Atrial FibrillationNCT05316753
BTE and Pulsed Waveforms for Cardioversion of Atrial Fibrillation - Escalation Strategy and Manual Pressure
A study testing Cardioversion with a pulsed biphasic waveform and Cardioversion with a biphasic truncated exponential waveform in people with Atrial Fibrillation, run by University National Heart Hospital at 1 site in Bulgaria.
Trial summary
Background: A Danish study raised the question of the usefulness of escalating energy protocols compared to fixed high-energy protocols. Maximal energies are usually the final choice of the physicians. Read more about this study
- Who is it for?
- Ages 18 years and older
- Conditions the study is for
- Atrial Fibrillation
- What is being tested?
- Cardioversion with a pulsed biphasic waveform and Cardioversion with a biphasic truncated exponential waveform
- Type of treatment
- Surgery or procedure
- Could I receive a placebo?
- No, everyone receives an active treatment
- How many people
- 94 participants
- Where
- 1 siteBulgaria
- Study dates
- 20 Jan 2022to 30 Jun 2024
- Phase
- Interventional
- Are there results?
- None posted in the registry
Key requirements
Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.
- Ages 18 years and older
- All sexes
- No healthy volunteers
- Adult, Older adult
Who may be able to join
- Indications for elective cardioversion of atrial fibrillation
- Patients > 18 years old and:
- Symptomatic AFIB with a duration of less than 12 months and EHRA score 2-4
- Symptomatic first detected AFIB and EHRA score 2-4
- Persistent AFIB after successful causal therapy
- Rare recurrences of AFIB with long periods of sinus rhythm
- Impossibility to reach a sustained normal ventricular rate in AFIB
Who cannot join
- Patients with atrial flutter
- Spontaneous HR <60/min
- Digitalis intoxication
- Impossibility to maintain sinus rhythm irrespective to antiarrhythmic therapy and frequent cardioversions
- Conduction disturbances (without fascicular block and AV block 1 degree) in patients without pacemaker
- Asymptomatic patients with AFIB for > 1 year
Show all 13Show fewer
- Thyroid dysfunction: euthyroid status of at least one month is required (TSH is measured).
- Thrombosis in cardiac cavities, assessment performed using Transesophageal echocardiography (TEE)
- Spontaneous echo contrast > 2 degree (TEE)
- Patients with planned cardiac operation in the next three months
- Patients with embolic event in the last three months
- Patients <18 years of age
- Pregnant woman
Trial locations and contacts
1 site in 1 country.
Bulgaria1 site, 1 recruiting
- Recruiting
University National Heart Hospital
Sofia, 1309
Elina G Trendafilova, MD, PhD, +359888718847, elitrendafilova@abv.bg
Elena S Dimitrova, MD, PhD, +359887394540, elena.sv@gmail.com
About this study
Official title: Comparison of Biphasic Truncated Exponential and Pulsed Waveforms for Cardioversion of Atrial Fibrillation-High Energy Escalation Strategy and Introduction to Manual Pressure Application
Background: A Danish study raised the question of the usefulness of escalating energy protocols compared to fixed high-energy protocols. Maximal energies are usually the final choice of the physicians. Some authors showed that decreasing impedance by manual pressure application (MPA) had a positive impact on cardioversion outcome. This is likely due to the impedance decrease linked to MPA.
Objective: This new clinical cardioversion study of atrial fibrillation (AF) patients aims to compare the efficacy and safety of a new high energy escalation strategy. The protocol combines high energy shocks at first shock, jumping to maximal defibrillator energy at second shock and finally complemented by MPA at third shock, if success is not reached using electric shocks only.
Read more
Experimental design: Patients will be recruited at the Intensive Cardiology Care Unit, Cardiology Clinic, National Cardiology Hospital (NCH), Sofia, Bulgaria. All eligible patients will sign a written informed consent prior to the cardioversion and will receive the standard hospital procedures during cardioversion.
AF patients will be alternatively randomized to cardioversion using one of the two defibrillators, following the strategy below:
DEFIGARD HD-7 arm: 3 consecutive shocks with escalating selected energy: 150J, 200J, 200J. The third shock is combined with MPA LIFEPAK15 arm: 3 consecutive shocks with escalating selected energy: 150J, 360J, 360J. The third shock is combined with MPA The statistical power analysis will consider a superiority comparison between the cumulative energy actually delivered by both defibrillators.
The secondary cardioversion efficacy outcome measures are: the cumulative success rate (measured at 1 minute post-shock), number of delivered shocks. Delivered energy will be measured during each shock with a dedicated pulse recording device (approved by the NCH Ethical Committee). Heart rhythm will be measured in continuously recorded peripheral ECG.
The secondary cardioversion safety outcome measures are: markers for myocardial necrosis (high sensitive troponin I, CK-MB) evaluated on blood samples taken before and 8-12 hours after cardioversion; ST-segment changes (post-shock - pre-shock) measured in lead II; Complications after cardioversion measured during 2 hours follow-up period in the ICCU - the presence of apnea, arrhythmias, bradycardia and the need for respective therapy at the discretion of attending physician.
Read the detailed description
Background Commercially available defibrillators generate direct current (DC) shock with various waveforms using distinct technologies. These waveforms are suggested to have different efficacy and safety. Although the superiority of biphasic over monophasic waveforms is well established, the relative efficacy and safety of the available biphasic waveforms is not clear.
Various BTE waveforms apply different potential gradients on the thorax that might produce various defibrillation effects. Apart from efficacy, the most important aspect is the patient safety, considering that larger potential gradients in the myocardium lasting longer could potentially induce an electroporation and then a fibrillation. A frequently observed effect of electroporation are the post-shock ST-segment deviations in the surface ECG, representing the potential difference between the normal tissue and sustained depolarized critical mass of myocardium closest to the associated origin of the electrical current. Although ST-segment changes are an easily ignored phenomenon, occurring acutely and resolving during the first few minutes post-shock, their presence in a short-term basis can identify electroporation by dangerously high potential gradients, while the sustained ST-changes in a long-term basis can identify cases with myocardial injury. The most reliable marker of shock-induced myocardial injuries is the high sensitive cardiac Troponin I (hsTnI).
