Clinical Accuracy of Vital Sign Pressure SensorNCT03321851

V-Sensors for Vitals: Assessment of the Accuracy of a Vital Sign Sensor Mounted on a Smartphone

A study testing V-sensor device user in people with Clinical Accuracy of Vital Sign Pressure Sensor, run by Léman Micro Devices SA.

Unknown status

Status not confirmed. The sponsor has not verified this record for over two years, so it may be out of date.

Official record
Search all trials

Trial summary

This trial seeks to demonstrate clinical accuracy and safety of the integrated V-Sensor when used by the general, healthy adult population interested in measuring their vital signs outside of the clinical setting. Read more about this study

Who is it for?
Ages 18 to 80 yearsHealthy volunteers can take part
What is being tested?
V-sensor device user
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
95 participants
Where
No sites listed yet
Study dates
Apr 2018to Sep 2018
Phase
Interventional
Are there results?
None posted in the registry

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 to 80 years
  • All sexes
  • Healthy volunteers accepted
  • Adult, Older adult

Who may be able to join

  • Informed Consent
  • 18 and 80 years old
  • no history or risk for developing lymphedema
  • no arteriovenous fistula in the right arm
  • no right limb amputation up to and including the right index finger
  • ability to obtain 3 readings from the device during the screening/training session.

Who cannot join

  • history of arrhythmia
  • arteriovenous fistula in the right arm
  • lymphedema or high risk for developing lymphedema in the right arm
  • inability to follow the procedures of the study
  • inability to hold the study device as instructed
  • inability to obtain 3 readings from the study device during the screening session
  • previous enrolment into the current study
  • enrolment of the Investigator or Sponsor, his/her family members, employees and other dependent persons.

Trial locations and contacts

No sites are listed yet.

Kelly A McClary, RN contact

+41 (0)797764954

km@leman-micro.com

Chris J Elliott, PhD contact

+41(0)791733306

cje@leman-micro.com

About this study

Official title: An Open Label, Interventional, Clinical Performance and Validation Study to Demonstrate the Clinical Accuracy of of the V-Sensor in the Healthy Adult Population; Single Site in Switzerland

This trial seeks to demonstrate clinical accuracy and safety of the integrated V-Sensor when used by the general, healthy adult population interested in measuring their vital signs outside of the clinical setting.

Read the detailed description

The five vital signs are blood pressure, pulse, respiration, SpO2, and temperature. The user tests two devices mounted on two smartphones.

  • Temperature: Each user will use the V-sensor mounted on a smartphone to obtain his/her temperature interfaced with a temperature specific 'app'. Before and after the the temperature is obtained using the V-Sensor, the temperature will be obtained using an equivalent thermometer.
  • Blood Pressure: Each user will test his/her blood pressure using the two sensors mounted on the two phones interfaced with a blood pressure specific 'app'. Before and after each measurement, a reference measurement using a mercury sphygmomanometer will be taken.
  • Pulse, Respiration, SpO2: Each user will test his/her pulse, respiration, and SpO2 using 2 V-sensors mounted on two separate smartphones interfaced with a Pulse, Respiration, and SpO2 'app'. Before and after each measurement, a reference measurement using equivalent reference devices will be taken.
Being tested
V-sensor device user

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • V-sensor device userDevice

    Participants use device to detect the 5 vital signs (Blood Pressure, Pulse, Respiration, SpO2, Temperature).

Study groups in NCT03321851
Study groupRoleWhat they receive
V-Sensor Device UserOtherThis diagnostic medical device is designed to detect the user's 5 vital signs. Each user tests two sensors, each sensor is mounted on a smartphone. Each vital sign is compared to an equivalent reference device obtained by the healthcare provider.
Receives: Device: V-sensor device user

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Blood Pressure accuracy to the reference device within the ranges specified by the ISO standard 81060-2

    For systolic and diastolic blood pressures, the mean error of determination of the individual paired determinations of the sphygmomanometer-under-test and the reference sphygmomanometer (Standby Baumanometer) for all participants shall not be greater than 8.0mmHg.

    Measured: 4 months

  2. Pulse Rate accuracy to the reference device within the ranges specified by the ISO 80601-2-61

    Pulse rate accuracy shall be stated as the root-mean-square (rms) difference between paired pulse rate data recorded with the pulse oximeter equipment and with the Nellcor N-PA reference device.

    Measured: 4 months

  3. Blood Oxygen accuracy to the reference device within the ranges specified by the ISO 80601-2-61.

    The SpO2 accuracy shall be stated as a root-mean-square (rms) difference of less than or equal to 4.0% SpO2 over the range of 70% to 100% SaO2

    Measured: 4 months

  4. Respiration rate accuracy

    Respiration rate accuracy shall be stated as the root-mean-square (rms) difference between paired respiratory rate data with the Passport Respiration Rate Sensor PS-2133 reference method; target range +/- 10% of participant's respiratory rate

    Measured: 4 months

  5. Temperature accuracy to the reference device with the ranges specified by the ISO 80601-2-56 & ATSM 1965 Standard.

    Temperature accuracy shall be stated as the root-mean-square (rms) difference between paired temperatures data recorded with the Exergen Temporal Scanner thermometer. LMD's target is a clinical bias <= 3.0°C and a clinical repeatability <= 3.0°C.

    Measured: 4 months

Secondary outcomes (1)
  1. No serious adverse device effects are reported.

    Zero serious adverse device effects are reported.

    Measured: 4 months

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted26 Oct 2017
  2. Study startApr 2018 (estimated)
  3. Primary completionAug 2018 (estimated)
  4. Study completionSep 2018 (estimated)

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Diagnosis
Allocation
Not applicable
Intervention model
Single group
Model description
Each participant tests two equivalent vital sign sensors (devices under test) mounted on two different smartphones, one after another to assess accuracy of the device compared to equivalent reference devices.
Masking
None (open label)
Enrolment
95 participants (estimated)
IDs, dates and publications
ClinicalTrials.gov ID
NCT03321851
Sponsor's protocol ID
LMD-CT-002
Status last verified
Oct 2017
First submitted
12 Oct 2017
Last updated
6 Mar 2018
Results posted
No
Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
No
Individual data sharing
No

Questions about this trial

What is this clinical trial studying?

A study testing V-sensor device user in people with Clinical Accuracy of Vital Sign Pressure Sensor, run by Léman Micro Devices SA.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 6 inclusion and 8 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: status not confirmed, as of the sponsor's last update on 6 Mar 2018. Status can change between updates, so confirm with the study contact before making plans.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

Results are not available in the source record.

How can a doctor refer a patient or contact the research team?

Contact Kelly A McClary, RN (+41 (0)797764954, km@leman-micro.com), or a site near the patient. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Blood Pressure accuracy to the reference device within the ranges specified by the ISO standard 81060-2 (time frame: 4 months). The study lists 5 primary outcomes in all. It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 6 Mar 2018.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in