TemperatureNCT03264833
Accuracy and Precision Comparison of Temptraq
An observational study of Temperature, run by The Cleveland Clinic.
Trial summary
Purpose The purpose of this feasibility study is to assess the accuracy and precision of the Temp-Traq thermometer for monitoring body temperature in adult patients under three conditions (hypothermia, normothermia, and hyperthermia). Specific Aims 1. Read more about this study
- Who is it for?
- Ages 21 years and older
- Conditions the study is for
- Temperature
- What is being tested?
- TempTraq
- How many people
- 60 participants
- Where
- No sites listed yet
- Study dates
- 15 Dec 2016to 1 Oct 2017
- Phase
- Observational
- Are there results?
- Yes, posted in the registrySummary in the doctors' view
Key requirements
Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.
- Ages 21 years and older
- All sexes
- Adult, Older adult
Study population: To address the specific aims of this study the investigators will identify at least 40 patients who have a pulmonary artery (PA) catheter in place (patients in intensive care units).
Who may be able to join
- Inclusion Criteria Adult patient (male or female) in intensive care unit who will have a PA catheter in place for at least the next 8 hours; and who have no visible skin condition to the axillary region upon inspection by the research nurse are eligible for this study
Who cannot join
- Patients younger than 21
- Patients who will not have a PA catheter in place for the next 8 hours
- Patients who have a visible skin condition to the axillary region upon inspection by the research nurse
- Patients who will be going for procedures or for some reason will not be available for the 6 hours of the study
About this study
Official title: Accuracy and Precision Test of TempTraq® Compared to Pulmonary Artery Catheter for Monitoring Temperature in Adults in Intensive Care Unit
Purpose The purpose of this feasibility study is to assess the accuracy and precision of the Temp-Traq thermometer for monitoring body temperature in adult patients under three conditions (hypothermia, normothermia, and hyperthermia).
Specific Aims
- To assess the accuracy of the Temp-Traq thermometer as compared to a gold standard (Core temp measured by Pulmonary Artery Catheter)
- To assess the precision of the Temp-Traq thermometer over repeated measures
- To determine if accuracy \&/or precision is consistent in three conditions (hypothermia, normothermia, or hyperthermia)
Read the detailed description
Design A repeated measures within-group comparative design will be used for this study.
Sample/Setting To address the specific aims of this study the investigators will identify patients who have a pulmonary artery (PA) catheter in place (patients in intensive care units). Letters of support are attached from the unit managers where the study will take place (see attached).
Procedure Once a patient is identified, the researcher will record basic demographic data (see data collection form attached) and the axillary area will be assessed for visible signs of any skin conditions. If there are no visible signs of a skin condition the researcher will apply the Temp-Traq thermometer to the right or left axilla of the subject and record the location of the placement. The researcher will record both the PA and Temp-Traq recordings of subject's temperature (taken at the same time) on the data collection form (see attachment) at four data points; baseline (5 minutes post application of Temp-Traq) and every two hours (+/- 15 minutes) after baseline X 3. Previous testing of the TempTraq® on children and healthy adults has not resulted in any adverse skin reactions. However, as this is a new thermometer format and there is not a significant body of literature on potential skin reactions, the investigators will assess skin before and after application. Thus, at the end of the last reading the Temp-Traq thermometer will be removed and the skin will be assessed for erythema. If erythema is present the patient's nurse and physician will be notified by the researcher.
Protection of Human Subjects Since non-invasive temperature monitoring poses no risk to patients and is part of usual care and since the investigators will not collect or record any protected health information (PHI), written informed consent will not be sought. Patients will be identified by their study ID number only. The data will be collected by a registered nurse data collector familiar with the intensive care setting and data will be stored in a locked file cabinet in the researchers locked office. Temperature data for the Temp-Traq is collected through the Temp-Traq Application (AP) which will be placed on a Cleveland Clinic Approved and encrypted i-pad. Once the study is completed and all data has been transferred to the study database, the Temp-Traq application and all of its data will be deleted from the i-pad.
