CataractNCT03500198

Clinical Investigation of the Next Generation Intraocular Lens

A study testing Investigational Intraocular Lens Device #1: Next Generation TECNIS IOL in people with Cataract, run by Johnson & Johnson Surgical Vision, Inc. at 10 sites in United States.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

An investigational IOL designed to provide improved intermediate and near visual acuity and decreased spectacle wear was evaluated for safety and effectiveness in a six-month study. Read more about this study

Who is it for?
Ages 22 years and olderHealthy volunteers can take part
Conditions the study is for
Cataract
What is being tested?
Investigational Intraocular Lens Device #1: Next Generation TECNIS IOL
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
247 participants
Where
10 sitesUnited States
Study dates
28 Jun 2018to 5 Jun 2019
Phase
Interventional
Are there results?
Yes, posted in the registrySummary in the doctors' view

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 22 years and older
  • All sexes
  • Healthy volunteers accepted
  • Adult, Older adult

Who may be able to join

  • Minimum 22 years of age
  • Bilateral cataracts for which posterior chamber IOL implantation has been planned
  • Corneal astigmatism:
  • Normal corneal topography
  • Availability, willingness and sufficient cognitive awareness to comply with examination procedures and study visits
  • Signed informed consent and HIPAA authorization or equivalent documentation necessary to comply with applicable privacy laws pertaining to medical treatment in the governing countries
  • Ability to understand and respond to a questionnaire in English

Who cannot join

  • Irregular corneal astigmatism
  • Inability to focus or fixate for prolonged periods of time (e.g., due to strabismus, nystagmus, etc.)
  • Prior corneal refractive (LASIK, LASEK, RK, PRK, etc.) or intraocular surgery, Including prophylactic peripheral iridotomies and peripheral laser retinal repairs
  • Poorly-controlled diabetes
  • Patient is pregnant, plans to become pregnant, is lactating or has another condition associated with the fluctuation of hormones that could lead to refractive changes
  • Concurrent participation or participation within 60 days prior to preoperative visit in any other clinical trial
  • Desire for monovision correction

Trial locations and contacts

10 sites in 1 country.

10 sites in 1 country

United States10 sites
  • The Eye Associates of Manatee, LLP

    Bradenton, Florida, 34209

  • Cincinnati Eye Institute

    Cincinnati, Ohio, 45242

  • Key-Whitman Eye Center

    Dallas, Texas, 11442

  • Berkeley Eye Institute, P.A.

    Houston, Texas, 77027

  • Slade & Baker Vision

    Houston, Texas, 77027

  • Levenson Eye Associates, Inc.

    Jacksonville, Florida, 32204

  • Eye Care Specialists

    Kingston, Pennsylvania, 18704

  • Jones Eye Clinic

    Sioux City, Iowa, 51104

  • Vance Thompson Vision Clinic, Prof LLC

    Sioux Falls, South Dakota, 57108

  • Wolstan & Goldberg Eye Associates

    Torrance, California, 90505

Other trials in United States

About this study

Official title: Clinical Investigation of the Safety and Effectiveness of the Next-Generation TECNIS® Symfony® Intraocular Lens Model ZHR00

An investigational IOL designed to provide improved intermediate and near visual acuity and decreased spectacle wear was evaluated for safety and effectiveness in a six-month study.

Read the detailed description

The next generation TECNIS IOL designed to provide improved intermediate and near visual acuity and decreased spectacle wear was evaluated for safety and effectiveness in a six-month interventional study and compared to a marketed monofocal IOL. Assessments included distance-corrected intermediate and near visual acuity, depth of focus at 0.2 logMAR and spectacle wear frequency.

Conditions
Cataract
Being tested
Investigational Intraocular Lens Device #1: Next Generation TECNIS IOL, Control Monofocal Intraocular Lens: TECNIS Monofocal IOL
Medical subject headings
Cataract

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Investigational Intraocular Lens Device #1: Next Generation TECNIS IOLDevice

    Intraocular lens replaces the natural lens removed during cataract surgery.

