TachycardiaCN_MRI_ICDNCT03828357

China (CN)_Magnetic Resonance Imaging (MRI)_Implantable Cardiac Defibrillator (ICD) [CN_MRI_ICD]

An observational study of Tachycardia, run by Abbott Medical Devices at 1 site in China.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

This clinical study is to evaluate the safety and effectiveness of the Ellipse VR/DR implantable cardiac defibrillators (ICDs) and the Quadra Assura MP cardiac resynchronization therapy defibrillators (CRT-Ds), with Durata or Optisure defibrillation leads, Tendril STS or Isoflex pacing leads, and the Quartet… Read more about this study

Who is it for?
Ages 18 years and older
Conditions the study is for
Tachycardia
What is being tested?
Study MRI scan
How many people
60 participants
Where
1 siteChina
Study dates
27 Mar 2019to 1 Dec 2020
Phase
Observational
Are there results?
Yes, posted in the registrySummary in the doctors' view

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 years and older
  • All sexes
  • No healthy volunteers
  • Adult, Older adult

Study population: The study population includes male and female Chinese nationals that meet the eligibility requirements and give consent for the study. Vulnerable subjects, such as minors or those unable to provide consent are excluded from participating.

Who may be able to join

  • To participate in this clinical study, the subject must meet all of the following inclusion criteria:
  • Have an approved indication for implantation of a ICD
  • Be a Chinese national
  • Will be implanted with one of the following device/lead combinations evaluated in this study:
  • Group 1: Ellipse VR single-chamber ICD with a Durata lead in the right ventricle (RV)
  • Group 2: Fortify Assura single-chamber ICD, with an Optisure lead
Show all 13Show fewer
  • Group 3: Ellipse DR dual-chamber ICD with a Durata lead in the RV and a Tendril STS lead in the right atrium (RA)
  • Group 4: Fortify Assura ICD and an Optisure lead in the RV and an Isoflex lead in the RA
  • Be willing to undergo an elective MRI scan without sedation ≥ 45 days after implant NOTE: Antianxiety agents (e.g. minor tranquilizers, etc.) may be used as long as the patient can communicate with site personnel during the MRI scan
  • Be able to provide informed consent for study participation (legal guardian or legally authorized representative is NOT acceptable)
  • Be willing and able to comply with the prescribed follow-up tests and procedures
  • Are not contraindicated for an MRI scan (per the MRI Screening Form)
  • Subjects who are at least 18 years of age (or older, required by local law)

Who cannot join

  • Subjects who meet any of the following exclusion criteria must be excluded from the clinical investigation:
  • Have a competitor's MRI compatible endocardial lead implanted or capped
  • Have another existing active implanted medical device, e.g., neurostimulator, infusion pump, etc. that has MR labeling that will not allow the MRI scans per this protocol to be completed.
  • Have other non-MRI compatible device or material implanted. The following examples may be included as long as labelling of these devices allow MRI scans conducted per this protocol:
  • MRI compatible knee replacements, hip replacements, stents, etc.
  • MRI compatible mechanical, prosthetic, and bio prosthetic heart valves
Show all 11Show fewer
  • Non-removable dental implants
  • Have a lead extender, adaptor, or capped/abandoned lead
  • Enrolled or intend to participate in a clinical drug and/or device study (investigational device, investigational drug, new indication for a device or drug or additional testing beyond standard of care procedures), which could confound the results of this study as determined by Abbott.
  • Pregnant or planning to become pregnant during the duration of the subject's participation in the study
  • Have a life expectancy of less than 12 months due to any condition

Trial locations and contacts

1 site in 1 country.

China1 site
  • The First Affiliated Hospital of Zhejiang University

    Hangzhou, Zhejiang, 310000

Other trials in China

About this study

Official title: The Evaluation of ICD Defibrillation Systems in a 1.5T Magnetic Resonance Imaging Environment in China

This clinical study is to evaluate the safety and effectiveness of the Ellipse VR/DR implantable cardiac defibrillators (ICDs) and the Quadra Assura MP cardiac resynchronization therapy defibrillators (CRT-Ds), with Durata or Optisure defibrillation leads, Tendril STS or Isoflex pacing leads, and the Quartet quadripolar leads in a 1.5T MRI environment for MR-conditional labeling expansion of these market-approved ICD/CRT-D systems in China.

