Type 1 Diabetes MellitusNCT04378114

Impact of Acetaminophen on Performance of Guardian™ Sensor (3) in Adults

A study testing Guardian™ Sensor (3) connected to a Guardian™ Connect Transmitter. C sensor algorithm applied retrospectively to raw sensor data., Guardian™ Sensor (3) connected to a Guardian™ Connect Transmitter. Zeus sensor algorithm applied retrospectively to raw sensor data. and Acetominophen in people with Type 1 Diabetes Mellitus and Type 2 Diabetes Mellitus, run by Medtronic MiniMed, Inc. at 6 sites in United States.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

The purpose of this study is to characterize the impact of acetaminophen ingestion on the performance of the Guardian™ Sensor (3) (i.e., C algorithm and Zeus algorithm) in subjects age 18 - 80 years. Read more about this study

Who is it for?
Ages 18 to 80 years
What is being tested?
Guardian™ Sensor (3) connected to a Guardian™ Connect Transmitter. C sensor algorithm applied retrospectively to raw sensor data., Guardian™ Sensor (3) connected to a Guardian™ Connect Transmitter. Zeus sensor algorithm applied retrospectively to raw sensor data. and Acetominophen
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
104 participants
Where
6 sitesUnited States
Study dates
18 May 2020to 17 Jul 2020
Phase
Interventional
Are there results?
Yes, posted in the registrySummary in the doctors' view

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 to 80 years
  • All sexes
  • No healthy volunteers
  • Adult, Older adult

Who may be able to join

  • Individual is 18 - 80 years of age at time of enrollment.
  • Subject has a clinical diagnosis of type 1 or type 2 diabetes for a minimum of 6 months duration as determined via medical record/ source documentation by an individual qualified to make a medical diagnosis.
  • Subject has adequate venous access as assessed by investigator or appropriate staff.

Who cannot join

  • Subject has history of allergy to acetaminophen or has been told by a health care provider they may not ingest acetaminophen
  • Subject reports history of liver cirrhosis or liver problems that a health care provider told them they should not use acetaminophen because of liver disorder.
  • Subject will not tolerate tape adhesive in the area of Guardian™ Sensor (3) placement as assessed by a qualified individual.
  • Subject has any unresolved adverse skin condition in the area of sensor or device placement (e.g., psoriasis, rash, Staphylococcus infection).
  • Subject is actively participating in an investigational study (e.g., drug or device) wherein he/she has received treatment from an investigational study (drug or device) in the last 2 weeks prior to Visit 1. (Please note participation in an observational study is acceptable.)
  • Subject is female of child-bearing potential and has a pregnancy screening test that is positive.
Show all 14Show fewer
  • Subject is female of child-bearing potential and who is sexually active should be excluded if she is not using a form of contraception deemed reliable by investigator.
  • Subject is female and plans to become pregnant during the course of the study.
  • Subject is breast feeding.
  • Subject has a chronic heavy alcohol use as determined by investigator.
  • Subject has a history of a seizure disorder.
  • Subject has a hematocrit (Hct) more than 10% below the lower limit of normal reference range (please note that patients may use prior blood draw from routine care as long as done within 6 months of screening and report of lab placed with subject source documents).
  • Subject has a history of adrenal insufficiency.
  • Subject is a member of the research staff involved with the study.

Trial locations and contacts

6 sites in 1 country.

6 sites in 1 country

United States6 sites
  • "Atlanta Diabetes Associates

    Atlanta, Georgia, 30318

  • Barbara Davis Center - Adults

    Aurora, Colorado, 80045

  • AMCR Institue

    Escondido, California, 92026

  • Rainier Clinical Research Center, Inc.

    Renton, Washington, 98057

  • Sansum Diabetes Research Institute

    Santa Barbara, California, 93105

  • Diablo Clinical Research

    Walnut Creek, California, 92025

Other trials in United States

About this study

The purpose of this study is to characterize the impact of acetaminophen ingestion on the performance of the Guardian™ Sensor (3) (i.e., C algorithm and Zeus algorithm) in subjects age 18 - 80 years.

Read the detailed description

The study is a multi-center, prospective, study without controls. The purpose of this study is to characterize the impact of acetaminophen ingestion on the performance of the Guardian™ Sensor (3) (i.e., C algorithm and Zeus algorithm) in subjects age 18 - 80 years.

Being tested
Guardian™ Sensor (3) connected to a Guardian™ Connect Transmitter. C sensor algorithm applied retrospectively to raw sensor data., Guardian™ Sensor (3) connected to a Guardian™ Connect Transmitter. Zeus sensor algorithm applied retrospectively to raw sensor data., Acetominophen
Medical subject headings
Diabetes Mellitus, Type 1, Diabetes Mellitus, Type 2

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Guardian™ Sensor (3) connected to a Guardian™ Connect Transmitter. C sensor algorithm applied retrospectively to raw sensor data.Device

    Characterize the impact of acetaminophen ingestion on the accuracy of Guardian™ Sensor (3) when C sensor algorithm is used.

