Vital USA SpO2 Accuracy Comparison to Arterial Blood CO-Oximetry

A study testing Accuracy of the VitalDetect™ pulse oximetry system in people with Arterial Disease, run by Vital USA, Inc. at 1 site in United States.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

Clinical Investigation Plan: Vital USA SpO2 Accuracy Comparison to Arterial Blood CO-Oximetry Study ID# PR 2019-353, Version: Rev 1, Date: 15 Oct 2019 Study Dates: November 4-5, 2019 Read more about this study

Who is it for?
Ages 18 to 52 yearsHealthy volunteers can take part
Conditions the study is for
Arterial Disease
What is being tested?
Accuracy of the VitalDetect™ pulse oximetry system
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
10 participants
Where
1 siteUnited States
Study dates
4 Nov 2019to 20 Nov 2019
Phase
Interventional
Are there results?
Yes, posted in the registrySummary in the doctors' view

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 to 52 years
  • All sexes
  • Healthy volunteers accepted
  • Adult

Who may be able to join

  • 10-15 Adults with a minimum of 3 males and a minimum of 3 females, with the balance made up of either
  • Subject must have the ability to understand and provide written informed consent
  • Subject is 18 to 50 years of age
  • Subject must be willing and able to comply with study procedures and duration
  • Subject is a non-smoker or who has not smoked within 2 days prior to the study

Who cannot join

  • Subject is considered as being morbidly obese (defined as BMI >39.5).
  • Compromised circulation, injury, or physical malformation of fingers, wrist, hands, ears or forehead/skull or other sensor sites which would limit the ability to test sites needed for the study. (Note: Certain malformations may still allow subjects to participate if the condition is noted and would not affect the particular sites utilized.)
  • Female subjects that are actively trying to get pregnant or are pregnant (confirmed by positive urine pregnancy test unless the subject is known to be not of child-bearing potential).
  • Smoker Subjects who have refrained will be screened for COPD levels >3% as assessed with a Radical 7 (Rainbow)
  • Subjects with known respiratory conditions such as: (self-reported):
  • uncontrolled / severe asthma,
Show all 37Show fewer
  • flu,
  • pneumonia / bronchitis,
  • shortness of breath / respiratory distress,
  • unresolved respiratory or lung surgery with continued indications of health issues,
  • emphysema, COPD, lung disease.
  • Subjects with known heart or cardiovascular conditions such as: (self-reported, except for blood pressure and ECG review)
  • hypertension: systolic >140mmHg, Diastolic >90mmHg on 3 consecutive readings (reviewed during health screen).
  • have had cardiovascular surgery
  • Chest pain (angina)
  • heart rhythms other than a normal sinus rhythm or with respiratory sinus arrhythmia (reviewed during health screen)
  • previous heart attack
  • blocked artery
  • unexplained shortness of breath
  • congestive heart failure (CHF)
  • history of stroke
  • transient ischemic attack
  • carotid artery disease
  • myocardial ischemia
  • myocardial infarction
  • cardiomyopathy
  • Self-reported health conditions as identified in the Health Assessment Form (self-reported)
  • diabetes,
  • uncontrolled thyroid disease,
  • kidney disease / chronic renal impairment,
  • history of seizures (except childhood febrile seizures),
  • epilepsy,
  • history of unexplained syncope,
  • recent history of frequent migraine headaches,
  • recent symptomatic head injury (within the last 2 months)
  • cancer / chemotherapy
  • Subjects with known clotting disorders (self-reported)

Trial locations and contacts

1 site in 1 country.

United States1 site
  • Clinimark Laboratory Services

    Louisville, Colorado, 80027

Other trials in United States

About this study

Official title: Results of the SpO2 and Pulse Rate Accuracy Comparison of VitalDetect™ to Arterial Blood CO-Oximetry and Reference ECG

Clinical Investigation Plan: Vital USA SpO2 Accuracy Comparison to Arterial Blood CO-Oximetry Study ID# PR 2019-353, Version: Rev 1, Date: 15 Oct 2019 Study Dates: November 4-5, 2019

Read the detailed description

An SpO2 accuracy comparison was conducted as part of the final validation of the VitalDetect™ pulse oximetry system. The study was conducted in accordance to Code of Federal Regulations (CFR) for Non-Significant Risk (NSR) investigational studies, following International Standards Organization (ISO) 14155:2011 as appropriate and the pulse oximetry guidelines of ISO 80601-2-61:2017 applicable sections, and Pulse Oximeters - Premarket Notifications Submissions [510(k)s] Guidance For Industry and Food and Drug Administration Staff (issued: March 4, 2013).

