AzTrials
AzTrials

Blood Pressure MeasurementNCT04132453

Continuous Blood Pressure and Waveform Measurements Comparison Between Invasive Arterial Line and PPG Sensor

An observational study of Blood Pressure Measurement, run by Sensifree Ltd. at 1 site in Israel.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

The purpose of this study is to compare blood pressure values (systolic, diastolic, mean arterial pressure) and waveform as recorded by a radial arterial line catheter to the waveform and values as calculated by the Sensifree's algorithm from a PPG sensor from a variety group of patients during elective surgeries. Read more about this study

Who is it for?
Ages 18 to 80 years
Conditions the study is for
Blood Pressure Measurement
What is being tested?
Arterial catheterization kit, GE Datex-Ohmeda Finger Clip Pulse Oximeter Sensor and NIBP Cuff (GE Healthcare)
Type of treatment
Medical device
How many people
100 participants
Where
1 siteIsrael
Study dates
19 Jun 2018to 31 Mar 2021
Phase
Observational
Are there results?
None posted in the registry

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 to 80 years
  • All sexes
  • No healthy volunteers
  • Adult, Older adult

Study population: Patients scheduled for elective surgery involving placement of radial artery A-line for BP monitoring

Who may be able to join

  • Signed inform consent
  • Age 18-20
  • Elective surgery that requires invasive BP monitoring (A-line)

Who cannot join

  • Emergency surgery
  • Subject has a condition that might prevent BP fluctuations
  • Subject is taking medicine that might prevent BP fluctuations
  • Arterial line not placed in the radial artery
  • Flush test has inadequate frequency response pre-trial
  • Subject's arterial line contralateral hand is not available for PPG sensor placement
  • Subject has 50% or more pacemaker dependency

Trial locations and contacts

1 site in 1 country.

Israel1 site
  • Wolfson Medical Center

    Holon, 58100

Other trials in Israel

About this study

Official title: Evaluation of Sensifree's Algorithm for Continuous Blood Pressure Measurements Based on a PPG Sensor's Signal Compared to Invasive Radial Arterial Line

The purpose of this study is to compare blood pressure values (systolic, diastolic, mean arterial pressure) and waveform as recorded by a radial arterial line catheter to the waveform and values as calculated by the Sensifree's algorithm from a PPG sensor from a variety group of patients during elective surgeries.

Read the detailed description

Before induction of anesthesia, an arterial line will be inserted to the patient radial artery using a designated short, stiff tubing kit according to catheter IFU (Instructions for Use), preferably under US guidance. Removal of air bubbles from the catheter-transducer system is mandatory prior to connecting the tubing to the catheter.

Study nurse will connect a Sensifree laptop with designated data acquisition software (Medicollector BEDSIDE) to the multi-parameter monitor communication port and start the recording of invasive blood pressure and pulse oximeter pleth (PPG sensor) per software IFU. Prior to connecting the arterial line system to the patient, a calibration (flush test) will be done and recorded. During the study, any deterioration in the waveform recorded by the intra-arterial catheter will be noted and appropriate corrective measures will be taken as soon as possible (e.g., flushing or adjusting the position of the catheter).

Preferable NIBP (Non Invasive Blood Pressure) measurement rate will be every 10 minutes, unless the medical procedure requires otherwise.

Calibration procedure will be repeated every 60 +/- 5 minutes and at the end of the surgery. Recording will take place for the duration of the surgery.

NIBP cuff will be placed over the arterial-line arm, NIBP air hose will be replaced with an air hose supplied by Sensifree and the pressure gauge in it will be connected to Sensifree laptop via analog to digital converter in order to capture the data.

Medication used during the surgery will be listed and documented in the CRF (Case Report Form) including drugs name, dose, time and method of administration.

Offline data analysis will compare the blood pressure values as calculated by the Sensifree's algorithm based on the PPG signal, to the values obtained from the arterial line. The comparison results will be measured against the accuracy performance requirements per ISO 81060-2.

Being tested
Arterial catheterization kit, GE Datex-Ohmeda Finger Clip Pulse Oximeter Sensor, NIBP Cuff (GE Healthcare)

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Arterial catheterization kitDevice

    Invasive monitoring of blood pressure via catheterization of the radial artery, displaying a continuous pressure waveform

  • GE Datex-Ohmeda Finger Clip Pulse Oximeter SensorDevice

    PPG based, fingertip mounted sensor, that measures the oxygen level (oxygen saturation) of the blood and generates a waveform

  • NIBP Cuff (GE Healthcare)Device

    A blood pressure cuff is used to measure blood pressure. The cuff has an inflatable rubber bladder that is typically fastened around the arm. A pressure meter indicates the cuff's pressure.

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Comparison of waveforms and BP values between PPG sensor and arterial line

    Blood pressure waveform will be acquired from the PPG while blood pressure waveform and values (systolic, diastolic, mean) will be acquired from A-line. PPG data will be analyzed by the Sensifree's algorithm and compared to the A-line measurement results according to ISO 81060-2 accuracy specifications

    Measured: 3 months after last patient enrolled

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted18 Oct 2019
  2. Study start19 Jun 2018
  3. Primary completion31 Mar 2021
  4. Study completion31 Mar 2021

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Observational
Observational model
Cohort
Time perspective
Prospective
Enrolment
100 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT04132453
Sponsor's protocol ID
CLI-0003.04
Status last verified
Oct 2020
First submitted
16 Oct 2019
Last updated
9 Jun 2022
Results posted
No
Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
Yes
Individual data sharing
No

Questions about this trial

What is this clinical trial studying?

An observational study of Blood Pressure Measurement, run by Sensifree Ltd. at 1 site in Israel.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 3 inclusion and 7 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 9 Jun 2022. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in Israel. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Comparison of waveforms and BP values between PPG sensor and arterial line (time frame: 3 months after last patient enrolled). It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 9 Jun 2022.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.