  • Guardian™ Sensor (3) connected to a Guardian™ Connect Transmitter. Zeus sensor algorithm applied retrospectively to raw sensor data.Device

    Characterize the impact of acetaminophen ingestion on the accuracy of Guardian™ Sensor (3) when Zeus sensor algorithm is used.

  • AcetominophenDrug

    One time administration of acetaminophen.

Study groups in NCT04378114
Study groupRoleWhat they receive
Subjects with diabetes wearing Guardian™ Sensor (3). C algorithm applied retrospectivelyExperimentalSubjects wear Guardian™ Sensor (3) over 7 days and undergo one frequent sample test (FST). C sensor algorithm applied retrospectively to raw sensor data.
Receives: Device: Guardian™ Sensor (3) connected to a Guardian™ Connect Transmitter. C sensor algorithm applied retrospectively to raw sensor data.; Drug: Acetominophen
Subjects with diabetes wearing Guardian™ Sensor (3). Zeus algorithm applied retrospectivelyExperimentalSubjects wear Guardian™ Sensor (3) over 7 days and undergo one FST. Zeus sensor algorithm applied retrospectively to raw sensor data.
Receives: Device: Guardian™ Sensor (3) connected to a Guardian™ Connect Transmitter. Zeus sensor algorithm applied retrospectively to raw sensor data.; Drug: Acetominophen

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Bias (mg/dL) Between the Guardian™ Sensor (3) Values and Yellow Springs Instrument™ (YSI™) Plasma Glucose

    Bias (mg/dL) between the Guardian™ Sensor (3) values and YSI™ plasma glucose values 1 hour before through 5 hours after ingestion of acetaminophen; results are reported hourly.

    Bias = Mean of [Guardian™ Sensor (3) values - YSI™ plasma glucose values].

    Measured: Frequent sample testing recordings from the 6-hour period leading to and after Acetaminophen ingestion on day 4 or day 5 reported

Secondary outcomes (2)
  1. Mean Absolute Relative Difference (MARD, %) Between the Guardian™ Sensor (3) Values and YSI™ Plasma Glucose

    Mean absolute relative difference (MARD, %) between the Guardian™ Sensor (3) values and YSI™ plasma glucose 1 hour before through 5 hours after ingestion of acetaminophen; results are reported hourly.

    Mean Absolute Relative Difference = Mean of ([absolute difference of Guardian™ Sensor (3) values and YSI™ plasma glucose / YSI™ plasma glucose] * 100%).

    Measured: Frequent sample testing recordings from the 6-hour period leading to and after Acetaminophen ingestion on day 4 or day 5 reported

  2. Percentage of Guardian™ Sensor (3) Values That Were Within 20% of YSI™ Plasma Glucose Values

    Percentage of Guardian™ Sensor (3) values within 20% of YSI™ plasma glucose value (±20 mg/dL [1.1 mmol/L] when YSI™ value less than or equal to (≤) 80 mg/dL [4.4 mmol/L]) during the 1 hour before through 5 hours after ingestion of acetaminophen; results are reported hourly.

    20% agreement = Number of Guardian™ Sensor (3) values within 20% (±20 mg/dL) of YSI™ plasma glucose value / Total number of paired Guardian™ Sensor (3) and YSI™ plasma glucose points * 100%.

    Measured: Frequent sample testing recordings from the 6-hour period leading to and after Acetaminophen ingestion on day 4 or day 5 reported

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted7 May 2020
  2. Study start18 May 2020
  3. Primary completion17 Jul 2020
  4. Study completion17 Jul 2020

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Other
Allocation
Non-randomised
Intervention model
Sequential
Masking
None (open label)
Enrolment
104 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT04378114
Sponsor's protocol ID
CIP333
Status last verified
Aug 2023
First submitted
4 May 2020
Last updated
8 Sep 2023
Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
Yes

Questions about this trial

What is this clinical trial studying?

A study testing Guardian™ Sensor (3) connected to a Guardian™ Connect Transmitter. C sensor algorithm applied retrospectively to raw sensor data., Guardian™ Sensor (3) connected to a Guardian™ Connect Transmitter. Zeus sensor algorithm applied retrospectively to raw sensor data. and Acetominophen in people with Type 1 Diabetes Mellitus and Type 2 Diabetes Mellitus, run by Medtronic MiniMed, Inc. at 6 sites in United States.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 3 inclusion and 14 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 8 Sep 2023. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 6 sites in United States. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

Yes. Results are posted in the registry (first posted 8 Sep 2023). A summary is in the For doctors view, and the full results are on ClinicalTrials.gov.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Bias (mg/dL) Between the Guardian™ Sensor (3) Values and Yellow Springs Instrument™ (YSI™) Plasma Glucose (time frame: Frequent sample testing recordings from the 6-hour period leading to and after Acetaminophen ingestion on day 4 or day 5 reported). It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 8 Sep 2023.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in