Sudden Cardiac ArrestNCT04840797

HeartStart HS1 Defibrillator* Event Registry

An observational study of Sudden Cardiac Arrest, run by Heartstream US LLC at 1 site in United States.

Recruiting

Looking for participants now. Contact the study team or a site below to ask about taking part.

Sites and contacts
Official record
Search all trials

Trial summary

A Prospective study on the performance and safety of the HeartStart OnSite (Model M5066A) \& HeartStart Home (Model M5068A) Defibrillator device data. Read more about this study

Who is it for?
Any age
Conditions the study is for
Sudden Cardiac Arrest
What is being tested?
HeartStart HS1
Type of treatment
Medical device
How many people
1,400 participants
Where
1 siteUnited States
Study dates
25 Feb 2021to 31 Dec 2030
How long will I take part?
1 Day
Phase
Observational
Are there results?
Not yet: the study is under way

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages Any age
  • All sexes
  • No healthy volunteers
  • Child, Adult, Older adult

Study population: Anyone who is suspected of having a sudden cardiac arrest.

Who may be able to join

  • Have been suspected of a circulatory arrest for any cause.
  • Have had HeartStart HS1 Defibrillator with electrodes:
  • Adult SMART Pads Cartridge [REF: M5071A], Infant/Child SMART Pads Cartridge (REF:M5072A) applied to their body and powered up with HS1 battery (Model M5070A), regardless of whether a defibrillation shock was delivered

Who cannot join

  • Subjects will be excluded if any of the following are present:
  • AED or pad use other than the HeartStart HS1 Defibrillator with above referenced Philips electrode pads. Note: Use of a device other than the Philips HeartStart HS1 AED does not preclude enrollment if the AED or an Advanced Life Support defibrillator was used after the HeartStart HS1 AED.
  • AED used for training purposes.

Trial locations and contacts

1 site in 1 country. Contact the study team or a site near you to ask about taking part.

United States1 site, 1 recruiting
  • Philips

    Pittsburgh, Pennsylvania, 15206

    Chase Wennick, 617-398-7347, chase.wennick@philips.com

    Gerrit Nordergraaf, MD PhD

    Recruiting

Other trials in United States

About this study

A Prospective study on the performance and safety of the HeartStart OnSite (Model M5066A) \& HeartStart Home (Model M5068A) Defibrillator device data.

Read the detailed description

The objective of this PMCF (Post Market Clinical Follow-up) prospective collection of post-event data registry is to confirm the safety and performance as expressed in the energy delivery by the HeartStart HS1 Defibrillator (HeartStart OnSite and Home models in North America) in the population for the device's intended use in reference to its commercial labeling, and in relationship to the type of pads attached to the device.

Being tested
HeartStart HS1
Keywords
AED, Post -Market, Cardiology
Medical subject headings
Death, Sudden, Cardiac

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • HeartStart HS1Device

    Automated External Defibrillator

Study groups in NCT04840797
Study groupRoleWhat they receive
Suspected Sudden Cardiac ArrestAll subjects with suspected of a circulatory arrest of any cause.
Receives: Device: HeartStart HS1

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Mean shock energy delivered in the pediatric population

    Mean of shock energy delivered, when the pediatric defibrillation pads were attached to the AED as intended for infants and children under 25 kilograms / 55 pounds or 0-8 years old.

    Measured: Up to 15 minutes during device placement

  2. Mean shock energy delivered in the adult population

    Mean of shock energy delivered, when the adult defibrillation pads were attached to the AED as intended for in children and adults over 25 kilograms / 55 pounds or greater than 8 years old.

    Measured: Up to 15 minutes during device placement

  3. The Patient Analysis System (PAS) shock advisory algorithm performance as determined by evaluation of the ECG as aligned with the device advisement.

    Performance is defined by the sensitivity and specificity; positive predictive value and false positive rate will be reported for each rhythm category (as defined by device labeling).

    Measured: Up to 15 minutes during device placement

  4. Accurate detection of shockable rhythms and facilitation of shocks.

    The percentage of events where a shockable rhythm was identified by the device, the device advised a shock, and the device delivered a shock when the shock button was pressed.

    Measured: Up to 15 minutes during device placement

  5. Defibrillation success

    Defibrillation success and the number of shocks required for success.

    Measured: Up to 15 minutes during device placement

Secondary outcomes (12)
  1. Frequency and severity of unexpected adverse events

    Frequency and severity of unexpected adverse events

    Measured: Up to 15 minutes during device placement

  2. Unanticipated adverse device effects (UADE)

    Unanticipated adverse device effects (UADE)

    Measured: Up to 15 minutes during device placement

  3. Median operating time intervals

    Median operating time intervals

    Measured: Up to 15 minutes during device placement

  4. Routine self-test performance/Battery Insertion self-test performance

    Routine self-test performance/Battery Insertion self-test performance

    Measured: Up to 15 minutes during device placement

  5. Device arming after shock advisement.

    The percentage of events where the device advised a shock and the device armed.

    Measured: Up to 15 minutes during device placement

  6. Energy delivered consistent with device labeling

    The percentage of defibrillations in which the energy delivered was consistent with device labeling.

    Measured: Up to 15 minutes during device placement

  7. Pad replacement notification

    Number of events in which pads were replaced at any time during the event after the user is prompted to do so by the device

    Measured: Up to 15 minutes during device placement

  8. Shock button effectively pressed after shock advisement

    The percentage of events where a shock was recommended by the device and the shock button was pressed.

    Measured: Up to 15 minutes during device placement

  9. Indicators of battery status/depletion (device logged data)

    Percentage of events that include indicators of battery status/depletion (device logged data)

    Measured: Up to 15 minutes during device placement

  10. Aborted shocks because of low or high impedance

    Percentage of aborted shocks because of low or high impedance

    Measured: Up to 15 minutes during device placement

  11. CPR guidance initiated (i-button pushed)

    Percentage of events where CPR guidance was initiated (i-button pushed)

    Measured: Up to 15 minutes during device placement

  12. CPR activity

    Percentage of events where CPR is performed during CPR guidance

    Measured: Up to 15 minutes during device placement

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted12 Apr 2021
  2. Study start25 Feb 2021
  3. Primary completion31 Dec 2030 (estimated)
  4. Study completion31 Dec 2030 (estimated)

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Observational
Observational model
Case only
Time perspective
Prospective
Enrolment
1,400 participants (estimated)
Target duration
1 Day
IDs, dates and publications
ClinicalTrials.gov ID
NCT04840797
Sponsor's protocol ID
TC_ECR_HS1_2020_10819
Status last verified
Aug 2026
First submitted
7 Apr 2021
Last updated
11 Aug 2026
Results posted
No
Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
Yes
Individual data sharing
No

Questions about this trial

What is this clinical trial studying?

An observational study of Sudden Cardiac Arrest, run by Heartstream US LLC at 1 site in United States.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 3 inclusion and 3 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: looking for participants now, as of the sponsor's last update on 11 Aug 2026. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in United States. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

This study is still under way. Results are not available in the source record yet.

How can a doctor refer a patient or contact the research team?

Contact Chase Wennick (617-398-7347, chase.wennick@philips.com), or a site near the patient. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Mean shock energy delivered in the pediatric population (time frame: Up to 15 minutes during device placement). The study lists 5 primary outcomes in all. It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 11 Aug 2026.

Free account

Keep searching with a free account

You’ve used your 3 free searches for this session. Log in or create a free AzTrials account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in
HeartStart HS1 Defibrillator* Event Registry | AzTrials