Apnea of PrematurityDREAMNCT05196646

Detection of CardioRespiratory Events Using Acoustic Monitoring in Preterm Infants on CPAP

An observational study of Apnea of Prematurity, run by McGill University Health Centre/Research Institute of the McGill University Health Centre at 1 site in Canada.

Active, not recruiting

Under way, not looking for participants. Everyone needed has enrolled. The study is still running.

Official record
Search all trials

Trial summary

This is an observational, proof-of-concept, feasibility study where 50 preterm infants with gestational age < 32+0 weeks will be recruited from the neonatal intensive care unit (NICU) at the Montreal Children's Hospital. Read more about this study

Who is it for?
Ages 72 hours and older
Conditions the study is for
Apnea of Prematurity
What is being tested?
Respiratory Acoustic Sensors, Nasal thermistor and Respiratory Inductive Plethysmography
How many people
50 participants
Where
1 siteCanada
Study dates
5 Dec 2022to 31 Dec 2026
Phase
Observational
Are there results?
Not yet: the study is under way

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 72 hours and older
  • All sexes
  • No healthy volunteers
  • Child, Adult, Older adult

Study population: Preterm infants admitted to the neonatal intensive care unit at the Montreal Children's Hospital.

Who may be able to join

  • Gestational age < 32+0 weeks
  • Postmenstrual age between 28+0 and 36+6 weeks.
  • Additional inclusion criteria for Groups 1 and 2:
  • Off any respiratory support and breathing in-room air
  • Less than 3 clinically significant cardiorespiratory events per calendar day
  • Additional inclusion criteria for Group 3:
Show all 9Show fewer
  • On the bubble CPAP device with the binasal prongs interface
  • Receiving CPAP levels of 5 to 7 cm H2O with gas flows not exceeding 10L/min
  • At least 3 clinically significant cardiorespiratory events per calendar day

Who cannot join

  • Major known congenital abnormalities
  • Known congenital heart disorders
  • Known neuromuscular disease
  • Known diaphragmatic paralysis or a diagnosed phrenic nerve injury
  • History of esophageal perforation in the 7 days preceding the study
  • History of pneumothorax requiring chest tube insertion in the 7 days preceding the study
Show all 12Show fewer
  • Receiving inotropes, narcotics, or sedative agents at the time of study recording
  • Additional exclusions at the time of the study recording:
  • Infants receiving ventilator-derived CPAP
  • Infants receiving CPAP via a nasal mask interface.
  • Infants receiving inotropes, narcotics or sedative agents
  • Infants deemed clinically unstable for the study by the attending neonatologist.

Trial locations and contacts

1 site in 1 country.

Canada1 site
  • McGill University Health Center

    Montreal, Quebec, H4A 3J1

Other trials in Canada

About this study

Official title: Detection of CardioRespiratory Events Using Acoustic Monitoring in Preterm Infants on Continuous Positive Airway Pressure: the DREAM Pilot Project

This is an observational, proof-of-concept, feasibility study where 50 preterm infants with gestational age < 32+0 weeks will be recruited from the neonatal intensive care unit (NICU) at the Montreal Children's Hospital.

The study's primary objective is to describe the relationship between respiratory acoustics and airflow and determine the reliability of a novel respiratory acoustic sensor at detecting breathing sounds in preterm infants.

Read more

The study's secondary objectives are:

  1. To compare transthoracic impedance, respiratory inductive plethysmography and an inertial measurement unit for the detection of respiratory efforts in preterm infants.
  2. To evaluate the feasibility and accuracy of a novel, non-invasive method for continuously detecting and differentiating cardiorespiratory events in preterm infants on CPAP by integrating measurements of respiratory effort with respiratory acoustic monitoring.
Read the detailed description

Cardiorespiratory events, defined by the occurrence of apneas, bradycardias, and desaturations, are almost ubiquitous in very preterm infants and are associated with numerous complications. Unfortunately, the current standard for monitoring cardiorespiratory events in the NICU, transthoracic impedance (TTI), does not permit for accurate differentiation of the different types of cardiorespiratory events; TTI cannot detect airflow and has low accuracy for detecting respiratory efforts. As a result, TTI does not detect obstructive apneas and may not reliably capture all central apneas.

Respiratory sounds are an attractive surrogate measure of airflow, and can be captured using respiratory acoustic technology (akin to a miniaturized electronic stethoscope). We hypothesize that respiratory acoustic monitoring can provide a continuous, non-invasive, and accurate representation of airflow and breathing sounds in preterm infants.

Altogether, we conjecture that the combination of respiratory acoustic monitoring with measurements of respiratory effort will improve the ability to differentiate and describe the nature of cardiorespiratory events in preterm infants.

Being tested
Respiratory Acoustic Sensors, Nasal thermistor, Respiratory Inductive Plethysmography, Pneumotachometer
Keywords
Airflow, Apnea, Cardiorespiratory events, Monitoring, Neonatal Intensive Care, Preterm infants, Respiratory acoustics, Prematurity, CPAP, Acoustic monitoring
Medical subject headings
Apnea, Premature Birth

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Respiratory Acoustic SensorsDevice

    Wireless sensor that contains a dual microphone and an inertial measurement unit (IMU) will capture the breathing sound and respiratory effect. Two wireless sensors will be used, with one placed on the suprasternal notch and the other placed on the right upper chest of the infant, in order to determine the sensor placement yielding the best respiratory signal. Data will be transmitted in real-time to a research-dedicated tablet using the Bluetooth Communication Controller (ISP1807, Insight SIP) and stored on the same device for future analysis.

