Arterial DiseaseNCT04285411
Vital USA SpO2 Accuracy Comparison to Arterial Blood CO-Oximetry
A study testing Accuracy of the VitalDetect™ pulse oximetry system in people with Arterial Disease, run by Vital USA, Inc. at 1 site in United States.
Trial summary
Clinical Investigation Plan: Vital USA SpO2 Accuracy Comparison to Arterial Blood CO-Oximetry Study ID# PR 2019-353, Version: Rev 1, Date: 15 Oct 2019 Study Dates: November 4-5, 2019 Read more about this study
- Who is it for?
- Ages 18 to 52 yearsHealthy volunteers can take part
- Conditions the study is for
- Arterial Disease
- What is being tested?
- Accuracy of the VitalDetect™ pulse oximetry system
- Type of treatment
- Medical device
- Could I receive a placebo?
- No, everyone receives an active treatment
- How many people
- 10 participants
- Where
- 1 siteUnited States
- Study dates
- 4 Nov 2019to 20 Nov 2019
- Phase
- Interventional
- Are there results?
- Yes, posted in the registrySummary in the doctors' view
Key requirements
Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.
- Ages 18 to 52 years
- All sexes
- Healthy volunteers accepted
- Adult
Who may be able to join
- 10-15 Adults with a minimum of 3 males and a minimum of 3 females, with the balance made up of either
- Subject must have the ability to understand and provide written informed consent
- Subject is 18 to 50 years of age
- Subject must be willing and able to comply with study procedures and duration
- Subject is a non-smoker or who has not smoked within 2 days prior to the study
Who cannot join
- Subject is considered as being morbidly obese (defined as BMI >39.5).
- Compromised circulation, injury, or physical malformation of fingers, wrist, hands, ears or forehead/skull or other sensor sites which would limit the ability to test sites needed for the study. (Note: Certain malformations may still allow subjects to participate if the condition is noted and would not affect the particular sites utilized.)
- Female subjects that are actively trying to get pregnant or are pregnant (confirmed by positive urine pregnancy test unless the subject is known to be not of child-bearing potential).
- Smoker Subjects who have refrained will be screened for COPD levels >3% as assessed with a Radical 7 (Rainbow)
- Subjects with known respiratory conditions such as: (self-reported):
- uncontrolled / severe asthma,
Show all 37Show fewer
- flu,
- pneumonia / bronchitis,
- shortness of breath / respiratory distress,
- unresolved respiratory or lung surgery with continued indications of health issues,
- emphysema, COPD, lung disease.
- Subjects with known heart or cardiovascular conditions such as: (self-reported, except for blood pressure and ECG review)
- hypertension: systolic >140mmHg, Diastolic >90mmHg on 3 consecutive readings (reviewed during health screen).
- have had cardiovascular surgery
- Chest pain (angina)
- heart rhythms other than a normal sinus rhythm or with respiratory sinus arrhythmia (reviewed during health screen)
- previous heart attack
- blocked artery
- unexplained shortness of breath
- congestive heart failure (CHF)
- history of stroke
- transient ischemic attack
- carotid artery disease
- myocardial ischemia
- myocardial infarction
- cardiomyopathy
- Self-reported health conditions as identified in the Health Assessment Form (self-reported)
- diabetes,
- uncontrolled thyroid disease,
- kidney disease / chronic renal impairment,
- history of seizures (except childhood febrile seizures),
- epilepsy,
- history of unexplained syncope,
- recent history of frequent migraine headaches,
- recent symptomatic head injury (within the last 2 months)
- cancer / chemotherapy
- Subjects with known clotting disorders (self-reported)
Trial locations and contacts
1 site in 1 country.
United States1 site
Clinimark Laboratory Services
Louisville, Colorado, 80027
About this study
Official title: Results of the SpO2 and Pulse Rate Accuracy Comparison of VitalDetect™ to Arterial Blood CO-Oximetry and Reference ECG
Clinical Investigation Plan: Vital USA SpO2 Accuracy Comparison to Arterial Blood CO-Oximetry Study ID# PR 2019-353, Version: Rev 1, Date: 15 Oct 2019 Study Dates: November 4-5, 2019
Read the detailed description
An SpO2 accuracy comparison was conducted as part of the final validation of the VitalDetect™ pulse oximetry system. The study was conducted in accordance to Code of Federal Regulations (CFR) for Non-Significant Risk (NSR) investigational studies, following International Standards Organization (ISO) 14155:2011 as appropriate and the pulse oximetry guidelines of ISO 80601-2-61:2017 applicable sections, and Pulse Oximeters - Premarket Notifications Submissions [510(k)s] Guidance For Industry and Food and Drug Administration Staff (issued: March 4, 2013).