The purpose of this study was to validate the SpO2 accuracy of the VitalDetect™ pulse oximetry system during non-motion conditions over the range of 70-100% SaO2 as compared to arterial blood samples assessed by CO-Oximetry. It was expected that the Accuracy Root Mean Square (ARMS) performance of the oximetry system would meet the required specification of ARMS of 3.0% or less.

A secondary goal was to evaluate the pulse rate performance simultaneously collected over the SpO2 range covered. The Pulse Rate Accuracy Root Mean Square (Arms) performance of the VitalDetect™ pulse oximetry system, were expected to meet a specification of 3 bpm.

Being tested
Accuracy of the VitalDetect™ pulse oximetry system

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Accuracy of the VitalDetect™ pulse oximetry systemDevice

    The purpose of this study was to validate the SpO2 accuracy of the VitalDetect™ pulse oximetry system during non-motion conditions over the range of 70-100% SaO2 as compared to arterial blood samples assessed by CO-Oximetry. It was expected that the Accuracy Root Mean Square (ARMS) performance of the oximetry system would meet the required specification of ARMS of 3.0% or less.

Study groups in NCT04285411
Study groupRoleWhat they receive
ARMS SpO2 70-100%ExperimentalComparison to Reference CO-Oximetry
Receives: Device: Accuracy of the VitalDetect™ pulse oximetry system

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Evaluate SpO2 Accuracy Per Standards for Pulse Oximetry.

    This study was a comparative, single-center, non-randomized study conducted to evaluate the SpO2 accuracy per standards and guidelines identified above for SpO2 accuracy for pulse oximetry equipment over the range of 70-100% SaO2 under non-motion conditions. Arterial blood sampling measured by functional SaO2 CO-Oximetry, was used as the basis for comparison. As a secondary goal, the pulse rate accuracy was compared to ECG Heart Rate realized over the range throughout the hypoxia procedure. Testing was conducted under normal office environment conditions.

    Measured: 90 Seconds

Secondary outcomes (1)
  1. The Pulse Rate Accuracy Compared to ECG Heart Rate.

    The pulse rate accuracy was compared to ECG Heart Rate as recorded by the S5 Multi-parameter monitor. Passing criteria was an ARMS of 3 bpm or better based measurements observed during the pulse oximetry evaluation.

    Measured: 90 Seconds

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted26 Feb 2020
  2. Study start4 Nov 2019
  3. Primary completion5 Nov 2019
  4. Study completion20 Nov 2019

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Device feasibility
Allocation
Not applicable
Intervention model
Single group
Model description
Testing was conducted under normal office environment conditions. The VitalDetect™ pulse oximetry was placed on the right index finger of all subjects to evaluate the SpO2 and pulse rate accuracy performance during steady state non-motion conditions. A Control Pulse Oximetry system was also placed on the subject to evaluate the stability of the draws and the real time oxygen saturation status. The subject was in a reclined position and connected to a breathing circuit, for administering medical grade oxygen and nitrogen. The gas flow delivery was adjusted for subject comfort. Arterial blood samples were drawn during simultaneous data collection from the control pulse oximeter and the test oximeters. The blood was immediately analyzed by Reference CO-Oximetry providing functional SpO2 for the basis of the SpO2 accuracy comparison. The pulse rate for the VitalDetect™ was compared to Reference ECG Heart Rate collected during the study.
Masking
None (open label)
Enrolment
10 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT04285411
Sponsor's protocol ID
PR 2019-353
Status last verified
May 2020
First submitted
23 Feb 2020
Last updated
2 Jun 2020
Data monitoring committee
Yes
FDA-regulated drug
No
FDA-regulated device
Yes
Individual data sharing
No

Questions about this trial

What is this clinical trial studying?

A study testing Accuracy of the VitalDetect™ pulse oximetry system in people with Arterial Disease, run by Vital USA, Inc. at 1 site in United States.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 5 inclusion and 37 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 2 Jun 2020. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in United States. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

Yes. Results are posted in the registry (first posted 2 Jun 2020). A summary is in the For doctors view, and the full results are on ClinicalTrials.gov.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Evaluate SpO2 Accuracy Per Standards for Pulse Oximetry. (time frame: 90 Seconds). It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 2 Jun 2020.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in