AtaCor Subcostal Temporary Extravascular Pacing III Study

A study testing AtaCor StealthTrac Lead in people with Conduction Defect, run by AtaCor Medical, Inc. at 4 sites in 3 countries.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

Third in-human study for the AtaCor Extravascular (EV) Temporary Pacing Lead System to collect initial safety and performance data for the latest AtaCor System. The objective of the study is to generate safety and performance data of the latest AtaCor EV Temporary Pacing Lead System 1) to support the development a… Read more about this study

Who is it for?
Ages 18 years and older
Conditions the study is for
Conduction Defect
What is being tested?
AtaCor StealthTrac Lead
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
33 participants
Where
4 sitesin 3 countries
Study dates
3 Oct 2020to 11 May 2022
Phase
Interventional
Are there results?
None posted in the registry

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 years and older
  • All sexes
  • No healthy volunteers
  • Adult, Older adult

Who may be able to join

  • At least 18 years old
  • Indicated for closed-chest cardiac invasive procedure (e.g. transcatheter valve replacement, balloon valvuloplasty, permanent pacemaker implantation and pacing lead extractions/revisions)

Who cannot join

  • NYHA IV functional class
  • Oxygen dependency
  • BMI ≥ 35 kg/m2
  • Circumstances that prevent data collection or follow-up (e.g., inability to perform a short walk with the Holter monitor)
  • Participation in any concurrent clinical study without prior written approval from the Sponsor
  • Inability to give an informed consent to participate in the Study
Show all 19Show fewer
  • Known prior history for any of the following:
  • Median or partial sternotomy
  • Surgery with disruption of the lung, pericardium or connective tissue between the sternum and pericardium
  • Significant anatomic derangement of or within the thorax (e.g., pectus excavatum, significant scoliosis)
  • Thoracic radiation therapy, pneumothorax, pneumomediastinum or other medical treatments/conditions which may complicate the AtaCor EV Temporary Pacing Lead System insertion procedure
  • Pericardial disease, pericarditis and mediastinitis
  • Medical treatments, surgeries or conditions that increase the potential for pericardial adhesions
  • Severe to very severe airflow limitation, defined as FEV1/FVC <0.7 AND FEV1 < 50% predicted
  • Symptomatic COPD exacerbation associated with either:
  • Modified MRC Dyspnea Scale Grade ≥2, OR
  • CAT Assessment
  • Surgically corrected congenital heart disease (not including catheter-based procedures)
  • Allergies to the device materials such as stainless steel, titanium, platinum, iridium, polyethylene, polyurethane, polycarbonate and silicone

Trial locations and contacts

4 sites in 3 countries.

4 sites in 3 countries

New Zealand2 sites
  • Christchurch Hospital

    Christchurch

  • Wellington Regional Hospital

    Wellington

Other trials in New Zealand

Czechia1 site
  • Na Homolce Hospital

    Prague

Other trials in Czechia

Paraguay1 site
  • Sanatorio Italiano

    Asunción

Other trials in Paraguay

About this study

Third in-human study for the AtaCor Extravascular (EV) Temporary Pacing Lead System to collect initial safety and performance data for the latest AtaCor System.

The objective of the study is to generate safety and performance data of the latest AtaCor EV Temporary Pacing Lead System 1) to support the development a future pivotal study with an indication limited to a maximum of 7 days, and 2) to obtain early clinical data for future research related to longer-term use.

Read the detailed description

Prior revisions of the study intended to evaluate multiple StealthTrac Leads (Models AC-1010, AC-1020, AC-2021 and AC-1030) to support the development of a future pivotal study. In the most recent revision of the study, an additional StealthTrac Lead (Model AC-1012) will be evaluated to support use in a larger pivotal study.

This study is a feasibility study serving dual purposes: 1) to generate developmental clinical data in support of a subsequent pivotal clinical study and marketing application for temporary pacing (in patients that retain the lead for a maximum of 7 days); and 2) to obtain early data for future research related to longer-term use (in patients that retain the lead for a maximum of 14 days).