Analysis The investigators will use the TOST (Two One Sided t-tests) method for equivalence testing. This method requires a defined range of mean differences between two test methods, an estimate of the precision of the measurement of the two systems, and an estimate of the size of the possible difference in the means of the two methods under consideration. The TOST null hypothesis is a joint null hypothesis that the mean measurement differences between the two methods is greater than a critical lower bound and less than a critical upper bound. If the null is rejected then the investigators can conclude that the absolute difference of the means between Temp Traq and PA catheter falls within the specified range. It was determined that to be considered equivalent with respect to accuracy the mean measurements of the two methods should be within ± 0.2 degrees of each other. Precision tests for Temp-Traq thermometer using accepted ASTM test methods for various combinations of temperature and humidity provided a range of measurement variation between .0000435 and .019928 with a mean of 0.000254500. If the investigators assume both test methods exhibit the same levels of precision then the variance of the differences between their two means will reduce to two times the values listed previously. This will result in estimates of the standard deviations of the mean differences of .0093, .032, and .063 respectively.
The two methods for body temperature that will be examined are core temperature measurements with a PA catheter and temperature measurements using the Temp-Traq (measured in degree Celsius simultaneously). Paired temperature measurements for both methods will be taken simultaneously every 2 hours for a 6 hour period. This will provide 4 repeated temperature measures per patient. The simultaneous measurements within a given patient should exhibit a high degree of correlation. If .7 is chosen as the lower bound for the correlation between the paired readings (higher correlations would result in an estimate of fewer patient samples thus the choice of .7 is conservative) and if the average standard deviation of the mean of the differences is used as an estimate of the expected variability, then it would be possible to declare measurement equivalence for a difference of ±.19 with a sample size of 40 patients. This assumes the ASTM measurements of instrument variation are representative of the within patient variance. Given the vagaries of patient-to-patient this may be too optimistic an estimate.
A sample size of at least 40 patients will provide 80% or greater power for testing for equivalence between methods where the mean difference is .1 or less for up to a ten-fold increase in the estimates of the variability of the differences based on the ASTM precision measures and for a mean difference of .15 for up to a five-fold increase in these same measures. If the sample size is 40 then the power for the test of equivalence, where the difference is .15 and the standard deviation is a ten-fold increase, is 35%. Agreement testing: The Bland-Altman test will be used to test the agreement of the two methods. This test will permit a check for significance of bias between the two measures (the average difference between the two measures) as well as a check for significant trending over the range of the measurements. If the bias is significant then the results of the two differ by an overall offset in their measurements. If the trending (slope of the regression line) is significant then the difference between the two measurements changes as their magnitude changes which means the two methods are not in agreement.
The only measurements of instrument precision are those provided by the ASTM assessment. It is reasonable to assume the within patient measurement variation will be greater than the measurements from the controlled testing. A sample size of at least 40 patients is, from a statistical standpoint, a reasonable number and this sample size provides acceptable power for a mean difference of .1 for a ten-fold increase in the estimate of minimum test variability. This sample size will also provide enough data for the Bland-Altman test for agreement. Because this is a repeated measures test and even one missing measurement will impact the results, the investigators plan to over-sample in order to assure the study has the power to answer the questions posed.
- Conditions
- Temperature
- Being tested
- TempTraq
What participants receive
What participants receive. In a randomised study each person is assigned to one group by chance.
TempTraqDiagnostic test
sensor-based external thermometer withmonitoring capability
What is being measured
The primary outcome is the main question the study is designed to answer.
Temperature Comparison
Core temperature monitoring (degree Celsius) with temp-Traq thermometer compared to Pulmonary Artery (PA) catheter
Measured: Multiple time points up to 6 hours post thermometer placement
Timeline
From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.
- First posted29 Aug 2017
- Study start15 Dec 2016
- Primary completion1 Sep 2017
- Study completion1 Oct 2017
Resources
- Official record on ClinicalTrials.govThe registry entry this page is built from, with its full change history.
- Posted resultsParticipant flow, outcome measures and adverse events reported by the sponsor.
- Publications (5)Papers the sponsor has linked to this study.
- Other trials for TemperatureStudies of the same condition, open and under way.
- More trials run by The Cleveland ClinicThe sponsor's other registered studies.
- What the trial phases meanHow a treatment moves from first tests in people to approval.
More details
For doctors and researchers: how the study is designed, who runs it, and its registry IDs.
