AzTrials
AzTrials

Mitral Annular CalcificationSITRALNCT02830204

Surgical Implantation of TRAnscatheter vaLve in Native Mitral Annular Calcification (SITRAL) Study

A study testing Mitral Valve Replacement (MVR) with Sapien3 in people with Mitral Annular Calcification, run by Baylor Research Institute at 6 sites in United States.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

The purpose of this study is to establish the safety and feasibility of the Edwards SAPIEN 3 valve in subjects with mitral annular calcification (MAC) associated with mitral stenosis (MS) and/or mitral regurgitation who are at high-risk for mitral valve surgery or deemed inoperable due to the extent of calcification. Read more about this study

Who is it for?
Ages 22 years and older
Conditions the study is for
Mitral Annular Calcification
What is being tested?
Mitral Valve Replacement (MVR) with Sapien3
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
25 participants
Where
6 sitesUnited States
Study dates
4 Nov 2016to 29 Nov 2023
Phase
Interventional
Are there results?
Yes, posted in the registrySummary in the doctors' view

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 22 years and older
  • All sexes
  • No healthy volunteers
  • Adult, Older adult

Who may be able to join

  • Subject has severe native mitral annular calcification associated with mitral stenosis and/or regurgitation. Qualifying echo must be within 60 days of the date of the procedure.
  • Subject has a clinical indication for mitral valve replacement, as demonstrated by reported New York Heart Association (NYHA) Functional Class II or greater.
  • The subject is at least 22 years old.
  • The Heart Team agrees that the subject is high risk or inoperable for surgical mitral valve repair or replacement (MVR), based on a conclusion that the probability of death or serious, irreversible morbidity exceeds the probability of meaningful improvement. The following concomitant procedures, are allowed: Maze, Tricuspid Valve Procedures (TVP), and Atrial Fibrillation (AF) ablation, coronary artery bypass grafting and septal myectomy.
  • The study subject has been informed of the nature of the study, agrees to its provisions and has provided written informed consent as approved by the Institutional Review Board (IRB) of the respective clinical site.
  • The study subject agrees to comply with all required post-procedure follow-up visits

Who cannot join

  • Evidence of an acute myocardial infarction (MI) ≤ 30 days before the intended treatment [defined as: Q wave MI, or non-Q wave MI with total creatine kinase (CK) elevation of creatine kinase-myocardial band (CK-MB) ≥ twice normal in the presence of myocardial band (MB) elevation and/or troponin level elevation (WHO definition)].
  • Any therapeutic invasive cardiac procedure resulting in a permanent implant that is performed within 30 days of the index procedure (unless part of planned strategy for treatment of concomitant coronary artery disease).
  • Leukopenia (white blood cell count < 2000 cell/mL), acute anemia (hemoglobin < 8 g/dL), or thrombocytopenia (platelet count < 50,000 cell/mL).
  • Hemodynamic or respiratory instability requiring vasoactive medications, mechanical ventilation at time of procedure.
  • Need for emergency surgery for any reason.
  • Severe left ventricular dysfunction with Left Ventricular Ejection Fraction (LVEF) < 30%.
Show all 14Show fewer
  • Severe right ventricular dysfunction
  • Pregnancy, lactation, or planning to become pregnant
  • Echocardiographic evidence of left ventricular mass, thrombus, or concerns of active infective endocarditis.
  • Active upper gastrointestinal (GI) bleeding within 3 months prior to procedure without treatment or 30 days prior to procedure with definitive treatment.
  • A known contraindication or hypersensitivity to all anticoagulation regimens, or inability to be maintained on oral anticoagulant following the study procedure.
  • End stage renal disease requiring dialysis
  • Clinically (by neurologist) or neuroimaging confirmed stroke or transient ischemic attack (TIA) within 30 days of the procedure.
  • Estimated life expectancy < 12 months

Trial locations and contacts

6 sites in 1 country.

6 sites in 1 country

United States6 sites
  • MHRI Maryland

    Baltimore, Maryland

  • University of Virginia

    Charlottesville, Virginia, 22903

  • Baylor College of Medicine

    Houston, Texas

  • St. Vincent Heart Center

    Indianapolis, Indiana, 46290

  • The Heart Hospital Baylor Plano

    Plano, Texas, 75093

  • Lankenau Institute for Medical Research

    Wynnewood, Pennsylvania, 19096

Other trials in United States

About this study

The purpose of this study is to establish the safety and feasibility of the Edwards SAPIEN 3 valve in subjects with mitral annular calcification (MAC) associated with mitral stenosis (MS) and/or mitral regurgitation who are at high-risk for mitral valve surgery or deemed inoperable due to the extent of calcification.