In a previous study Trendafilova et al. were compared fixed-high energy protocols (200J-200J-200J) delivered by two different defibrillation waveforms: Pulsed (BTE-PE) and High-energy BTE (BTE-HE). This study concluded on non-inferiority between devices. Cumulative energy, efficiency of the waveforms and safety (measured by High-sensitive Troponin) were not reaching significant difference for both waveforms. Delivered energy of the first efficient shock was significantly lower for BTE-PE and although BTE-PE delivered almost 28J lower energy after the whole procedure, significance could not be reached due to wide variance.
In a Danish study Schmidt et al. compared two cardioversion protocols - one is fixed-high energy (360J-360J-360J), the other is low energy escalating (125J-150J-200J). The overall efficacy of fixed-energy protocol was better than escalating. Safety is non significantly different. Schmidt et al. raised the question of the usefulness of escalating energy protocols compared to fixed high-energy protocols.
There is an ethical consideration to apply maximal energy shocks to all patients without considering patient individuality. Ramirez et al. remind that failed shocks might depend on other factors like transthoracic impedance. Lavignasse et al. found that cumulative energy setting differs with patient characteristics, having significant association with: AF duration, gender, BSA, LV TDD, valvular disease and chronic respiratory disease. Indeed, some authors (Voskoboinik et al. and Ramirez et al.) showed that decreasing impedance by manual pressure application (MPA) had a positive impact on cardioversion outcome.
On most defibrillators, increasing the energy level is leading to an increase of the voltage delivered to the patient. This voltage increase implies a current increase in the thorax of the patient. In fact, applying a force on the defibrillation pads or increasing shock voltage by increasing the energy setting would lead to increase the current in the thorax, associated with expectations for higher shock efficiency. MPA is performed by pressing on the pads using a non-connected pair of paddles in order to assure isolation of the practitioner from the patient during the shock.
Objective:
This new clinical cardioversion study of atrial fibrillation (AF) patients aims to compare the efficacy and safety of a new high energy escalation strategy. The protocol combines reasonably high energy shocks complemented by MPA when necessary.
Two biphasic defibrillators - a standard BTE-HE waveform (Lifepak 15, Physio Control) and a BTE-PE waveform (Defigard HD 7, Schiller Medical) with different maximal energy settings of 360J and 200J, respectively are compared.
The proposed high energy escalation strategy is based on the hypothesis that most of the patients do not implicitly require the maximal levels of energy for cardioversion. Thus patients who could be easily cardioverted are removed from the pool that would need a more aggressive treatment.
Methods:
Study population All patients admitted for elective cardioversion of atrial fibrillation in ICCU- NCH will be potentially eligible for the study. Informed consent will be obtained from all eligible patients prior to cardioversion. Patients declining to participate in the study will receive treatment according to the hospital protocol.
Cardiologists will check the eligibility of each patient following the detailed list of inclusion and exclusion criteria, based on the documented pre-treatment data.
Additionally, eligibility will be verified through pre-cardioversion medical exams (blood test within 24 hours before cardioversion): routine blood chemistry, glucose, creatinine, eGFR, K, Na; CK-MB, high sensitive troponin I measurements on the day of cardioversion.
Study design This will be a prospective randomized trial (alternating design) where eligible patients will be randomized to treatment with two different defibrillators. Following the order of patient admittance in ICCU-NCH, the attending cardiologist will assign the odd and even eligible patients to the defibrillators in arm 1 and arm 2, respectively. The cardiologist cannot control the order of patient admittance in ICCU-NCH. The alternative randomization design will equalize the number of subjects on each treatment. This study is not designed to control for sex, age, comorbidity, type of device used for cardioversion, cumulative energy delivered during shocks, number of shocks administered. Randomization is performed to limit these and any other confounding factors.
Patient preparation Patients will be consulted prior to cardioversion (CV) in a quiet setting. Any questions will be answered by the attending cardiologist. Patients will be asked to manually sign their informed consent for study participation before further preparations and data collection. If the patient declines to participate, then he/she will receive treatment according to the hospital protocol without data documentation, as further defined in the study.
Demographic data will be taken. Pre-treatment data will be read from the patient files: concomitant diseases; medications during the previous 7 days; Left atrium dimensions, left ventricle dimensions and volumes, ejection fraction by transthoracic echocardiography during index hospitalization; TEE will be performed for all patients before CV - presence of cardiac thrombus and spontaneous echo contrast will be checked; duration of index AF in days; European Heart Rhythm Association class score of symptoms from index AF; standard blood test will be documented.
Appropriate standard anticoagulation with unfractionated heparin (UFH) or acenocumarol or direct oral anticoagulants (DOAC) is required before and after the CV.
The anesthesia will be conducted by an anesthesiologist with slow intravenous injection of Propofol, adjusted individually to reach deep sedation (Cook' s scale points <7).
Cardioversion protocol:
The CV procedure is performed in ICCU by a cardiologist and anesthesiologist. Defibrillator pads are placed according to the user guides on the pads and respecting >40 cm distance between them. Standard self-adhesive defibrillation pads, recommended by specific defibrillator's manufacturer will be used.
Defibrillator use: The time-interval interval between consecutive shocks is 1 min; The energy of consecutive shocks follows pre-defined protocol; The protocol stops at successful cardioversion (sinus rhythm at 1 min post-shock), otherwise at the last shock of the protocol.
Data collection and Follow up:
- Defibrillator pulse recorder: The shock waveforms delivered by the studied defibrillators is recorded in real-time during the shock by a dedicated measurement device (Defimpulse recorder DR2) connected in the patient circuit. This allows accurate measurement of the delivered energy, which is the primary measured outcome. Additionally, currents, voltages, patient impedances at each shock can be also measured. The device is fully automatic and expert assistance is not need during the CV intervention. The use of the measurement device (Defimpulse recorder DR2) has been approved by the NCH Ethical Committee.
