Thoracoabdominal Aortic AneurysmsNCT03246126
Visceral Manifold Study for the Repair of Thoracoabdominal Aortic Aneurysms
A study testing Valiant™Thoracoabdominal Stent Graft System in people with Thoracoabdominal Aortic Aneurysms, run by NYU Langone Health at 1 site in United States.
Trial summary
The purpose of this study is to evaluate the use of an investigational device called the ValiantTM Visceral Manifold Thoracoabdominal Stent Graft System for the repair of thoracoabdominal aortic aneurysms (TAAA), which is a balloon-like bulge in the aorta (major artery leading away from your heart) that originates in… Read more about this study
- Who is it for?
- Ages 18 years and older
- Conditions the study is for
- Thoracoabdominal Aortic Aneurysms
- What is being tested?
- Valiant™Thoracoabdominal Stent Graft System
- Type of treatment
- Medical device
- Could I receive a placebo?
- No, everyone receives an active treatment
- How many people
- 17 participants
- Where
- 1 siteUnited States
- Study dates
- 1 Nov 2017to 7 Jun 2023
- Phase
- Interventional
Key requirements
Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.
- Ages 18 years and older
- All sexes
- No healthy volunteers
- Adult, Older adult
Primary Arm Inclusion Criteria:
A patient may be entered into the study if the patient has at least one of the following
- An aneurysm with a maximum diameter of > 5.5 cm or 2 times the normal diameter just proximal to the aneurysm using orthogonal (i.e., perpendicular to the centerline) measurements
- Aneurysm with a history of growth > 0.5 cm in 6 months
- Saccular aneurysm deemed at significant risk for rupture
- Symptomatic aneurysm greater than or equal to 4.5 cm
Additional Primary Arm Inclusion Criteria:
- Axillary or brachial and iliac or femoral access vessel morphology that is compatible with vascular access techniques, devices or accessories, with or without use of a surgical conduit
- Proximal landing zone for the thoracic bifurcation stent graft that has: (a) ≥ 2.5 cm of nonaneurysmal aortic segment including previously placed graft material (neck) distal to the left subclavian artery (LSA); (b) Diameter in the range of 26-42 mm; (c) Adequate distance from the celiac artery, in order to accommodate cannulation from the antegrade access point when considering the total deployed length of the thoracic bifurcation and visceral manifold
- Iliac artery or aortic distal fixation site, including both native tissue and previously placed graft, greater than or equal to 15 mm in length and diameter in the range of 8 - 25 mm
- Life expectancy: > 1 year
Expanded Use Arm Inclusion Criteria:
- Branch vessel diameter <5mm
- Urgent or emergent presentation
- Patient has a contained rupture
- Patient has a ruptured aneurysm
- Patient has a type B dissection (subacute or chronic) in the portion of the aorta intended to be treated
- Poor performance status including two major system failures (cardiovascular, pulmonary, renal, hepatobiliary, and neuromuscular)
- Baseline creatinine greater than 2.0 mg/dL
- Anatomy that does not allow maintenance of at least one hypogastric artery
- Anatomy that does not allow primary or assisted patency of the left subclavian artery
- Prior aneurysm repair that would involve relining of the previously placed graft material requiring placement of the investigational system in a landing zone that expands beyond any limits of the previously placed graft material
- Obstructive stenting of any or all of the visceral vessels
- Treatment of the patient population listed above will be considered for this expanded arm of the study if they are not a candidate for open surgical repair, cannot be treated with approved devices, and do not meet inclusion into the primary study arm as per opinion of the Principal Investigator, with the concurrence of the IRB.
Primary Arm Exclusion Criteria:
- Patient is a good candidate for and elects for open surgical repair
- Can be treated in accordance with the instructions for use with a legally marketed endovascular prosthesis
- Is eligible for enrollment in a manufacturer- sponsored IDE at the investigational site
- Unwilling to comply with the follow-up schedule
- Inability or refusal to give informed consent by patient or legal representative
- Urgent or emergent presentation
- Patient is pregnant or breastfeeding
- Patient has a contained rupture
- Patient has a ruptured aneurysm
- Patient has a dissection in the portion of the aorta intended to be treated
- Obstructive stenting of any or all of the visceral vessels
- Poor performance status including two major system failures (cardiovascular, pulmonary, renal, hepatobiliary, and neuromuscular)
- Prior aneurysm repair that would involve relining of the previously placed graft material requiring placement of the investigational system in a landing zone that expands beyond any limits of the previously placed graft material
- Known sensitivities or allergies to the materials of construction of the devices, including nitinol (Nickel: Titanium), polyester, platinum-iridium, polytetrafluoroethylene (PTFE), platinum, gold, polyethylene, or stainless steel.
- Known hypersensitivity or contraindication to anticoagulation or contrast media that cannot be adequately medically managed
- Uncorrectable coagulopathy
- Body habitus that would inhibit x-ray visualization of the aorta or exceeds the safe capacity of the equipment
- Patient has had a major surgical or interventional procedure unrelated to the treatment of the aneurysm planned < 30 days of the endovascular repair
- Unstable angina (defined as angina with a progressive increase in symptoms, new onset at rest or nocturnal angina)
- Systemic or local infection that may increase the risk of endovascular graft infection
- Baseline creatinine greater than 2.0 mg/dL
- History of connective tissue disorders (e.g., Marfan Syndrome, Ehler's Danlos Syndrome)
- ranch vessel diameter less than 5 mm
- Thrombus or excessive calcification within the neck of the aneurysm
- Anatomy that would not allow maintenance of at least one patent hypogastric artery
- Anatomy that would not allow primary or assisted patency of the left subclavian artery
Trial locations and contacts
1 site in 1 country.