Note: Protocol updated: Fortify Assura VR/DR ICD removed; Quadra Assura MP CRT-D and Quartet leads added. The Detailed Description Section below was updated upon protocol amendment.

Read the detailed description

Endpoints, and contents of clinical study

Objectives:

The objective of this clinical study is to evaluate the safety and effectiveness of the Ellipse VR/DR implantable cardiac defibrillators (ICDs) and the Quadra Assura MP cardiac resynchronization therapy defibrillators (CRT-Ds), with Durata or Optisure defibrillation leads, Tendril STS or Isoflex pacing leads, and the Quartet quadripolar leads in a 1.5T MRI environment for MR-conditional labeling expansion of these market-approved ICD/CRT-D systems in China.

There are four device combination groups in this clinical study as outlined below:

  • Group 1: Ellipse VR single-chamber ICD with a Durata lead in the right ventricle (RV).
  • Group 2: Ellipse DR dual-chamber ICD, with a Durata lead in the RV and a Tendril STS lead in the right atrium (RA).
  • Group 3: Ellipse DR dual-chamber ICD with a Optisure lead in the RV and a Isoflex lead in the RA.
  • Group 4: Quadra Assura MP CRT-D with a Durata or Optisure lead in the RV, a Tendril STS or Isoflex lead in the RA, and a Quartet lead in the left ventricle (LV).

Endpoints

Primary Endpoints

  • Safety: Freedom from MRI-scan related complications related to the ICD device and/or leads from the time of the MRI scan to 1-month post-MRI scan testing.
  • Effectiveness: Proportion of leads with a capture threshold increase of ≤ 0.5V at 0.5ms for RA and RV leads and ≤1.0V at 0.5ms for LV leads from pre-MRI scan to 1-month post-MRI scan testing.
  • Effectiveness: Proportion of leads with a sensing amplitude decrease of ≤ 50% from pre-MRI scan testing to 1-month post-MRI scan.

Descriptive Endpoint(s)

Descriptive endpoints are reported using only summary statistics and no hypothesis tests will be performed.

The following data will be collected:

  1. Demographics: gender, age, ethnicity, race, cardiac disease history, arrhythmia history, indication for ICD implant, history of smoking, etc.
  2. Device electrical measurements at the MRI Scan Visit (pre- and post-scan) and at the 1 Month Post Scan Visit
  3. ADE, SADE, USADE
  4. Mortality
  5. Number of non-study MRI scans
  6. Summarize number of subjects that returned to usual programming after the MRI scan and number of subjects, if any, experiencing delays in reprogramming

This is a prospective, multi-center, single-arm study to evaluate the safety and effectiveness of the Ellipse VR/DR ICDs, and the Quadra Assura MP CRT-D, with Durata or Optisure defibrillation leads, Tendril STS or Isoflex pacing leads, and the Quartet quadripolar LV lead in a 1.5T MRI environment for MR-conditional labeling expansion in China.

This study will be conducted in at least 6 centers in China and at least 60 subjects who provide consent for participation will be enrolled. Subjects will be considered enrolled after providing consent. Subjects will be implanted with any one of the device/lead combination groups outlined in Figure 1, depending on their clinical need as determined by the Investigators. At least 15 subjects will be enrolled in each group.

  • Group 1 will have at least 15 enrolled subjects with the Ellipse VR single-chamber ICD with a Durata lead in the RV.
  • Group 2 will have at least 15 enrolled subjects with the Ellipse DR dual-chamber ICD with a Durata lead in the RV and a Tendril STS lead in the RA.
  • Group 3 will have at least 15 enrolled subjects with the Ellipse DR dual-chamber ICD with a Optisure lead in the RV and a Isoflex lead in the RA.
  • Group 4 will have at least 15 enrolled subjects with a Quadra Assura MP CRT-D and an Optisure or Durata lead in the RV, an Isoflex or Tendril STS lead in the RA, and a Quartet lead in the LV.

