Death, Sudden, CardiacACE-CONVERTNCT04132466
ASSURE WCD Clinical Evaluation - Conversion Efficacy Study
A study testing Defibrillation using the ASSURE wearable cardioverter defibrillator therapy delivered via a waveform delivery test system in people with Death, Sudden, Cardiac, run by Kestra Medical Technologies, Inc. at 4 sites in United States.
Trial summary
A multicenter single arm open label evaluation of the ASSURE Wearable Cardioverter Defibrillator (WCD) defibrillation waveform in adult cardiac patients Read more about this study
- Who is it for?
- Ages 18 years and older
- Conditions the study is for
- Death, Sudden, Cardiac
- What is being tested?
- Defibrillation using the ASSURE wearable cardioverter defibrillator therapy delivered via a waveform delivery test system
- Type of treatment
- Medical device
- Could I receive a placebo?
- No, everyone receives an active treatment
- How many people
- 13 participants
- Where
- 4 sitesUnited States
- Study dates
- 25 Nov 2019to 19 Mar 2020
- Phase
- Interventional
- Are there results?
- Yes, posted in the registrySummary in the doctors' view
Key requirements
Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.
- Ages 18 years and older
- All sexes
- No healthy volunteers
- Adult, Older adult
Who may be able to join
- Males or females, age ≥ 18 years
- Able and willing to provide written informed consent before undergoing any study-related procedures
- Scheduled for any of the following procedures:
- Electrophysiology study for induction of ventricular arrhythmias
- Non-invasive electrophysiology testing using an existing implantable defibrillator
- Implantable cardioverter-defibrillator (ICD) replacement procedure during which induction of a ventricular arrhythmia is planned
- Ablation of ventricular tachycardia (patients undergoing ventricular tachycardia ablation in which only a substrate modification approach is planned, with no intention of inducing a ventricular arrhythmia, should not be included)
Who cannot join
- Any condition that by the judgement of the physician investigator precludes the subject's ability to comply with the study requirements
- Pregnancy
- Use of mechanical circulatory support (e.g. Left Ventricular Assist Device, Total Artificial Heart, intraaortic balloon pump or Impella)
- Documented nonchronic cardiac thrombus
- Atrial fibrillation or atrial flutter without therapeutic systemic anticoagulation
- Critical aortic stenosis
Show all 16Show fewer
- Unstable coronary artery disease (CAD)
- Recent stroke or transient ischemic attack (TIA)
- Hemodynamic instability
- Currently implanted Boston Scientific S-ICD (due to location of implant relative to test system)
- Unstable angina
- New York Heart Association (NYHA) Class IV
- Left Ventricular Ejection Fraction (LVEF) < 20%
- Any medical condition that by the judgement of the physician investigator, patient participation in this study is not in the best interest of the patient
- History of difficulty of ventricular arrhythmia induction
- Amiodarone use within 3 months before the study procedure
Trial locations and contacts
4 sites in 1 country.
United States4 sites
Alaska Cardiovascular Research Foundation, Inc.
Anchorage, Alaska, 99508
University of Michigan
Ann Arbor, Michigan, 48109
Washington University St Louis
St Louis, Missouri, 63110
Institute for Research and Innovation MultiCare Health System
Tacoma, Washington, 98405
About this study
Official title: ASSURE WCD Clinical Evaluation - Conversion Efficacy Study (ACE-CONVERT)
A multicenter single arm open label evaluation of the ASSURE Wearable Cardioverter Defibrillator (WCD) defibrillation waveform in adult cardiac patients
Read the detailed description
The objective of this study is to evaluate Ventricular Tachycardia/Ventricular Fibrillation (VT/VF) conversion efficacy in adult human subjects using the ASSURE Wearable Cardioverter Defibrillator (WCD) defibrillation waveform.
- Conditions
- Death, Sudden, Cardiac
- Being tested
- Defibrillation using the ASSURE wearable cardioverter defibrillator therapy delivered via a waveform delivery test system
- Medical subject headings
- Death, Sudden, Cardiac
What participants receive
What participants receive. In a randomised study each person is assigned to one group by chance.
Defibrillation using the ASSURE wearable cardioverter defibrillator therapy delivered via a waveform delivery test systemDevice
defibrillation using a 170 joule biphasic truncated exponential waveform designed for the ASSURE wearable cardioverter defibrillator
| Study group | Role | What they receive |
|---|---|---|
| Enrolled Subjects | Experimental | Adult subjects who met eligibility criteria and provided written informed consent to participate Receives: Device: Defibrillation using the ASSURE wearable cardioverter defibrillator therapy delivered via a waveform delivery test system |
What is being measured
The primary outcome is the main question the study is designed to answer.
Cumulative First and Second Shock VT/VF Conversion Rate
The number of subjects with successful (first or second shock) arrhythmia conversion using the Test System divided by the total number of inductions attempted using the Test System
A successful arrhythmia conversion is defined as termination of an induced ventricular rhythm by first or second shock from the Test System to a non-shockable rhythm (rhythms other than VT or VF).
