small children and toddlers undergoing surgery. CI/PS identification code: DHF-TaSC Device Name: Tcore™ system (Tcore-Adapter MP00999 \& Tcore-Sensor MP 00989) Manufacturer: Drägerwerk AG \& CO KGaA Study design… Read more about this study
- Who is it for?
- Ages 0 to 7 years
- Conditions the study is for
- Hypothermia; Anesthesia
- What is being tested?
- Tcore™ system
- Type of treatment
- Medical device
- How many people
- 58 participants
- Where
- 1 siteAustria
- Study dates
- 1 Oct 2022to 26 Feb 2023
- Phase
- Observational
- Are there results?
- None posted in the registry
Key requirements
Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.
- Ages 0 to 7 years
- All sexes
- No healthy volunteers
- Child
Study population: Study population: toddlers and small children undergoing surgery at a tertiary referral, university-affiliated hospital
Who may be able to join
- 0-7 years
- elective surgical procedures with a procedure time of more than 30 min
- operations for which insertion of a rectal temperature probe is clinically indicated as a routine method, (e.g., operations in regional anaesthesia)
Who cannot join
- rash or infection that prevents the application of the Heat flux thermometers.
- Fragile forehead skin state;
- known allergy to the probe adhesive or any constituent components
- maxillofacial trauma or lesions
- procedures impeding proper placement of the Tcore™ sensor
- neurologically impaired children with abnormal thermoregulation;
Show all 13Show fewer
- hemodynamic instability; need for vasoactive medication;
- procedures associated with extended use of abdomino/thoracic rinsing fluids;
- thoracoscopic/thoracotomy procedures;
- malignant hyperthermia or family history of malignant hyperthermia
- patient with fever or infection;
- all conditions that might be judged to abnormally alter skin perfusion
- patients or their legal guardian are not willing to participate in the trial.
Trial locations and contacts
1 site in 1 country.
About this study
Title: Evaluation of a licensed double-sensor-heat-flux (DHF) non-invasive core temperature sensor in small children and toddlers undergoing surgery.
CI/PS identification code: DHF-TaSC Device Name: Tcore™ system (Tcore-Adapter MP00999 \& Tcore-Sensor MP 00989) Manufacturer: Drägerwerk AG \& CO KGaA Study design: Prospective, single arm, clinical study
Read more
Patients:
- Number: 72
- Age / gender: females and males between 0 and 7 years
- Patients of a tertiary referral, university-affiliated hospital undergoing surgery
Exclusion Criterions:
- Operation site, rash or infection that prevents the application of the heat flux thermometers.
- Patients and/or legal guardians not willing to participate in the trial.
- Patients older than 7 years Setting: ORs of a tertiary referral, university-affiliated hospital. Study variables: measurement triplets measured via two double-sensor-heat-flux (DHF) thermometer the Tcore™ (Dräger, Drägerwerk AG \& Co KG, 23558 Lübeck, Germany) one placed on the forehead and one on the upper belly and temperature measured by a rectally placed temperature probe.
Read the detailed description
Body temperature is important for maintaining normal physiological functions. Inadvertent hypothermia as well as pyrexia are associated with higher morbidity and mortality. Consequently, continuous measurement of body temperature is standard of care at intensive care units (ICUs) and in operation rooms (ORs). The use of core temperature measurements such as pulmonary catheter (gold standard), esophageal or urinary probes is recommended because of higher accuracy than non-invasive measuring methods. The double-sensor-heat-flux (DHF) thermometer the Tcore™ (Dräger, Drägerwerk AG \& Co KG, 23558 Lübeck, Germany). The DHF thermometers consist of two sensors separated by a known thermal resistance. One side is placed directly at the patient's skin and the other is facing the environment. After some equilibration time the core temperature can then be calculated. To date no study, exist for the use of the Tcore™ thermometer in children or toddlers. Children are at higher risk for hypothermia and perioperative hypothermia is still common in children.(8) The objective is to find a non-invasive core temperature measurement for children to guide perioperative temperature management.
Primary Objective (Hypothesis):
Evaluation of the accuracy of the licensed double-sensor-heat-flux (DHF) thermometer the Tcore™ (Dräger, Drägerwerk AG \& Co KG, 23558 Lübeck, Germany) when used for intraoperative core temperature measurement in small children and toddlers.
The Hypothesis is that each of the two temperature measurements obtained with the Tcore™ (forehead, upper belly) have a high level of agreement(<0.5°C) with the standard of care temperature measurements via rectal temperature probe.