- Blood samples, standard ECG, blood pressure and heart rate at baseline.
- Continuous ECG in 3 peripheral leads is recorded during the whole CV procedure. ST deviation is measured on the first cardiac complex after 10 s post-shock in standard manner (0.080 seconds after J point) in lead II.
- Standard ECG is recorded in 5 min after CV; 24 h after CV; at discharge if the discharge day is different from 24 h after CV.
- Heart rate and blood pressure are measured 3 times after CV during follow-up in ICCU
- Follow-up period in ICCU (2 hours after CV): Potential complications after CV are recorded
- Follow-up period in the Cardiology Clinic (24 hours after CV): blood samples to analyze CK-MB and troponin (8 to 12 h after CV); Clinical exam with heart rate and blood pressure; ECG; Any complications before discharge.
Statistics:
For arm1 (BTE PE waveform), the expected efficacy is 50% at the first shock (150J), 80% at the second shock (200J) and 85% at the third shock (200J combined with MPA).
For arm2 (BTE HE waveform), the expected efficacy is 50% at the first shock (150J), 80% at the second shock (360J) and 85% at the third shock (360J combined with MPA).
Taking into account that BTE PE waveform typically delivers 180J for energy setting of 200J, the estimated cumulative energy at the end of the CV procedure is 258J (+/-120J) and 366J (+/-240J) for arm1 and arm2, respectively.
The statistical analysis is designed for a superiority comparison between the two defibrillation strategies in respect of their delivered cumulative energies hypothesizing similar cardioversion efficacy rate. It is estimated that a total of 94 patients (47 per arm) would allow to reach a statistical power of 80% in a superiority study design with one-sided type I error rate of 0.05.
Continuous variables will be expressed as mean value ± standard deviation or median values (inter-quartile range) and compared with Students t-test or equivalent non-parametric test, respectively. Categorical variables are expressed as percentages and compared using the Chi-square or Fisher's exact test . P<0.05 is considered statistically significant for all comparisons.
Ethics All patients will sign a written informed consent prior to the cardioversion and will receive the standard hospital procedures during cardioversion. Both defibrillators included in this study are approved for clinical use and both are used on a daily basis.
Data anonymization policy will be respected. In order to ensure the medical confidentiality, no information about the patient will be entered in the defibrillators. An identification number will be given to each patient by the dedicated measurement device (Defimpulse recorder DR2). This number and the recording data are properties of the principal investigator. Approval from the NCH Ethical Committee concerning energy strategy and the use of the Defimpulse recorder during the cardioversion intervention has been obtained.
- Conditions
- Atrial Fibrillation
- Being tested
- Cardioversion with a pulsed biphasic waveform, Cardioversion with a biphasic truncated exponential waveform
- Keywords
- electrical cardioversion
- Medical subject headings
- Atrial Fibrillation
What participants receive
What participants receive. In a randomised study each person is assigned to one group by chance.
Cardioversion with a pulsed biphasic waveformProcedure or surgery
Also called Schiller Defigard HD- 7 - DGHD7, Schiller Medical, France
Cardioversion with a pulsed biphasic waveform
Cardioversion with a biphasic truncated exponential waveformProcedure or surgery
Also called LIFEPAK 15 monitor/defibrillator 41577-000267
Cardioversion with a biphasic truncated exponential waveform
| Study group | Role | What they receive |
|---|---|---|
| Schiller Defigard HD- 7 - DGHD7 | Experimental | Device: Cardioversion with a pulsed biphasic waveform Cardioversion is performed by a pulsed biphasic (Multipulse Biowave®) waveform (Schiller Defigard HD- 7 - DGHD7, Schiller Medical, France) with adult pads (0-21-0003 Schiller) following an energy protocol of 3 consecutive shocks with escalating selected energy: 150J, 200J, 200J. The third shock is combined with MAP (Manual Pressure Application). The protocol is stopped at successfull cardioversion (sinus rhythm at 1 min post-shock), otherwise after the 3rd shock Other Name: DGHD7 Receives: Procedure: Cardioversion with a pulsed biphasic waveform |
| LIFEPAK 15, Physio-Control - LP15 | Active comparator | Device: Cardioversion with a biphasic truncated exponential waveform Cardioversion is performed by a biphasic truncated exponential waveform (LIFEPAK 15, Physio-Control Inc., Redmond, WA, USA) with recommended by the manufacturer adult pads (Redipak QUICK COMBO, Physio-Control) following an energy protocol of 3 consecutive shocks with escalating selected energy: 150J, 360J, 360J. The third shock is combined with MAP (Manual Pressure Application). The protocol is stopped at successfull cardioversion (sinus rhythm at 1 minute post-shock), otherwise after the 3rd shock.
Other Name: LP15 Receives: Procedure: Cardioversion with a biphasic truncated exponential waveform |
What is being measured
The primary outcome is the main question the study is designed to answer.
Defibrillator Efficacy
The cumulative delivered energy by the consecutive defibrillation shocks during the successful cardioversion procedure.
Measured: Delivered energy (Joules) will be measured during each shock with a dedicated pulse recording device
Secondary outcomes (7)
The cumulative success rate at 1 minute
The cumulative success rate (at 1 minute) of the consecutive defibrillation shocks during the cardioversion procedure is defined as the conversion to sinus rhythm at 1 minute post shock.
Measured: Peripheral ECG will be continuously recorded during cardioversion and the presence of sinus rhythm will be read by a cardiologist at the first minute after each shock
Number of shocks
Number of shocks
Measured: During the whole cardioversion procedure, each electrical shock delivered to the patient will be counted
Changes in hsTnI at 8-12 hours
Changes in hsTnI before and after the cardioversion procedure
Measured: the blood samples will be taken before cardioversion (on the same day) and after cardioversion (from 8 to 12 hours after the intervention)
Changes in CK and CK-MB at 8-12 hours
Changes in CK and CK-MB before and after the cardioversion procedure before and after the cardioversion procedure
Measured: the blood samples will be taken before cardioversion (on the same day) and after cardioversion (from 8 to 12 hours after the intervention)
The cumulative success rate (at 2 hours) of the consecutive defibrillation shocks during the cardioversion procedure
The cumulative success rate (at 2 hours) of the consecutive defibrillation shocks during the cardioversion procedure, is defined as the conversion to sinus rhythm at 2 hours after the last shock of the procedure.
