Validation of Point Partial User Needs With Partial Finger Amputees

A study testing Point Partial in people with Amputation; Traumatic, Hand, run by Point Designs at 2 sites in United States.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

The objective of the device feasibility study will be to validate the user needs of the Point Partial system. This study will be a single group intervention model where one group of 5 partial finger amputees will be asked to perform several tasks. Successful completion of a task results in a fulfilled user need. Read more about this study

Who is it for?
Ages 18 years and older
Conditions the study is for
Amputation; Traumatic, Hand
What is being tested?
Point Partial
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
5 participants
Where
2 sitesUnited States
Study dates
3 Nov 2020to 31 Dec 2022
Phase
Interventional
Are there results?
Yes, posted in the registrySummary in the doctors' view

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 years and older
  • All sexes
  • No healthy volunteers
  • Adult, Older adult

Who may be able to join

  • Partial finger amputees with partial absence of index and/or middle fingers and presence of thumb. Additional finger loss is acceptable if all other criterion are met.
  • Fluent in English
  • Individuals aged 18 or greater
  • Patients may present with either fused or mobile MCP joints.

Who cannot join

  • Significant cognitive deficits as determined upon clinical evaluation
  • Significant neurological deficits as determined upon clinical evaluation
  • Significant physical deficits of the residual limb impacting full participation in the study as determined upon clinical evaluation
  • Uncontrolled pain or phantom pain impacting full participation in the study as determined upon OT evaluation
  • Serious uncontrolled medical problems as judged by the project therapist.

Trial locations and contacts

2 sites in 1 country.

United States2 sites
  • University of Colorado Denver | Anschutz Medical Campus

    Aurora, Colorado, 80045

  • Point Designs

    Lafayette, Colorado, 80026

Other trials in United States

About this study

The objective of the device feasibility study will be to validate the user needs of the Point Partial system. This study will be a single group intervention model where one group of 5 partial finger amputees will be asked to perform several tasks. Successful completion of a task results in a fulfilled user need. Failure to complete a task results in an unfulfilled user need.

Being tested
Point Partial

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Point PartialDevice

    The Point Partial partial finger prosthetic system consists of 1-4 ratcheting mechanical partial digits and mounting brackets

Study groups in NCT05012657
Study groupRoleWhat they receive
ProsthesisExperimentalPatient is temporarily fit with Point Partial partial finger prosthetic system
Receives: Device: Point Partial

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. VAL-1: System Comfort Validation Test

    User wears Point Partial(s) for a full day (Pass/Fail). This validation method was used to confirm the following user need of the Point Partial has been met: Point Partial shall be comfortable for a full day's use. A simple pass/fail result was recorded for each user need for each participant. An overall passing rate above 80% across all user needs and participants is considered a success. Each participant attended a single-day experimental session. A day-use prosthetic socket was fabricated and Point Partials were mounted to the socket. Then, each participant attempted the validation test.

    Measured: Day 1 (8 hours)

  2. VAL-2: Unilateral Function Validation Test

    User flexes and extends the Point Partial(s) without using contralateral hand (Pass/Fail). This validation method was used to confirm the following user need of the Point Partial has been met: Point Partial shall allow the user to flex and extend finger without their second hand. A simple pass/fail result was recorded for each user need for each participant. An overall passing rate above 80% across all user needs and participants is considered a success. Each participant attended a single-day experimental session. A day-use prosthetic socket was fabricated and Point Partials were mounted to the socket. Then, each participant attempted the validation test.

    Measured: Day 1 (10 minutes)

  3. VAL-3.1: System Robustness Validation Test With 25 lb Bag

    User lifts a 25 lb. bag with Point Partial(s) (Pass/Fail). This validation method was used to confirm the following user need of the Point Partial has been met: Point Partial shall be robust enough to operate in challenging environments. A simple pass/fail result was recorded for each user need for each participant. An overall passing rate above 80% across all user needs and participants is considered a success. Each participant attended a single-day experimental session. A day-use prosthetic socket was fabricated and Point Partials were mounted to the socket. Then, each participant attempted the validation test.

    Measured: Day 1 (10 minutes)

  4. VAL-3.2: Robustness Validation Test With Hammer

    User grips a hammer with Point Partial(s) and drives a nail into a piece of wood (Pass/Fail). This validation method was used to confirm the following user need of the Point Partial has been met: Point Partial shall be robust enough to operate in challenging environments. A simple pass/fail result was recorded for each user need for each participant. An overall passing rate above 80% across all user needs and participants is considered a success. Each participant attended a single-day experimental session. A day-use prosthetic socket was fabricated and Point Partials were mounted to the socket. Then, each participant attempted the validation test.

