A Comparative Clinical Investigation of Two Hearing Instrument Digital Signal Processing

A study testing Polaris miniRITE BTE and Viron miniRiTE BTE in people with Hearing Loss, run by Bernafon AG at 1 site in Switzerland.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

The purpose of this study is to show that the performance of the new digital signal processing system is better than the system used in the currently CE marked devices. Speech in narrowband noise and acceptable levels of noise should be improved with the new system, and speech understanding overall should not be… Read more about this study

Who is it for?
Ages 18 years and olderHealthy volunteers can take part
Conditions the study is for
Hearing Loss
What is being tested?
Polaris miniRITE BTE and Viron miniRiTE BTE
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
19 participants
Where
1 siteSwitzerland
Study dates
3 Aug 2020to 30 Nov 2020
Phase
Interventional
Are there results?
None posted in the registry

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 years and older
  • All sexes
  • Healthy volunteers accepted
  • Adult, Older adult

Who may be able to join

  • Hearing loss (sensorineural, conductive, mixed-must have doctor approval for mixed and conductive Hearing loss)
  • Severity of Hearing loss ranging from mild to profound
  • German speaking
  • Current Hearing aid users
  • Ability and willingness to sign the consent form

Who cannot join

  • Contraindications for amplification
  • New Hearing aid users
  • Inability to follow the procedures
  • Inability to attend Weekly study Appointments
  • Strongly reduced dexterity
  • Central Hearing disorders
  • Bernafon employees
  • Family members of Bernafon employees

Trial locations and contacts

1 site in 1 country.

Switzerland1 site
  • Bernafon AG

    Bern, 3018

Other trials in Switzerland

About this study

Official title: A Comparative Clinical Investigation of Two Hearing Instrument Digital Signal Processing Strategies and Their Effect on Speech Understanding

The purpose of this study is to show that the performance of the new digital signal processing system is better than the system used in the currently CE marked devices. Speech in narrowband noise and acceptable levels of noise should be improved with the new system, and speech understanding overall should not be negatively affected. There should be no artefacts or unwanted noises caused by the new system.

Read the detailed description

This is a controlled, double-blinded, comparative clinical evaluation conducted monocentric at the premised of Bernafon AG in Bern, Switzerland.

For this study, Bernafon AG will carry out testing with participants with hearing loss to validate the performance of a new digital signal processing (DSP). System testing of the DSP shows an improved signal, yet it must be tested with humans to be sure of an improvement in real-life situations. The current study will compare the new DSP to the one that is used in the Bernafon hearing aids that are CE- marked and currently available on the market. The goal is to determine whether the new DSP is better in real-life situations than the current system.

Primary Endpoint: The primary objective is to assess the new DSP system with various hearing losses and levels of amplification. Speech testing will be used to analyse the speech perception of participants.

Secondary Endpoint: The secondary objective is to verify the overall performance of the new device. Acceptable noise levels will be measured as well as further speech testing. Questionnaires will be used to measure subjective responses to the performance.

Other Endpoints: The subjective performance of an optimized fitting versus a standard fitting and the acceptance of new domes will be measured with questionnaires.

Safety Endpoint: The study aims to identify any new risk factors and to ensure the safety of the device with the newly implemented DSP system before releasing it to the market.

Conditions
Hearing Loss
Being tested
Polaris miniRITE BTE, Viron miniRiTE BTE
Medical subject headings
Hearing Loss

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Polaris miniRITE BTEDevice

    The Polaris device is programmed with computer software to match the Hearing loss of the participant.

  • Viron miniRiTE BTEDevice

    The Viron Hearing device is programmed with computer software to match the Hearing loss of the participant.

Study groups in NCT04374851
Study groupRoleWhat they receive
New DeviceExperimentalThe new Hearing aid is essentially the same as the current device (i.e., hardware, use) but with an improved digital signal processing (DSP).
Receives: Device: Polaris miniRITE BTE
Current DeviceActive comparatorThe current device is the Hearing aid that is currently sold on the market. It is used as a normal Hearing aid that is worn daily to amplify sounds for Hearing-impaired people.
Receives: Device: Viron miniRiTE BTE

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Speech perception with narrowband noise

    Speech perception is tested using a standardized speech test that requires subjects to repeat sentences presented among narrowband background noise. The test is adaptive with the background noise getting louder or softer so that the subject answers 50% correct. The scores are measured in dB SNR (signal-to-noise ratio) and a lower score is better. The lowest possible score is -15 dB SNR and the highest is +20 dB SNR. It is measured in the unaided condition.