Measured: Peripheral ECG will be continuously recorded during cardioversion and the presence of sinus rhythm will be read by a cardiologist at two hours after each shock
The cumulative success rate (at 24 hours) of the consecutive defibrillation shocks during the cardioversion procedure
The cumulative success rate (at 24 hours) of the consecutive defibrillation shocks during the cardioversion procedure, is defined as the conversion to sinus rhythm at 24 hours after the last shock of the procedure.
Measured: Peripheral ECG will be continuously recorded during cardioversion and the presence of sinus rhythm will be read by a cardiologist at twenty four hours after each shock
ST-segment changes after each defibrillation shock
ST-segment changes after each defibrillation shock
Measured: ST-segment will be measured in the continuously recorded lead II 10 seconds after each defibrillation shock (0.080 seconds after J point in the first QRS at 10 seconds post-shock)
Timeline
From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.
- First posted7 Apr 2022
- Study start20 Jan 2022
- Primary completion30 May 2024 (estimated)
- Study completion30 Jun 2024 (estimated)
Resources
- Official record on ClinicalTrials.govThe registry entry this page is built from, with its full change history.
- Publications (6)Papers the sponsor has linked to this study.
- Other trials for Atrial FibrillationStudies of the same condition, open and under way.
- What the trial phases meanHow a treatment moves from first tests in people to approval.
More details
For doctors and researchers: how the study is designed, who runs it, and its registry IDs.
Study design
- Study type
- Interventional
- Phase
- Phase N/A
- Primary purpose
- Treatment
- Allocation
- Randomised
- Intervention model
- Parallel groups
- Model description
- Randomized Interventional study, Alternating Assignment, Open Label
- Masking
- None (open label)
- Enrolment
- 94 participants (estimated)
Sponsor and investigators
- Lead sponsor
- University National Heart Hospital
- Sponsor type
- University, hospital or other
- Responsible party
- Principal investigator, Elina Trendafilova, University National Heart Hospital
- Investigators
- Elina G Trendafilova, MD, PhD, Principal investigator, UniversityNationalHeartH, ICCU
IDs, dates and publications
- ClinicalTrials.gov ID
- NCT05316753
- Sponsor's protocol ID
- UniversityNationalHeartH 1
- Status last verified
- Jul 2022
- First submitted
- 8 Mar 2022
- Last updated
- 25 Jul 2022
- Results posted
- No
- Data monitoring committee
- No
- FDA-regulated drug
- No
- FDA-regulated device
- No
- Individual data sharing
- No
Publications
Lavignasse D, Trendafilova E, Dimitrova E, Krasteva V. Cardioversion of Atrial Fibrillation and Flutter: Comparative Study of Pulsed vs. Low Energy Biphasic Truncated Exponential Waveforms. J Atr Fibrillation. 2019 Oct 31;12(3):2172. doi: 10.4022/jafib.2172. eCollection 2019 Oct-Nov.
PubMed 32435331Coats AJS, Anker SD, Baumbach A, Alfieri O, von Bardeleben RS, Bauersachs J, Bax JJ, Boveda S, Celutkiene J, Cleland JG, Dagres N, Deneke T, Farmakis D, Filippatos G, Hausleiter J, Hindricks G, Jankowska EA, Lainscak M, Leclercq C, Lund LH, McDonagh T, Mehra MR, Metra M, Mewton N, Mueller C, Mullens W, Muneretto C, Obadia JF, Ponikowski P, Praz F, Rudolph V, Ruschitzka F, Vahanian A, Windecker S, Zamorano JL, Edvardsen T, Heidbuchel H, Seferovic PM, Prendergast B. The management of secondary mitral regurgitation in patients with heart failure: a joint position statement from the Heart Failure Association (HFA), European Association of Cardiovascular Imaging (EACVI), European Heart Rhythm Association (EHRA), and European Association of Percutaneous Cardiovascular Interventions (EAPCI) of the ESC. Eur Heart J. 2021 Mar 31;42(13):1254-1269. doi: 10.1093/eurheartj/ehab086.
PubMed 33734354Ramirez FD, Sadek MM, Boileau I, Cleland M, Nery PB, Nair GM, Redpath CJ, Green MS, Davis DR, Charron K, Henne J, Zakutney T, Beanlands RSB, Hibbert B, Wells GA, Birnie DH. Evaluation of a novel cardioversion intervention for atrial fibrillation: the Ottawa AF cardioversion protocol. Europace. 2019 May 1;21(5):708-715. doi: 10.1093/europace/euy285.
PubMed 30535367Schmidt AS, Lauridsen KG, Torp P, Bach LF, Rickers H, Lofgren B. Maximum-fixed energy shocks for cardioverting atrial fibrillation. Eur Heart J. 2020 Feb 1;41(5):626-631. doi: 10.1093/eurheartj/ehz585.
PubMed 31504412Trendafilova E, Dimitrova E, Didon JP, Krasteva V. A Randomized Comparison of Delivered Energy in Cardioversion of Atrial Fibrillation: Biphasic Truncated Exponential Versus Pulsed Biphasic Waveforms. Diagnostics (Basel). 2021 Jun 17;11(6):1107. doi: 10.3390/diagnostics11061107.
PubMed 34204498Voskoboinik A, Moskovitch J, Plunkett G, Bloom J, Wong G, Nalliah C, Prabhu S, Sugumar H, Paramasweran R, McLellan A, Ling LH, Goh CY, Noaman S, Fernando H, Wong M, Taylor AJ, Kalman JM, Kistler PM. Cardioversion of atrial fibrillation in obese patients: Results from the Cardioversion-BMI randomized controlled trial. J Cardiovasc Electrophysiol. 2019 Feb;30(2):155-161. doi: 10.1111/jce.13786. Epub 2018 Nov 14.
