Visceral and Digestive SurgeryTELESENSENCT05357729

Safety and Performance Study of the MultiSense® Ambulatory Telemonitoring System for Non-critical Patient Management

A study testing MultiSense® remote monitoring in people with Visceral and Digestive Surgery and Post-surgical Monitoring, run by IHU Strasbourg at 1 site in France.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

In this clinical investigation, the investigators will assess the safety, performances and clinical added value of the MultiSense® ambulatory telemonitoring solution in real-life situations. Read more about this study

Who is it for?
Ages 18 years and older
What is being tested?
MultiSense® remote monitoring
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
20 participants
Where
1 siteFrance
Study dates
14 Jun 2022to 2 Sep 2022
Phase
Interventional
Are there results?
None posted in the registry

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 years and older
  • All sexes
  • No healthy volunteers
  • Adult, Older adult

Who may be able to join

  • Adult (male or female over 18 years) in non-critical care settings
  • Patient programmed for an elective surgery
  • Need to monitor at home at least one physiological parameter: heart rate, respiratory rate, SpO2, temperature, physical activity
  • Patient able to use a smartphone
  • Patient with Wi-Fi and/or Cellular connectivity at home
  • Patient with a personal phone which allows to receive text messages
  • Patient able to receive and understand information related to the study and give written informed consent
  • Patient affiliated to the French social security system

Who cannot join

  • Patient with known skin allergy (adhesive or silicon) or skin disease that would not allow the use of an adhesive
  • Patient with an active implantable device (such as a pacemaker or automatic defibrillator)
  • Patient with an elective imagery planned during the use of MultiSense
  • Pregnant or lactating patient
  • Patient in exclusion period (determined by a previous or a current study)
  • Patient under guardianship or trusteeship
  • Patient under the protection of justice or deprived of liberty
  • Patient in situation of emergency

Trial locations and contacts

1 site in 1 country.

France1 site
  • Service de Chirurgie Digestive et Endocrinienne

    Strasbourg, 67000

Other trials in France

About this study

In this clinical investigation, the investigators will assess the safety, performances and clinical added value of the MultiSense® ambulatory telemonitoring solution in real-life situations.

Read the detailed description

The MultiSense® solution is a medical device for telemonitoring for in and outpatient use.

It includes 5 physiological data sensors and communicates via Bluetooth with a smartphone. The data is centralized in the web and a web application allows healthcare professionals to scrutinize patient's data.

Among its potential benefits, the MultiSense® solution offers continuous, non-invasive and wireless physiological data recording over 5 days whenever patient monitoring is needed. There is a special focus on the post-operative monitoring, especially during the initial convalescence phase outside the hospital. An automatic notification system can be set up by a healthcare professional to notify abnormal data and potential need for phone call and/or physical home intervention.

The aim of this study is to evaluate the safety, the performance and added value of the MultiSense® solution in real-life situations. The study population includes 20 post-surgical patients having a genuine need for post-hospital monitoring and be likely to have variations in physiological variables monitored, without being in a critical condition. The patients included will be equipped with the MultiSense® solution prior to the hospital discharge. The device will be used for 6 days from the date of actual deployment.

Being tested
MultiSense® remote monitoring
Keywords
Vital signs, Connected device, real-time monitoring, post-surgery monitoring, cardio-respiratory monitoring, telemonitoring

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • MultiSense® remote monitoringDevice

    At hospital, the patient will be monitored with conventional monitoring devices as well as the MultiSense® solution (at least one hour with a reference monitor). After hospital discharge, the patients will be remotely monitored by their healthcare professional using the MultiSense® solution.

Study groups in NCT05357729
Study groupRoleWhat they receive
MultiSense® remote monitoringExperimentalThe patients included will be equipped with the MultiSense® solution prior to the hospital discharge. The device will be used for 6 days from the date of actual deployment.
Receives: Device: MultiSense® remote monitoring

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Determination of the safety and performance of the MultiSense® solution in patient monitoring, in terms of heart rate accuracy by comparison with a reference monitor.

    Accuracy is defined as an average measurement variation less than 5 bpm of the mean value, with respect to the reference standard

    Measured: From patch placement to hospital discharge, assessed up to 3 hours

  2. Determination of the safety and performance of the MultiSense® solution in patient monitoring, in terms of data availability.

    Percentage of missing data due to technical issues (non-behavioral events) over the monitoring period below 15%.

    Measured: From patch placement to removal, assessed up to 7 days

  3. Determination of the safety and performance of the MultiSense® solution in patient monitoring, in terms of data artefacts.

    Percentage of artefacts due to technical issues (non-behavioral events) over the monitoring period below 10% for heart rate, 20% for the respiratory rate or 30% for SpO2.