Measured: Through study procedure completion, average of 2 hours
Secondary outcomes (1)
First Shock VT/VF Conversion Rate
The number of subjects with successful (first shock) arrhythmia conversion using the Test System divided by the total number of inductions attempted with using the Test System
A successful arrhythmia conversion is defined as termination of an induced ventricular rhythm by first shock from the Test System to a non-shockable rhythm (rhythms other than VT or VF).
Measured: Through study procedure completion, average of 2 hours
Timeline
From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.
- First posted18 Oct 2019
- Study start25 Nov 2019
- Primary completion19 Mar 2020
- Study completion19 Mar 2020
Resources
- Official record on ClinicalTrials.govThe registry entry this page is built from, with its full change history.
- Posted resultsParticipant flow, outcome measures and adverse events reported by the sponsor.
- Publications (1)Papers the sponsor has linked to this study.
- Other trials for Death, Sudden, CardiacStudies of the same condition, open and under way.
- What the trial phases meanHow a treatment moves from first tests in people to approval.
More details
For doctors and researchers: how the study is designed, who runs it, and its registry IDs.
Study design
- Study type
- Interventional
- Phase
- Phase N/A
- Primary purpose
- Treatment
- Allocation
- Not applicable
- Intervention model
- Single group
- Model description
- Multicenter single arm open label
- Masking
- None (open label)
- Enrolment
- 13 participants (actual)
Sponsor and investigators
- Lead sponsor
- Kestra Medical Technologies, Inc.
- Sponsor type
- Industry (pharma, biotech, devices)
- Collaborators
- Regulatory and Clinical Research Institute Inc
- Responsible party
- Sponsor
- Investigators
- Marye Gleva, MD, Principal investigator, Washington University School of Medicine
IDs, dates and publications
- ClinicalTrials.gov ID
- NCT04132466
- Sponsor's protocol ID
- 3333934_C
- Acronym
- ACE-CONVERT
- Status last verified
- Apr 2024
- First submitted
- 15 Oct 2019
- Last updated
- 5 Apr 2024
- Results posted
- Yes, on ClinicalTrials.gov
- Data monitoring committee
- No
- FDA-regulated drug
- No
- FDA-regulated device
- Yes
- Individual data sharing
- No
Publications
Gleva MJ, Sullivan J, Crawford TC, Walcott G, Birgersdotter-Green U, Branch KR, Doshi RN, Kivilaid K, Brennan K, Rowbotham RK, Gustavson LM, Poole JE. Defibrillation effectiveness and safety of the shock waveform used in a contemporary wearable cardioverter defibrillator: Results from animal and human studies. PLoS One. 2023 Mar 14;18(3):e0281340. doi: 10.1371/journal.pone.0281340. eCollection 2023.
PubMed 36917566
Study at a glance
From the registry record, last updated by the sponsor on 5 Apr 2024. Recruitment status is as the sponsor reported it: confirm with the study team.
- Official title
- ASSURE WCD Clinical Evaluation - Conversion Efficacy Study (ACE-CONVERT)
- ClinicalTrials.gov ID
- NCT04132466
- Acronym / sponsor ID
- ACE-CONVERT / 3333934_C
- Conditions
- Death, Sudden, Cardiac
- Study type
- Interventional
- Recruitment status
- Completed
- Enrolment
- 13 participants (actual)
- Lead sponsor
- Kestra Medical Technologies, Inc.Industry (pharma, biotech, devices)
- Collaborators
- Regulatory and Clinical Research Institute Inc
- Study start
- 25 Nov 2019 (actual)
- Primary completion
- 19 Mar 2020 (actual)
- Study completion
- 19 Mar 2020 (actual)
- First posted
- 18 Oct 2019 (actual)
- Results
- Posted 5 Apr 2024
Clinical summary
The research question as the sponsor registered it. This describes what the study plans to test, not what it found.
A multicenter single arm open label evaluation of the ASSURE Wearable Cardioverter Defibrillator (WCD) defibrillation waveform in adult cardiac patients
Detailed description
The objective of this study is to evaluate Ventricular Tachycardia/Ventricular Fibrillation (VT/VF) conversion efficacy in adult human subjects using the ASSURE Wearable Cardioverter Defibrillator (WCD) defibrillation waveform.
Study design
- Primary purpose
- Treatment
- Allocation
- Not applicable
- Intervention model
- Single group
- Masking
- None (open label)
- Number of arms
- 1
- Model description
- Multicenter single arm open label
| Arm | Type | Receives |
|---|---|---|
| Enrolled Subjects | Experimental | Device: Defibrillation using the ASSURE wearable cardioverter defibrillator therapy delivered via a waveform delivery test systemAdult subjects who met eligibility criteria and provided written informed consent to participate |
Who may qualify
- Ages
- 18 years and older
- Sex
- All
- Healthy volunteers
- Not accepted
- Age groups
- Adult, Older adult
Key inclusion criteria (7 listed)
- Males or females, age ≥ 18 years
- Able and willing to provide written informed consent before undergoing any study-related procedures
- Scheduled for any of the following procedures:
Key exclusion criteria (16 listed)
- Any condition that by the judgement of the physician investigator precludes the subject's ability to comply with the study requirements
- Pregnancy
- Use of mechanical circulatory support (e.g. Left Ventricular Assist Device, Total Artificial Heart, intraaortic balloon pump or Impella)
- Documented nonchronic cardiac thrombus
- Atrial fibrillation or atrial flutter without therapeutic systemic anticoagulation
Eligibility is decided by the study team after screening; this overview does not establish that a patient qualifies. Full inclusion and exclusion criteria.