Subjects undergoing a qualifying index procedure will have a Model AC-1012 StealthTrac Lead inserted, evaluated for a Maximum Lead Insertion Time prior to removal and a final post-removal follow-up 27-33 days after lead insertion. The following elements will change after the absence of AtaCor-related cardiac tamponade and no more than one (1) serious AtaCor-related pericardial effusion has been confirmed from the initial eight (8) experiences with the Model AC-1012 StealthTrac Lead, including insertion and removal procedures performed by multiple operators:

  • Maximum Lead Insertion Time will change from 7 to 14 Days
  • Follow-up will change from in-hospital follow-up only to allowing hospital discharge with the StealthTrac lead inserted, but capped and inactive after Day 2 provided in-person wound checks occur at least every 3 days
  • Echocardiography requirements will transition from 3 distinct days to procedure day and removal day only.
Being tested
AtaCor StealthTrac Lead
Keywords
Temporary Ventricular Pacing
Medical subject headings
Arrhythmias, Cardiac

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • AtaCor StealthTrac LeadDevice

    Subjects are assigned non-randomly to receive one of the four AtaCor StealthTrac leads being evaluated in the study.

Study groups in NCT04538287
Study groupRoleWhat they receive
AtaCor StealthTrac Lead Model AC-1010Active comparatorSubjects inserted with the AtaCor StealthTrac Lead Model AC-1010. This was the first model evaluated in the study.
Receives: Device: AtaCor StealthTrac Lead
AtaCor StealthTrac Lead Model AC-1020Active comparatorSubjects inserted with the AtaCor StealthTrac Lead Model AC-1020. This was the second model evaluated in the study.
Receives: Device: AtaCor StealthTrac Lead
AtaCor StealthTrac Lead Model AC-1021Active comparatorSubjects inserted with the AtaCor StealthTrac Lead Model AC-1021. This was the third model evaluated in the study.
Receives: Device: AtaCor StealthTrac Lead
AtaCor StealthTrac Lead Model AC-1012Active comparatorSubjects inserted with the AtaCor StealthTrac Lead Model AC-1012. This is the fourth model currently being evaluated in the study.
Receives: Device: AtaCor StealthTrac Lead

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Safety - Freedom from Adverse Device Effects

    Freedom from Adverse Device Effects (ADE)

    Measured: Lead removal (up to 14 Days)

  2. Performance - Pacing Capture Threshold

    Summary statistics for pacing capture threshold for each StealthTrac Lead model from insertion through removal

    Measured: Lead removal (up to 14 Days)

  3. Performance - Pacing Impedance

    Summary statistics for pacing impedance for each StealthTrac Lead model from insertion through removal for each StealthTrac Lead model from insertion through removal

    Measured: Lead removal (up to 14 Days)

  4. Performance - Sensed R-wave Amplitude

    Summary statistics for sensed R-wave amplitude for each StealthTrac Lead model from insertion through removal

    Measured: Lead removal (up to 14 Days)

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted4 Sep 2020
  2. Study start3 Oct 2020
  3. Primary completion25 Jan 2022
  4. Study completion11 May 2022

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Basic science
Allocation
Non-randomised
Intervention model
Parallel groups
Masking
None (open label)
Enrolment
33 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT04538287
Sponsor's protocol ID
DOC-10092
Acronym
STEP III
Status last verified
Jul 2022
First submitted
25 Aug 2020
Last updated
7 Jul 2022
Results posted
No
Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
No
Individual data sharing
No

Questions about this trial

What is this clinical trial studying?

A study testing AtaCor StealthTrac Lead in people with Conduction Defect, run by AtaCor Medical, Inc. at 4 sites in 3 countries.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 2 inclusion and 19 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 7 Jul 2022. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 4 sites in New Zealand, Czechia, Paraguay. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Safety - Freedom from Adverse Device Effects (time frame: Lead removal (up to 14 Days)). The study lists 4 primary outcomes in all. It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 7 Jul 2022; AzTrials last refreshed its copy on 30 Oct 2026.

Free account

Keep searching with a free account

You’ve used your 3 free searches for this session. Log in or create a free AzTrials account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in