The enrollment group and group 4 lead pairing will be at the discretion of the Investigator, dependent on patient needs in conjunction with study needs to fulfill study goals.

At least forty-five (45) days after system implantation or after a system revision, subjects will have a baseline visit followed by an MRI visit; subjects will undergo a study-related MRI scan at 1.5T. Subjects will have a 1-month follow-up after the study MRI scan. The minimum duration of each subject's participation is approximately 2.5 months from enrollment. The expected duration of enrollment is approximately 14 months. The total duration of clinical study is expected to be 16.5 months.

The study population includes male and female Chinese nationals that meet the eligibility requirements and give consent for the study. Vulnerable subjects, such as minors or those unable to provide consent are excluded from participating.

Inclusion criteria

Assessment for general eligibility criteria is based on medical records of the site and interview with a candidate patient. If some of the clinical and laboratory tests are not included in site standard tests, they must be done but after written informed consent is obtained. Patients must meet ALL the inclusion criteria to be considered for the clinical study. If ANY of the exclusion criteria are met, the patient is excluded from the clinical study and cannot be enrolled.

To participate in this clinical study, the subject must meet all the following inclusion criteria:

  • Have an approved indication for implantation of an ICD, or CRT-D
  • Be a Chinese national
  • Will be implanted with one of the following device/lead combinations evaluated in this study:
  • Group 1: Ellipse VR single-chamber ICD with a Durata lead in the RV
  • Group 2: Ellipse DR dual-chamber ICD, with a Durata lead in the RV and a Tendril STS lead in the RA
  • Group 3: Ellipse DR dual-chamber ICD with a Optisure lead in the RV and a Isoflex lead in the RA
  • Group 4: Quadra Assura MP CRT-D and an Optisure or Durata lead in the RV, an Isoflex or Tendril STS lead in the RA, and a Quartet lead in the LV
  • Be willing to undergo an elective MRI scan without sedation ≥ 45 days after implant NOTE: Antianxiety agents (e.g. minor tranquilizers, etc.) may be used as long as the patient can communicate with site personnel during the MRI scan
  • Be able to provide informed consent for study participation (legal guardian or legally authorized representative is NOT acceptable)
  • Be willing and able to comply with the prescribed follow-up tests and procedures
  • Are not contraindicated for an MRI scan (per the MRI Screening Form)
  • Subjects who are at least 18 years of age (or older, required by local law)

Exclusion criteria

Subjects who meet any of the following exclusion criteria must be excluded from the clinical investigation:

  • Have a competitor's MRI compatible endocardial lead implanted or capped
  • Have another existing active implanted medical device, e.g., neurostimulator, infusion pump, etc. that has MR labeling that will not allow the MRI scans per this protocol to be completed.
  • Have other non-MRI compatible device or material implanted. The following examples may be included as long as labelling of these devices allow MRI scans conducted per this protocol:
  • MRI compatible knee replacements, hip replacements, stents, etc.
  • MRI compatible mechanical, prosthetic, and bio prosthetic heart valves
  • Non-removable dental implants
  • Have a lead extender, adaptor, or capped/abandoned lead
  • Enrolled or intend to participate in a clinical drug and/or device study (investigational device, investigational drug, new indication for a device or drug or additional testing beyond standard of care procedures), which could confound the results of this study as determined by Abbott.
  • Pregnant or planning to become pregnant during the duration of the subject's participation in the study
  • Have a life expectancy of less than 12 months due to any condition

Effectiveness evaluation method

There are two effectiveness endpoints to be evaluated in this study.

  1. Description of the effectiveness endpoints
  • Proportion of leads with a capture threshold increase of ≤ 0.5V at 0.5ms for RA and RV leads and ≤ 1.0V at 0.5ms for LV leads from pre-MRI scan to 1-month post-MRI scan testing.
  • Proportion leads with a sensing amplitude decrease of ≤ 50% from pre-MRI scan testing to 1-month post-MRI scan.
  1. Selection of method and time for evaluation, recording and analysis of the effectiveness endpoints Since the objective is to evaluate the effect of MRI exposure for the ICD/ CRT-D device groups included in this study, the effectiveness parameters were selected to compare abnormal changes in device performance from pre- to 1-month post- MRI scan. The pre-MRI measurements are collected within the same day prior to the MRI scan. The 1-month post-MRI scan measurements are collected at 30 days post-MRI scan (within +14 days) to ensure any effect of the MRI exposure is captured.