Study design
- Study type
- Observational
- Observational model
- Cohort
- Time perspective
- Prospective
- Enrolment
- 60 participants (actual)
Sponsor and investigators
- Lead sponsor
- The Cleveland Clinic
- Sponsor type
- University, hospital or other
- Collaborators
- Blue Spark Technologies
- Responsible party
- Principal investigator, Sandra Siedlecki, The Cleveland Clinic
IDs, dates and publications
- ClinicalTrials.gov ID
- NCT03264833
- Sponsor's protocol ID
- NCT-03264833
- Status last verified
- Jan 2026
- First submitted
- 2 Feb 2017
- Last updated
- 12 Feb 2026
- Results posted
- Yes, on ClinicalTrials.gov
- Data monitoring committee
- No
- FDA-regulated drug
- No
- FDA-regulated device
- No
- Individual data sharing
- No: IPD will not be shared
Publications
Eliades, A, Stoner, K. and Frisone, J. (unpublished paper) Use of skin patch device to obtain temperature measurements
Farnell S, Maxwell L, Tan S, Rhodes A, Philips B. Temperature measurement: comparison of non-invasive methods used in adult critical care. J Clin Nurs. 2005 May;14(5):632-9. doi: 10.1111/j.1365-2702.2004.00916.x.
PubMed 15840078Lawson L, Bridges EJ, Ballou I, Eraker R, Greco S, Shively J, Sochulak V. Accuracy and precision of noninvasive temperature measurement in adult intensive care patients. Am J Crit Care. 2007 Sep;16(5):485-96.
PubMed 17724246Pinsky MR, Vincent JL. Let us use the pulmonary artery catheter correctly and only when we need it. Crit Care Med. 2005 May;33(5):1119-22. doi: 10.1097/01.ccm.0000163238.64905.56.
PubMed 15891346Sund-Levander M, Grodzinsky E. Assessment of body temperature measurement options. Br J Nurs. 2013 Aug 8-Sep 12;22(15):880, 882-8. No abstract available.
PubMed 24052967
Study at a glance
From the registry record, last updated by the sponsor on 12 Feb 2026. Recruitment status is as the sponsor reported it: confirm with the study team.
- Official title
- Accuracy and Precision Test of TempTraq® Compared to Pulmonary Artery Catheter for Monitoring Temperature in Adults in Intensive Care Unit
- ClinicalTrials.gov ID
- NCT03264833
- Acronym / sponsor ID
- NCT-03264833
- Conditions
- Temperature
- Study type
- Observational
- Recruitment status
- Completed
- Enrolment
- 60 participants (actual)
- Lead sponsor
- The Cleveland ClinicUniversity, hospital or other
- Collaborators
- Blue Spark Technologies
- Study start
- 15 Dec 2016 (actual)
- Primary completion
- 1 Sep 2017 (actual)
- Study completion
- 1 Oct 2017 (actual)
- First posted
- 29 Aug 2017 (actual)
- Results
- Posted 12 Feb 2026
Clinical summary
The research question as the sponsor registered it. This describes what the study plans to test, not what it found.
Purpose The purpose of this feasibility study is to assess the accuracy and precision of the Temp-Traq thermometer for monitoring body temperature in adult patients under three conditions (hypothermia, normothermia, and hyperthermia).
Specific Aims
- To assess the accuracy of the Temp-Traq thermometer as compared to a gold standard (Core temp measured by Pulmonary Artery Catheter)
- To assess the precision of the Temp-Traq thermometer over repeated measures
- To determine if accuracy \&/or precision is consistent in three conditions (hypothermia, normothermia, or hyperthermia)
Detailed description
Design A repeated measures within-group comparative design will be used for this study.
Sample/Setting To address the specific aims of this study the investigators will identify patients who have a pulmonary artery (PA) catheter in place (patients in intensive care units). Letters of support are attached from the unit managers where the study will take place (see attached).
Procedure Once a patient is identified, the researcher will record basic demographic data (see data collection form attached) and the axillary area will be assessed for visible signs of any skin conditions. If there are no visible signs of a skin condition the researcher will apply the Temp-Traq thermometer to the right or left axilla of the subject and record the location of the placement. The researcher will record both the PA and Temp-Traq recordings of subject's temperature (taken at the same time) on the data collection form (see attachment) at four data points; baseline (5 minutes post application of Temp-Traq) and every two hours (+/- 15 minutes) after baseline X 3. Previous testing of the TempTraq® on children and healthy adults has not resulted in any adverse skin reactions. However, as this is a new thermometer format and there is not a significant body of literature on potential skin reactions, the investigators will assess skin before and after application. Thus, at the end of the last reading the Temp-Traq thermometer will be removed and the skin will be assessed for erythema. If erythema is present the patient's nurse and physician will be notified by the researcher.