    Measured: Day 1 (10 minutes)

  5. VAL-4: Object Release Validation Test

    User releases Point Partial(s) while performing an active grasp by pressing the dorsal button (Pass/Fail). This validation method was used to confirm the following user need of the Point Partial has been met: Point Partial shall be releasable while performing active grasps. A simple pass/fail result was recorded for each user need for each participant. An overall passing rate above 80% across all user needs and participants is considered a success. Each participant attended a single-day experimental session. A day-use prosthetic socket was fabricated and Point Partials were mounted to the socket. Then, each participant attempted the validation test.

    Measured: Day 1 (10 minutes)

  6. VAL-5: Position Function Validation Test

    User positions Point Partial(s) into each of the locking positions (Pass/Fail). This validation method was used to confirm the following user need of the Point Partial has been met: Point Partial shall have a smooth ratcheting mechanism. A simple pass/fail result was recorded for each user need for each participant. An overall passing rate above 80% across all user needs and participants is considered a success. Each participant attended a single-day experimental session. A day-use prosthetic socket was fabricated and Point Partials were mounted to the socket. Then, each participant attempted the validation test.

    Measured: Day 1 (10 minutes)

  7. VAL-6: Extension Function Validation Test

    User activates spring-back mechanism by using (1) contralateral hand, (2) tabletop, and (3) side or thigh throughout day without failed extension (Pass/Fail). This validation method was used to confirm the following user need of the Point Partial has been met: Point Partial shall have a reliable auto spring-back mechanism. A simple pass/fail result was recorded for each user need for each participant. An overall passing rate above 80% across all user needs and participants is considered a success. Each participant attended a single-day experimental session. A day-use prosthetic socket was fabricated and Point Partials were mounted to the socket. Then, each participant attempted the validation test.

    Measured: Day 1 (10 minutes)

  8. VAL-7: Grip Functions Validation Test

    User grips a 5 lb. cylindrical smooth object using Point Partial(s) for 10 seconds (Pass/Fail). This validation method was used to confirm the following user need of the Point Partial has been met: Point Partial shall have high grip surface friction. A simple pass/fail result was recorded for each user need for each participant. An overall passing rate above 80% across all user needs and participants is considered a success. Each participant attended a single-day experimental session. A day-use prosthetic socket was fabricated and Point Partials were mounted to the socket. Then, each participant attempted the validation test.

    Measured: Day 1 (10 minutes)

  9. VAL-8: Fine Motor Function Validation Test

    User picks up 1 coin from a smooth tabletop (Pass/Fail). This validation method was used to confirm the following user need of the Point Partial has been met: Point Partial shall have a pronounced fingernail. A simple pass/fail result was recorded for each user need for each participant. An overall passing rate above 80% across all user needs and participants is considered a success. Each participant attended a single-day experimental session. A day-use prosthetic socket was fabricated and Point Partials were mounted to the socket. Then, each participant attempted the validation test.

    Measured: Day 1 (10 minutes)

  10. VAL-9: Overall Function & Robustness Validation Test

    User performs VAL-1 through VAL-8 (Pass/Fail). This validation method was used to confirm the following user need of the Point Partial has been met: Point Partial shall be robustly attached to mounting system. A simple pass/fail result was recorded for each user need for each participant. An overall passing rate above 80% across all user needs and participants is considered a success. Each participant attended a single-day experimental session. A day-use prosthetic socket was fabricated and Point Partials were mounted to the socket. Then, each participant attempted the validation test.

    Measured: Day 1 (8 hours)

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted19 Aug 2021
  2. Study start3 Nov 2020
  3. Primary completion31 Dec 2022
  4. Study completion31 Dec 2022

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Device feasibility
Allocation
Not applicable
Intervention model
Single group
Masking
None (open label)
Enrolment
5 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT05012657
Sponsor's protocol ID
299089
Status last verified
Aug 2024
First submitted
12 Aug 2021
Last updated
8 Nov 2024
FDA-regulated drug
No
FDA-regulated device
Yes
Individual data sharing
No

Questions about this trial

What is this clinical trial studying?

A study testing Point Partial in people with Amputation; Traumatic, Hand, run by Point Designs at 2 sites in United States.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 4 inclusion and 5 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 8 Nov 2024. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 2 sites in United States. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

Yes. Results are posted in the registry (first posted 8 Nov 2024). A summary is in the For doctors view, and the full results are on ClinicalTrials.gov.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

VAL-1: System Comfort Validation Test (time frame: Day 1 (8 hours)). The study lists 10 primary outcomes in all. It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 8 Nov 2024.

Free account

Keep searching with a free account

You’ve used your 3 free searches for this session. Log in or create a free AzTrials account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in