Our primary objective is to investigate the agreement between the two temperature measurements obtained with the Tcore™ (Dräger, Drägerwerk AG \& Co KGaA, 23558 Lübeck, Germany) placed on the forehead and the upper belly and temperature measured via a rectally placed Medical Level 1 disposable General Purpose Temperature Probe (Smiths Medical Österreich GmbH, Brunn am Gebirge, Austria). Measure triplets of the two Tcore™ sensors and the rectally placed probe are obtained every 3 minutes for up to 12 consecutive hours or for the duration the patient spends in the OR and the recovery room. If the observed limits of agreement (± 1.96 SD around the mean difference), within which 95% of the differences are expected to fall, are clinically acceptable, the methods are considered to agree. We define deviations of +/-0.5°C, a priori, to be acceptable for the Bland Altman method comparison analysis as was done in previous studies.
Intended use The Tcore system is a thermometer intended for non-invasive, continuous measurement of the core body temperature of children from 5 years of age, adolescents, and adults. The Tcore system consists of a sensor and an adapter. The adapter is reusable. The sensor is intended for single use. The Tcore system works in adjusted mode, i.e., the core body temperature is determined by a dual sensor.
Off-Label/Investigational Use In context of this clinical study the Tcore-system will be investigated outside its approved and intended use, namely in children below the age of five and toddlers.
Descriptive statistics:
- For nominal and ordinal data absolute frequencies and percentages, fraction, rate, and incidence will be calculated, if appropriate.
- Continuous data are described by mean ± standard deviation (SD) in case of approximate normal distribution, or by median, interquartile range otherwise.
Endpoint analysis:
For the primary objective, a scatterplot of differences between Tcore and standard measurements and their mean will be produced for each of the two Tcore measurements in relation to the measurement standard (rectal temperature). If the differences prove to be approximately independent for the observed range of temperatures, limits of agreement will be calculated using the method by Zou (2013).(23) If these limits are contained within the acceptable deviation of +/-0.5°C the respective Tcore measurement method will be considered acceptable. Irrespective of the success of the investigation of agreement the results for both Tcore measurement locations will be reported.
For the secondary objective sensitivity and specificity will be calculated with 95% confidence intervals for each Tcore measurement in relation to the measurement standard.
All study data will be queried from the ICCA and from the database of electronic health records the Vienna General Hospital information management system (AKIM; "Allgemeines Krankenhaus Informationsmanagement") (Siemens AG Österreich, Vienna, Austria). Intellispace Critical Care and Anesthesia. After acquisition, patient data will be anonymized, cleaned, and stored in a study database.
Patient identification will contain name, sex, and birthdate. This data will be documented in an excel sheet, protected with a password. All patients will be given a number for pseudonymization. The verification key will be stored on a separate computer in our laboratory. Only specially authorized persons (Dr. Sebastian Zeiner and Prof. Oliver Kimberger) will have access to the data.
All data will be stored on a password secured computer in the laboratory of our department (Department for Anesthesia) at the Medical University of Vienna.
At any time, medical confidentiality will be fulfilled by all persons involved in this project.
- Conditions
- Hypothermia; Anesthesia
- Being tested
- Tcore™ system
- Medical subject headings
- Hypothermia
What participants receive
What participants receive. In a randomised study each person is assigned to one group by chance.
Tcore™ systemDevice
The Tcore sensor is a double-sensor-heat-flux (DHF) thermometer and consist of two sensors separated by a known thermal resistance. One side is placed directly at the patient's skin and the other is facing the environment. After some equilibration time the core temperature can then be calculated. The sensors are placed above the eye and the sinus frontalis.
Intended use The Tcore system is a thermometer intended for non-invasive, continuous measurement of the core body temperature of children from 5 years of age, adolescents, and adults. The Tcore system consists of a sensor and an adapter. The adapter is reusable. The sensor is intended for single use. The Tcore system works in adjusted mode, i.e., the core body temperature is determined by a dual sensor.
Off-Label/Investigational Use In context of this clinical study the Tcore-system will be investigated outside its approved and intended use, namely in children below the age of five and toddlers.
| Study group | Role | What they receive |
|---|---|---|
| group 1 | toddlers and small children undergoing surgery at a tertiary referral, university-affiliated hospital Receives: Device: Tcore™ system |
What is being measured
The primary outcome is the main question the study is designed to answer.