Read the detailed description

Technical Success:

Alive, with

  • Successful access, delivery and retrieval of the device delivery system, and
  • Deployment and correct positioning (including repositioning/recapture if needed) of the single intended device, and
  • No need for additional unplanned or emergency surgery or re-intervention related to the device or access procedure

Device Success:

Alive and stroke free, with

  • Original intended device in place, and
  • No additional surgical or interventional procedures related to access or the device since completion of the original procedure (i.e., exit from the cath lab/Operating Room(OR)), and
  • Intended performance of the device:
  • Structural performance: No migration, embolization, detachment, fracture, hemolysis, thrombosis (including reduced leaflet motion) or endocarditis, etc, and
  • Hemodynamic performance: Maintenance of relief of stenosis or insufficiency without producing the opposite (Stenosis = Mitral Valve Area (MVA) < 1.5cm2 and Mitral Valve (MV) gradient > 5mmHg, Insufficiency = Mitral Regurgitation (MR) >1+), and
  • Absence of para-device complications (e.g., Paravalvular Leak (PVL) > mild, need for a Permanent Pacemaker (PPM), erosion, Annular rupture or Aortic Valve (AV) Groove disruption, Left Ventricular Outflow Tract (LVOT) gradient increase > 10mmHg)

Procedural Success:

Device success, and

  • No device or procedure related Serious Adverse Events (SAEs) (Life threatening bleed; major vascular or cardiac structural complications requiring unplanned reintervention or surgery; stage 2 or 3 Acute Kidney Injury (includes new dialysis); Myocardial Infarction (MI) or need for percutaneous coronary intervention (PCI)/coronary artery bypass graft (CABG); severe heart failure (HF) or hypotension requiring intravenous (IV) inotrope, ultrafiltration or mechanical circulatory support; prolonged intubation ( > 48 hours)

6.2 Secondary objective(s)

Device Success (at 6 months and 1 year)

  • Subject success 1 year

Device Success:

Alive and stroke free, with

  • Original intended device in place, and
  • No additional surgical or interventional procedures related to access or the device since completion of the original procedure (i.e., exit from the cath lab/Operating Room (OR)), and
  • Intended performance of the device:
  • Structural performance: No migration, embolization, detachment, fracture, hemolysis, thrombosis (including reduced leaflet motion) or endocarditis, etc, and
  • Hemodynamic performance: Maintenance of relief of stenosis or insufficiency without producing the opposite (Stenosis = MVA < 1.5cm2 and MV gradient > 5mmHg, Insufficiency = MR >1+), and
  • Absence of para-device complications (e.g., PVL > mild, need for a PPM, erosion, Annular rupture or AV Groove disruption, LVOT gradient increase > 10mmHg)
Being tested
Mitral Valve Replacement (MVR) with Sapien3
Keywords
mitral valve regurgitation, mitral valve stenosis
Medical subject headings
Mitral Valve Insufficiency, Mitral Valve Stenosis

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Mitral Valve Replacement (MVR) with Sapien3Device

    Also called MVR

    subjects with surgical MVR with Sapien3

Study groups in NCT02830204
Study groupRoleWhat they receive
Mitral Valve Replacement (MVR) with Sapien3Experimentalsubjects with surgical MVR with Sapien3
Receives: Device: Mitral Valve Replacement (MVR) with Sapien3

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Procedural Success

    Procedural Success is defined as:

    1. Device success, and
    2. No device or procedure related serious adverse events (SAEs) (Life threatening bleed; major vascular or cardiac structural complications requiring unplanned reintervention or surgery; stage 2 or 3 acute kidney injury (AKI) (includes new dialysis); myocardial infarction (MI) or need for percutaneous coronary intervention (PCI)/coronary artery bypass graft (CABG); severe heart failure (HF) or hypotension requiring intravenous (IV) inotrope, ultrafiltration or mechanical circulatory support; prolonged intubation ( > 48 hours)

    Measured: 30 days

  2. Technical Success

    Technical Success is defined as:

    1. Alive, with
    2. Successful access, delivery and retrieval of the device delivery system, and
    3. Deployment and correct positioning (including repositioning/recapture if needed) of the single intended device, and
    4. No need for additional unplanned or emergency surgery or re-intervention related to the device or access procedure

    Measured: Procedure Stop Time

  3. Device Success

    Device Success is defined as:

    1. Alive and stroke free, with
    2. Original intended device in place, and
    3. No additional surgical or interventional procedures related to access or the device since completion of the original procedure (i.e., exit from the cath lab/operating room (OR)), and
    4. Intended performance of the device:
    5. Structural performance: No migration, embolization, detachment, fracture, hemolysis, thrombosis (including reduced leaflet motion) or endocarditis, etc, and
    6. Hemodynamic performance: Maintenance of relief of stenosis or insufficiency without producing the opposite (Stenosis = mitral valve area (MVA) < 1.5cm2 and mitral valve (MV) gradient > 5mmHg, Insufficiency = mitral regurgitation (MR) >1+), and
    7. Absence of para-device complications (e.g., paravalvular leak (PVL) > mild, need for a permanent pacemaker (PPM), erosion, Annular rupture or aortic valve (AV) Groove disruption, left ventricular outflow tract (LVOT) gradient increase > 10mmHg)

    Measured: 30 days

Secondary outcomes (3)
  1. Device Success

    Device Success is defined as:

    1. Alive and stroke free, with
    2. Original intended device in place, and
    3. No additional surgical or interventional procedures related to access or the device since completion of the original procedure (i.e., exit from the cath lab/operating room (OR)), and
    4. Intended performance of the device:
    5. Structural performance: No migration, embolization, detachment, fracture, hemolysis, thrombosis (including reduced leaflet motion) or endocarditis, etc, and
    6. Hemodynamic performance: Maintenance of relief of stenosis or insufficiency without producing the opposite (Stenosis = MVA < 1.5cm2 and MV gradient > 5mmHg, Insufficiency = MR >1+), and
    7. Absence of para-device complications (e.g., PVL > mild, need for a PPM, erosion, Annular rupture or AV Groove disruption, LVOT gradient increase > 10mmHg)

    Measured: 6 months

  2. Device Success

    Device Success is defined as:

    1. Alive and stroke free, with
    2. Original intended device in place, and
    3. No additional surgical or interventional procedures related to access or the device since completion of the original procedure (i.e., exit from the cath lab/OR), and
    4. Intended performance of the device:
    5. Structural performance: No migration, embolization, detachment, fracture, hemolysis, thrombosis (including reduced leaflet motion) or endocarditis, etc, and
    6. Hemodynamic performance: Maintenance of relief of stenosis or insufficiency without producing the opposite (Stenosis = MVA < 1.5cm2 and MV gradient > 5mmHg, Insufficiency = MR >1+), and
    7. Absence of para-device complications (e.g., PVL > mild, need for a PPM, erosion, Annular rupture or AV Groove disruption, LVOT gradient increase > 10mmHg)

    Measured: 1 year

  3. Subject Success

    Subject Success is defined as:

    All of the following must be present:

    1. Device success (either optimal or acceptable), and II. Patient returned to the pre-procedural setting: and III. No rehospitalizations or reinterventions for the underlying condition (e.g., mitral regurgitation, heart failure); and IV. Improvement from baseline in symptoms (e.g., NYHA improvement by > or = 1 functional class); and V. Improvement from baseline in functional status (e.g., 6-min walk test improvement by > or = 50 m); and VI. Improvement from baseline in quality-of-life (e.g., Kansas City Cardiomyopathy Questionnaire improvement by > or = 10)

    Measured: 1 year

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted12 Jul 2016 (estimated)
  2. Study start4 Nov 2016
  3. Primary completion3 Jan 2023
  4. Study completion29 Nov 2023

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Treatment
Allocation
Not applicable
Intervention model
Single group
Masking
None (open label)
Enrolment
25 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT02830204
Sponsor's protocol ID
016-100
Acronym
SITRAL
Status last verified
Feb 2025
First submitted
4 May 2016
Last updated
24 Feb 2025
Data monitoring committee
Yes
FDA-regulated drug
No
FDA-regulated device
Yes
Individual data sharing
No: There is currently no plan to share individual participant data

Questions about this trial

What is this clinical trial studying?

A study testing Mitral Valve Replacement (MVR) with Sapien3 in people with Mitral Annular Calcification, run by Baylor Research Institute at 6 sites in United States.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 6 inclusion and 14 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 24 Feb 2025. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 6 sites in United States. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

Yes. Results are posted in the registry (first posted 24 Feb 2025). A summary is in the For doctors view, and the full results are on ClinicalTrials.gov.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Procedural Success (time frame: 30 days). The study lists 3 primary outcomes in all. It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 24 Feb 2025.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in