  • Nasal thermistorDevice

    Also called MLT415/A®, ADInstruments®, Bella Vista, Australia

    The nasal temperature probe that detects changes in temperature between inhaled and exhaled gases allows for the surrogate measure of airflow. It will be placed in one naris and secured with tape at the upper lip or cheek. The nasal temperature signal will be acquired using the Power Lab analog-digital acquisition system and stored for later analysis.

  • Respiratory Inductive PlethysmographyDevice

    Also called SleepSense®, Scientific Laboratory Products, Elgin, USA

    Two respiratory bands will be placed circumferentially around the infant's chest (at the level of nipple line) and around the abdomen (just above the level of the umbilicus) in order to measure chest and abdominal wall movements, respectively. These movements will be recorded using Respiratory Inductive Plethysmography (Respitrace QDC®, Viasys® Healthcare, USA). The Respitrace® signals will be acquired using the Power Lab data acquisition system and stored for later analysis.

  • PneumotachometerDevice

    Also called Hans Rudolph, Inc, Shawnee, KS

    The pneumotachometer is a pressure-differential based flow sensor that is used to measure respiratory flow. It will be connected to a standard face mask that is gently applied to cover the infant's mouth and nose. The face mask will be similar to the masks used as part of standard of care in the NICU for infants who require continuous positive pressure, with or without ventilation. The flow measurements will be recorded using the Power Lab data acquisition system and stored for later analysis.

Study groups in NCT05196646
Study groupRoleWhat they receive
(1) 10 preterm infants spontaneously breathing in-room air with no respiratory supportGroup 1 will consist of 10 preterm infants spontaneously breathing in room air, with no respiratory support, in whom respiratory acoustic signals from the acoustic sensor will be compared with airflow measurements obtained using a pneumotachometer, i.e. the gold standard. Data will be acquired for 10 minutes.
Receives: Device: Respiratory Acoustic Sensors; Device: Pneumotachometer
(2) 20 preterm infants spontaneously breathing in-room air with no respiratory supportGroup 2 will consist of 20 preterm infants spontaneously breathing in room air, with no respiratory support, in whom respiratory acoustic signals from the acoustic sensor will be compared with airflow measurements obtained using a nasal temperature sensor. In addition, measurements of respiratory efforts will be obtained using the Respiratory Inductance Plethysmography (RIP), an inertial measurement unit (IMU) integrated within the acoustic sensor, and the Transthoracic Impedance (TTI) from the bedside monitor. Data will be continuously recorded for 3 hours.
Receives: Device: Respiratory Acoustic Sensors; Device: Nasal thermistor; Device: Respiratory Inductive Plethysmography
(3) 20 preterm infants on continuous positive airway pressure (CPAP) with cardiorespiratory eventsGroup 3 will consist of 10 preterm infants on CPAP with established cardiorespiratory events, in whom respiratory acoustic signals from the acoustic sensor will be continuously measured for 3 hours. In addition, measurements of respiratory efforts will be obtained using the Respiratory Inductance Plethysmography (RIP), an inertial measurement unit (IMU) integrated within the acoustic sensor, and the Transthoracic Impedance (TTI). Data will be continuously recorded for 3 hours.
Receives: Device: Respiratory Acoustic Sensors; Device: Respiratory Inductive Plethysmography

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Reliability of respiratory acoustics at detecting airflow compared to airflow measurements obtained from a pneumotachometer.

    The airflow signal derived from the respiratory acoustic sensor will be compared with the airflow signal derived from the pneumotachometer.

    Measured: 10 minutes (group 1) or 3 hours (groups 2 and 3)

Secondary outcomes (2)
  1. Reliability of the inertial measurement unit (IMU) at detecting respiratory efforts compared to Respiratory Inductance Plethysmography (RIP).

    The chest wall movement signal derived from the respiratory acoustic sensor will be compared with the chest wall movement signal derived from RIP.

    Measured: 3 hours (groups 2 and 3 only)

  2. Reliability of the inertial measurement unit (IMU) at detecting respiratory efforts compared to Transthoracic Impedance (TTI).

    The chest wall movement signal derived from the respiratory acoustic sensor will be compared with the chest wall movement signal derived from TTI.

    Measured: 3 hours (groups 2 and 3 only)

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted19 Jan 2022
  2. Study start5 Dec 2022
  3. Primary completion28 Aug 2026 (estimated)
  4. Study completion31 Dec 2026 (estimated)

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Observational
Observational model
Cohort
Time perspective
Prospective
Enrolment
50 participants (estimated)
IDs, dates and publications
ClinicalTrials.gov ID
NCT05196646
Sponsor's protocol ID
2022-7444
Acronym
DREAM
Status last verified
Mar 2026
First submitted
12 Dec 2021
Last updated
30 Mar 2026
Results posted
No
Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
No
Individual data sharing
No

Questions about this trial

What is this clinical trial studying?

An observational study of Apnea of Prematurity, run by McGill University Health Centre/Research Institute of the McGill University Health Centre at 1 site in Canada.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 9 inclusion and 12 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: under way, not looking for participants, as of the sponsor's last update on 30 Mar 2026. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in Canada. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

This study is still under way. Results are not available in the source record yet.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Reliability of respiratory acoustics at detecting airflow compared to airflow measurements obtained from a pneumotachometer. (time frame: 10 minutes (group 1) or 3 hours (groups 2 and 3)). It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 30 Mar 2026.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in