Accuracy of Tcore core temperature measurements
A scatterplot of differences between Tcore and standard measurements of body temperature,limits of agreement will be calculated using the method by Zou (2013).(23)
Measured: For the duration of Anesthesia and a maximum of 12 hours
Secondary outcomes (1)
sensitivity and specificity
For the secondary objective sensitivity and specificity will be calculated with 95% confidence intervals for each Tcore measurement in relation to the measurement standard
Measured: For the duration of Anesthesia and a maximum of 12 hours
Timeline
From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.
- First posted26 Aug 2022
- Study start1 Oct 2022
- Primary completion26 Feb 2023
- Study completion26 Feb 2023
Resources
- Official record on ClinicalTrials.govThe registry entry this page is built from, with its full change history.
- Other trials for Hypothermia; AnesthesiaStudies of the same condition, open and under way.
- What the trial phases meanHow a treatment moves from first tests in people to approval.
More details
For doctors and researchers: how the study is designed, who runs it, and its registry IDs.
Study design
- Study type
- Observational
- Observational model
- Cohort
- Time perspective
- Prospective
- Enrolment
- 58 participants (actual)
Sponsor and investigators
- Lead sponsor
- Oliver Kimberger
- Sponsor type
- University, hospital or other
- Responsible party
- Sponsor-investigator, Oliver Kimberger, Medical University of Vienna
IDs, dates and publications
- ClinicalTrials.gov ID
- NCT05516771
- Sponsor's protocol ID
- DHF-TaSC
- Status last verified
- Mar 2023
- First submitted
- 21 Aug 2022
- Last updated
- 20 Mar 2023
- Results posted
- No
- Data monitoring committee
- No
- FDA-regulated drug
- No
- FDA-regulated device
- Yes
- Individual data sharing
- No
Study at a glance
From the registry record, last updated by the sponsor on 20 Mar 2023. Recruitment status is as the sponsor reported it: confirm with the study team.
- Official title
- Evaluation of a Licensed Double-sensor-heat-flux (DHF) Non-invasive Core Temperature Sensor in Small Children and Toddlers Undergoing Surgery
- ClinicalTrials.gov ID
- NCT05516771
- Acronym / sponsor ID
- DHF-TaSC
- Conditions
- Hypothermia; Anesthesia
- Study type
- Observational
- Recruitment status
- Completed
- Enrolment
- 58 participants (actual)
- Lead sponsor
- Oliver KimbergerUniversity, hospital or other
- Study start
- 1 Oct 2022 (actual)
- Primary completion
- 26 Feb 2023 (actual)
- Study completion
- 26 Feb 2023 (actual)
- First posted
- 26 Aug 2022 (actual)
- Results
- Not posted in the source record
Clinical summary
The research question as the sponsor registered it. This describes what the study plans to test, not what it found.
Title: Evaluation of a licensed double-sensor-heat-flux (DHF) non-invasive core temperature sensor in small children and toddlers undergoing surgery.
CI/PS identification code: DHF-TaSC Device Name: Tcore™ system (Tcore-Adapter MP00999 \& Tcore-Sensor MP 00989) Manufacturer: Drägerwerk AG \& CO KGaA Study design: Prospective, single arm, clinical study
Patients:
- Number: 72
- Age / gender: females and males between 0 and 7 years
- Patients of a tertiary referral, university-affiliated hospital undergoing surgery
Exclusion Criterions:
- Operation site, rash or infection that prevents the application of the heat flux thermometers.
- Patients and/or legal guardians not willing to participate in the trial.
- Patients older than 7 years Setting: ORs of a tertiary referral, university-affiliated hospital. Study variables: measurement triplets measured via two double-sensor-heat-flux (DHF) thermometer the Tcore™ (Dräger, Drägerwerk AG \& Co KG, 23558 Lübeck, Germany) one placed on the forehead and one on the upper belly and temperature measured by a rectally placed temperature probe.
Detailed description
Body temperature is important for maintaining normal physiological functions. Inadvertent hypothermia as well as pyrexia are associated with higher morbidity and mortality. Consequently, continuous measurement of body temperature is standard of care at intensive care units (ICUs) and in operation rooms (ORs). The use of core temperature measurements such as pulmonary catheter (gold standard), esophageal or urinary probes is recommended because of higher accuracy than non-invasive measuring methods. The double-sensor-heat-flux (DHF) thermometer the Tcore™ (Dräger, Drägerwerk AG \& Co KG, 23558 Lübeck, Germany). The DHF thermometers consist of two sensors separated by a known thermal resistance. One side is placed directly at the patient's skin and the other is facing the environment. After some equilibration time the core temperature can then be calculated. To date no study, exist for the use of the Tcore™ thermometer in children or toddlers. Children are at higher risk for hypothermia and perioperative hypothermia is still common in children.(8) The objective is to find a non-invasive core temperature measurement for children to guide perioperative temperature management.