PubMed 30375104
Study at a glance
From the registry record, last updated by the sponsor on 25 Jul 2022. Recruitment status is as the sponsor reported it: confirm with the study team.
- Official title
- Comparison of Biphasic Truncated Exponential and Pulsed Waveforms for Cardioversion of Atrial Fibrillation-High Energy Escalation Strategy and Introduction to Manual Pressure Application
- ClinicalTrials.gov ID
- NCT05316753
- Acronym / sponsor ID
- UniversityNationalHeartH 1
- Conditions
- Atrial Fibrillation
- Keywords
- electrical cardioversion
- Study type
- Interventional
- Recruitment status
- Unknown status
- Enrolment
- 94 participants (estimated)
- Lead sponsor
- University National Heart HospitalUniversity, hospital or other
- Study start
- 20 Jan 2022 (actual)
- Primary completion
- 30 May 2024 (estimated)
- Study completion
- 30 Jun 2024 (estimated)
- First posted
- 7 Apr 2022 (actual)
- Results
- Not posted in the source record
Clinical summary
The research question as the sponsor registered it. This describes what the study plans to test, not what it found.
Background: A Danish study raised the question of the usefulness of escalating energy protocols compared to fixed high-energy protocols. Maximal energies are usually the final choice of the physicians. Some authors showed that decreasing impedance by manual pressure application (MPA) had a positive impact on cardioversion outcome. This is likely due to the impedance decrease linked to MPA.
Objective: This new clinical cardioversion study of atrial fibrillation (AF) patients aims to compare the efficacy and safety of a new high energy escalation strategy. The protocol combines high energy shocks at first shock, jumping to maximal defibrillator energy at second shock and finally complemented by MPA at third shock, if success is not reached using electric shocks only.
Experimental design: Patients will be recruited at the Intensive Cardiology Care Unit, Cardiology Clinic, National Cardiology Hospital (NCH), Sofia, Bulgaria. All eligible patients will sign a written informed consent prior to the cardioversion and will receive the standard hospital procedures during cardioversion.
AF patients will be alternatively randomized to cardioversion using one of the two defibrillators, following the strategy below:
DEFIGARD HD-7 arm: 3 consecutive shocks with escalating selected energy: 150J, 200J, 200J. The third shock is combined with MPA LIFEPAK15 arm: 3 consecutive shocks with escalating selected energy: 150J, 360J, 360J. The third shock is combined with MPA The statistical power analysis will consider a superiority comparison between the cumulative energy actually delivered by both defibrillators.
The secondary cardioversion efficacy outcome measures are: the cumulative success rate (measured at 1 minute post-shock), number of delivered shocks. Delivered energy will be measured during each shock with a dedicated pulse recording device (approved by the NCH Ethical Committee). Heart rhythm will be measured in continuously recorded peripheral ECG.
The secondary cardioversion safety outcome measures are: markers for myocardial necrosis (high sensitive troponin I, CK-MB) evaluated on blood samples taken before and 8-12 hours after cardioversion; ST-segment changes (post-shock - pre-shock) measured in lead II; Complications after cardioversion measured during 2 hours follow-up period in the ICCU - the presence of apnea, arrhythmias, bradycardia and the need for respective therapy at the discretion of attending physician.
Detailed description
Background Commercially available defibrillators generate direct current (DC) shock with various waveforms using distinct technologies. These waveforms are suggested to have different efficacy and safety. Although the superiority of biphasic over monophasic waveforms is well established, the relative efficacy and safety of the available biphasic waveforms is not clear.
Various BTE waveforms apply different potential gradients on the thorax that might produce various defibrillation effects. Apart from efficacy, the most important aspect is the patient safety, considering that larger potential gradients in the myocardium lasting longer could potentially induce an electroporation and then a fibrillation. A frequently observed effect of electroporation are the post-shock ST-segment deviations in the surface ECG, representing the potential difference between the normal tissue and sustained depolarized critical mass of myocardium closest to the associated origin of the electrical current. Although ST-segment changes are an easily ignored phenomenon, occurring acutely and resolving during the first few minutes post-shock, their presence in a short-term basis can identify electroporation by dangerously high potential gradients, while the sustained ST-changes in a long-term basis can identify cases with myocardial injury. The most reliable marker of shock-induced myocardial injuries is the high sensitive cardiac Troponin I (hsTnI).
In a previous study Trendafilova et al. were compared fixed-high energy protocols (200J-200J-200J) delivered by two different defibrillation waveforms: Pulsed (BTE-PE) and High-energy BTE (BTE-HE). This study concluded on non-inferiority between devices. Cumulative energy, efficiency of the waveforms and safety (measured by High-sensitive Troponin) were not reaching significant difference for both waveforms. Delivered energy of the first efficient shock was significantly lower for BTE-PE and although BTE-PE delivered almost 28J lower energy after the whole procedure, significance could not be reached due to wide variance.
In a Danish study Schmidt et al. compared two cardioversion protocols - one is fixed-high energy (360J-360J-360J), the other is low energy escalating (125J-150J-200J). The overall efficacy of fixed-energy protocol was better than escalating. Safety is non significantly different. Schmidt et al. raised the question of the usefulness of escalating energy protocols compared to fixed high-energy protocols.
There is an ethical consideration to apply maximal energy shocks to all patients without considering patient individuality. Ramirez et al. remind that failed shocks might depend on other factors like transthoracic impedance. Lavignasse et al. found that cumulative energy setting differs with patient characteristics, having significant association with: AF duration, gender, BSA, LV TDD, valvular disease and chronic respiratory disease. Indeed, some authors (Voskoboinik et al. and Ramirez et al.) showed that decreasing impedance by manual pressure application (MPA) had a positive impact on cardioversion outcome.
