InfusionNCT04851782

Effects of ivAED™ Device on "Air-in-line" Alarms and Workflow Disruption

A study testing ivEAD tubing kit in people with Infusion, run by Herrick Medical LLC at 1 site in United States.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

The investigators propose to test the hypothesis, that is, the effectiveness of the ivAED to eliminate air-in-line, comparing two commercially available, standard IV infusion pumps: the Braun Infusomat® Space P and the Becton-Dickenson Alaris™ model 8100 pump (the latter is currently in use as the standard IV infusion… Read more about this study

Who is it for?
Ages 18 years and older
Conditions the study is for
Infusion
What is being tested?
ivEAD tubing kit
Could I receive a placebo?
No, everyone receives an active treatment
How many people
55 participants
Where
1 siteUnited States
Study dates
29 Mar 2021to 1 Jun 2024
Phase
Interventional
Are there results?
None posted in the registry

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 years and older
  • All sexes
  • No healthy volunteers
  • Adult, Older adult

Who may be able to join

  • Surgical ICU patients that are receiving IV infusions
  • Patients or their surrogates have provided written, informed consent

Who cannot join

  • Patients who do not give consent to our researchers
  • Children (patients <18 years of age)
  • Pregnancy
  • Prisoners

Trial locations and contacts

1 site in 1 country.

United States1 site
  • USC Keck Hospital

    Los Angeles, California, 90033

Other trials in United States

About this study

Official title: Optimizing Value in Critical Care: Effect of Braun Infusomat®Space I IV Infusion Pump and ivEAD™ Air Elimination Device on "Air-in-line" Alarms and Workflow Disruption

The investigators propose to test the hypothesis, that is, the effectiveness of the ivAED to eliminate air-in-line, comparing two commercially available, standard IV infusion pumps: the Braun Infusomat® Space P and the Becton-Dickenson Alaris™ model 8100 pump (the latter is currently in use as the standard IV infusion pump at Keck Hospital of USC). Testing of an IV infusion pump other the Keck Hospital of USC standard is indicated, as the independent laboratory testing of the Alaris™ 8100 pump found that this model pump is prone to developing air bubbles as the infusion passes through the pump mechanism. These air bubbles can be sufficient to trigger the air-in-line alarm. The Braun Infusomat® Space P IV infusion pump is a market leader like the Alaris™ 8100 pump, was evaluated in independent testing with and without the ivAED™ device, and was NOT found to create bubbles.

Read the detailed description

There are very few controlled human studies in the literature that assess various existing preventative measures for air embolism. After extensive, independent pre-clinical laboratory testing of the ivAED™ device demonstrating efficacy in removing air-in-line from IV infusions (see Appendix 1), this is the first-in-humans, clinical application of the ivAED™ device.

Testing of the ivAED™ device in patients is justified as that is the only means to determine the impact of the device on clinical IV infusion and nursing workflow disruptions (the primary outcomes of the study). Testing of the device in animals, or in healthy volunteers, or in computer simulations cannot provide this real-world information. In addition, the FDA typically requires that laboratory pre-clinical testing be coupled with clinical testing when applying for approval of new, investigational devices.

The investigators propose to test the hypothesis, that is, the effectiveness of the ivAED to eliminate air-in-line, comparing two commercially available, standard IV infusion pumps: the Braun Infusomat® Space P and the Becton-Dickenson Alaris™ model 8100 pump (the latter is currently in use as the standard IV infusion pump at Keck Hospital of USC). Testing of an IV infusion pump other the Keck Hospital of USC standard is indicated, as the independent laboratory testing of the Alaris™ 8100 pump found that this model pump is prone to developing air bubbles as the infusion passes through the pump mechanism. These air bubbles can be sufficient to trigger the air-in-line alarm. The Braun Infusomat® Space P IV infusion pump is a market leader like the Alaris™ 8100 pump, was evaluated in independent testing with and without the ivAED™ device, and was NOT found to create bubbles.

Conditions
Infusion
Being tested
ivEAD tubing kit
Keywords
Infusion, Alarms

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • ivEAD tubing kitDevice

    Also called Braun Infusomat Pump, Standard Infusion Pump

    ivEAD tubing designed to reduce infusion pump alarms do to air in infusion line

Study groups in NCT04851782
Study groupRoleWhat they receive
Braun Infusomat Pump and ivEAD tubingExperimentalivEAD tubing
Receives: Device: ivEAD tubing kit
Braun Infusomat Pump and standard tubingNo interventionStandard tubing
Standard pump and ivEAD tubingExperimentalivEAD tubing
Receives: Device: ivEAD tubing kit
Standard infusion pump and standard tubingNo interventionStandard tubing

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Number of air-in-line IV infusion alarms (as detected by the IV infusion pump)

    Number of air-in-line IV infusion alarms (as detected by the IV infusion pump)

    Measured: Patients will participate in the study from the time of enrollment in the ICU for up to 3 days

  2. Disruption of nursing workflow, measured as number of minutes required to attend to alarm and restore infusion flow.

    Disruption of nursing workflow, measured as number of minutes required to attend to alarm and restore infusion flow.