United States1 site
New York University School of Medicine
New York, New York, 10016
About this study
The purpose of this study is to evaluate the use of an investigational device called the ValiantTM Visceral Manifold Thoracoabdominal Stent Graft System for the repair of thoracoabdominal aortic aneurysms (TAAA), which is a balloon-like bulge in the aorta (major artery leading away from your heart) that originates in your chest and extends to your abdomen and also includes the branch arteries that supply blood to the liver, spleen, intestine, kidneys and other organs in your abdomen. The word "investigational" means the device is still being tested and is not approved by the Food and Drug Administration (FDA) for sale in the United States.
Read the detailed description
The primary objective of the clinical investigation is to assess the use of the thoracic bifurcation and the visceral manifold to repair thoracoabdominal aortic aneurysms in patients having appropriate anatomy. The primary intent of the study is to assess safety (i.e. freedom from major adverse events (MAE) at 30 days) and preliminary effectiveness (i.e., treatment success and technical success) of the device (i.e., the proportion of treatment group subjects that achieve and maintain treatment success at one year). Additionally, the study will assess technical success and treatment success at each follow-up interval.
- Conditions
- Thoracoabdominal Aortic Aneurysms
- Being tested
- Valiant™Thoracoabdominal Stent Graft System
- Medical subject headings
- Aortic Aneurysm, Thoracoabdominal
What participants receive
What participants receive. In a randomised study each person is assigned to one group by chance.
Valiant™Thoracoabdominal Stent Graft SystemDevice
Also called Endurant Stent Graft System
The Visceral Manifold System is comprised of the Thoracic Bifurcation and Visceral Manifold to treat a thoracoabdominal aortic aneurysm or TAAA. The experimental procedure to implant the stent graft will be done in an operating room and will last anywhere from four to eight hours, but may take longer.
| Study group | Role | What they receive |
|---|---|---|
| Primary Arm: Treatment with Valiant Thoracoabdominal Stent Graft System | Experimental | The implantation of the Valiant Thoracoabdominal Stent Graft System is conducted under fluoroscopic/angiographic guidance. Receives: Device: Valiant™Thoracoabdominal Stent Graft System |
| Expanded Use Arm: Treatment with Valiant Thoracoabdominal Stent Graft System | Experimental | The implantation of the Valiant Thoracoabdominal Stent Graft System is conducted under fluoroscopic/angiographic guidance. Receives: Device: Valiant™Thoracoabdominal Stent Graft System |
What is being measured
The primary outcome is the main question the study is designed to answer.
Freedom From Major Adverse Events (MAE) at Day 30
Freedom from MAE is defined as the number of participants who have not experienced a major adverse event (including death, bowel ischemia, myocardial infarction, paraplegia, renal failure, respiratory failure, and stroke) at 30 days.
Measured: Up to Day 30
Number of Participants Who Achieve Technical Success
Technical success is defined as successful access of the aneurysm site and deployment of the Visceral Manifold investigational devices in intended locations.
Measured: Up to Year 1
Secondary outcomes (1)
Freedom From Paraplegia at Day 30
Number of participants who have not experienced paraplegia at Day 30.
Measured: Up to Day 30
Timeline
From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.
- First posted11 Aug 2017
- Study start1 Nov 2017
- Primary completion7 Jun 2023
- Study completion7 Jun 2023
Resources
- Official record on ClinicalTrials.govThe registry entry this page is built from, with its full change history.
- Posted resultsParticipant flow, outcome measures and adverse events reported by the sponsor.
- Other trials for Thoracoabdominal Aortic AneurysmsStudies of the same condition, open and under way.
- More trials run by NYU Langone HealthThe sponsor's other registered studies.
- What the trial phases meanHow a treatment moves from first tests in people to approval.
More details
For doctors and researchers: how the study is designed, who runs it, and its registry IDs.
Study design
- Study type
- Interventional
- Phase
- Phase N/A
- Primary purpose
- Treatment
- Allocation
- Non-randomised
- Intervention model
- Parallel groups
- Masking
- None (open label)
- Enrolment
- 17 participants (actual)
Sponsor and investigators
- Lead sponsor
- NYU Langone Health
- Sponsor type
- University, hospital or other
- Responsible party
- Sponsor
- Investigators
- Thomas Maldonado, Principal investigator, NYU Langone Health
IDs, dates and publications
- ClinicalTrials.gov ID
- NCT03246126
- Sponsor's protocol ID
- 16-01441
- Status last verified
- Mar 2024
- First submitted
- 2 Aug 2017
- Last updated
- 10 Apr 2024
- Results posted
- Yes, on ClinicalTrials.gov
- Data monitoring committee
- No
- FDA-regulated drug
- No
- FDA-regulated device
- Yes
- Individual data sharing
- Undecided