Primary Objective (Hypothesis):
Evaluation of the accuracy of the licensed double-sensor-heat-flux (DHF) thermometer the Tcore™ (Dräger, Drägerwerk AG \& Co KG, 23558 Lübeck, Germany) when used for intraoperative core temperature measurement in small children and toddlers.
The Hypothesis is that each of the two temperature measurements obtained with the Tcore™ (forehead, upper belly) have a high level of agreement(<0.5°C) with the standard of care temperature measurements via rectal temperature probe.
Our primary objective is to investigate the agreement between the two temperature measurements obtained with the Tcore™ (Dräger, Drägerwerk AG \& Co KGaA, 23558 Lübeck, Germany) placed on the forehead and the upper belly and temperature measured via a rectally placed Medical Level 1 disposable General Purpose Temperature Probe (Smiths Medical Österreich GmbH, Brunn am Gebirge, Austria). Measure triplets of the two Tcore™ sensors and the rectally placed probe are obtained every 3 minutes for up to 12 consecutive hours or for the duration the patient spends in the OR and the recovery room. If the observed limits of agreement (± 1.96 SD around the mean difference), within which 95% of the differences are expected to fall, are clinically acceptable, the methods are considered to agree. We define deviations of +/-0.5°C, a priori, to be acceptable for the Bland Altman method comparison analysis as was done in previous studies.
Intended use The Tcore system is a thermometer intended for non-invasive, continuous measurement of the core body temperature of children from 5 years of age, adolescents, and adults. The Tcore system consists of a sensor and an adapter. The adapter is reusable. The sensor is intended for single use. The Tcore system works in adjusted mode, i.e., the core body temperature is determined by a dual sensor.
Off-Label/Investigational Use In context of this clinical study the Tcore-system will be investigated outside its approved and intended use, namely in children below the age of five and toddlers.
Descriptive statistics:
- For nominal and ordinal data absolute frequencies and percentages, fraction, rate, and incidence will be calculated, if appropriate.
- Continuous data are described by mean ± standard deviation (SD) in case of approximate normal distribution, or by median, interquartile range otherwise.
Endpoint analysis:
For the primary objective, a scatterplot of differences between Tcore and standard measurements and their mean will be produced for each of the two Tcore measurements in relation to the measurement standard (rectal temperature). If the differences prove to be approximately independent for the observed range of temperatures, limits of agreement will be calculated using the method by Zou (2013).(23) If these limits are contained within the acceptable deviation of +/-0.5°C the respective Tcore measurement method will be considered acceptable. Irrespective of the success of the investigation of agreement the results for both Tcore measurement locations will be reported.
For the secondary objective sensitivity and specificity will be calculated with 95% confidence intervals for each Tcore measurement in relation to the measurement standard.
All study data will be queried from the ICCA and from the database of electronic health records the Vienna General Hospital information management system (AKIM; "Allgemeines Krankenhaus Informationsmanagement") (Siemens AG Österreich, Vienna, Austria). Intellispace Critical Care and Anesthesia. After acquisition, patient data will be anonymized, cleaned, and stored in a study database.
Patient identification will contain name, sex, and birthdate. This data will be documented in an excel sheet, protected with a password. All patients will be given a number for pseudonymization. The verification key will be stored on a separate computer in our laboratory. Only specially authorized persons (Dr. Sebastian Zeiner and Prof. Oliver Kimberger) will have access to the data.
All data will be stored on a password secured computer in the laboratory of our department (Department for Anesthesia) at the Medical University of Vienna.
At any time, medical confidentiality will be fulfilled by all persons involved in this project.