    Measured: Week 1

  2. Speech perception with narrowband noise

    Speech perception is tested using a standardized speech test that requires subjects to repeat sentences presented among narrowband background noise. The test is adaptive with the background noise getting louder or softer so that the subject answers 50% correct. The scores are measured in dB SNR (signal-to-noise ratio) and a lower score is better. The lowest possible score is -15 dB SNR and the highest is +20 dB SNR. It is measured in the aided condition with the new device or the current device depending on the randomization.

    Measured: Week 2

  3. Speech perception with narrowband noise

    Speech perception is tested using a standardized speech test that requires subjects to repeat sentences presented among narrowband background noise. The test is adaptive with the background noise getting louder or softer so that the subject answers 50% correct. The scores are measured in dB SNR (signal-to-noise ratio) and a lower score is better. The lowest possible score is -15 dB SNR and the highest is +20 dB SNR. It is measured in the aided condition with the new device or the current device depending on the randomization.

    Measured: Week 3

Secondary outcomes (6)
  1. Speech perception with broadband noise

    Speech perception is tested using a standardized speech test that requires subjects to repeat sentences presented among broadband background noise. The test is adaptive with the background noise getting louder or softer so that the subject answers 50% correct. The scores are measured in dB SNR (signal-to-noise ratio) and a lower score is better. The lowest possible score is -15 dB SNR and the highest is +20 dB SNR. It is measured in the unaided condition.

    Measured: Week 1

  2. Speech perception with broadband noise

    Speech perception is tested using a standardized speech test that requires subjects to repeat sentences presented among broadband background noise. The test is adaptive with the background noise getting louder or softer so that the subject answers 50% correct. The scores are measured in dB SNR (signal-to-noise ratio) and a lower score is better. The lowest possible score is -15 dB SNR and the highest is +20 dB SNR. It is measured in the aided condition with the new device or the current device depending on the randomization.

    Measured: Week 2

  3. Speech perception with broadband noise

    Speech perception is tested using a standardized speech test that requires subjects to repeat sentences presented among broadband background noise. The test is adaptive with the background noise getting louder or softer so that the subject answers 50% correct. The scores are measured in dB SNR (signal-to-noise ratio) and a lower score is better. The lowest possible score is -15 dB SNR and the highest is +20 dB SNR. It is measured in the aided condition with the new device or the current device depending on the randomization.

    Measured: Week 3

  4. Acceptance of noise

    Acceptance of noise is tested using a standardized noise test that requires the subjects to adjust the background noise to the maximum Levels that they can tolerate it and to the Levels at which they can understand speech. It is measured in dB SNR (signal-to-noise ratio) and a lower score is better. The lowest score possible is -25 dB SNR and the highest is +20 dB SNR. It is measured in the unaided condition.

    Measured: Week 1

  5. Acceptance of noise

    Acceptance of noise is tested using a standardized noise test that requires the subjects to adjust the background noise to the maximum Levels that they can tolerate it and to the Levels at which they can understand speech. It is measured in dB SNR (signal-to-noise ratio) and a lower score is better. The lowest score possible is -25 dB SNR and the highest is +20 dB SNR. It is measured in the aided condition with the new device or the current device depending on the randomization.

    Measured: Week 2

  6. Acceptance of noise

    Acceptance of noise is tested using a standardized noise test that requires the subjects to adjust the background noise to the maximum Levels that they can tolerate it and to the Levels at which they can understand speech. It is measured in dB SNR (signal-to-noise ratio) and a lower score is better. The lowest score possible is -25 dB SNR and the highest is +20 dB SNR. It is measured in the aided condition with the new device or the current device depending on the randomization.

    Measured: Week 3

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted5 May 2020
  2. Study start3 Aug 2020
  3. Primary completion30 Nov 2020
  4. Study completion30 Nov 2020

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Treatment
Allocation
Randomised
Intervention model
Crossover
Model description
Subjects will randomly wear the new device or the current device for the first period and then switch to the other device for the second period.
Masking
Double blind
Who is masked
Participant, Investigator
Enrolment
19 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT04374851
Sponsor's protocol ID
BF005-2000
Status last verified
Jan 2021
First submitted
16 Apr 2020
Last updated
7 Jan 2021
Results posted
No
Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
No
Individual data sharing
No: There is no plan to share individual data.

Questions about this trial

What is this clinical trial studying?

A study testing Polaris miniRITE BTE and Viron miniRiTE BTE in people with Hearing Loss, run by Bernafon AG at 1 site in Switzerland.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 5 inclusion and 8 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 7 Jan 2021. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in Switzerland. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Speech perception with narrowband noise (time frame: Week 1). The study lists 3 primary outcomes in all. It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 7 Jan 2021.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in