PULSTA Transcatheter Pulmonary Valve Pre-Approval Study

A study testing PULSTA TPV System in people with Heart Diseases and Pulmonary Valve Regurgitation, run by Taewoong Medical Co., Ltd. at 11 sites in 6 countries.

Active, not recruiting

Under way, not looking for participants. Everyone needed has enrolled. The study is still running.

Official record
Search all trials

Trial summary

The purpose of the study is to evaluate the safety and effectiveness of the PULSTA Transcatheter Pulmonary Valve (TPV) System for the treatment of congenital or acquired pulmonary valve stenosis and/or regurgitation who require pulmonary valve replacement. Read more about this study

Who is it for?
Ages 10 years and older
What is being tested?
PULSTA TPV System
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
58 participants
Where
11 sitesin 6 countries
Study dates
7 Oct 2019to Dec 2026
Phase
Interventional
Are there results?
Not yet: the study is under way

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 10 years and older
  • All sexes
  • No healthy volunteers
  • Child, Adult, Older adult

Who may be able to join

  • Body weight greater than or equal to 30 kilograms
  • Pulmonary regurgitation moderate or severe pulmonary regurgitation or pulmonary valve stenosis with mean gradient >35mmHg by echocardiography
  • Main pulmonary artery trunk of ≥16 mm and ≤30 mm
  • Patients willing to provide written informed consent and comply with follow-up evaluations.

Who cannot join

  • Females of child-bearing potential who are unable to take adequate contraceptive precautions, are known to be pregnant, or are currently breastfeeding an infant up until 6 months after TPV implantation
  • Pre-existing mechanical valve in pulmonary position or require concomitant repair of other cardiac valves
  • Obstruction of the central veins to be approached for the TPV implantation
  • Coronary artery compression confirmed by angiography
  • A known severe allergy to Nickel
  • A known hypersensitivity to any anticoagulation therapy, contrast agent or a hemorrhagic disease
  • Suspected active infectious disease (e.g. endocarditis, meningitis)
  • Life expectancy of less than 6 months

Trial locations and contacts

11 sites in 6 countries.

11 sites in 6 countries

South Korea3 sites
  • Sejong General Hospital

    Bucheon-si, Hohyun-ro, Sosa-gu, 14754

  • Seoul National University Hospital

    Seoul, Daehak-ro Jongno-gu, 03080

  • Severance Hospital

    Seoul, Yonsei-ro, Seodaemun-gu, 03722

Other trials in South Korea

Netherlands2 sites
  • Erasmus Medical Center

    Rotterdam, Wytemaweg, 3015

  • Utrecht University Wilhelmina

    Utrecht, Utrecht, 3508

Other trials in Netherlands

Spain2 sites
  • Gregorio Marañon hospital

    Madrid, Madrid, 28007

  • University Hospital La Paz

    Madrid, Madrid, 28046

Other trials in Spain

Turkey (Türkiye)2 sites
  • Koc University Hospital

    Istanbul, Topkapı, Zetinburnu/İstanbul, 34010

  • Siyami Ersek Hospital

    Istanbul, Üsküdar/İstanbul, 34668

Other trials in Turkey (Türkiye)

Germany1 site
  • Deutsches Herzzentrum Munchen

    München, München, 80636

Other trials in Germany

Italy1 site
  • Policlinico San Donato

    Milan, San Donato Milanese, 20097

Other trials in Italy

About this study

Official title: The PULSTA Transcatheter Pulmonary Valve (TPV) Pre-Approval Study

The purpose of the study is to evaluate the safety and effectiveness of the PULSTA Transcatheter Pulmonary Valve (TPV) System for the treatment of congenital or acquired pulmonary valve stenosis and/or regurgitation who require pulmonary valve replacement.

Read the detailed description

The PULSTA TPV is a self-expandable valve with flared-ends to adapt to the larger native right ventricular outflow tract (RVOT) and is using a relatively low profile delivery catheter from knitted nitinol wire backbone and trileaflets made from treated porcine pericardial tissue, and a delivery system, which provides access to the right ventricular outflow tract through blood vessels. Consecutive subject data should be collected at discharge, 1, 6 months, and 1-5 years post PULSTA TPV implantation.

Being tested
PULSTA TPV System
Medical subject headings
Heart Diseases, Pulmonary Valve Insufficiency, Pulmonary Valve Stenosis, Tetralogy of Fallot

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • PULSTA TPV SystemDevice

    PULSTA transcatheter pulmonary valve replacement.