The purpose of this study was to validate the SpO2 accuracy of the VitalDetect™ pulse oximetry system during non-motion conditions over the range of 70-100% SaO2 as compared to arterial blood samples assessed by CO-Oximetry. It was expected that the Accuracy Root Mean Square (ARMS) performance of the oximetry system would meet the required specification of ARMS of 3.0% or less.
A secondary goal was to evaluate the pulse rate performance simultaneously collected over the SpO2 range covered. The Pulse Rate Accuracy Root Mean Square (Arms) performance of the VitalDetect™ pulse oximetry system, were expected to meet a specification of 3 bpm.
- Conditions
- Arterial Disease
- Being tested
- Accuracy of the VitalDetect™ pulse oximetry system
What participants receive
What participants receive. In a randomised study each person is assigned to one group by chance.
Accuracy of the VitalDetect™ pulse oximetry systemDevice
The purpose of this study was to validate the SpO2 accuracy of the VitalDetect™ pulse oximetry system during non-motion conditions over the range of 70-100% SaO2 as compared to arterial blood samples assessed by CO-Oximetry. It was expected that the Accuracy Root Mean Square (ARMS) performance of the oximetry system would meet the required specification of ARMS of 3.0% or less.
| Study group | Role | What they receive |
|---|---|---|
| ARMS SpO2 70-100% | Experimental | Comparison to Reference CO-Oximetry Receives: Device: Accuracy of the VitalDetect™ pulse oximetry system |
What is being measured
The primary outcome is the main question the study is designed to answer.
Evaluate SpO2 Accuracy Per Standards for Pulse Oximetry.
This study was a comparative, single-center, non-randomized study conducted to evaluate the SpO2 accuracy per standards and guidelines identified above for SpO2 accuracy for pulse oximetry equipment over the range of 70-100% SaO2 under non-motion conditions. Arterial blood sampling measured by functional SaO2 CO-Oximetry, was used as the basis for comparison. As a secondary goal, the pulse rate accuracy was compared to ECG Heart Rate realized over the range throughout the hypoxia procedure. Testing was conducted under normal office environment conditions.
Measured: 90 Seconds
Secondary outcomes (1)
The Pulse Rate Accuracy Compared to ECG Heart Rate.
The pulse rate accuracy was compared to ECG Heart Rate as recorded by the S5 Multi-parameter monitor. Passing criteria was an ARMS of 3 bpm or better based measurements observed during the pulse oximetry evaluation.
Measured: 90 Seconds
Timeline
From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.
- First posted26 Feb 2020
- Study start4 Nov 2019
- Primary completion5 Nov 2019
- Study completion20 Nov 2019
Resources
- Official record on ClinicalTrials.govThe registry entry this page is built from, with its full change history.
- Posted resultsParticipant flow, outcome measures and adverse events reported by the sponsor.
- Other trials for Arterial DiseaseStudies of the same condition, open and under way.
- What the trial phases meanHow a treatment moves from first tests in people to approval.
More details
For doctors and researchers: how the study is designed, who runs it, and its registry IDs.
Study design
- Study type
- Interventional
- Phase
- Phase N/A
- Primary purpose
- Device feasibility
- Allocation
- Not applicable
- Intervention model
- Single group
- Model description
- Testing was conducted under normal office environment conditions. The VitalDetect™ pulse oximetry was placed on the right index finger of all subjects to evaluate the SpO2 and pulse rate accuracy performance during steady state non-motion conditions. A Control Pulse Oximetry system was also placed on the subject to evaluate the stability of the draws and the real time oxygen saturation status. The subject was in a reclined position and connected to a breathing circuit, for administering medical grade oxygen and nitrogen. The gas flow delivery was adjusted for subject comfort. Arterial blood samples were drawn during simultaneous data collection from the control pulse oximeter and the test oximeters. The blood was immediately analyzed by Reference CO-Oximetry providing functional SpO2 for the basis of the SpO2 accuracy comparison. The pulse rate for the VitalDetect™ was compared to Reference ECG Heart Rate collected during the study.
- Masking
- None (open label)
- Enrolment
- 10 participants (actual)
Sponsor and investigators
- Lead sponsor
- Vital USA, Inc.
- Sponsor type
- Industry (pharma, biotech, devices)
- Collaborators
- Clinimark, LLC
- Responsible party
- Sponsor
- Investigators
- Arthur Ruiz, MD, Principal investigator, Avista Adventist Hospital
IDs, dates and publications
- ClinicalTrials.gov ID
- NCT04285411
- Sponsor's protocol ID
- PR 2019-353
- Status last verified
- May 2020
- First submitted
- 23 Feb 2020
- Last updated
- 2 Jun 2020
- Results posted
- Yes, on ClinicalTrials.gov
- Data monitoring committee
- Yes
- FDA-regulated drug
- No
- FDA-regulated device
- Yes
- Individual data sharing
- No
Study at a glance
From the registry record, last updated by the sponsor on 2 Jun 2020. Recruitment status is as the sponsor reported it: confirm with the study team.