On most defibrillators, increasing the energy level is leading to an increase of the voltage delivered to the patient. This voltage increase implies a current increase in the thorax of the patient. In fact, applying a force on the defibrillation pads or increasing shock voltage by increasing the energy setting would lead to increase the current in the thorax, associated with expectations for higher shock efficiency. MPA is performed by pressing on the pads using a non-connected pair of paddles in order to assure isolation of the practitioner from the patient during the shock.
Objective:
This new clinical cardioversion study of atrial fibrillation (AF) patients aims to compare the efficacy and safety of a new high energy escalation strategy. The protocol combines reasonably high energy shocks complemented by MPA when necessary.
Two biphasic defibrillators - a standard BTE-HE waveform (Lifepak 15, Physio Control) and a BTE-PE waveform (Defigard HD 7, Schiller Medical) with different maximal energy settings of 360J and 200J, respectively are compared.
The proposed high energy escalation strategy is based on the hypothesis that most of the patients do not implicitly require the maximal levels of energy for cardioversion. Thus patients who could be easily cardioverted are removed from the pool that would need a more aggressive treatment.
Methods:
Study population All patients admitted for elective cardioversion of atrial fibrillation in ICCU- NCH will be potentially eligible for the study. Informed consent will be obtained from all eligible patients prior to cardioversion. Patients declining to participate in the study will receive treatment according to the hospital protocol.
Cardiologists will check the eligibility of each patient following the detailed list of inclusion and exclusion criteria, based on the documented pre-treatment data.
Additionally, eligibility will be verified through pre-cardioversion medical exams (blood test within 24 hours before cardioversion): routine blood chemistry, glucose, creatinine, eGFR, K, Na; CK-MB, high sensitive troponin I measurements on the day of cardioversion.
Study design This will be a prospective randomized trial (alternating design) where eligible patients will be randomized to treatment with two different defibrillators. Following the order of patient admittance in ICCU-NCH, the attending cardiologist will assign the odd and even eligible patients to the defibrillators in arm 1 and arm 2, respectively. The cardiologist cannot control the order of patient admittance in ICCU-NCH. The alternative randomization design will equalize the number of subjects on each treatment. This study is not designed to control for sex, age, comorbidity, type of device used for cardioversion, cumulative energy delivered during shocks, number of shocks administered. Randomization is performed to limit these and any other confounding factors.
Patient preparation Patients will be consulted prior to cardioversion (CV) in a quiet setting. Any questions will be answered by the attending cardiologist. Patients will be asked to manually sign their informed consent for study participation before further preparations and data collection. If the patient declines to participate, then he/she will receive treatment according to the hospital protocol without data documentation, as further defined in the study.
Demographic data will be taken. Pre-treatment data will be read from the patient files: concomitant diseases; medications during the previous 7 days; Left atrium dimensions, left ventricle dimensions and volumes, ejection fraction by transthoracic echocardiography during index hospitalization; TEE will be performed for all patients before CV - presence of cardiac thrombus and spontaneous echo contrast will be checked; duration of index AF in days; European Heart Rhythm Association class score of symptoms from index AF; standard blood test will be documented.
Appropriate standard anticoagulation with unfractionated heparin (UFH) or acenocumarol or direct oral anticoagulants (DOAC) is required before and after the CV.
The anesthesia will be conducted by an anesthesiologist with slow intravenous injection of Propofol, adjusted individually to reach deep sedation (Cook' s scale points <7).
Cardioversion protocol:
The CV procedure is performed in ICCU by a cardiologist and anesthesiologist. Defibrillator pads are placed according to the user guides on the pads and respecting >40 cm distance between them. Standard self-adhesive defibrillation pads, recommended by specific defibrillator's manufacturer will be used.
Defibrillator use: The time-interval interval between consecutive shocks is 1 min; The energy of consecutive shocks follows pre-defined protocol; The protocol stops at successful cardioversion (sinus rhythm at 1 min post-shock), otherwise at the last shock of the protocol.
Data collection and Follow up:
- Defibrillator pulse recorder: The shock waveforms delivered by the studied defibrillators is recorded in real-time during the shock by a dedicated measurement device (Defimpulse recorder DR2) connected in the patient circuit. This allows accurate measurement of the delivered energy, which is the primary measured outcome. Additionally, currents, voltages, patient impedances at each shock can be also measured. The device is fully automatic and expert assistance is not need during the CV intervention. The use of the measurement device (Defimpulse recorder DR2) has been approved by the NCH Ethical Committee.
- Blood samples, standard ECG, blood pressure and heart rate at baseline.
- Continuous ECG in 3 peripheral leads is recorded during the whole CV procedure. ST deviation is measured on the first cardiac complex after 10 s post-shock in standard manner (0.080 seconds after J point) in lead II.
- Standard ECG is recorded in 5 min after CV; 24 h after CV; at discharge if the discharge day is different from 24 h after CV.
- Heart rate and blood pressure are measured 3 times after CV during follow-up in ICCU
- Follow-up period in ICCU (2 hours after CV): Potential complications after CV are recorded
- Follow-up period in the Cardiology Clinic (24 hours after CV): blood samples to analyze CK-MB and troponin (8 to 12 h after CV); Clinical exam with heart rate and blood pressure; ECG; Any complications before discharge.
Statistics:
For arm1 (BTE PE waveform), the expected efficacy is 50% at the first shock (150J), 80% at the second shock (200J) and 85% at the third shock (200J combined with MPA).
For arm2 (BTE HE waveform), the expected efficacy is 50% at the first shock (150J), 80% at the second shock (360J) and 85% at the third shock (360J combined with MPA).
Taking into account that BTE PE waveform typically delivers 180J for energy setting of 200J, the estimated cumulative energy at the end of the CV procedure is 258J (+/-120J) and 366J (+/-240J) for arm1 and arm2, respectively.
The statistical analysis is designed for a superiority comparison between the two defibrillation strategies in respect of their delivered cumulative energies hypothesizing similar cardioversion efficacy rate. It is estimated that a total of 94 patients (47 per arm) would allow to reach a statistical power of 80% in a superiority study design with one-sided type I error rate of 0.05.