    Measured: From patch placement to removal, assessed up to 7 days

  4. Determination of the safety and performance of the MultiSense® solution in patient monitoring, in terms of data transmission.

    Data transmission is defined as more than 90% of data with transmission time from patch to server less than 5 min.

    Measured: From patch placement to removal, assessed up to 7 days

Secondary outcomes (13)
  1. Patient's tolerance to prolonged use of the MultiSense® solution (Safety in use, side effects)

    Accounting and classification of unexpected local and distant event due to the use of the MultiSense® solution (vigilance data) and Patient's questionnaire

    Measured: From patch placement to its removal (up to 7 days)

  2. Device defects

    Accounting of device deficiencies reported by the investigator, the patient or identified by MultiSense data analysis

    Measured: From patch placement to its removal (up to 7 days)

  3. Clinical relevance of the notifications provided by MultiSense® solution

    Relevance of the notifications generated by MultiSense® will be judged on the relevance of the notification's functionality by the healthcare professional in real-life situation by the mean of a questionnaire (4-levels Likert scale)

    Measured: From patch placement to its removal (up to 7 days)

  4. Clinical relevance of the notifications provided by MultiSense® solution

    Relevance of the notifications generated by MultiSense® will be judged on the relevance of the notification's functionality by the healthcare professional in real-life situation by the mean of notification treatment through MultiSense® web application

    Measured: From patch placement to its removal (up to 7 days)

  5. Remote monitoring facilitation for the healthcare professional through the solution's ease of use and satisfaction gained

    Determination of healthcare professionals' practical interest in the solution and satisfaction of use assessed via a questionnaire (4-levels Likert scale)

    Measured: From patch placement to its removal (up to 7 days)

  6. Patient's satisfaction and reassurance when monitored by the device

    Determination of patient satisfaction and reassurance by an end-study questionnaire (4-levels Likert scale)

    Measured: From patch placement to its removal (up to 7 days)

  7. Patch autonomy

    Percentage of patches with an autonomy > 5 days including skin adhesion and battery life (MultiSense® data and patient's questionnaire)

    Measured: From patch placement to its removal (up to 7 days)

  8. Respiratory rate accuracy in terms of impedance

    Respiratory rate accuracy by comparison of MultiSense® data with the reference monitor in terms of impedance

    Measured: From patch placement to hospital discharge, assessed up to 3 hours

  9. Respiratory rate accuracy in terms of EtCO2

    Respiratory rate accuracy by comparison of MultiSense® data with the reference monitor in terms of EtCO2

    Measured: From patch placement to hospital discharge, assessed up to 3 hours

  10. Estimation of posture and activity level reported by the MultiSense® solution

    Determination of the posture and activity level consistency by accounting the percentage of lying position and low activity at night and the percentage of high activity during day (MultiSense® data)

    Measured: From patch placement to its removal (up to 7 days)

  11. MultiSense® data quality after moderate level of activity

    Distribution of activity level across monitoring period, number of showers taken over the monitoring period and evolution of the artefact's metric (as defined in the primary endpoint) over time (MultiSense® data)

    Measured: From patch placement to its removal (up to 7 days)

  12. Correlation between body temperature and skin temperature

    Correlation between the body temperature measured with a standard device and the skin temperature measured with the MultiSense® solution

    Measured: From patch placement to hospital discharge, assessed up to 2 days

  13. End of study hardware recovery rate

    Percentage of MultiSense® patch and gateway returned to RDS at the end of the study

    Measured: From patch removal to material return, assessed up to 1 month

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted3 May 2022
  2. Study start14 Jun 2022
  3. Primary completion2 Sep 2022
  4. Study completion2 Sep 2022

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Other
Allocation
Not applicable
Intervention model
Single group
Masking
None (open label)
Enrolment
20 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT05357729
Sponsor's protocol ID
21-004
Acronym
TELESENSE
Status last verified
Jan 2024
First submitted
27 Apr 2022
Last updated
5 Jan 2024
Results posted
No
Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
No
Individual data sharing
No

Questions about this trial

What is this clinical trial studying?

A study testing MultiSense® remote monitoring in people with Visceral and Digestive Surgery and Post-surgical Monitoring, run by IHU Strasbourg at 1 site in France.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 8 inclusion and 8 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 5 Jan 2024. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in France. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Determination of the safety and performance of the MultiSense® solution in patient monitoring, in terms of heart rate accuracy by comparison with a reference monitor. (time frame: From patch placement to hospital discharge, assessed up to 3 hours). The study lists 4 primary outcomes in all. It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 5 Jan 2024.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in