Protection of Human Subjects Since non-invasive temperature monitoring poses no risk to patients and is part of usual care and since the investigators will not collect or record any protected health information (PHI), written informed consent will not be sought. Patients will be identified by their study ID number only. The data will be collected by a registered nurse data collector familiar with the intensive care setting and data will be stored in a locked file cabinet in the researchers locked office. Temperature data for the Temp-Traq is collected through the Temp-Traq Application (AP) which will be placed on a Cleveland Clinic Approved and encrypted i-pad. Once the study is completed and all data has been transferred to the study database, the Temp-Traq application and all of its data will be deleted from the i-pad.
Analysis The investigators will use the TOST (Two One Sided t-tests) method for equivalence testing. This method requires a defined range of mean differences between two test methods, an estimate of the precision of the measurement of the two systems, and an estimate of the size of the possible difference in the means of the two methods under consideration. The TOST null hypothesis is a joint null hypothesis that the mean measurement differences between the two methods is greater than a critical lower bound and less than a critical upper bound. If the null is rejected then the investigators can conclude that the absolute difference of the means between Temp Traq and PA catheter falls within the specified range. It was determined that to be considered equivalent with respect to accuracy the mean measurements of the two methods should be within ± 0.2 degrees of each other. Precision tests for Temp-Traq thermometer using accepted ASTM test methods for various combinations of temperature and humidity provided a range of measurement variation between .0000435 and .019928 with a mean of 0.000254500. If the investigators assume both test methods exhibit the same levels of precision then the variance of the differences between their two means will reduce to two times the values listed previously. This will result in estimates of the standard deviations of the mean differences of .0093, .032, and .063 respectively.
The two methods for body temperature that will be examined are core temperature measurements with a PA catheter and temperature measurements using the Temp-Traq (measured in degree Celsius simultaneously). Paired temperature measurements for both methods will be taken simultaneously every 2 hours for a 6 hour period. This will provide 4 repeated temperature measures per patient. The simultaneous measurements within a given patient should exhibit a high degree of correlation. If .7 is chosen as the lower bound for the correlation between the paired readings (higher correlations would result in an estimate of fewer patient samples thus the choice of .7 is conservative) and if the average standard deviation of the mean of the differences is used as an estimate of the expected variability, then it would be possible to declare measurement equivalence for a difference of ±.19 with a sample size of 40 patients. This assumes the ASTM measurements of instrument variation are representative of the within patient variance. Given the vagaries of patient-to-patient this may be too optimistic an estimate.
A sample size of at least 40 patients will provide 80% or greater power for testing for equivalence between methods where the mean difference is .1 or less for up to a ten-fold increase in the estimates of the variability of the differences based on the ASTM precision measures and for a mean difference of .15 for up to a five-fold increase in these same measures. If the sample size is 40 then the power for the test of equivalence, where the difference is .15 and the standard deviation is a ten-fold increase, is 35%. Agreement testing: The Bland-Altman test will be used to test the agreement of the two methods. This test will permit a check for significance of bias between the two measures (the average difference between the two measures) as well as a check for significant trending over the range of the measurements. If the bias is significant then the results of the two differ by an overall offset in their measurements. If the trending (slope of the regression line) is significant then the difference between the two measurements changes as their magnitude changes which means the two methods are not in agreement.
The only measurements of instrument precision are those provided by the ASTM assessment. It is reasonable to assume the within patient measurement variation will be greater than the measurements from the controlled testing. A sample size of at least 40 patients is, from a statistical standpoint, a reasonable number and this sample size provides acceptable power for a mean difference of .1 for a ten-fold increase in the estimate of minimum test variability. This sample size will also provide enough data for the Bland-Altman test for agreement. Because this is a repeated measures test and even one missing measurement will impact the results, the investigators plan to over-sample in order to assure the study has the power to answer the questions posed.
Study design
- Observational model
- Cohort
- Time perspective
- Prospective
Who may qualify
- Ages
- 21 years and older
- Sex
- All
- Age groups
- Adult, Older adult
- Study population
- To address the specific aims of this study the investigators will identify at least 40 patients who have a pulmonary artery (PA) catheter in place (patients in intensive care units).
