AzTrials
AzTrials

Port-site Injury Caused by Stray Energy TransferNCT03477201

Characterizing Stray Energy Injuries During Robotic Surgery

A study testing Laparoscopic robotic DaVinci assisted inguinal hernia repair and Standard laparoscopic inguinal hernia repair in people with Port-site Injury Caused by Stray Energy Transfer, run by University of Colorado, Denver at 1 site in United States.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

Stray energy transfer during laparoscopic surgery is recognized as a cause of potentially serious complications. This will be the first study to demonstrate clinical evidence of tissue injury due to stray energy transfer during robotic surgery. Read more about this study

Who is it for?
Any age, men only
What is being tested?
Laparoscopic robotic DaVinci assisted inguinal hernia repair and Standard laparoscopic inguinal hernia repair
Type of treatment
Surgery or procedure
Could I receive a placebo?
No, everyone receives an active treatment
How many people
42 participants
Where
1 siteUnited States
Study dates
1 Feb 2019to 1 Apr 2023
Phase
Interventional
Are there results?
None posted in the registry

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages Any age
  • Men only
  • No healthy volunteers
  • Child, Adult, Older adult

Who may be able to join

  • Patients will be recruited in the general surgery clinic once the presence of an operable, inguinal hernia has been identified.
  • All patients with a unilateral, symptomatic, inguinal hernia, and
  • Who are surgical candidates for a laparoscopic inguinal hernia repair.

Who cannot join

  • Under 18 years,
  • Emergent hernia repairs,
  • Recurrent hernias,
  • Bilateral hernias, and
  • Incarcerated bowel.

Trial locations and contacts

1 site in 1 country.

United States1 site
  • RMR Denver VA

    Aurora, Colorado, 80016

Other trials in United States

About this study

Stray energy transfer during laparoscopic surgery is recognized as a cause of potentially serious complications. This will be the first study to demonstrate clinical evidence of tissue injury due to stray energy transfer during robotic surgery. This information can then be used to define surgeon modifiable factors that can reduce the risk of patient injury. In addition, these data can guide the development of future robotic and laparoscopic platforms.

Read the detailed description

BACKGROUND: Stray energy transfer from monopolar instruments during laparoscopic surgery is recognized as a cause of potentially catastrophic complications. The investigators have published multiple studies on the varied mechanisms of stray energy transfer during laparoscopic procedures. In addition, a preliminary study by the investigators has confirmed stray energy transfer during robotic surgery in an in vivo model. However, there are no data available on the potential clinical impact of stray energy in robotic surgery. Furthermore, there are no studies directly comparing stray energy transfer between laparoscopic and robotic procedures.

PURPOSE: The purpose of this study is to assess for thermal injury due to stray energy transfer during elective laparoscopic and robotic inguinal hernia repairs.

HYPOTHESIS: The Investigators hypothesize that thermal injury to the skin surrounding the working ports occurs during robotic and laparoscopic procedures. Based on prior data, the Investigators hypothesize these injuries will occur more frequently at the camera port during laparoscopy, and at the assistant port during robotic surgery.

METHODS: The Investigators plan to conduct a prospective, randomized controlled trial of patients undergoing elective laparoscopic or robotic unilateral inguinal hernia repair. Patients will be randomized into two groups, with one group undergoing laparoscopic transabdominal preperitoneal (TAPP) technique and a second group undergoing robotic inguinal hernia repair (TAPP technique) with the Xi DaVinci System (Intuitive, Sunnyvale, CA). The monopolar instrument will deliver via standard instruments on 30W coagulation mode (ForceTriad electrosurgical generator, Covidien, Boulder, CO). At the completion of the procedure, skin biopsies will be taken at each of the three port incisions. These samples will be examined for visual and histologic evidence of thermal injury by a blinded pathologist.

IMPORTANCE: This will be the first study to demonstrate clinical evidence of tissue injury due to stray energy transfer during robotic surgery. The inclusion of a laparoscopic arm will allow confirmation of prior study findings as well as direct comparison of stray energy transfer in both modalities. This information can then be used to define surgeon modifiable factors that can reduce the risk of patient injury. In addition, these data can guide the development of future robotic and laparoscopic platforms.