Study design
- Observational model
- Cohort
- Time perspective
- Prospective
- Number of arms
- 1
| Arm | Type | Receives |
|---|---|---|
| group 1 | – | Device: Tcore™ systemtoddlers and small children undergoing surgery at a tertiary referral, university-affiliated hospital |
Who may qualify
- Ages
- 0 to 7 years
- Sex
- All
- Healthy volunteers
- Not accepted
- Age groups
- Child
- Study population
- Study population: toddlers and small children undergoing surgery at a tertiary referral, university-affiliated hospital
Key inclusion criteria (3 listed)
- 0-7 years
- elective surgical procedures with a procedure time of more than 30 min
- operations for which insertion of a rectal temperature probe is clinically indicated as a routine method, (e.g., operations in regional anaesthesia)
Key exclusion criteria (13 listed)
- rash or infection that prevents the application of the Heat flux thermometers.
- Fragile forehead skin state;
- known allergy to the probe adhesive or any constituent components
- maxillofacial trauma or lesions
- procedures impeding proper placement of the Tcore™ sensor
Eligibility is decided by the study team after screening; this overview does not establish that a patient qualifies. Full inclusion and exclusion criteria.
Interventions
| Intervention | Type | Arms |
|---|---|---|
| Tcore™ systemThe Tcore sensor is a double-sensor-heat-flux (DHF) thermometer and consist of two sensors separated by a known thermal resistance. One side is placed directly at the patient's skin and the other is facing the environment. After some equilibration time the core temperature can then be calculated. The sensors are placed above the eye and the sinus frontalis. Intended use The Tcore system is a thermometer intended for non-invasive, continuous measurement of the core body temperature of children from 5 years of age, adolescents, and adults. The Tcore system consists of a sensor and an adapter. The adapter is reusable. The sensor is intended for single use. The Tcore system works in adjusted mode, i.e., the core body temperature is determined by a dual sensor. Off-Label/Investigational Use In context of this clinical study the Tcore-system will be investigated outside its approved and intended use, namely in children below the age of five and toddlers. | Device | group 1 |
Primary outcome
Accuracy of Tcore core temperature measurements
Time frame: For the duration of Anesthesia and a maximum of 12 hours
A scatterplot of differences between Tcore and standard measurements of body temperature,limits of agreement will be calculated using the method by Zou (2013).(23)
1 secondary outcome
sensitivity and specificity
Time frame: For the duration of Anesthesia and a maximum of 12 hours
Results and safety
The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.
Research links
- Official record on ClinicalTrials.govLast updated by the sponsor 20 Mar 2023.
- Individual participant dataSharing planned: No
Related trials for Hypothermia; Anesthesia
Recruiting
Goal-Directed Temperature Management During Prolonged Abdominal Surgery
Hypothermia Following Anesthesia, Intraoperative Complications
Enrolling by invitationObservational study
Study of the Prevalence of Postoperative Hypothermia in Adults Undergoing Non-Cardiac Surgery in Spanish Post-Anesthesia Care Units
Hypothermia Following Anesthesia
Active, not recruitingObservational study
Effect of Hypothermia on BIS and PSI Correlation During Cardiopulmonary Bypass
Cardiopulmonary Bypass, Hypothermia; Anesthesia
Active, not recruitingObservational study
Incidence and Risk Factors of Intraoperative Hypothermia in Adult Patients After Protocol Implementation
Hypothermia Following Anesthesia
Recruiting
The Effect of Heating the Intensive Care Room in the Early Postoperative Period on Patient Outcomes
Abdominal Surgery Complications, Hypothermia Following Anesthesia
Referral and contact
AzTrials does not enrol participants or take referrals. Contact the study team or a site listed below; they decide who can take part. This study is not recruiting at the moment.
The record lists no contact. Use the official record to reach the sponsor.
1 site in 1 country: see every site and its status.
Questions about this trial
What is this clinical trial studying?
An observational study of Hypothermia; Anesthesia, run by Oliver Kimberger at 1 site in Austria.
The full registered description is in the About this study section and the official record.
Who may be eligible to take part?
The sponsor lists 3 inclusion and 13 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.
Is this clinical trial currently recruiting?
The record says: finished, as of the sponsor's last update on 20 Mar 2023. Status can change between updates, so confirm with the study contact before making plans.
Where is the study taking place?
The record lists 1 site in Austria. Whether a site is taking part right now should be confirmed with that site. See every site.
What are the possible benefits and risks?
They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.
Are results available for this study?
The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.
How can a doctor refer a patient or contact the research team?
The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.
What does the primary outcome measure?
Accuracy of Tcore core temperature measurements (time frame: For the duration of Anesthesia and a maximum of 12 hours). It is what the study was designed to measure, not a finding.
Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?
No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.
How often is this information updated?
AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 20 Mar 2023.