Study groups in NCT03983512
Study groupRoleWhat they receive
PULSTA TPVExperimentalPULSTA Transcatheter Pulmonary Valve (TPV) System
Receives: Device: PULSTA TPV System

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Procedural/Device related serious adverse events at 6 months

    Measured: 6 months

  2. Hemodynamic functional improvement at 6 months

    Hemodynamic functional improvement is a composite of mean RVOT gradient ≤30 mmHg by continuous wave Doppler and a pulmonary regurgitant fraction <20% by cardiac magnetic resonance (MR)

    Measured: 6 months

Secondary outcomes (10)
  1. Procedural Success

    Procedural success is defined as the TPV implant within desired position, Right ventricle-Pulmonary artery (RV-PA) peak systolic pressure gradient <35mmHg by catheter, less mild pulmonary regurgitation by angiography, and freedom from explantation of the TPV at 24 hours post-implant.

    Measured: 7 days

  2. Hemodynamic Function

    Hemodynamic function will be assessed including peak RVOT pressure gradient, mean RVOT pressure gradient, right ventricle (RV) end-diastolic volume index by echocardiography or cardiac MR.

    Measured: 5 years

  3. Pulmonary Regurgitant Fraction

    Pulmonary regurgitant fraction (PRF) will be assessed via cardiac MRI at 6 months after the PULSTA TPV implantation.

    Measured: 6months

  4. Severity of Pulmonary Regurgitation

    Severity of pulmonary regurgitation will be assessed via Doppler echocardiography.

    Measured: 5 years

  5. NYHA functional classification

    Clinical outcome will be assessed as per the New York Heart Association (NYHA) functional classification t all follow-up visits and will be compared to baseline values.

    Measured: 5 years

  6. Catheter reintervention on TPV

    Catheter re-intervention on PULSTA TPV is defined as any interventional catheter procedure (e.g. balloon angioplasty) that alters or adjusts, or replaces a previously implanted PULSTA TPV during a 5-year follow-up period, and the frequency will be evaluated.

    Measured: 5 years

  7. Reoperation

    Reoperation is defined as open surgery that alters or adjusts, or replaces a previously implanted PULSTA TPV during a 5-year follow-up period, and the frequency will be evaluated.

    Measured: 5 years

  8. Procedural / Device related serious adverse events

    Measured: 5 years

  9. Death (All cause / procedural/device-related)

    Measured: 5 years

  10. Other adverse events

    Measured: 5 years

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted12 Jun 2019
  2. Study start7 Oct 2019
  3. Primary completion21 Sep 2022
  4. Study completionDec 2026 (estimated)

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Treatment
Allocation
Not applicable
Intervention model
Single group
Masking
None (open label)
Enrolment
58 participants (estimated)
IDs, dates and publications
ClinicalTrials.gov ID
NCT03983512
Sponsor's protocol ID
TWTPV-E1001
Status last verified
Apr 2026
First submitted
7 Jun 2019
Last updated
13 Apr 2026
Results posted
No
Data monitoring committee
Yes
FDA-regulated drug
No
FDA-regulated device
No
Individual data sharing
No

Publications

  • Lee SY, Kim GB, Carminati M, Giugno L, Celebi A, Yucel IK, Eicken A, Ewert P, Georgiev S, Choi JY, Kim CS, Gutierrez-Larraya F, Balbacid Domingo E, Odemis E, Kizilkaya MH, Luis Zunzunegui J, Ballesteros Tejerizo F, Kim SH, Jang SI, Krasemann T, Krings G. Prospective Midterm Outcomes of the PULSTA Self-Expandable Transcatheter Pulmonary Valve: The PULSTA CE Approval Study. Circ Cardiovasc Interv. 2026 Jun 18:e016616. doi: 10.1161/CIRCINTERVENTIONS.126.016616. Online ahead of print.

    PubMed 42312408

Questions about this trial

What is this clinical trial studying?

A study testing PULSTA TPV System in people with Heart Diseases and Pulmonary Valve Regurgitation, run by Taewoong Medical Co., Ltd. at 11 sites in 6 countries.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 4 inclusion and 8 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: under way, not looking for participants, as of the sponsor's last update on 13 Apr 2026. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 11 sites in South Korea, Netherlands, Spain, Turkey (Türkiye), Germany and 1 more country. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

This study is still under way. Results are not available in the source record yet.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Procedural/Device related serious adverse events at 6 months (time frame: 6 months). The study lists 2 primary outcomes in all. It is what the study was designed to measure, not a finding.

Where can I read the original study information?

On the ClinicalTrials.gov record for NCT03983512, which also lists 1 linked publication.

Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 13 Apr 2026.

Free account

Keep searching with a free account

You’ve used your 3 free searches for this session. Log in or create a free AzTrials account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in
Clinical Trial NCT03983512 | AzTrials