- Official title
- Results of the SpO2 and Pulse Rate Accuracy Comparison of VitalDetect™ to Arterial Blood CO-Oximetry and Reference ECG
- ClinicalTrials.gov ID
- NCT04285411
- Acronym / sponsor ID
- PR 2019-353
- Conditions
- Arterial Disease
- Study type
- Interventional
- Recruitment status
- Completed
- Enrolment
- 10 participants (actual)
- Lead sponsor
- Vital USA, Inc.Industry (pharma, biotech, devices)
- Collaborators
- Clinimark, LLC
- Study start
- 4 Nov 2019 (actual)
- Primary completion
- 5 Nov 2019 (actual)
- Study completion
- 20 Nov 2019 (actual)
- First posted
- 26 Feb 2020 (actual)
- Results
- Posted 2 Jun 2020
Clinical summary
The research question as the sponsor registered it. This describes what the study plans to test, not what it found.
Clinical Investigation Plan: Vital USA SpO2 Accuracy Comparison to Arterial Blood CO-Oximetry Study ID# PR 2019-353, Version: Rev 1, Date: 15 Oct 2019 Study Dates: November 4-5, 2019
Detailed description
An SpO2 accuracy comparison was conducted as part of the final validation of the VitalDetect™ pulse oximetry system. The study was conducted in accordance to Code of Federal Regulations (CFR) for Non-Significant Risk (NSR) investigational studies, following International Standards Organization (ISO) 14155:2011 as appropriate and the pulse oximetry guidelines of ISO 80601-2-61:2017 applicable sections, and Pulse Oximeters - Premarket Notifications Submissions [510(k)s] Guidance For Industry and Food and Drug Administration Staff (issued: March 4, 2013).
The purpose of this study was to validate the SpO2 accuracy of the VitalDetect™ pulse oximetry system during non-motion conditions over the range of 70-100% SaO2 as compared to arterial blood samples assessed by CO-Oximetry. It was expected that the Accuracy Root Mean Square (ARMS) performance of the oximetry system would meet the required specification of ARMS of 3.0% or less.
A secondary goal was to evaluate the pulse rate performance simultaneously collected over the SpO2 range covered. The Pulse Rate Accuracy Root Mean Square (Arms) performance of the VitalDetect™ pulse oximetry system, were expected to meet a specification of 3 bpm.
Study design
- Primary purpose
- Device feasibility
- Allocation
- Not applicable
- Intervention model
- Single group
- Masking
- None (open label)
- Number of arms
- 1
- Model description
- Testing was conducted under normal office environment conditions. The VitalDetect™ pulse oximetry was placed on the right index finger of all subjects to evaluate the SpO2 and pulse rate accuracy performance during steady state non-motion conditions. A Control Pulse Oximetry system was also placed on the subject to evaluate the stability of the draws and the real time oxygen saturation status. The subject was in a reclined position and connected to a breathing circuit, for administering medical grade oxygen and nitrogen. The gas flow delivery was adjusted for subject comfort. Arterial blood samples were drawn during simultaneous data collection from the control pulse oximeter and the test oximeters. The blood was immediately analyzed by Reference CO-Oximetry providing functional SpO2 for the basis of the SpO2 accuracy comparison. The pulse rate for the VitalDetect™ was compared to Reference ECG Heart Rate collected during the study.
| Arm | Type | Receives |
|---|---|---|
| ARMS SpO2 70-100% | Experimental | Device: Accuracy of the VitalDetect™ pulse oximetry systemComparison to Reference CO-Oximetry |
Who may qualify
- Ages
- 18 to 52 years
- Sex
- All
- Healthy volunteers
- Accepted
- Age groups
- Adult
Key inclusion criteria (5 listed)
- 10-15 Adults with a minimum of 3 males and a minimum of 3 females, with the balance made up of either
- Subject must have the ability to understand and provide written informed consent
- Subject is 18 to 50 years of age
- Subject must be willing and able to comply with study procedures and duration
- Subject is a non-smoker or who has not smoked within 2 days prior to the study
Key exclusion criteria (37 listed)
- Subject is considered as being morbidly obese (defined as BMI >39.5).
- Compromised circulation, injury, or physical malformation of fingers, wrist, hands, ears or forehead/skull or other sensor sites which would limit the ability to test sites needed for the study. (Note: Certain malformations may still allow subjects to participate if the condition is noted and would not affect the particular sites utilized.)