The effectiveness parameters are recorded directly from the device. The device measurements will be obtained using the Merlin programmer to interrogate the device during study visits. The analysis of the effectiveness parameters will be conducted on all subjects with complete data from pre- to 1-month post-MRI scan.

Safety evaluation method There is one safety primary endpoint to be evaluated in this study.

  1. Description of the safety endpoint Freedom from MRI-scan related complications from the time of the MRI scan to 1-month post-MRI scan testing.
  2. Selection of method and time for evaluation, recording and analysis of the safety endpoints Since the objective is to evaluate the effect of MRI exposure for the ICD/ CRT-D device groups included in this study, the safety parameters were selected to include only MRI-scan related complications from the time of device programming for an MRI scan to 1-month post-MRI scan. An independent physician CEC will adjudicate all adverse device effects during this period to determine the relatedness to the MRI scan.

Safety surveillance within this study and the safety reporting, both performed by the investigator, starts as soon as the subject completes the pre-MRI visit testing and the MRI Settings have been programmed according to the protocol. The safety surveillance and the safety reporting will continue until the last study visit has been performed, the subject is deceased, the subject/investigator concludes his participation into the clinical study or the subject withdrawal from the clinical study.

Monitoring plan

Sponsor and/or designee will monitor the clinical study over its duration according to the CIP-specific monitoring plan which will include the planned extent of source data verification.

Prior to initiating any procedure, the Sponsor monitor (or delegate) will ensure that the following criteria are met:

  • The investigator understands and accepts the obligation to conduct the clinical study according to the CIP and applicable regulations, and has signed the Clinical Trial Agreement.
  • The Investigator and his/her staff should have sufficient time and facilities to conduct the clinical study and should have access to an adequate number of appropriate subjects to conduct the clinical study.
  • Source documentation (including original medical records) must be available to substantiate proper informed consent procedures, adherence to CIP procedures, adequate reporting and follow-up of adverse events, accuracy of data collected on case report forms, and device information.
  • The Investigator/site will permit access to such records. A monitoring visit sign-in log will be maintained at the site. The Investigator will agree to dedicate an adequate amount of time to the monitoring process. The Investigator and/or research coordinator will be available for monitoring visits. It is expected that the Investigator will provide the monitor with a suitable working environment for review of clinical study-related documents.

Statistical considerations

This study is an observational study to meet the requirements for device approval by the National Medical Products Administration (NMPA, previously China FDA). There is no endpoint hypotheses testing planned in this study. The endpoints, demographics, safety and device data will be summarized using descriptive statistics as outlined in the Statistical Analysis Plan (SAP).

Any major changes to the SAP will be documented in an amendment to the SAP. Less significant changes to the planned analyses will be documented in the final report(s).

Data management

Study data will be handled according to China Device GCP (Order 25). The Leading PI and his/her associated organization (leading Unit) will be responsible for compiling and submitting all required reports to governmental agencies.

Before the submission, Leading PI and his/her associated organization (leading Unit) should allow sponsor to have a copy of the above reports for reviewing, and providing comments for Leading PI and his/her associated organization consideration.

Data will be analyzed by the Leading PI and his/her associated organization (leading Unit) and/or its affiliates. The Leading PI and his/her associated organization (leading Unit)'s also agree that data may be transferred to the Sponsor's locations worldwide and/or any other worldwide regulatory authority in support of a market-approval application.

Conditions
Tachycardia
Being tested
Study MRI scan
Medical subject headings
Tachycardia

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Study MRI scanOther

    At least forty-five (45) days after system implantation or after a system revision, Subjects will undergo a study-related MRI scan at 1.5T.