Key inclusion criteria (1 listed)
- Inclusion Criteria Adult patient (male or female) in intensive care unit who will have a PA catheter in place for at least the next 8 hours; and who have no visible skin condition to the axillary region upon inspection by the research nurse are eligible for this study
Key exclusion criteria (4 listed)
- Patients younger than 21
- Patients who will not have a PA catheter in place for the next 8 hours
- Patients who have a visible skin condition to the axillary region upon inspection by the research nurse
- Patients who will be going for procedures or for some reason will not be available for the 6 hours of the study
Eligibility is decided by the study team after screening; this overview does not establish that a patient qualifies. Full inclusion and exclusion criteria.
Interventions
| Intervention | Type | Arms |
|---|---|---|
| TempTraqsensor-based external thermometer withmonitoring capability | Diagnostic test | – |
Primary outcome
Temperature Comparison
Time frame: Multiple time points up to 6 hours post thermometer placement
Core temperature monitoring (degree Celsius) with temp-Traq thermometer compared to Pulmonary Artery (PA) catheter
Results and safety
Loading the posted results from ClinicalTrials.gov…
Research links
- Official record on ClinicalTrials.govLast updated by the sponsor 12 Feb 2026.
- Posted resultsParticipant flow, baseline, all outcome measures and adverse events.
- BackgroundEliades, A, Stoner, K. and Frisone, J. (unpublished paper) Use of skin patch device to obtain temperature measurements
- Background: PubMed 15840078Farnell S, Maxwell L, Tan S, Rhodes A, Philips B. Temperature measurement: comparison of non-invasive methods used in adult critical care. J Clin Nurs. 2005 May;14(5):632-9. doi: 10.1111/j.1365-2702.2004.00916.x.
- Background: PubMed 17724246Lawson L, Bridges EJ, Ballou I, Eraker R, Greco S, Shively J, Sochulak V. Accuracy and precision of noninvasive temperature measurement in adult intensive care patients. Am J Crit Care. 2007 Sep;16(5):485-96.
- Background: PubMed 15891346Pinsky MR, Vincent JL. Let us use the pulmonary artery catheter correctly and only when we need it. Crit Care Med. 2005 May;33(5):1119-22. doi: 10.1097/01.ccm.0000163238.64905.56.
- Background: PubMed 24052967Sund-Levander M, Grodzinsky E. Assessment of body temperature measurement options. Br J Nurs. 2013 Aug 8-Sep 12;22(15):880, 882-8. No abstract available.
- Individual participant dataSharing planned: No. IPD will not be shared
Related trials for Temperature
Not yet recruiting
Accuracy of Moni-Patch Compared With Tympanic Temperature Monitoring in Adults at Risk of Fever
Fever, Temperature Monitoring
Recruiting
Moni-Patch Temperature Data Collection for Algorithm Development During Surgery
Body Temperature Monitoring
Recruiting
Thermal Therapies in the Treatment of Muscle Injuries
Hamstring Injuries, Hamstring Muscle Repair
Not yet recruiting
Stimulation-Temperature-Oxygen-Positioning Multimodal Neuroprotection for Acute Ischemic Stroke (STOP)
Acute Ischemic Stroke
RecruitingObservational study
Host and Parasite Factors That Influence Susceptibility to Malaria Infection and Disease During Pregnancy and Early Childhood in Ouelessebougou and Bamako, Mali
Malaria
Referral and contact
AzTrials does not enrol participants or take referrals. Contact the study team or a site listed below; they decide who can take part. This study is not recruiting at the moment.
The record lists no contact. Use the official record to reach the sponsor.
Questions about this trial
What is this clinical trial studying?
An observational study of Temperature, run by The Cleveland Clinic.
The full registered description is in the About this study section and the official record.
Who may be eligible to take part?
The sponsor lists 1 inclusion and 4 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.
Is this clinical trial currently recruiting?
The record says: finished, as of the sponsor's last update on 12 Feb 2026. Status can change between updates, so confirm with the study contact before making plans.
What are the possible benefits and risks?
They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.
Are results available for this study?
Yes. Results are posted in the registry (first posted 12 Feb 2026). A summary is in the For doctors view, and the full results are on ClinicalTrials.gov.
How can a doctor refer a patient or contact the research team?
The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.
What does the primary outcome measure?
Temperature Comparison (time frame: Multiple time points up to 6 hours post thermometer placement). It is what the study was designed to measure, not a finding.
Where can I read the original study information?
On the ClinicalTrials.gov record for NCT03264833, which also lists 5 linked publications.
Does a listing mean the treatment is approved or proven to work?
No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.
How often is this information updated?
AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 12 Feb 2026.