Being tested
Laparoscopic robotic DaVinci assisted inguinal hernia repair, Standard laparoscopic inguinal hernia repair

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Laparoscopic robotic DaVinci assisted inguinal hernia repairProcedure or surgery

    Also called Dissection use of monopolar energy provided bt ForceTriad System (Covidien)

    Laparoscopic inguinal hernia repair using a robotic assisted laparoscopic procedure.

  • Standard laparoscopic inguinal hernia repairProcedure or surgery

    Also called Dissection use of monopolar energy provided by ForceTriad System (Covidien)

    Laparoscopic inguinal hernia repair, an accepted safe method of repairing inguinal hernia.

Study groups in NCT03477201
Study groupRoleWhat they receive
Laparoscopic robotic DaVinci assisted inguinal hernia repairExperimentalIntervention: 30 patients will undergo a robotic assisted laparoscopic inguinal hernia repair. This will be done using the DaVinci Robotic Platform by Intuitive Surgical. This an accepted safe method of repairing inguinal hernia. This platform uses special robotic ports produced and supplied by Intuitive Surgical required to dock the machine to the patient. Monopolar energy will be provided by a ForceTriad system (Covidien, Boulder, CO), standard an common system used in most ORs. This will be used for dissection. The investigators will obtain small biopsies from port site to assess stray energy transfer injury, a model described by earlier studies.
Receives: Procedure: Laparoscopic robotic DaVinci assisted inguinal hernia repair
Standard Laparoscopic inguinal hernia repairActive comparatorIntervention: 30 patients will undergo a laparoscopic inguinal hernia repair, an accepted safe method of repairing inguinal hernia. This platform uses standard laparoscopic ports. In our institution we use plastic ports made by Covidien, Boulder, CO. The operation will require two 5mm VersaPort (Covidien) and a Hassan Port. As in the robotic arm, monopolar energy will be provided by a ForceTriad system (Covidien, Boulder, CO), standard an common system used in most ORs. This will be used for dissection. The investigators will obtain small biopsies from port site to assess stray energy transfer injury, a model described by earlier studies.
Receives: Procedure: Standard laparoscopic inguinal hernia repair

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Histologic analysis of port-site skin biopsies

    Samples will be taken at the time of surgery from each of the three ports and analyzed by a board certified pathologist to assess for injury on histologic level.

    Measured: Biopsies will be taken right away and analysed within 1-2 weeks.

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted26 Mar 2018
  2. Study start1 Feb 2019
  3. Primary completion1 Apr 2023
  4. Study completion1 Apr 2023

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Prevention
Allocation
Randomised
Intervention model
Parallel groups
Model description
The investigators plan to conduct a prospective, randomized controlled trial of patients undergoing elective laparoscopic or robotic unilateral inguinal hernia repair. Patients will be randomized into two groups, with one group undergoing laparoscopic transabdominal preperitoneal (TAPP) technique and a second group undergoing robotic inguinal hernia repair (TAPP technique) with the Xi DaVinci System (Intuitive, Sunnyvale, CA). The monopolar instrument will deliver via standard instruments on 30W coagulation mode (ForceTriad electrosurgical generator, Covidien, Boulder, CO). At the completion of the procedure, skin biopsies will be taken at each of the three port incisions. These samples will be examined for visual and histologic evidence of thermal injury by a blinded pathologist.
Masking
Single blind
Who is masked
Participant
Enrolment
42 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT03477201
Sponsor's protocol ID
18-0527
Status last verified
Jan 2024
First submitted
9 Mar 2018
Last updated
1 Feb 2024
Results posted
No
Data monitoring committee
Yes
FDA-regulated drug
No
FDA-regulated device
No
Individual data sharing
No: There is not a plan to make IPD available.

Questions about this trial

What is this clinical trial studying?

A study testing Laparoscopic robotic DaVinci assisted inguinal hernia repair and Standard laparoscopic inguinal hernia repair in people with Port-site Injury Caused by Stray Energy Transfer, run by University of Colorado, Denver at 1 site in United States.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 3 inclusion and 5 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 1 Feb 2024. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in United States. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Histologic analysis of port-site skin biopsies (time frame: Biopsies will be taken right away and analysed within 1-2 weeks.). It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 1 Feb 2024.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in