  • Control Monofocal Intraocular Lens: TECNIS Monofocal IOLDevice

    Intraocular lens replaces the natural lens removed during cataract surgery.

Study groups in NCT03500198
Study groupRoleWhat they receive
Investigational Device: Next Generation TECNIS IOLExperimentalInvestigational Intraocular Lens Device #1: Next Generation TECNIS IOL
Receives: Device: Investigational Intraocular Lens Device #1: Next Generation TECNIS IOL
Control Device: TECNIS Monofocal IOLActive comparatorControl Monofocal Intraocular Lens: TECNIS Monofocal IOL
Receives: Device: Control Monofocal Intraocular Lens: TECNIS Monofocal IOL

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Monocular, Photopic Distance Corrected Intermediate Visual Acuity (DCIVA) at 66 Centimeter (cm)

    Visual Acuity (VA) was assessed monocularly (each eye separately) under photopic (well-lit) conditions using best distance correction (distance refraction) and high contrast, Early Treatment Diabetic Retinopathy Study (ETDRS) chart at 66 cm from spectacle plane. It was measured in logarithm of the minimum angle of resolution (logMAR), on an ETDRS chart. A lower numeric value represented better VA. This analysis was pre-specified for the first operative eye only.

    Measured: 6 months (postoperative)

  2. Monocular Photopic Distance Corrected Depth of Focus Assessed by the Mean Defocus Curve

    Depth of focus was assessed using best corrected distance refraction, and 100% contrast ETDRS letters. VA was measured between +2.00 Diopter (D) and -4.00 D in 0.5 D defocus steps. VA was measured in logMAR. The depth of focus was estimated as the dioptric range between zero defocus and the first point on the negative lens induced mean defocus curve that crosses the 0.2 logMAR using a linear interpolation. A lower numeric value represents better VA. This analysis was pre-specified to be reported for the first operative eye only.

    Measured: 6 months (postoperative)

Secondary outcomes (2)
  1. Monocular, Photopic Distance Corrected Near Visual Acuity (DCNVA) at 40 cm

    Visual acuity was tested monocularly under photopic conditions using best distance correction (distance refraction) and high contrast, ETDRS chart set at 40 cm from the spectacle plane using the near point rod. VA was measured in logMAR on an ETDRS chart. A lower numeric value represented better VA. This analysis was pre-specified to be reported for the first operative eye only.

    Measured: 6 months (postoperative)

  2. Overall Spectacle Wear

    Overall spectacle wear was determined for participants wearing glasses or contact lenses using Patient-Reported Spectacle Independence Questionnaire (PRSIQv2) asking if their response is 'none of the time' in all the 4 conditions (distance vision, intermediate vision, near vision and overall vision) of the spectacle wear questions were reported. PRSIQv2 questionnaire responses were divided into two response groups: reports of "none of the time" and the combined reports of "a little of the time," "some of the time," "most of the time," or "all of the time".

    Measured: 6 months (postoperative)

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted18 Apr 2018
  2. Study start28 Jun 2018
  3. Primary completion5 Jun 2019
  4. Study completion5 Jun 2019

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Treatment
Allocation
Randomised
Intervention model
Parallel groups
Masking
Double blind
Who is masked
Participant, Outcomes assessor
Enrolment
247 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT03500198
Sponsor's protocol ID
SUR-IOL-652-2001
Status last verified
Jun 2020
First submitted
9 Apr 2018
Last updated
25 Jun 2020
Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
Yes
Individual data sharing
No

Questions about this trial

What is this clinical trial studying?

A study testing Investigational Intraocular Lens Device #1: Next Generation TECNIS IOL in people with Cataract, run by Johnson & Johnson Surgical Vision, Inc. at 10 sites in United States.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 7 inclusion and 7 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 25 Jun 2020. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 10 sites in United States. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

Yes. Results are posted in the registry (first posted 25 Jun 2020). A summary is in the For doctors view, and the full results are on ClinicalTrials.gov.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Monocular, Photopic Distance Corrected Intermediate Visual Acuity (DCIVA) at 66 Centimeter (cm) (time frame: 6 months (postoperative)). The study lists 2 primary outcomes in all. It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.