Interventions
| Intervention | Type | Arms |
|---|---|---|
| Defibrillation using the ASSURE wearable cardioverter defibrillator therapy delivered via a waveform delivery test systemdefibrillation using a 170 joule biphasic truncated exponential waveform designed for the ASSURE wearable cardioverter defibrillator | Device | Enrolled Subjects |
Primary outcome
Cumulative First and Second Shock VT/VF Conversion Rate
Time frame: Through study procedure completion, average of 2 hours
The number of subjects with successful (first or second shock) arrhythmia conversion using the Test System divided by the total number of inductions attempted using the Test System A successful arrhythmia conversion is defined as termination of an induced ventricular rhythm by first or second shock from the Test System to a non-shockable rhythm (rhythms other than VT or VF).
1 secondary outcome
First Shock VT/VF Conversion Rate
Time frame: Through study procedure completion, average of 2 hours
Results and safety
Loading the posted results from ClinicalTrials.gov…
Research links
- Official record on ClinicalTrials.govLast updated by the sponsor 5 Apr 2024.
- Posted resultsParticipant flow, baseline, all outcome measures and adverse events.
- Results publication: PubMed 36917566Gleva MJ, Sullivan J, Crawford TC, Walcott G, Birgersdotter-Green U, Branch KR, Doshi RN, Kivilaid K, Brennan K, Rowbotham RK, Gustavson LM, Poole JE. Defibrillation effectiveness and safety of the shock waveform used in a contemporary wearable cardioverter defibrillator: Results from animal and human studies. PLoS One. 2023 Mar 14;18(3):e0281340. doi: 10.1371/journal.pone.0281340. eCollection 2023.
- Individual participant dataSharing planned: No
Related trials for Death, Sudden, Cardiac
RecruitingObservational study
Observational Study of Cardiac Arrhythmias During Treatment With BTK Inhibitors or Venetoclax
Chronic Lymphocytic Leukemia (CLL), Waldenstr(SqrRoot)(Delta)m s Macroglobulinemia
Recruiting
Study of the Link Between Structural Alterations in the Hypothalamus and Cardiac Dysautonomia in Drug-resistant Epilepsy
Epilepsy, Hypothalamus
Not yet recruitingObservational study
Cardiovascular Pre-participation Screening in Children and Adolescents
Sudden Cardiac Death, Cardiac Arrhythmias
Recruiting
Speed-up the Diagnosis and Evaluation of anoMalous Coronary ARTery From the Aorta
AAOCA, ACAOS
RecruitingObservational study
Hypertrophic Cardiomyopathy Registry, Biobank and Imaging Data Repository
Hypertrophic Cardiomyopathy, Hypertrophic Obstructive Cardiomyopathy
Referral and contact
AzTrials does not enrol participants or take referrals. Contact the study team or a site listed below; they decide who can take part. This study is not recruiting at the moment.
Marye Gleva, MD principal investigator
Washington University School of Medicine
4 sites in 1 country: see every site and its status.
Questions about this trial
What is this clinical trial studying?
A study testing Defibrillation using the ASSURE wearable cardioverter defibrillator therapy delivered via a waveform delivery test system in people with Death, Sudden, Cardiac, run by Kestra Medical Technologies, Inc. at 4 sites in United States.
The full registered description is in the About this study section and the official record.
Who may be eligible to take part?
The sponsor lists 7 inclusion and 16 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.
Is this clinical trial currently recruiting?
The record says: finished, as of the sponsor's last update on 5 Apr 2024. Status can change between updates, so confirm with the study contact before making plans.
Where is the study taking place?
The record lists 4 sites in United States. Whether a site is taking part right now should be confirmed with that site. See every site.
What are the possible benefits and risks?
They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.
Are results available for this study?
Yes. Results are posted in the registry (first posted 5 Apr 2024). A summary is in the For doctors view, and the full results are on ClinicalTrials.gov.
How can a doctor refer a patient or contact the research team?
The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.
What does the primary outcome measure?
Cumulative First and Second Shock VT/VF Conversion Rate (time frame: Through study procedure completion, average of 2 hours). It is what the study was designed to measure, not a finding.
Where can I read the original study information?
On the ClinicalTrials.gov record for NCT04132466, which also lists 1 linked publication.
Does a listing mean the treatment is approved or proven to work?
No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.
How often is this information updated?
AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 5 Apr 2024.