- Female subjects that are actively trying to get pregnant or are pregnant (confirmed by positive urine pregnancy test unless the subject is known to be not of child-bearing potential).
- Smoker Subjects who have refrained will be screened for COPD levels >3% as assessed with a Radical 7 (Rainbow)
- Subjects with known respiratory conditions such as: (self-reported):
Eligibility is decided by the study team after screening; this overview does not establish that a patient qualifies. Full inclusion and exclusion criteria.
Interventions
| Intervention | Type | Arms |
|---|---|---|
| Accuracy of the VitalDetect™ pulse oximetry systemThe purpose of this study was to validate the SpO2 accuracy of the VitalDetect™ pulse oximetry system during non-motion conditions over the range of 70-100% SaO2 as compared to arterial blood samples assessed by CO-Oximetry. It was expected that the Accuracy Root Mean Square (ARMS) performance of the oximetry system would meet the required specification of ARMS of 3.0% or less. | Device | ARMS SpO2 70-100% |
Primary outcome
Evaluate SpO2 Accuracy Per Standards for Pulse Oximetry.
Time frame: 90 Seconds
This study was a comparative, single-center, non-randomized study conducted to evaluate the SpO2 accuracy per standards and guidelines identified above for SpO2 accuracy for pulse oximetry equipment over the range of 70-100% SaO2 under non-motion conditions. Arterial blood sampling measured by functional SaO2 CO-Oximetry, was used as the basis for comparison. As a secondary goal, the pulse rate accuracy was compared to ECG Heart Rate realized over the range throughout the hypoxia procedure. Testing was conducted under normal office environment conditions.
1 secondary outcome
The Pulse Rate Accuracy Compared to ECG Heart Rate.
Time frame: 90 Seconds
Results and safety
Loading the posted results from ClinicalTrials.gov…
Research links
- Official record on ClinicalTrials.govLast updated by the sponsor 2 Jun 2020.
- Posted resultsParticipant flow, baseline, all outcome measures and adverse events.
- Individual participant dataSharing planned: No
Related trials for Arterial Disease
Not yet recruitingPhase 1
Hepatic Artery Infusion of Carfilzomib in Participants With Liver Metastatic Disease Previously Treated With Hepatic Artery Infusion Pump Therapy
Colorectal Neoplasms, Neoplasms
Not yet recruiting
Effect of Stress Ball Use During Femoral Arterial Sheath Removal After Coronary Angiography
Femoral Arterial Sheath Removal, Stress Ball
Active, not recruiting
Shockwave™ vs Surgical Endarterectomy for Calcified Severe Common Femoral Artery Stenosis: Comparison of Efficacy, Safety and Long-Term Outcomes
Common Femoral Artery Stenosis, Calcification; Heart
Recruiting
PAW-Hydrogel for Chronic Wound Healing
Diabetic Foot Ulcer (DFU), Pressure Ulcer (PU)
Active, not recruitingPhase 4
Effects of Sequential Therapy With Mexidol® (Intravenous) and Mexidol® FORTE 250 (Oral) on Functional Brain Activity Assessed by Functional Magnetic Resonance Imaging and Biochemical Parameters in Patients With Chronic Cerebral Ischemia: A Comparison of Patients With and Without Metabolic Disorders
Metabolic Abnormalities, Obesity & Overweight
Referral and contact
AzTrials does not enrol participants or take referrals. Contact the study team or a site listed below; they decide who can take part. This study is not recruiting at the moment.
Arthur Ruiz, MD principal investigator
Avista Adventist Hospital
1 site in 1 country: see every site and its status.
Questions about this trial
What is this clinical trial studying?
A study testing Accuracy of the VitalDetect™ pulse oximetry system in people with Arterial Disease, run by Vital USA, Inc. at 1 site in United States.
The full registered description is in the About this study section and the official record.
Who may be eligible to take part?
The sponsor lists 5 inclusion and 37 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.
Is this clinical trial currently recruiting?
The record says: finished, as of the sponsor's last update on 2 Jun 2020. Status can change between updates, so confirm with the study contact before making plans.
Where is the study taking place?
The record lists 1 site in United States. Whether a site is taking part right now should be confirmed with that site. See every site.
What are the possible benefits and risks?
They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.
Are results available for this study?
Yes. Results are posted in the registry (first posted 2 Jun 2020). A summary is in the For doctors view, and the full results are on ClinicalTrials.gov.
How can a doctor refer a patient or contact the research team?
The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.
What does the primary outcome measure?
Evaluate SpO2 Accuracy Per Standards for Pulse Oximetry. (time frame: 90 Seconds). It is what the study was designed to measure, not a finding.