Study groups in NCT03828357
Study groupRoleWhat they receive
Ellipse VR with DurataEllipse VR single-chamber ICD with a Durata defibrillation lead
Receives: Other: Study MRI scan
Ellipse DR with Durata & Tendril STSEllipse DR dual-chamber ICD with a Durata lead in the right ventricle (RV) and a Tendril STS lead in the right atrium (RA).
Receives: Other: Study MRI scan
Ellipse DR with Optisure and IsoflexEllipse DR dual-chamber ICD with an Optisure defibrillation lead in the right ventricle (RV) and an Isoflex lead in the right atrium (RA).
Receives: Other: Study MRI scan
Quadra Assura MP CRT-DQuadra Assura MP CRT-D with an Optisure or Durata defibrillation lead in the RV, an Isoflex or Tendril STS lead in the RA, and a Quartet lead in the LV
Receives: Other: Study MRI scan

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Number of Participants With Freedom From Magnetic Resonance Imaging (MRI) Scan Related Complications

    Freedom from Magnetic Resonance Imaging (MRI) scan related complications related to the Implantable Cardiac Difibrillator (ICD) device and/or leads from the time of the MRI scan to 1-month post-MRI scan testing.

    Measured: 1 month from MRI scan

  2. Effectiveness Outcomes: Percentage of Leads With a Capture Threshold Increase

    Proportion of leads with a capture threshold increase of ≤ 0.5V at 0.5ms from pre-MRI scan to 1-month post-MRI scan testing.

    Measured: pre-MRI scan to 1 month post-MRI scan

  3. The Percentage of Leads With a Sensing Amplitude Decrease of ≤ 50% From Pre-MRI Scan Testing to 1-month Post-MRI Scan.

    Proportion of subjects with a sensing amplitude decrease ≤ 50% from pre-MRI scan to 1-month post MRI scan % (n/N)

    Measured: pre-MRI scan to 1 month post-MRI scan

Secondary outcomes (3)
  1. Device Electrical Measurements

    Summaries of electrical measurements taken in the Analysis Cohort at Pre-MRI scan, Post-MRI scan, and 1-month post MRI scan. The stable electrical measurements from pre-MRI scan to post-MRI scan and 1-month post MRI scan indicate that the MRI scan did not impact the electrical performance of the Ellipse ICD or Quadra Assura CRT-D system

    Measured: Pre-MRI scan, Post MRI scan, 1 month follow-up

  2. Number of Participants With Whole Body (WB) SAR Values Measured ≥ 1.8 W/kg

    Whole body (WB) SAR values from the RF intensive MRI scans.

    Abbott is seeking MR-conditional labeling for MRI scans in Normal Operating Mode (WB Specific Absorption Rate (SAR) up to 2 W/kg). Achieving a WB SAR of exactly 2 W/kg is not feasible since the MR technologist must adjust multiple interdependent scab parameters to increase the WB SAR. Therefore, an acceptable limit of ≥ 1.8 W/kg (within 10%) was established as part of the MRI scan sequence guideline document for this study prior to site enrollments to account for expected user variability.

    Measured: during MRI scan

  3. Summary of Subjects That Returned to Usual Programming

    Subjects that returned to usual programming after the MRI scan and the number of subjects experiencing delays in disabling MRI settings after the MRI scan

    Measured: MRI scan visit

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted4 Feb 2019
  2. Study start27 Mar 2019
  3. Primary completion1 Dec 2020
  4. Study completion1 Dec 2020

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Observational
Observational model
Case only
Time perspective
Prospective
Enrolment
60 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT03828357
Sponsor's protocol ID
CRD_930
Acronym
CN_MRI_ICD
Status last verified
Mar 2023
First submitted
8 Jan 2019
Last updated
7 Dec 2023
Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
No
Individual data sharing
No

Questions about this trial

What is this clinical trial studying?

An observational study of Tachycardia, run by Abbott Medical Devices at 1 site in China.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 13 inclusion and 11 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 7 Dec 2023. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in China. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

Yes. Results are posted in the registry (first posted 7 Dec 2023). A summary is in the For doctors view, and the full results are on ClinicalTrials.gov.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Number of Participants With Freedom From Magnetic Resonance Imaging (MRI) Scan Related Complications (time frame: 1 month from MRI scan). The study lists 3 primary outcomes in all. It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.