    Measured: Patients will participate in the study from the time of enrollment in the ICU for up to 3 days

Secondary outcomes (2)
  1. Cost

    Cost of disruption of nursing workflow, as measured by time-driven, activity-based costing

    Measured: Patients will participate in the study from the time of enrollment in the ICU for up to 3 days

  2. Loss of Infusate

    Loss of infusate secondary to clearance of air in line

    Measured: Patients will participate in the study from the time of enrollment in the ICU for up to 3 days

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted20 Apr 2021
  2. Study start29 Mar 2021
  3. Primary completion1 Apr 2024
  4. Study completion1 Jun 2024

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Prevention
Allocation
Randomised
Intervention model
Parallel groups
Model description
Participants either have Braun Infusomat Pump with ivEAD tubing or standard pump OR another pump with or without ivEAD tubing or standard tubing
Masking
None (open label)
Enrolment
55 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT04851782
Sponsor's protocol ID
HM 001
Status last verified
Jul 2024
First submitted
16 Apr 2021
Last updated
11 Jul 2024
Results posted
No
Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
Yes
Individual data sharing
No

Publications

  • Kaplan RS, Anderson SR. Time-driven activity-based costing. Harv Bus Rev. 2004 Nov;82(11):131-8, 150.

    PubMed 15559451
  • Kaplan RS, Witkowski M, Abbott M, Guzman AB, Higgins LD, Meara JG, Padden E, Shah AS, Waters P, Weidemeier M, Wertheimer S, Feeley TW. Using time-driven activity-based costing to identify value improvement opportunities in healthcare. J Healthc Manag. 2014 Nov-Dec;59(6):399-412.

    PubMed 25647962
  • Infusion pump market is anticipated to exceed US$ 49 billion by 2025. https://www.marketwatch.com/press-release/infusion-pump-market-is-anticipated-to-exceed-us-49-billion-by-2025-2019-03-28. Updated 2019. Accessed August 23, 2019.

  • Intravascular air-in-line and air embolism risks associated with infusion pumps, fluid warmers, and rapid infusers: FDA safety communication. U.S. Food and Drug Administration. 2019

  • Mirski MA, Lele AV, Fitzsimmons L, Toung TJ. Diagnosis and treatment of vascular air embolism. Anesthesiology. 2007 Jan;106(1):164-77. doi: 10.1097/00000542-200701000-00026.

    PubMed 17197859
  • Brull SJ, Prielipp RC. Vascular air embolism: A silent hazard to patient safety. J Crit Care. 2017 Dec;42:255-263. doi: 10.1016/j.jcrc.2017.08.010. Epub 2017 Aug 7.

    PubMed 28802790
  • Kizer JR, Devereux RB. Clinical practice. Patent foramen ovale in young adults with unexplained stroke. N Engl J Med. 2005 Dec 1;353(22):2361-72. doi: 10.1056/NEJMcp043981. No abstract available.

    PubMed 16319385
  • Agency for Healthcare Research and Quality. Patient safety primer: Never events. https://psnet.ahrq.gov/primers/primer/3/Never-Events. Accessed August 23, 2019.

  • Lee PT, Thompson F, Thimbleby H. Analysis of infusion pump error logs and their significance for health care. Br J Nurs. 2012 Apr 26-May 9;21(8):S12, S14, S16-20. doi: 10.12968/bjon.2012.21.Sup8.S12.

    PubMed 22629592
  • Matocha D. Reducing infusion pump alarms through structured interventions. Journal of the Association for Vascular Access. 2018;23(2):87-95.

Questions about this trial

What is this clinical trial studying?

A study testing ivEAD tubing kit in people with Infusion, run by Herrick Medical LLC at 1 site in United States.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 2 inclusion and 4 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 11 Jul 2024. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in United States. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Number of air-in-line IV infusion alarms (as detected by the IV infusion pump) (time frame: Patients will participate in the study from the time of enrollment in the ICU for up to 3 days). The study lists 2 primary outcomes in all. It is what the study was designed to measure, not a finding.

Where can I read the original study information?

On the ClinicalTrials.gov record for NCT04851782, which also lists 10 linked publications.

Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 11 Jul 2024.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in