Continuous variables will be expressed as mean value ± standard deviation or median values (inter-quartile range) and compared with Students t-test or equivalent non-parametric test, respectively. Categorical variables are expressed as percentages and compared using the Chi-square or Fisher's exact test . P<0.05 is considered statistically significant for all comparisons.
Ethics All patients will sign a written informed consent prior to the cardioversion and will receive the standard hospital procedures during cardioversion. Both defibrillators included in this study are approved for clinical use and both are used on a daily basis.
Data anonymization policy will be respected. In order to ensure the medical confidentiality, no information about the patient will be entered in the defibrillators. An identification number will be given to each patient by the dedicated measurement device (Defimpulse recorder DR2). This number and the recording data are properties of the principal investigator. Approval from the NCH Ethical Committee concerning energy strategy and the use of the Defimpulse recorder during the cardioversion intervention has been obtained.
Study design
- Primary purpose
- Treatment
- Allocation
- Randomised
- Intervention model
- Parallel groups
- Masking
- None (open label)
- Number of arms
- 2
- Model description
- Randomized Interventional study, Alternating Assignment, Open Label
| Arm | Type | Receives |
|---|---|---|
| Schiller Defigard HD- 7 - DGHD7 | Experimental | Procedure: Cardioversion with a pulsed biphasic waveformDevice: Cardioversion with a pulsed biphasic waveform Cardioversion is performed by a pulsed biphasic (Multipulse Biowave®) waveform (Schiller Defigard HD- 7 - DGHD7, Schiller Medical, France) with adult pads (0-21-0003 Schiller) following an energy protocol of 3 consecutive shocks with escalating selected energy: 150J, 200J, 200J. The third shock is combined with MAP (Manual Pressure Application). The protocol is stopped at successfull cardioversion (sinus rhythm at 1 min post-shock), otherwise after the 3rd shock Other Name: DGHD7 |
| LIFEPAK 15, Physio-Control - LP15 | Active comparator | Procedure: Cardioversion with a biphasic truncated exponential waveformDevice: Cardioversion with a biphasic truncated exponential waveform Cardioversion is performed by a biphasic truncated exponential waveform (LIFEPAK 15, Physio-Control Inc., Redmond, WA, USA) with recommended by the manufacturer adult pads (Redipak QUICK COMBO, Physio-Control) following an energy protocol of 3 consecutive shocks with escalating selected energy: 150J, 360J, 360J. The third shock is combined with MAP (Manual Pressure Application). The protocol is stopped at successfull cardioversion (sinus rhythm at 1 minute post-shock), otherwise after the 3rd shock. Other Name: LP15 |
Who may qualify
- Ages
- 18 years and older
- Sex
- All
- Healthy volunteers
- Not accepted
- Age groups
- Adult, Older adult
Key inclusion criteria (7 listed)
- Indications for elective cardioversion of atrial fibrillation
- Patients > 18 years old and:
- Symptomatic AFIB with a duration of less than 12 months and EHRA score 2-4
- Symptomatic first detected AFIB and EHRA score 2-4
- Persistent AFIB after successful causal therapy
Key exclusion criteria (13 listed)
- Patients with atrial flutter
- Spontaneous HR <60/min
- Digitalis intoxication
- Impossibility to maintain sinus rhythm irrespective to antiarrhythmic therapy and frequent cardioversions
- Conduction disturbances (without fascicular block and AV block 1 degree) in patients without pacemaker
Eligibility is decided by the study team after screening; this overview does not establish that a patient qualifies. Full inclusion and exclusion criteria.
Interventions
| Intervention | Type | Arms |
|---|---|---|
| Cardioversion with a pulsed biphasic waveformAlso: Schiller Defigard HD- 7 - DGHD7, Schiller Medical, FranceCardioversion with a pulsed biphasic waveform | Procedure or surgery | Schiller Defigard HD- 7 - DGHD7 |
| Cardioversion with a biphasic truncated exponential waveformAlso: LIFEPAK 15 monitor/defibrillator 41577-000267Cardioversion with a biphasic truncated exponential waveform | Procedure or surgery | LIFEPAK 15, Physio-Control - LP15 |
Primary outcome
Defibrillator Efficacy
Time frame: Delivered energy (Joules) will be measured during each shock with a dedicated pulse recording device
The cumulative delivered energy by the consecutive defibrillation shocks during the successful cardioversion procedure.
7 secondary outcomes
The cumulative success rate at 1 minute
Time frame: Peripheral ECG will be continuously recorded during cardioversion and the presence of sinus rhythm will be read by a cardiologist at the first minute after each shock
Number of shocks
Time frame: During the whole cardioversion procedure, each electrical shock delivered to the patient will be counted
Changes in hsTnI at 8-12 hours
Time frame: the blood samples will be taken before cardioversion (on the same day) and after cardioversion (from 8 to 12 hours after the intervention)
Changes in CK and CK-MB at 8-12 hours
Time frame: the blood samples will be taken before cardioversion (on the same day) and after cardioversion (from 8 to 12 hours after the intervention)
The cumulative success rate (at 2 hours) of the consecutive defibrillation shocks during the cardioversion procedure
Time frame: Peripheral ECG will be continuously recorded during cardioversion and the presence of sinus rhythm will be read by a cardiologist at two hours after each shock
The cumulative success rate (at 24 hours) of the consecutive defibrillation shocks during the cardioversion procedure
Time frame: Peripheral ECG will be continuously recorded during cardioversion and the presence of sinus rhythm will be read by a cardiologist at twenty four hours after each shock
ST-segment changes after each defibrillation shock
Time frame: ST-segment will be measured in the continuously recorded lead II 10 seconds after each defibrillation shock (0.080 seconds after J point in the first QRS at 10 seconds post-shock)
Results and safety
Results are not available in the source record.
Research links
- Official record on ClinicalTrials.govLast updated by the sponsor 25 Jul 2022.
- Background: PubMed 32435331Lavignasse D, Trendafilova E, Dimitrova E, Krasteva V. Cardioversion of Atrial Fibrillation and Flutter: Comparative Study of Pulsed vs. Low Energy Biphasic Truncated Exponential Waveforms. J Atr Fibrillation. 2019 Oct 31;12(3):2172. doi: 10.4022/jafib.2172. eCollection 2019 Oct-Nov.
- Background: PubMed 33734354Coats AJS, Anker SD, Baumbach A, Alfieri O, von Bardeleben RS, Bauersachs J, Bax JJ, Boveda S, Celutkiene J, Cleland JG, Dagres N, Deneke T, Farmakis D, Filippatos G, Hausleiter J, Hindricks G, Jankowska EA, Lainscak M, Leclercq C, Lund LH, McDonagh T, Mehra MR, Metra M, Mewton N, Mueller C, Mullens W, Muneretto C, Obadia JF, Ponikowski P, Praz F, Rudolph V, Ruschitzka F, Vahanian A, Windecker S, Zamorano JL, Edvardsen T, Heidbuchel H, Seferovic PM, Prendergast B. The management of secondary mitral regurgitation in patients with heart failure: a joint position statement from the Heart Failure Association (HFA), European Association of Cardiovascular Imaging (EACVI), European Heart Rhythm Association (EHRA), and European Association of Percutaneous Cardiovascular Interventions (EAPCI) of the ESC. Eur Heart J. 2021 Mar 31;42(13):1254-1269. doi: 10.1093/eurheartj/ehab086.
- Background: PubMed 30535367Ramirez FD, Sadek MM, Boileau I, Cleland M, Nery PB, Nair GM, Redpath CJ, Green MS, Davis DR, Charron K, Henne J, Zakutney T, Beanlands RSB, Hibbert B, Wells GA, Birnie DH. Evaluation of a novel cardioversion intervention for atrial fibrillation: the Ottawa AF cardioversion protocol. Europace. 2019 May 1;21(5):708-715. doi: 10.1093/europace/euy285.
- Background: PubMed 31504412Schmidt AS, Lauridsen KG, Torp P, Bach LF, Rickers H, Lofgren B. Maximum-fixed energy shocks for cardioverting atrial fibrillation. Eur Heart J. 2020 Feb 1;41(5):626-631. doi: 10.1093/eurheartj/ehz585.
- Background: PubMed 34204498Trendafilova E, Dimitrova E, Didon JP, Krasteva V. A Randomized Comparison of Delivered Energy in Cardioversion of Atrial Fibrillation: Biphasic Truncated Exponential Versus Pulsed Biphasic Waveforms. Diagnostics (Basel). 2021 Jun 17;11(6):1107. doi: 10.3390/diagnostics11061107.
- Background: PubMed 30375104Voskoboinik A, Moskovitch J, Plunkett G, Bloom J, Wong G, Nalliah C, Prabhu S, Sugumar H, Paramasweran R, McLellan A, Ling LH, Goh CY, Noaman S, Fernando H, Wong M, Taylor AJ, Kalman JM, Kistler PM. Cardioversion of atrial fibrillation in obese patients: Results from the Cardioversion-BMI randomized controlled trial. J Cardiovasc Electrophysiol. 2019 Feb;30(2):155-161. doi: 10.1111/jce.13786. Epub 2018 Nov 14.
- Individual participant dataSharing planned: No
Related trials for Atrial Fibrillation
Active, not recruiting
Horizon 360 Protocol for the Treatment of Paroxysmal Atrial Fibrillation With the Sphere-360™ Catheter and Affera™ Mapping and Ablation System (Horizon 360)
Paroxysmal AF
Active, not recruitingObservational study
Concomitant Left Atrial Appendage Closure and Pulsed Field Ablation-Asia
Concomitant Procedures, Atrial Fibrillation
Active, not recruiting
CONVERGE Post-Approval Study (PAS)
Chronic Atrial Fibrillation
Recruiting
Early Closure of Left Atrial Appendage for Patients With Atrial Fibrillation and Ischemic Stroke Despite Anticoagulation Therapy
Ischemic Stroke, Atrial Fibrillation
RecruitingPhase 2
Safety in Adult Participants With Atrial Fibrillation Who Are Treated With Anticoagulation
Atrial Fibrillation (AF)
Referral and contact
AzTrials does not enrol participants or take referrals. Contact the study team or a site listed below; they decide who can take part. This study is not recruiting at the moment.
Elina G Trendafilova, MD, PhD principal investigator
UniversityNationalHeartH, ICCU
1 site in 1 country: see every site and its status.
Questions about this trial
What is this clinical trial studying?
A study testing Cardioversion with a pulsed biphasic waveform and Cardioversion with a biphasic truncated exponential waveform in people with Atrial Fibrillation, run by University National Heart Hospital at 1 site in Bulgaria.
The full registered description is in the About this study section and the official record.
Who may be eligible to take part?
The sponsor lists 7 inclusion and 13 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.
Is this clinical trial currently recruiting?
The record says: status not confirmed, as of the sponsor's last update on 25 Jul 2022. Status can change between updates, so confirm with the study contact before making plans.
Where is the study taking place?
The record lists 1 site in Bulgaria. Whether a site is taking part right now should be confirmed with that site. See every site.
What are the possible benefits and risks?
They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.
Are results available for this study?
Results are not available in the source record.
How can a doctor refer a patient or contact the research team?
Contact Elina G Trendafilova, MD, PhD (+359888718847, elitrendafilova@abv.bg), or a site near the patient. AzTrials does not enrol participants or take referrals.
What does the primary outcome measure?
Defibrillator Efficacy (time frame: Delivered energy (Joules) will be measured during each shock with a dedicated pulse recording device). It is what the study was designed to measure, not a finding.
Where can I read the original study information?
On the ClinicalTrials.gov record for NCT05316753, which also lists 6 linked publications.
Does a listing mean the treatment is approved or proven to work?
No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.
How often is this information updated?
AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 25 Jul 2022.